Index: AlarmDefs.h =================================================================== diff -u -r965807a3ebb89a23fbf9799122fe974c8b4b7cc3 -r30d3d5b6cf73c708c6d63fc321d73eade44d1b49 --- AlarmDefs.h (.../AlarmDefs.h) (revision 965807a3ebb89a23fbf9799122fe974c8b4b7cc3) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 30d3d5b6cf73c708c6d63fc321d73eade44d1b49) @@ -125,7 +125,7 @@ ALARM_ID_DG_HEATING_INVALID_CAL_RECORD = 87, ///< DG heating invalid calibration record ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE = 88, ///< DG concentrate pumps hall sensor out of range ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR = 89, ///< HD BP occlusion sensor error - ALARM_ID____AVAILABLE_25 = 90, ///< This alarm ID is available for use + ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR = 90, ///< DG load cells FPGA read error. ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE = 91, ///< HD No dialysate flow data receive in the last 3 seconds ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS = 92, ///< DG conductivity sensor bad status. ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE = 93, ///< DG outlet primary conductivity out of range @@ -152,10 +152,10 @@ ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 114, ///< HD venous air bubble detector self-test failure ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE = 115, ///< DG temperature sensor out of range ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE = 116, ///< DG temperature sensor ADC out of range - ALARM_ID_DG_PRIMARY_HEATER_INTERNAL_TEMP_OUT_OF_RANGE = 117, ///< DG primary heater internal temperature out of range // TODO remove when DEN-11750 was merged - ALARM_ID_DG_PRIMARY_HEATER_CJ_TEMP_OUT_OF_RANGE = 118, ///< DG primary heater cold junction temperature out of range // TODO remove when DEN-11750 was merged - ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE = 119, ///< DG trimmer heater internal temperature out of range // TODO remove when DEN-11750 was merged - ALARM_ID_DG_TRIMMER_HEATER_CJ_TEMP_OUT_OF_RANGE = 120, ///< DG trimmer heater cold junction temperature out of range // TODO remove when DEN-11750 was merged + ALARM_ID____AVAILABLE_33 = 117, ///< This alarm ID is available for use + ALARM_ID____AVAILABLE_32 = 118, ///< This alarm ID is available for use + ALARM_ID____AVAILABLE_31 = 119, ///< This alarm ID is available for use + ALARM_ID____AVAILABLE_30 = 120, ///< This alarm ID is available for use ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE = 121, ///< DG main primary heater voltage out of range ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE = 122, ///< DG small primary heater voltage out of range ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE = 123, ///< DG trimmer heater voltage out of range @@ -233,7 +233,7 @@ ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE = 195, ///< DG RO pump pressure out of range ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT = 196, ///< DG temperature sensor error flag fault. ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT = 197, ///< DG temperature sensors ADC fault - ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE = 198, ///< DG heaters cold junction temperature out of range + ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE = 198, ///< DG heaters cold junction temperature out of range // TODO remove when DEN-12224 merged ALARM_ID_DG_HEATERS_FAULT = 199, ///< DG heaters fault ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE = 200, ///< DG thermistors/sensors temperature out of range ALARM_ID____AVAILABLE_29 = 201, ///< This alarm ID is available for use @@ -450,8 +450,8 @@ { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_WATCHDOG_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_COMM_POST_FAILED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RTC_CONFIG_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_RTC_OR_TIMER_ACCURACY_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, @@ -522,20 +522,20 @@ // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_PUMP_TRACK_LATCH_OPENED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_PUMP_TRACK_LATCH_OPENED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID____AVAILABLE_25, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -561,15 +561,15 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_INTERNAL_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_CJ_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_CJ_TEMP_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID____AVAILABLE_33, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID____AVAILABLE_32, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID____AVAILABLE_31, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID____AVAILABLE_30, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 4, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_HIGH, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 5, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK, 0, ALARM_ID_NO_ALARM }, @@ -580,7 +580,7 @@ { ALARM_PRIORITY_HIGH, 11, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_VENOUS_PRESSURE_HIGH, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 12, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_ARTERIAL_PRESSURE_LOW, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 13, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_ARTERIAL_PRESSURE_HIGH, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 14, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_DG_FLUID_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 14, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE, ALARM_ID_DG_FLUID_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 15, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_FLUID_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 16, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_SHOCK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 16, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_SHOCK, 0, ALARM_ID_NO_ALARM }, @@ -615,9 +615,9 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM, 0, ALARM_ID_NO_ALARM }, @@ -628,37 +628,37 