Index: AlarmDefs.h =================================================================== diff -u -r9c1dea2b9e72dd1f7eebd0be2a7b279e58f9eee0 -ra2b9ebe48e6d25c002a183f9a33d7690be0b87ed --- AlarmDefs.h (.../AlarmDefs.h) (revision 9c1dea2b9e72dd1f7eebd0be2a7b279e58f9eee0) +++ AlarmDefs.h (.../AlarmDefs.h) (revision a2b9ebe48e6d25c002a183f9a33d7690be0b87ed) @@ -1164,330 +1164,330 @@ const ALARM_MANUAL_T ALARM_MANUAL_TABLE[] = { // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition - { 0 , ALARM_ID_NO_ALARM , "N/A | | | | Indication for when no alarms is a possible situation | N/A | "}, - { 1 , ALARM_ID_HD_STUCK_BUTTON_TEST_FAILED , "HD | Stuck Button | A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. | HD POST: Stuck Button | Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up | Stop or Off button are pressed for more than 1 sec during POST. | "}, - { 2 , ALARM_ID_HD_FPGA_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: FPGA Compatibility | HD FPGA POST failure | Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. | "}, - { 3 , ALARM_ID_DG_FPGA_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: FPGA Compatibility | DG FPGA POST failure | Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header | "}, - { 4 , ALARM_ID_HD_WATCHDOG_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: WD Safestate | HD Watchdog POST failure | If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA | "}, - { 5 , ALARM_ID_DG_WATCHDOG_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: WD Safestate | DG Watchdog POST failure | If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V | "}, - { 6 , ALARM_ID_HD_UI_COMM_POST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Proc- UI Comm | UI communication POST failure. UI failed to communicate within a reasonable time after power up | If the UI failed to checkin via the CAN bus within 40 seconds | "}, - { 7 , ALARM_ID_HD_RTC_BATTERY_LOW , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): RTC Bat Low | Real-time clock configuration error | If the RTC is not functional as indicated by the bits in the first 3 control registers. | "}, - { 8 , ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Acc Vector | HD accelerometer failed POST | If the calculated vector length of the accelerometer x, y, z is not the nominal length. | "}, - { 9 , ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Acc Vector | DG accelerometer failed POST | If the calculated vector length of the accelerometer x, y, z is not the nominal length. | "}, - { 10 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Config Emb Mode | HD blood leak sensor embedded mode communication failed | If the blood leak sensor embedded mode is unresponsive. | "}, - { 11 , ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP , "HD | Low Dialysate Temperature | Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. | Dial. Gen: Dialysate Temp Low | Dialysate temperature is below the target temperature | If the dialysate temperature is below the target temperature for a certain period of time. | "}, - { 12 , ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: HD Image | HD firmware image integrity POST test failed | If the calculated CRC of the HD image does not match the expected CRC. | "}, - { 13 , ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: DG Image | DG firmware image integrity POST test failed | If the calculated CRC of the DG firmware image does not matched the expected CRC | "}, - { 14 , ALARM_ID_DG_INVALID_USAGE_RECORD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Usage Rec NVRAM CRC | DG non-volatile usage information group invalid record CRC | If the DG usage information record's CRC does not match the calculated CRC in POST. | "}, - { 15 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP , "HD | High Dialysate Temperature | Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. | Dial. Gen: Dialysate Temp High | Dialysate temperature is out of high safety range | If the dialysate temperature is > 46 C for a certain period of time. | "}, - { 16 , ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Primary Alarm Current | HD alarm audio failed POST | If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output | "}, - { 17 , ALARM_ID_HD_UI_POST_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI POST Not Complete | HD UI POST failed | If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. | "}, - { 18 , ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_DG , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (DG ACK) | HD to DG comm error. | HD message to DG that requires acknowledge was not acknowledged within a certain time. | "}, - { 19 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP , "HD | High Dialysate Temperature | Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. | Dial. Gen: Dialysate Temp High | Dialysate temperature is above the target temperature | If the dialysate temperature is above the target temperature for a certain period of time. | "}, - { 20 , ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conductivity Cal | DG conductivity sensors invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, - { 21 , ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Drain Line Vol Cal | DG drain line volume invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, - { 22 , ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Reservoir Vol Cal | DG reservoirs invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, - { 23 , ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Acid Conc Cal | DG acid concentrate calibration record | If the non-volatile data is valid and if not raise the alarm | "}, - { 24 , ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Bicarb Conc Cal | DG bicarb concentrate calibration record | If the non-volatile data is valid and if not raise the alarm | "}, - { 25 , ALARM_ID_DG_ACCELEROMETERS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Acc Cal | DG accelerometer invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, - { 26 , ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Acc Cal | HD accelerometer invalid calibration record | If the accel calibration data that was received from NVDataMgmt is NOT legitimate. | "}, - { 27 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED , "HD | Self Test: Blood Leak | A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD BLD sensor failed zero sequence | BLD fails to complete zero sequence. | "}, - { 28 , ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Conductivity RTD Comm | DG two wire sensors FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 29 , ALARM_ID_HD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: SP DAC Cal | HD heparin force sensor invalid calibration record | If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. | "}, - { 30 , ALARM_ID_HD_SOFTWARE_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: SW Invalid State | HD Software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, - { 31 , ALARM_ID_HD_BLOOD_PUMP_MC_CURRENT_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Current | Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running | If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. | "}, - { 32 , ALARM_ID_HD_BLOOD_PUMP_OFF_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Speed | Blood pump failed motor off check. Measured speed while commanded off | If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM | "}, - { 33 , ALARM_ID_HD_BLOOD_PUMP_MC_DIRECTION_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Direction | Blood pump failed motor direction check. Measured vs. commanded | If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. | "}, - { 34 , ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_CHECK , "HD | Blood Pump Error | A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. | HD Fault: BP Rotor | Blood pump failed rotor speed check. Mismatch with rotor and motor speeds | If the delta of the measured motor speed and the commanded motor speed is not within tolerance. | "}, - { 35 , ALARM_ID_HD_DIAL_IN_PUMP_MC_CURRENT_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPi Current | Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running | If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. | "}, - { 36 , ALARM_ID_HD_DIAL_IN_PUMP_OFF_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPi Speed | Dialysis inlet pump failed motor off check. Measured speed while commanded off | If during command off state, the DPi pump motor speed is measured > 100.0 RMP. | "}, - { 37 , ALARM_ID_HD_DIAL_IN_PUMP_MC_DIRECTION_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPi Direction | Dialysis inlet pump failed motor direction check. Measured vs. commanded | If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. | "}, - { 38 , ALARM_ID_HD_DIAL_IN_PUMP_ROTOR_SPEED_CHECK , "HD | Dialysate Pump Error | A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. | HD Fault: DPi Rotor | Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds | If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. | "}, - { 39 , ALARM_ID_HD_DIAL_OUT_PUMP_MC_CURRENT_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPo Current | Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running | If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. | "}, - { 40 , ALARM_ID_HD_DIAL_OUT_PUMP_OFF_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPo Speed | Dialysis outlet pump failed motor off check. Measured speed while commanded off | If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. | "}, - { 41 , ALARM_ID_HD_DIAL_OUT_PUMP_MC_DIRECTION_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPo Direction | Dialysis outlet pump failed motor direction check. Measured vs. commanded | If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. | "}, - { 42 , ALARM_ID_HD_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK , "HD | Dialysate Pump Error | A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. | HD Fault: DPo Rotor | Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds | If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed | "}, - { 43 , ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Iso ADC Comm | Main primary heater FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 44 , ALARM_ID_HD_UI_COMM_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UI Comm | User interface communication timeout | If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. | "}, - { 45 , ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN CRC Com | Too many bad CRCs detected on received system messages | If too many bad messages CRCs are received within a set period of time | "}, - { 46 , ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_UI , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN ACK UI Com | System message that required acknowledgment was not acknowledged | If no retries left in the pending acknowlegement list. | "}, - { 47 , ALARM_ID_DG_MAX_RO_PUMP_PWM_EXCEEDED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Pump High | DG RO pump out of upper range | If the RO pump > 95% PWM for 12 secs. | "}, - { 48 , ALARM_ID_HD_UF_VOLUME_ACCURACY_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UF Rate Tare Err | Ultrafiltration volume accuracy error during treatment | If the UF volume (refUFvolume - measureVolume) > 100.0 mL. | "}, - { 49 , ALARM_ID_HD_FPGA_COMM_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: FPGA Com | HD FPGA communication down for too long | If there are more than 3 comm errors in a 1 min period. | "}, - { 50 , ALARM_ID_DG_VALVE_CONTROL_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Valve Control | DG FPGA not accepting commanded valve states | If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. | "}, - { 51 , ALARM_ID_HD_BLOOD_PUMP_MOTOR_SPEED_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Speed | Blood pump failed motor speed check. Measured vs. commanded | If measured motor speed versus commanded rotor speed exceeded limits. | "}, - { 52 , ALARM_ID_HD_DIAL_IN_PUMP_MOTOR_SPEED_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPi Speed | Dialysate inlet pump failed motor speed check. Measured vs. commanded | If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. | "}, - { 53 , ALARM_ID_HD_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPo Speed | Dialysate outlet pump failed motor speed check. Measured vs. commanded | If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. | "}, - { 54 , ALARM_ID_HD_CRITICAL_DATA_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Critical Data Err | HD critical data integrity check failed | If the critical data failed range check. | "}, - { 55 , ALARM_ID_DG_CRITICAL_DATA_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Critical Data Err | DG critical data integrity check failed | If the critical data failed range check. | "}, - { 56 , ALARM_ID_HD_ACCELEROMETER_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Acc Com | HD accelerometer error (no readings or FPGA reports error) | If no fresh sample of accelerometer data has been detected for more than 100 msec. | "}, - { 57 , ALARM_ID_DG_ACCELEROMETER_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Acc Com | DG accelerometer error (no readings or FPGA reports error) | If no fresh sample of accelerometer data has been detected for more than 100 msec. | "}, - { 58 , ALARM_ID_HD_VALVE_HOMING_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Home | HD valve homing failed | If the number of failed homing > = 3 counts. | "}, - { 59 , ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Transition Err | HD valve transition time out | if the valve's transition time has timed out (3000msec) | "}, - { 60 , ALARM_ID_HD_VALVE_NOT_FUNCTIONAL , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Control | HD valve not functional | If the valve bit masks are not as expected in different valve control mode. | "}, - { 61 , ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Current | HD valve current out of range | If the valve current is over the threshold of 1.0 A for 1000 msec. | "}, - { 62 , ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Position | HD valve position out of target | If the current position has deviated from the expected position for more than 1 second. | "}, - { 63 , ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Pre- Tx Hep. Pump Timeout. | HD syringe pump timeout alarm | If the syringe pump priming > 5 seconds | "}, - { 64 , ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Range | Barometric pressure is out of range | If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. | "}, - { 65 , ALARM_ID_HD_DG_COMMAND_INVALID_PARAMETER_FAULT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Reject) | HD requests DG command with invalid parameter fault | If the latest DG command response indicated invalid parameter | "}, - { 66 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Config Set Point | HD blood leak sensor set point set failure | If the set point of the blood leak sensor does not match the set point in the calibration record. | "}, - { 67 , ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD blood pump occlusion self-test failure alarm | If the blood pump pressure > 5000 | "}, - { 68 , ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE , "HD | Dialysate Generation Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. | Dial. Gen: Dialysate not ready | HD active reservoir recirculation out of range | If the recirculation level has exceeded the max allowed | "}, - { 69 , ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Cal | HD blood leak sensor invalid calibration record | If the BLD data that was received from NVDataMgmt is NOT legitimate. | "}, - { 70 , ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD arterial pressure self-test failure alarm | If arterial pressure < -300 or > 100 mmHg | "}, - { 71 , ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD venous pressure self-test failure alarm | If the venous pressure < -100 or > 600 mmHg | "}, - { 72 , ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (LC) | HD No load cell data message receive at least once every 2 seconds | No new load cell data for 2 seconds. | "}, - { 73 , ALARM_ID_HD_NEW_DIALYSATE_TEMP_DATA_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (TDI/TRO) | HD No dialysate temperature data message received at least once every 2 seconds | No new dialysate temperature data for 2 seconds. | "}, - { 74 , ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVi On No Flow | DG inlet UV reactor is on while there is no flow | If the inlet UV reactor is on while VPi is closed for a certain period of time. | "}, - { 75 , ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe DAC | HD syringe pump self-test failure alarm | If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. | "}, - { 76 , ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Voltage Monitor | HD monitored voltage is out of range | If the HD monitored voltages are out of range for a certain period of time. | "}, - { 77 , ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Voltage Monitor | DG monitored voltage is out of range | If the DG monitored voltages are out of range for a certain period of time. | "}, - { 78 , ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Enc Dir | HD syringe pump direction (from encoder) error | If the measured encoder direction of the syringe pump does not agree with the expected direction | "}, - { 79 , ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Drv Dir | HD syringe pump direction (from controller) error | If the measured direction of the syringe pump motor controller does not agree with the expected direction | "}, - { 80 , ALARM_ID_HD_SYRINGE_PUMP_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Driver | HD syringe pump fault reported by FPGA | If the FPGA reported the syringe pump fault | "}, - { 81 , ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump OT | HD syringe pump over travel error | If the syringe pump position is deemed to have overtraveled | "}, - { 82 , ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump DAC Err | HD syringe pump DAC write failure | If the syringe pump DAC write error is reported by the FPGA | "}, - { 83 , ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump On Err | HD syringe pump is running while the BP is off | If the syringe pump is on while the blood pump is off | "}, - { 84 , ALARM_ID_DG_SET_RTC_YEAR_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: DG RTC Time | DG RTC set year is invalid | If DG RTC set year is < 2021. | "}, - { 85 , ALARM_ID_HD_PUMP_TRACK_LATCH_OPENED , "HD | Cartridge Lever is Open | A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press Resume/OK button to continue. | Blood Stop: Cartridge Lever Open | HD pump track lever is not fully latched | If the pump track lever is detected as unlatched when it should be. | "}, - { 86 , ALARM_ID_HD_SET_RTC_YEAR_INVALID , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: HD RTC Time | HD RTC set year is invalid | If HD RTC set year is < 2021. | "}, - { 87 , ALARM_ID_DG_HEATING_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Heat Cal | DG heating parameters invalid calibration record | If the heating paramters' non-volatile record's CRC does not match the calculated CRC. | "}, - { 88 , ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CP1 CP2 Speed Range | DG concentrate pump hall sensor out of range. | Hall sensor pulse to pulse timing is < 1ms (speed too fast). | "}, - { 89 , ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVo On No Flow | DG outlet UV reactor is on while there is no flow | If the outlet UV reactor is on while the RO flow is less than 200 mL/min. | "}, - { 90 , ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC A1/B1 Comm | DG load cells A1/B1 FPGA fault | If the FPGA does not report fresh data for a certain period of time. | "}, - { 91 , ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (FMD) | HD No dialysate flow data received in the last 3 seconds | If not receiving new dialysate flow readings in expected interval (3 seconds) | "}, - { 92 , ALARM_ID_DG_TEMPERATURE_SENSORS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Temp Snsr Cal | DG temperature sensors invalid calibration | If the calibration date of the temperature sensors is 0. | "}, - { 93 , ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Permeate Cond High | DG outlet primary conductivity sensor out of range | If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. | "}, - { 94 , ALARM_ID_DG_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Pressure Sensor Range | DG pressure sensor out of range | If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. | "}, - { 95 , ALARM_ID_DG_WATCHDOG_EXPIRED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: WD Error | DG watchdog expired | If the watchdog signal is low while not in POST. | "}, - { 96 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE , "DG | Incoming Water Temperature High | Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Temp High | DG inlet water temperature out of range | If the inlet water temperature is > 37 C. | "}, - { 97 , ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Out of Range | Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Dialysate Cond. | DG fill conductivity out of range | If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. | "}, - { 98 , ALARM_ID_HD_BATTERY_COMM_FAULT , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Battery Com | HD battery or battery charger communication fault | If battery I2C communication fault persistence count > 5. | "}, - { 99 , ALARM_ID_HD_SYRINGE_PUMP_STALL , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Stall | HD syringe pump stall alarm | If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. | "}, - { 100 , ALARM_ID_DG_COND_SENSORS_INVALID_TEMP_COMP_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conductivity Cal | DG conductivity sensors temperatrue compensation invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, - { 101 , ALARM_ID_UI_POST_HD_COMM , "UI | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI- Proc Comm | UI POST FW Communication failure | If the HD is not communicating with the UI during POST. | "}, - { 102 , ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Tpo THD Temp Delta | DG heat disinfection temperature gradient out of range | If temperature gradient between Tpo and Thd exceeds maximum 15°C. | "}, - { 103 , ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Cal NVRAM CRC | HD non-volatile calibration group invalid record CRC | If the computed CRC of the calibration group record is not equal to the record CRC. | "}, - { 104 , ALARM_ID_HD_AIR_TRAP_ILLEGAL_LEVELS , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Air Trap Sensor | Air trap level sensors reporting illegal combination of air/fluid | If the air trap level readings are not valid for more than 2 seconds. | "}, - { 105 , ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Cal NVRAM CRC | DG non-volatile system group invalid record CRC | If the DG NVDATAMGMT CAL group calculated CRC does not matched the expected CRC. | "}, - { 106 , ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Flow Sensor Cal | DG flow sensors calibration error | If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. | "}, - { 107 , ALARM_ID_HD_DG_RESTARTED_FAULT , "HD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DG Restarted | HD reports DG restarted fault | If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. | "}, - { 108 , ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump ADC Com | HD syringe pump ADC error | If the ADC read is not fresh after POST init and the ADC read counter has not changed. | "}, - { 109 , ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Vol | HD syringe pump volume check error | If the volume delivered is off by more than 5% or 0.1mL. | "}, - { 110 , ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Speed | HD syringe pump speed check error | If the syringe measured rate is off by more than 5% or 0.1mL. | "}, - { 111 , ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Speed | HD syringe pump not stopped in off state error | If the syringe pump position is not equal to its last position. | "}, - { 112 , ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC A2/B2 Comm | DG load cells A2/B2 FPGA fault | If the FPGA does not report fresh data for a certain period of time. | "}, - { 113 , ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Current | DG drain pump current out of range | If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. | "}, - { 114 , ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE , "HD | Self Test: Venous Air Detector | A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD venous air bubble detector self-test failure | If the venous air bubble self test did not detect bubble for more than 500 msec. | "}, - { 115 , ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Temp Sensor Range | DG temperature sensor out of range | If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. | "}, - { 116 , ALARM_ID_HD_UI_SDCARD_FAILURE , "UI HD | Service Required: Dialysate Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): SD- Card Error | SD card is non-operational | SD card is non-operational | "}, - { 117 , ALARM_ID_DG_CAN_MESSAGE_NOT_ACKED , "DG | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of HD Comm (HD ACK) | System message that required acknowledgment was not acknowledged | If no retries left in the pending acknowlegement list. | "}, - { 118 , ALARM_ID_DG_RTC_CONFIG_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RTC Config | DG RTC configuration error | If the RTC configuration registers report an error. | "}, - { 119 , ALARM_ID_DG_RTC_BATTERY_LOW , "DG | Service Required: Dialysate Device | A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. | DG Alarm (Low): RTC Bat Low | DG RTC battery low | If the RTC battery bit report a low alarm. | "}, - { 120 , ALARM_ID_HD_PRE_TREATMENT_WET_PRIME_TEST_FAILURE , "HD | Self Test: Venous Air Detector | A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD pre-treatment wet prime test failure | If the HD pre-treatment wet self test failed. | "}, - { 121 , ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Prim Heater Lg Control | DG main primary heater voltage out of range | If the main primary heater's voltage is out of range. | "}, - { 122 , ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Prim Heater Sm Control | DG small primary heater voltage out of range | If the small primary heater's voltage is out of range. | "}, - { 123 , ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Trimmer Heater Control | DG trimmer heater voltage out of range | If the trimmer voltage is out of range. | "}, - { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, - { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, - { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. | Blood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, - { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- If there has been an access disconnection or needle\n dislodgement, follow facility policy to address and\nterminate treatment without rinseback.\n- If access connections are properly secured, \npress Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, - { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, - { 129 , ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "HD | Blood Leak: Bypass | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. | Blood Stop: Dialyzer Blood Leak Bypass | HD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "}, - { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, - { 131 , ALARM_ID_HD_ARTERIAL_PRESSURE_LOW , "HD | Low Arterial Pressure | Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. Low | Arterial pressure too low during treatment | If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, - { 132 , ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH , "HD | High Arterial Pressure | Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. High | Arterial pressure too high during treatment | If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, - { 133 , ALARM_ID_DG_FLUID_LEAK_DETECTED , "DG | Dialysate Device Fluid Leak | Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Fluid Leak in DG | DG fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, - { 134 , ALARM_ID_HD_FLUID_LEAK_DETECTED , "HD | Fluid Leak | Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. | Blood Stop: Fluid Leak in HD | HD fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, - { 135 , ALARM_ID_HD_SHOCK , "HD | Blood Pump Stopped: Excessive Movement | Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Blood Stop: HD Shock Detected | HD experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, - { 136 , ALARM_ID_DG_SHOCK , "DG | Dialysate Device: Excessive Movement | Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Dial. Gen: Shock- Vibration | DG experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, - { 137 , ALARM_ID_HD_EXCESSIVE_TILT , "HD | Blood Pump Stopped: Excessive Tilt | Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Blood Stop: HD Tilted | HD tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 3 deg for more than 1 sec. | "}, - { 138 , ALARM_ID_DG_EXCESSIVE_TILT , "DG | Dialysate Device: Excessive Tilt | Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. | Dial. Gen: Tilt | DG tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 3 deg for more than 1 sec. | "}, - { 139 , ALARM_ID_HD_AC_POWER_LOST , "HD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. | "}, - { 140 , ALARM_ID_HD_DG_COMM_TIMEOUT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Timeout) | Dialysate generator communication timeout | If DG communication timeout is detected for more than 2 seconds. | "}, - { 141 , ALARM_ID_HD_AIR_TRAP_FILL_DURING_TREATMENT , "HD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Fill | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "}, - { 142 , ALARM_ID_HD_OCCLUSION_BLOOD_PUMP , "HD | Blood Pump Stopped: Occlusion | Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Blood Pump Occlusion | Blood pump occlusion detected | If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall | "}, - { 143 , ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: TRo TDi Delta | DG dialysate temperature sensors drift timeout. | If TDi and TRo have drifted for more 1 C for a certain period of time. | "}, - { 144 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_HIGH , "DG | Disinfection: Water Conductivity High | Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Cond High | DG cleaning mode inlet water conductivity out of high range | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode | "}, - { 145 , ALARM_ID_DG_ACID_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Error During Prime | Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Acid Conc. Cond. | DG Concentrate conductivity after adding acid out of range alarm | If the conductivity value for the acid concentrate is out of range for more than 45 secs. | "}, - { 146 , ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: DG RTC Timer | DG RTC vs. Timer accuracy error | If RTC and timer are not aligned on 1 second elapsed time. | "}, - { 147 , ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT , "DG | Dialysate Generation Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. | Dial. Gen: Dialysate not ready | DG dialysate generation conductivity fault alarm | If dialysate conductivity is not with 5% of the TCD | "}, - { 148 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_HIGH , "DG | Disinfection: Water Temperature High | Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Temp High | Cleaning Mode Inlet water temperature too high | Inlet temperature > 37 C. | "}, - { 149 , ALARM_ID_HD_WATCHDOG_EXPIRED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: WD Error | HD watchdog expired | If the watchdog signal is low while not in POST. | "}, - { 150 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE , "DG | Incoming Water Conductivity High | Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Cond High | Inlet water conductivity too high | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). | "}, - { 151 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE , "DG | Incoming Water Conductivity Low | Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Cond Low | Inlet water conductivity too low | Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). | "}, - { 152 , ALARM_ID_DG_INLET_WATER_PRESSURE_IN_LOW_RANGE , "DG | Incoming Water Pressure Low | Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Pres Low | Inlet water pressure too low | Inlet water pressure < 25 PSI. | "}, - { 153 , ALARM_ID_HD_PRIME_COMPLETED_HIGH , "HD | Pre-Treatment Process Terminated | The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press Resume/OK for continue with next steps to remove \ncartridge from the device. | Process: Pre- Tx Terminated | HD prime completed high priority alarm | In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) | "}, - { 154 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_LOW , "DG | Disinfection: Water Conductivity Low | Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Cond Low | Cleaning mode inlet water conductivity too low | Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode | "}, - { 155 , ALARM_ID_HD_NEW_RESERVOIRS_DATA_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Reservoir) | HD No reservoir data message receive at least once every 2 seconds | No new reservoir data for 2 seconds. | "}, - { 156 , ALARM_ID_HD_DG_NEW_OPERATION_MODE_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Op Mode) | HD No DG op mode message receive at least once every 2 seconds | No DG op mode for 2 seconds. | "}, - { 157 , ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT , "DG | Chemical Disinfectant Not Detected | Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Chem Disinf PAA Prime | Chemical disinfect priming the acid line timeout | If the disinfectant acid line is not primed within a certain period of time. | "}, - { 158 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_LOW_RANGE , "DG | Incoming Water Temperature Low | Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Temp Low | Inlet water temperature low | Inlet temperature < 24 °C for 5 seconds. | "}, - { 159 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_REMOVE_ACID , "DG | Remove Chemical Disinfectant | The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. | Cleaning: Chem Disinf Connect Cap | Close concentrate line be able to flush it | Start of chemical disinfect flush. | "}, - { 160 , ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Rotor Max Speed | Blood pump rotor speed too high | If blood pump motor speed exceeded 100.0 RPM. | "}, - { 161 , ALARM_ID_DG_COMM_TOO_MANY_BAD_CRCS , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: HD- DG Comm | Too many bad CRCs detected on received system messages | If >= 5 bad CRCs detected on received CAN messages within 10 minutes. | "}, - { 162 , ALARM_ID_DG_FPGA_CLOCK_SPEED_CHECK_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Proc- FPGA Clock | FPGA clock speed failure | If the FPGA clock speed is relatively close to the processor clock for a certain period of time. | "}, - { 163 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM , "HD | Ultrafiltration Error | A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. | Dial. Gen: UF Rate Check | HD sees primary load cell for reservoir 1 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, - { 164 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM , "HD | Ultrafiltration Error | A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. | Dial. Gen: UF Rate Check | HD sees primary load cell for reservoir 2 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, - { 165 , ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_ALARM , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | HD in treatment rinseback operation timeout | If the maximum time inrinseback operation has been exceeded. | "}, - { 166 , ALARM_ID_TEST_ALARM_NO_BLOOD_RECIRC_CLR_TOP_ONLY , " | Test Alarm - Alarm 166 | | | | | "}, - { 167 , ALARM_ID_HD_CARTRIDGE_DOOR_OPENED , "HD | Blood Pump Stopped: Door Open | The device door is open.\n Close the device door. \n- Press Resume/OK to continue. | Blood Stop: Door Open | HD cartridge door opened alarm | If the cartridge door is opened as reported by the FPGA. | "}, - { 168 , ALARM_ID_TEST_ALARM_NO_DIAL_RECIRC_CLR_TOP_ONLY , " | Test Alarm - Alarm 168 | | | | | "}, - { 169 , ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Dialysate Flow Sens Rg | Dialysate flow rate out of range | If the dialysate flow rate > 2.00 L/min. | "}, - { 170 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY , "HD | Heparin Syringe Empty | Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. | Heparin Syringe Empty | HD syringe empty alarm | If the syringe pump position is near empty position. | "}, - { 171 , ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION , "HD | Heparin Pump Occlusion | A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. | Process: Pre- Tx Hep. Pump Occ. | HD syringe pump occlusion alarm | If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. | "}, - { 172 , ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM , "HD | Not Enough Heparin | There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. | Process: Pre- Tx Not Enough Hep. | HD syringe pump not enough Heparin alarm | If the estimated syringe volume needed for treatment is not sufficient | "}, - { 173 , ALARM_ID_HD_RTC_CONFIG_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: RTC Config | HD RTC configuration error | If the RTC configuration registers report an error. | "}, - { 174 , ALARM_ID_HD_RTC_OR_TIMER_ACCURACY_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: HD RTC Timer | HD RTC vs. Timer accuracy error | If RTC and timer are not aligned on 1 second elapsed time. | "}, - { 175 , ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Pump Commutation Err | HD pump commutation error | If the set pump direction does not agree with the hall sensor direction. | "}, - { 176 , ALARM_ID_TEST_ALARM_NO_RECIRC_CLR_TOP_ONLY , " | Test Alarm - Alarm 176 | | | | | "}, - { 177 , ALARM_ID_DG_SOFTWARE_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Software Invalid State | DG software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, - { 178 , ALARM_ID_HD_COMM_TIMEOUT , "UI | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UI- Proc Comm | HD communication timeout | If the HD stops communicating with the UI. | "}, - { 179 , ALARM_ID_DG_FPGA_COMM_TIMEOUT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: FPGA Com | DG FPGA communication down for too long | If retries for commands exceeds limit or FPGA reports comm error. | "}, - { 180 , ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Flow Sensor Range | RO flow rate out of range | If the RO flow rate > 2.00 L/min. | "}, - { 181 , ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Tare Out of Range | DG load cells weight out of range for tare | If the load cell weight is out of range for tare. | "}, - { 182 , ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: LC Cal | DG load cells invalid calibration | If the CRC of load cells calibration data does not match the calculated CRC. | "}, - { 183 , ALARM_ID_DG_INVALID_LOAD_CELL_VALUE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Invalid Resv Transfer | DG invalid load cell value | If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. | "}, - { 184 , ALARM_ID_DG_INLET_UV_REACTOR_NOT_HEALTHY , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVi Error | Inlet reactor not healthy | If the inlet reactor are not healthy. | "}, - { 185 , ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. | DG Alarm (Low): Fan Speed | DG fan RPM out of range | If the fan RPM is out of Min/Max range. | "}, - { 186 , ALARM_ID_DG_CONCENTRATE_PUMP_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CP1/CP2 Fault | DG concentrate pump fault | If the concentrate pumps fault is reported by the FPGA. | "}, - { 187 , ALARM_ID_DG_CP1_SPEED_CONTROL_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CP1 Speed Fault | DG concentrate pump CP1 speed control error | If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. | "}, - { 188 , ALARM_ID_DG_CP2_SPEED_CONTROL_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CP2 Speed Fault | DG concentrate pump CP2 speed control error | If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. | "}, - { 189 , ALARM_ID_DG_DRAIN_PUMP_RPM_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Speed Fault | DG drain pump RPM out of range | If the drain pump RPM > 100.0 of the target RPM. | "}, - { 190 , ALARM_ID_DG_DRAIN_PUMP_OFF_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Speed Not Off | DG drain pump off fault | If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. | "}, - { 191 , ALARM_ID_DG_FLOW_RATE_OUT_OF_UPPER_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Flow High | DG flow rate out of upper range | If the RO flow rate > 10% from the target flow rate. | "}, - { 192 , ALARM_ID_DG_FLOW_RATE_OUT_OF_LOWER_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Flow Low | DG flow rate out of lower range | If the RO flow rate < 10% from the target flow rate. | "}, - { 193 , ALARM_ID_HD_PARTIAL_OCCLUSION_BLOOD_PUMP , "HD | Blood Pump Stopped: Partial Occlusion | Partial occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Blood Pump Partial Occlusion | Blood pump partial occlusion detected | If blood occlusion is detected > PARTIAL_OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionDuringTreatement | "}, - { 194 , ALARM_ID_DG_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Driver Feedback | DG RO pump duty cycle out of range | If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. | "}, - { 195 , ALARM_ID_DG_RO_PUMP_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Pump High Pressure | DG RO pump pressure out of range | If the DG RO pump pressure > 135 psi for a certain period of time. | "}, - { 196 , ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CPi/CPo Comm | DG CPi/CPo conductivity sensors FPGA fault | If the FPGA reports error for a certain period of time. | "}, - { 197 , ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CD1/CD2 Comm | DG CD1/CD2 conductivity sensors FPGA fault | If the FPGA reports error for a certain period of time. | "}, - { 198 , ALARM_ID_DG_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: HP On With No Flow | DG primary heater while RO flow is low | If the RO flow is < 0.2 L/min for a certain period of time. | "}, - { 199 , ALARM_ID_DG_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: HR On With No Flow | DG trimmer heater while dialysate flow is low | If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. | "}, - { 200 , ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Thermistor Range | DG thermistor or temperature sensor out of range | If thermistor > 80.0 C or < 0.0 C. | "}, - { 201 , ALARM_ID_HD_PRE_TREATMENT_WET_FLOW_TEST_FAILURE , "HD | Self Test: Dialysate Flow Rate | A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD pre-treatment wet flow test failure | If the pre-treatment wet flow test has failed. | "}, - { 202 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE , "HD | Self Test: Pressure | A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. | Process: Pre- Tx Dry Self- Test | HD pre-treatment dry pressure normal test failure | If the pre-treatment dry pressure normal test failed. | "}, - { 203 , ALARM_ID_DG_FILL_CONDUCTIVITIES_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Fill Cond Cal | DG fill acid/bicarb invalid calibration record | If the fill target acid/bicarb conductivities' non-volatile record's calibration time is not 0. | "}, - { 204 , ALARM_ID_DG_RO_REJECTION_RATIO_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Rejection | RO rejection ratio out of range | If RO rejection ratio > 0.1 | "}, - { 205 , ALARM_ID_DG_CONDUCTIVITY_SENSOR_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Cond Sensors Error | Conductivity sensor fault | If conductivity sensor reports 0x11 for more than 2 consecutive reads. | "}, - { 206 , ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Fill Timeout | DG dialysate fill runs out of time | If DG dialysate fill > 5 minutes. | "}, - { 207 , ALARM_ID_DG_FLOW_METER_CHECK_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: FMP Flow Rate Error | DG flow meter check failure alarm | If integrated Volume To Load Cell Reading Percent > 0.1. | "}, - { 208 , ALARM_ID_HD_VENOUS_LINE_OCCLUSION , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Venous Pres. High | HD venous line occlusion fault | If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. | "}, - { 209 , ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Flush Circ Error | DG drain circulation line timeout | If the integrated volume < 0.1 for more than 60 secs. | "}, - { 210 , ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Battery Fault | HD battery pack detected an error | If the battery pack status error is detected from the slave device I2C address 0x0B. | "}, - { 211 , ALARM_ID_HD_BLOOD_SITTING_WARNING , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. | "}, - { 212 , ALARM_ID_HD_END_OF_TREATMENT_ALARM , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment end paused state, if the blood is in the line for more than 4 minutes. | "}, - { 213 , ALARM_ID_HD_PRIME_COMPLETED_MEDIUM , "HD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre- Tx Prime Complete | HD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) | "}, - { 214 , ALARM_ID_AVAILABLE_30 , " | | | | | | "}, - { 215 , ALARM_ID_AVAILABLE_31 , " | | | | | | "}, - { 216 , ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM , "HD | Treatment Complete | Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. | Process: End of Dialysis Tx | HD end treatment sub-mode timeout alarm | If the treatment ended successfully. | "}, - { 217 , ALARM_ID_AVAILABLE_32 , " | | | | | | "}, - { 218 , ALARM_ID_HD_SYRINGE_DETECTED , "HD | Syringe Detected | Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. | Process: Remove Syringe | HD syringe detected alarm | If the syringe detection switch > 2.0 volts. | "}, - { 219 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED , "HD | Blood Stopped: Syringe Not Detected | There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. | Blood Stop: Syringe Not Detected | HD syringe pump syringe removed alarm | If the syringe detection switch < 2.0 volts. | "}, - { 220 , ALARM_ID_DG_THD_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: THD RTD Comm | DG THd sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 221 , ALARM_ID_HD_EMPTY_SALINE_BAG , "HD | Blood Pump Stopped: Saline Bag Empty | The saline bag appears empty.\n- Check the saline bag and replace if empty. | Blood Stop: Saline Bag Empty | HD Saline bag is empty | If the measured arterial pressure < -300.0 mmHg for more than 250 msec. | "}, - { 222 , ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Occ Com | HD occlusion sensor FPGA fault | If the FPGA does not report fresh data or indicates an error for a certain period of time. | "}, - { 223 , ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Art. Pres. Com | HD arterial sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 224 , ALARM_ID_HD_TREATMENT_STOPPED_BY_USER , "HD | Blood Pump Stopped: Treatment Paused | Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. | Blood Stopped: Treatment Pause | Treatment stopped by user action - pressed stop button | If the user has pressed the stop button. | "}, - { 225 , ALARM_ID_HD_END_OF_TREATMENT_WARNING , "HD | Treatment Complete | Dialysis treatment is complete.\n- If blood sampling is required, draw blood at\n the arterial sample port\n- Press RINSEBACK to end treatment and rinseback. | Process: End of Treatment | HD end of treatment warning | If the treatment duration has exceeded the prescribed treatment time entered by the user. | "}, - { 226 , ALARM_ID_HD_PRIME_COMPLETED_LOW_PRIORITY , "HD | Priming Complete | Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre- Tx Priming Complete | HD prime completed low priority alarm | Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection | "}, - { 227 , ALARM_ID_AVAILABLE_15 , " | | | | | | "}, - { 228 , ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. | Process: Pre- Tx Prime Timeout | HD time out on prime sub-mode purge air state. | During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. | "}, - { 229 , ALARM_ID_HD_PRIME_DIALYSATE_DIALYZER_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. | Process: Pre- Tx Prime Timeout | HD prime dialysate dialyzer time out alarm | If the priming of the dialyzer dialysate flow path has exceeded the time limit. | "}, - { 230 , ALARM_ID_HD_PRIME_DIALYSATE_BYPASS_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. | Process: Pre- Tx Prime Timeout | HD prime dialysate bypass time out alarm | If the priming of the dialyzer bypass flow path has exceeded the time limit. | "}, - { 231 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "HD | Self Test: Cartridge Check | A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. | Process: Pre- Tx Dry Self- Test | HD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, - { 232 , ALARM_ID_HD_PRE_TREATMENT_WET_LC_TEST_FAILURE , "HD | Self Test: Dialysate Flow Rate | A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD pre-treatment mode wet self-tests failure | If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState | "}, - { 233 , ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Cal CRC | DG barometric pressure sensor bad coefficients CRC | If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. | "}, - { 234 , ALARM_ID_DG_INLET_WATER_PRESSURE_IN_HIGH_RANGE , "DG | Incoming Water Pressure High | Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Pres High | DG inlet water pressure is too high | If the inlet water pressure is > 80 psig. | "}, - { 235 , ALARM_ID_HD_TREATMENT_STOPPED_AFTER_RINSEBACK , "HD | Rinseback Complete | Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press Resume/OK to continue. | Process: Rinseback Complete | HD in treatment stopped sub-mode after rinseback completed (no escalation) | If there is no active alarm and rinseback has completed. | "}, - { 236 , ALARM_ID_HD_INSTALL_NEW_CARTRIDGE , "HD | Self Test: Fluid Present | A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume. | Process: Pre- Tx Dry Self- Test | HD needs new cartridge to be installed | If the air trap sensor reported AIR and bubble sensor detected bubbles. | "}, - { 237 , ALARM_ID_AVAILABLE_13 , " | | | | | | "}, - { 238 , ALARM_ID_HD_NO_CARTRIDGE_LOADED , "HD | Cartridge Not Detected | A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed. | Process: Pre- Tx Dry Self- Test | HD no cartridge loaded or installed improperly alarm | If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. | "}, - { 239 , ALARM_ID_HD_CARTRIDGE_REMOVAL_FAILURE , "HD | Cartridge Still Detected | A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed. | Process: Post- Tx Remove Cartridge | HD fail to remove cartridge alarm | In the post treatment state, if the cartridge has not been removed. | "}, - { 240 , ALARM_ID_DG_BICARB_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Error During Prime | Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Bicarb Conc. Cond. | Bicarb conductivity out of range during bicarb pump check alarm | If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. | "}, - { 241 , ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Resv Drain Timeout | DG reservoir drain time out | If the DG reservoir drain is not complete in 120 secs. | "}, - { 242 , ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Resv Fill Timeout | DG reservoir fill time out | If the DG reservoir fill is not complete in 120 secs. | "}, - { 243 , ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Fluid Loss Detected | DG reservoir leak time out | If the reservoir level is drifted by 100 mL for more than 30 seconds. | "}, - { 244 , ALARM_ID_DG_CLEANING_MODE_TEMP_SENSORS_DIFF_OUT_OF_RANGE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Temp Sensor Check | DG temperature sensors difference out of range | If any of the temperature sensors are out of the average of the sensors by more than 3 C. | "}, - { 245 , ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Heat Disinfect Temp TO | DG heat disinfect target temperature time out (could not reach to temperature) | If heating up to minimum temperature for heat disinfect failed after 4 hours. | "}, - { 246 , ALARM_ID_DG_CLEANING_MODE_COND_SENSORS_OUT_OF_RANGE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: RO Water Cond Error | DG chemical disinfect inlet conducitivty out of range | If CPo and CD2 conductivity values are above 100 uS/cm. | "}, - { 247 , ALARM_ID_AVAILABLE_7 , " | | | | | | "}, - { 248 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Chem Disinfect Proc | DG chemical disinfect out of target temperature | If TPo > 50 C after 1 second. | "}, - { 249 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE , "DG | Chemical Disinfectant Not Detected | Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Chem Disinf PAA Cond | DG chemical disinfect out of target conductivity | If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. | "}, - { 250 , ALARM_ID_DG_CHEM_DISINFECT_INSERT_ACID , "DG | Connect Chemical Disinfectant | Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press Resume/OK to continue. | Cleaning: Chem Disinf Connect PAA | Insert the acid to start chemical disinfect | Start of chemical disinfect | "}, - { 251 , ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Sys Rec NVRAM CRC | HD invalid system record | If the calculated CRC of the HD system record image does not match the expected CRC. | "}, - { 252 , ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Serv Rec NVRAM CRC | HD invalid service record | If the calculated CRC of the HD service record image does not match the expected CRC. | "}, - { 253 , ALARM_ID_DG_INVALID_SYSTEM_RECORD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Sys Rec NVRAM CRC | DG invalid system record | If the calculated CRC of the DG system record image does not match the expected CRC. | "}, - { 254 , ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Serv Rec NVRAM CRC | DG invalid service record | If the calculated CRC of the DG service record image does not match the expected CRC. | "}, - { 255 , ALARM_ID_HD_UI_COMPATIBILITY_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Compatibility | HD and UI software builds are not compatible | If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. | "}, - { 256 , ALARM_ID_HD_DISINFECT_CHEM_FLUSH , "HD | Chemical Flush in Progress | Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Chem Flush In Progress | DG flush started | If DG flush successfully started from the UI. | "}, - { 257 , ALARM_ID_HD_INVALID_INSTITUTIONAL_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Institutional Rec NVRAM CRC | HD invalid institutional record | If the calculated CRC of the HD institutional record image does not match the expected CRC. | "}, - { 258 , ALARM_ID_HD_UI_POST_FAILURE_OS_VERSION , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: OS Version | UI POST OS Version compatibility failure | If the UI POST reports the OS Version compatibility failure. | "}, - { 259 , ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Temp Sensors Error | HD temperatures out of range | If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. | "}, - { 260 , ALARM_ID_HD_UI_POST_FAILURE_SHASUM , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Image | UI POST Application Integrity (Sha256Sum) failure | If the UI POST reports a shasum failure. | "}, - { 261 , ALARM_ID_HD_UI_POST_FAILURE_CANBUS , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI CAN Driver | UI POST CANBus failure | If the UI POST reports a can bus failure. | "}, - { 262 , ALARM_ID_HD_UI_POST_FAILURE_DISPLAY , " | | | | UI POST Display failure | If the UI POST reports a display failure. | "}, - { 263 , ALARM_ID_HD_UI_POST_FAILURE_TOUCH , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Touch Driver | UI POST Touch failure | If the UI POST reports a touch failure. | "}, - { 264 , ALARM_ID_HD_UI_POST_FAILURE_SDCARD , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI SD- Card | UI POST SD-Card failure | If the UI POST reports SD card failure. | "}, - { 265 , ALARM_ID_HD_UI_POST_FAILURE_RTC , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI RTC | UI POST RTC failure | If the UI POST reports RTC failure. | "}, - { 266 , ALARM_ID_HD_UI_POST_FAILURE_WIFI , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI WiFi | UI POST WiFi failure | If the UI POST reports WiFi failure. | "}, - { 267 , ALARM_ID_HD_UI_POST_FAILURE_BLUETOOTH , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Bluetooth | UI POST Bluetooth failure | If the UI POST reports bluetooth failure. | "}, - { 268 , ALARM_ID_HD_UI_POST_FAILURE_ETHERNET , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Ethernet | UI POST Ethernet failure | If the UI POST reports ethernet failure. | "}, - { 269 , ALARM_ID_HD_UI_POST_FAILURE_SOUND , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Sound | UI POST Sound failure | If the UI POST reports sound failure. | "}, - { 270 , ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Proc Safestate | HD POST Safety Shutdown failure | If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. | "}, - { 271 , ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Proc Safestate | DG POST Safety Shutdown failure | If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. | "}, - { 272 , ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Fan Speed | HD Fan RPM out of range | If the fan RPM is out of Min/Max range. | "}, - { 273 , ALARM_ID_AVAILABLE_10 , " | | | | | | "}, - { 274 , ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Fault | DG idle reservoir weight out of range | If the reservoir weight > 200.0 of the target mL. | "}, - { 275 , ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Arterial Press Range | HD arterial pressure sensor is reading out of range | If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, - { 276 , ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Venous Press Range | HD venous pressure sensor is reading out of range | If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, - { 277 , ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Occlusion Sens Range | HD BP occlusion sensor is reading out of range | If the blood pump occlusion count is < 2000. | "}, - { 278 , ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "HD | Dialysate Reservoir Error | A problem was detected with the dialysate generation. \n- Treatment is paused and will be need to be resumed\n when dialysate is ready for use. | Dial Gen: Reservoir Overfill | HD active reservoir weight out of range. | If the active reservoir weight is > 1900.0 g. | "}, - { 279 , ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Drain | DG dialysate drain time out | If the DG dialysate drain exceeded 120 seconds. | "}, - { 280 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Direction Fault | DG drain pump invalid direction | If the drain pump's direction is not 1 when draining. | "}, - { 281 , ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME , "DG | Acid Concentrate Low | The acid concentrate is low.\n- Replace acid concentrate container. | Dial. Gen: Acid Conc. Low | DG Acid concentration bottle low volume alarm | If the acid concentration bottle volume > 3240 (3340 - 100). | "}, - { 282 , ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME , "DG | Bicarb Concentrate Low | Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. | Dial. Gen: Bicarb Conc. Low | DG Bicarbonate concentration bottle low volume alarm | If the bicarbonate concentration bottle volume > 3680 (3780 - 100). | "}, - { 283 , ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC Range | DG load cells weight out of range | If the weight of the DG load cells > 4500.0 g or < 0.0 g. | "}, - { 284 , ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Redundant LC Check | DG load cells primary/back up drift out of range | If redundant load cells deviate by more than 60g after tare for 5 seconds. | "}, - { 285 , ALARM_ID_HD_RO_PERMEATE_SAMPLE , "HD | RO Permeate Sample in Progress | RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: RO Permeate Sample In Progress | DG RO permeate sample started | If DG RO permeate sample successfully started from UI. | "}, - { 286 , ALARM_ID_DG_CONCENTRATE_CAP_NOT_IN_PROPER_POSITION , "DG | Concentrate Connection Cap Removed | The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. | Cleaning: Conc Cap Removed | DG cap in wrong state | If the concentrate cap is not in the right position in the cleaning modes. | "}, - { 287 , ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Proc- FPGA Clock | HD processor clock speed checks against FPGA clock failure | If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. | "}, - { 288 , ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE , "HD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC Drift Check | HD load cell inconsistency | If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. | "}, - { 289 , ALARM_ID_DG_DIALYSATE_CAP_NOT_IN_PROPER_POSITION , "DG | Dialysate Connection Cap Removed | The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. | Cleaning: Dialysate Cap Removed | DG cap in wrong state | If the dialysate cap is not in the right position in the cleaning modes. | "}, - { 290 , ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Ven. Pres. Com | HD venous sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 291 , ALARM_ID_HD_DISINFECT_FLUSH , "HD | Flush In Progress | System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Flush In Progress | DG flush started | If DG flush successfully started from the UI. | "}, - { 292 , ALARM_ID_HD_DISINFECT_HEAT , "HD | Heat Disinfection in Progress | Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Heat Disinf In Progress | DG heat disinfect started | If DG heat disinfect successfully started from the UI. | "}, - { 293 , ALARM_ID_HD_DISINFECT_CHEM , "HD | Chemical Disinfection in Progress | Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Chem Disinfect In Progress | DG chem disinfect started | If DG flush successfully started from the UI. | "}, - { 294 , ALARM_ID_HD_DISINFECT_HEAT_COOL , "HD | Active Cool in Progress | Active Cool is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Active Cool In Progress | DG active cool started | If DG active cool successfully started. | "}, - { 295 , ALARM_ID_HD_UI_POST_FAILURE_CLOUDSYNC , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI CloudSync | UI cloud sync POST failure | If the UI POST reports cloud sync failure. | "}, - { 296 , ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: TDi RTD Comm | DG TDi sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 297 , ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: TRo RTD Comm | DG TRo sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 298 , ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Pres Comm | DG barometric sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 299 , ALARM_ID_DG_INVALID_SERIAL_NUMBER , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Serial Number Check | DG invalid serial number | If the DG serial number is empty and does not have any characters. | "}, - { 300 , ALARM_ID_HD_INVALID_SERIAL_NUMBER , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Serial Number Check | HD invalid serial number | If the HD serial number is empty and does not have any characters. | "}, - { 301 , ALARM_ID_DG_RO_PERMEATE_SAMPLE_REMOVE_DIA_CAP , "DG | Remove Dialysate Cap | Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. | Cleaning: RO Permeate Sample Connect Cap | Open dialysate cap to be able to collecte the dispensing fluid | RO permeate sample collection state | "}, - { 302 , ALARM_ID_DG_CPU_RAM_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Processor RAM error | DG_CPU RAM error | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, - { 303 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Hall Commutation | DG drain pump direction FPGA fault | If the FPGA indicates an error for a certain period of time. | "}, - { 304 , ALARM_ID_HD_INVALID_USAGE_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Usage Rec NVRAM CRC | HD invalid usage reocord | If the CRC of the HD usage record does not match the calculated CRC. | "}, - { 305 , ALARM_ID_HD_CPU_RAM_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Processor RAM error | HD_CPU RAM error | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, - { 306 , ALARM_ID_AVAILABLE_50 , " | | | | | | "}, - { 307 , ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT , "HD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. | "}, - { 308 , ALARM_ID_DG_CPI_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CPi Comm | DG CPi conductivity sensor transmits invalid characters | If the CPi conductivity sensor transmits invalid characters for a certain period of time. | "}, - { 309 , ALARM_ID_DG_CPO_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CPo Comm | DG CPo conductivity sensor transmits invalid characters | If the CPo conductivity sensor transmits invalid characters for a certain period of time. | "}, - { 310 , ALARM_ID_DG_CD1_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CD1 Comm | DG CD1 conductivity sensor transmits invalid characters | If the CD1 conductivity sensor transmits invalid characters for a certain period of time. | "}, - { 311 , ALARM_ID_DG_CD2_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CD2 Comm | DG CD2 conductivity sensor transmits invalid characters | If the CD2 conductivity sensor transmits invalid characters for a certain period of time. | "}, - { 312 , ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE , "HD | Low Dialysate Flow | Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nResume/OK button will be available.\n- Press Resume/OK to continue. | Dial. Gen: Dialysate In Flow | The dialysate inlet flow is too low. | If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. | "}, - { 313 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_LOW , "DG | Disinfection: Water Temperature Low | Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Temp Low | DG cleaning mode inlet water temperature too low | If inlet water temperature is < 24 C. | "}, - { 314 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_HIGH , "DG | Disinfection: Water Pressure High | Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Press High | DG cleaning mode inlet water pressure too high | If inlet water temperature is > 80 psig. | "}, - { 315 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_LOW , "DG | Disinfection: Water Pressure Low | Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water pressure\n before initiating disinfection again. | Cleaning: Inlet Water Press Low | DG cleaning mode inlet water pressure too low | If inlet water pressure is < 14 psig. | "}, - { 316 , ALARM_ID_AVAILABLE_20 , " | | | | | | "}, - { 317 , ALARM_ID_AVAILABLE_21 , " | | | | | | "}, - { 318 , ALARM_ID_AVAILABLE_2 , " | | | | | | "}, - { 319 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_FLUSH_SAMPLE , "DG | Residual Chemical Sampling | A water sample is needed to confirm the \nabsence of residual chemical following disinfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. | Cleaning: Chem Residual Sampling | Collect sample to check the chemical disinfectant acid | Chemical disinfect flush sample alarm is