Index: AlarmDefs.h =================================================================== diff -u -r90961069b30c0188cdaae90923e22f16ca78e3d5 -r4396595671c3b2136f36092304c22304602188fc --- AlarmDefs.h (.../AlarmDefs.h) (revision 90961069b30c0188cdaae90923e22f16ca78e3d5) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 4396595671c3b2136f36092304c22304602188fc) @@ -283,8 +283,8 @@ ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT = 245, ///< DG heat disinfect target temperature time out (could not reach to temperature) ALARM_ID_DG_CLEANING_MODE_COND_SENSORS_OUT_OF_RANGE = 246, ///< DG cleaning mode conductivity sensors out of range ALARM_ID_DG_HEAT_DISINFECT_ACTIVE_COOL_TIME_OUT = 247, ///< DG heat disinfect active cool time tour - ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT = 248, ///< DG chemical disinfect could not reach to target temperature - ALARM_ID_AVAILABLE_4 = 249, ///< Alarm ID available for use + ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE = 248, ///< DG chemical disinfect could not reach to target temperature out of range + ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE = 249, ///< DG chemical disinfect target conductivity out of range ALARM_ID_AVAILABLE_5 = 250, ///< Alarm ID available for use ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC = 251, ///< HD invalid system record ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC = 252, ///< HD invalid service record @@ -635,7 +635,7 @@ { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_LOW, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NEW_RESERVOIRS_DATA_MESSAGE_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NEW_DG_OPERATION_MODE_MESSAGE_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH, 0, ALARM_ID_NO_ALARM }, @@ -729,8 +729,8 @@ { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CLEANING_MODE_COND_SENSORS_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_ACTIVE_COOL_TIME_OUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_4, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_5, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, @@ -956,7 +956,7 @@ { 3, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE }, { 3, ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE }, { 3, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE }, - { 3, ALARM_ID_AVAILABLE_4 }, + { 3, ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE }, { 3, ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT }, { 3, ALARM_ID_AVAILABLE_5 }, { 3, ALARM_ID_DG_CLEANING_MODE_TEMP_SENSORS_DIFF_OUT_OF_RANGE }, @@ -995,7 +995,7 @@ { 3, ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT }, { 3, ALARM_ID_DG_CLEANING_MODE_COND_SENSORS_OUT_OF_RANGE }, { 3, ALARM_ID_DG_HEAT_DISINFECT_ACTIVE_COOL_TIME_OUT }, - { 3, ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT }, + { 3, ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE }, { 3, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE }, { 3, ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT }, { 3, ALARM_ID_HD_RTC_BATTERY_LOW }, @@ -1394,8 +1394,8 @@ { 245 , ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT , "DG | Heat Disinfection Error | Call service to schedule a repair. | DG heat disinfect target temperature time out (could not reach to temperature) | If heating up to minimum temperature for heat disinfect failed after 4 hours. | "}, { 246 , ALARM_ID_DG_CLEANING_MODE_COND_SENSORS_OUT_OF_RANGE , "DG | Chemical Disinfection Error | Call service to schedule a repair. | DG chemical disinfect inlet conducitivty out of range | If CPo and CD2 conductivity values are above 100 uS/cm. | "}, { 247 , ALARM_ID_DG_HEAT_DISINFECT_ACTIVE_COOL_TIME_OUT , "DG | Heat Disinfection Active Cool Error | Call service to schedule a repair. | DG heat disinfect active cool time out | If THd > 44 C after 90 minutes of cooling. | "}, - { 248 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT , "DG | Chemical Disinfection Error | Call service to schedule a repair. | DG chemical disinfect could not reach to target temperature | If heating up to minimum temperature for chemical disinfect failed | "}, - { 249 , ALARM_ID_AVAILABLE_4 , "N/A | N/A | N/A | N/A | N/A | "}, + { 248 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE , "DG | Chemical Disinfection Error | Call service to schedule a repair. | DG chemical disinfect out of target temperature | If TPo > 50 C after 1 second. | "}, + { 249 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE , "DG | Chemical Disinfection Error | Call service to schedule a repair. | DG chemical disinfect out of target conductivity | If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. | "}, { 250 , ALARM_ID_AVAILABLE_5 , "N/A | N/A | N/A | N/A | N/A | "}, { 251 , ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC , "HD | Machine Error | Call service to schedule a repair. | HD invalid system record | If the calculated CRC of the HD system record image does not match the expected CRC. | "}, { 252 , ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC , "HD | Machine Error | Call service to schedule a repair. | HD invalid service record | If the calculated CRC of the HD service record image does not match the expected CRC. | "},