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_SOFTWARE_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_LOAD_CELL_VALUE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UV_REACTOR_NOT_HEALTHY, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UV_REACTOR_NOT_HEALTHY, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP1_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP2_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_OFF_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_OFF_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID____AVAILABLE_29, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_INLET_WATER_PRESSURE_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_METER_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 800, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_SITTING_WARNING, 0, ALARM_ID_NO_ALARM }, @@ -690,7 +690,7 @@ { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_INSTALL_NEW_CARTRIDGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_NO_CARTRIDGE_LOADED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_REMOVAL_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_REMOVAL_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT, 0, ALARM_ID_NO_ALARM }, @@ -709,7 +709,7 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_UI_COMPATIBILITY_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DG_COMPATIBILITY_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_POWER_OUT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_POWER_OUT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_POWER_OUT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SHASUM, 0, ALARM_ID_NO_ALARM }, @@ -734,14 +734,14 @@ { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_HD_TREATMENT_RECIRC_TIMEOUT_WARNING, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_WARNING, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION, 0, ALARM_ID_NO_ALARM }, }; // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To @@ -848,7 +848,7 @@ { 3, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD }, { 3, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE }, { 3, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR }, - { 3, ALARM_ID____AVAILABLE_25 }, + { 3, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR }, { 3, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE }, { 3, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS }, { 3, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE }, @@ -874,10 +874,10 @@ { 3, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE }, { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE }, - { 3, ALARM_ID_DG_PRIMARY_HEATER_INTERNAL_TEMP_OUT_OF_RANGE }, - { 3, ALARM_ID_DG_PRIMARY_HEATER_CJ_TEMP_OUT_OF_RANGE }, - { 3, ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE }, - { 3, ALARM_ID_DG_TRIMMER_HEATER_CJ_TEMP_OUT_OF_RANGE }, + { 3, ALARM_ID____AVAILABLE_33 }, + { 3, ALARM_ID____AVAILABLE_32 }, + { 3, ALARM_ID____AVAILABLE_31 }, + { 3, ALARM_ID____AVAILABLE_30 }, { 3, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE }, { 3, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE }, { 3, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE }, @@ -913,6 +913,18 @@ { 3, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE }, { 3, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME }, { 3, ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, + { 3, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE }, + { 3, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE }, + { 3, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE }, + { 3, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, + { 3, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, + { 3, ALARM_ID_UV_REACTOR_NOT_HEALTHY }, + { 3, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT }, + { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT }, + { 3, ALARM_ID_DG_LOAD_CELL_ADC_ERROR }, + { 3, ALARM_ID_DG_FPGA_COMM_TIMEOUT }, + { 3, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT }, + { 3, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT }, { 4, ALARM_ID_END_OF_TREATMENT_HIGH }, { 5, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, { 6, ALARM_ID_HD_BLOOD_LEAK_DETECTED }, @@ -951,7 +963,6 @@ { 799, ALARM_ID_PUMP_TRACK_LATCH_OPENED }, { 799, ALARM_ID_CARTRIDGE_DOOR_OPENED }, { 799, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT }, - { 799, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE }, { 799, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY }, { 799, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION }, { 799, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM }, @@ -961,32 +972,21 @@ { 799, ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH }, { 799, ALARM_ID_DG_SOFTWARE_FAULT }, { 799, ALARM_ID_HD_COMM_TIMEOUT }, - { 799, ALARM_ID_DG_FPGA_COMM_TIMEOUT }, - { 799, ALARM_ID_DG_LOAD_CELL_ADC_ERROR }, { 799, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE }, { 799, ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION }, - { 799, ALARM_ID_UV_REACTOR_NOT_HEALTHY }, { 799, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT }, { 799, ALARM_ID_CP1_SPEED_CONTROL_ERROR }, { 799, ALARM_ID_CP2_SPEED_CONTROL_ERROR }, - { 799, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, - { 799, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, - { 799, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE }, - { 799, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE }, { 799, ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE }, { 799, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE }, - { 799, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT }, - { 799, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT }, { 799, ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE }, { 799, ALARM_ID_DG_HEATERS_FAULT }, { 799, ALARM_ID____AVAILABLE_29 }, { 799, ALARM_ID_INLET_WATER_PRESSURE_FAULT }, { 799, ALARM_ID_PRESSURE_SENSOR_FAULT }, { 799, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT }, - { 799, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT }, { 799, ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT }, { 799, ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED }, - { 799, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT }, { 799, ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT }, { 799, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED }, { 799, ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION },