Index: AlarmDefs.h =================================================================== diff -u -r8f7d37b13b80fbf0bf90227ebe736bdcb5573df7 -r4f151adda6a8e1ea9cf7fdd8806a81b3491d2baa --- AlarmDefs.h (.../AlarmDefs.h) (revision 8f7d37b13b80fbf0bf90227ebe736bdcb5573df7) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 4f151adda6a8e1ea9cf7fdd8806a81b3491d2baa) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) Dara Navaei -* @date (last) 16-Jul-2023 +* @author (last) Sean Nash +* @date (last) 20-Jul-2023 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -1389,7 +1389,7 @@ { 228 , ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre- Tx Prime Timeout | HD time out on prime sub-mode purge air state. | During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. | "}, { 229 , ALARM_ID_HD_PRIME_DIALYSATE_DIALYZER_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre- Tx Prime Timeout | HD prime dialysate dialyzer time out alarm | If the priming of the dialyzer dialysate flow path has exceeded the time limit. | "}, { 230 , ALARM_ID_HD_PRIME_DIALYSATE_BYPASS_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre- Tx Prime Timeout | HD prime dialysate bypass time out alarm | If the priming of the dialyzer bypass flow path has exceeded the time limit. | "}, - { 231 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "HD | Self Test: Pressure | A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press OK to resume. \n- If problem persists, install a new cartridge.\n\nProcess: Pre- Tx Dry Self- Test | HD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, + { 231 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "HD | Self Test: Cartridge Check | A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press OK to resume. \n- If problem persists, install a new cartridge.\n\nProcess: Pre- Tx Dry Self- Test | HD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, { 232 , ALARM_ID_HD_PRE_TREATMENT_WET_LC_TEST_FAILURE , "HD | Self Test: Dialysate Flow Rate | A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre- Tx Wet Self- Test | HD pre-treatment mode wet self-tests failure | If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState | "}, { 233 , ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Cal CRC | DG barometric pressure sensor bad coefficients CRC | If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. | "}, { 234 , ALARM_ID_DG_INLET_WATER_PRESSURE_IN_HIGH_RANGE , "DG | Incoming Water Pressure High | Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nRESUME button will be available.\n- Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres High | DG inlet water pressure is too high | If the inlet water pressure is > 80 psig. | "},