prompted. | "}, - { 320 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_SAMPLE_TIMEOUT , "DG | Chemical Residual Timeout | A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. | Cleaning: Chem Residual Timeout | The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes | Chemical disinfect flush sample was not collected form more than 30 minutes. | "}, - { 321 , ALARM_ID_DG_OUTLET_UV_REACTOR_NOT_HEALTHY , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVo Error | Outlet reactor not healthy | If the outlet reactor are not healthy. | "}, - { 322 , ALARM_ID_HD_UI_POST_FAILURE_INVALID_YEAR , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST Alarm: UI Invalid Year | UI POST Invalid Year | If the UI POST reports a invalid year. | "}, - { 323 , ALARM_ID_HD_UI_POST_FAILURE_SETTINGS_BAD , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST Alarm: UI Bad Settings | UI POST Bad Settings | If the UI POST reports a bad settings. (e.g. no settings read, empty settings file, invalid/empty parms, CRC failure) | "}, + { 0 , ALARM_ID_NO_ALARM , "N/A | | | | Indication for when no alarms is a possible situation | N/A | "}, + { 1 , ALARM_ID_HD_STUCK_BUTTON_TEST_FAILED , "HD | Stuck Button | A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. | HD POST: Stuck Button | Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up | Stop or Off button are pressed for more than 1 sec during POST. | "}, + { 2 , ALARM_ID_HD_FPGA_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: FPGA Compatibility | HD FPGA POST failure | Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. | "}, + { 3 , ALARM_ID_DG_FPGA_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: FPGA Compatibility | DG FPGA POST failure | Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header | "}, + { 4 , ALARM_ID_HD_WATCHDOG_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: WD Safestate | HD Watchdog POST failure | If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA | "}, + { 5 , ALARM_ID_DG_WATCHDOG_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: WD Safestate | DG Watchdog POST failure | If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V | "}, + { 6 , ALARM_ID_HD_UI_COMM_POST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Proc- UI Comm | UI communication POST failure. UI failed to communicate within a reasonable time after power up | If the UI failed to checkin via the CAN bus within 40 seconds | "}, + { 7 , ALARM_ID_HD_RTC_BATTERY_LOW , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): RTC Bat Low | Real-time clock configuration error | If the RTC is not functional as indicated by the bits in the first 3 control registers. | "}, + { 8 , ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Acc Vector | HD accelerometer failed POST | If the calculated vector length of the accelerometer x, y, z is not the nominal length. | "}, + { 9 , ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Acc Vector | DG accelerometer failed POST | If the calculated vector length of the accelerometer x, y, z is not the nominal length. | "}, + { 10 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Config Emb Mode | HD blood leak sensor embedded mode communication failed | If the blood leak sensor embedded mode is unresponsive. | "}, + { 11 , ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP , "HD | Low Dialysate Temperature | Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. | Dial. Gen: Dialysate Temp Low | Dialysate temperature is below the target temperature | If the dialysate temperature is below the target temperature for a certain period of time. | "}, + { 12 , ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: HD Image | HD firmware image integrity POST test failed | If the calculated CRC of the HD image does not match the expected CRC. | "}, + { 13 , ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: DG Image | DG firmware image integrity POST test failed | If the calculated CRC of the DG firmware image does not matched the expected CRC | "}, + { 14 , ALARM_ID_DG_INVALID_USAGE_RECORD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Usage Rec NVRAM CRC | DG non-volatile usage information group invalid record CRC | If the DG usage information record's CRC does not match the calculated CRC in POST. | "}, + { 15 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP , "HD | High Dialysate Temperature | Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. | Dial. Gen: Dialysate Temp High | Dialysate temperature is out of high safety range | If the dialysate temperature is > 46 C for a certain period of time. | "}, + { 16 , ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Primary Alarm Current | HD alarm audio failed POST | If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output | "}, + { 17 , ALARM_ID_HD_UI_POST_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI POST Not Complete | HD UI POST failed | If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. | "}, + { 18 , ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_DG , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (DG ACK) | HD to DG comm error. | HD message to DG that requires acknowledge was not acknowledged within a certain time. | "}, + { 19 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP , "HD | High Dialysate Temperature | Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. | Dial. Gen: Dialysate Temp High | Dialysate temperature is above the target temperature | If the dialysate temperature is above the target temperature for a certain period of time. | "}, + { 20 , ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conductivity Cal | DG conductivity sensors invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, + { 21 , ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Drain Line Vol Cal | DG drain line volume invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, + { 22 , ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Reservoir Vol Cal | DG reservoirs invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, + { 23 , ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Acid Conc Cal | DG acid concentrate calibration record | If the non-volatile data is valid and if not raise the alarm | "}, + { 24 , ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Bicarb Conc Cal | DG bicarb concentrate calibration record | If the non-volatile data is valid and if not raise the alarm | "}, + { 25 , ALARM_ID_DG_ACCELEROMETERS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Acc Cal | DG accelerometer invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, + { 26 , ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Acc Cal | HD accelerometer invalid calibration record | If the accel calibration data that was received from NVDataMgmt is NOT legitimate. | "}, + { 27 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED , "HD | Self Test: Blood Leak | A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD BLD sensor failed zero sequence | BLD fails to complete zero sequence. | "}, + { 28 , ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Conductivity RTD Comm | DG two wire sensors FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, + { 29 , ALARM_ID_HD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: SP DAC Cal | HD heparin force sensor invalid calibration record | If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. | "}, + { 30 , ALARM_ID_HD_SOFTWARE_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: SW Invalid State | HD Software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, + { 31 , ALARM_ID_HD_BLOOD_PUMP_MC_CURRENT_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Current | Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running | If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. | "}, + { 32 , ALARM_ID_HD_BLOOD_PUMP_OFF_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Speed | Blood pump failed motor off check. Measured speed while commanded off | If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM | "}, + { 33 , ALARM_ID_HD_BLOOD_PUMP_MC_DIRECTION_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Direction | Blood pump failed motor direction check. Measured vs. commanded | If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. | "}, + { 34 , ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_CHECK , "HD | Blood Pump Error | A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. | HD Fault: BP Rotor | Blood pump failed rotor speed check. Mismatch with rotor and motor speeds | If the delta of the measured motor speed and the commanded motor speed is not within tolerance. | "}, + { 35 , ALARM_ID_HD_DIAL_IN_PUMP_MC_CURRENT_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPi Current | Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running | If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. | "}, + { 36 , ALARM_ID_HD_DIAL_IN_PUMP_OFF_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPi Speed | Dialysis inlet pump failed motor off check. Measured speed while commanded off | If during command off state, the DPi pump motor speed is measured > 100.0 RMP. | "}, + { 37 , ALARM_ID_HD_DIAL_IN_PUMP_MC_DIRECTION_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPi Direction | Dialysis inlet pump failed motor direction check. Measured vs. commanded | If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. | "}, + { 38 , ALARM_ID_HD_DIAL_IN_PUMP_ROTOR_SPEED_CHECK , "HD | Dialysate Pump Error | A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. | HD Fault: DPi Rotor | Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds | If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. | "}, + { 39 , ALARM_ID_HD_DIAL_OUT_PUMP_MC_CURRENT_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPo Current | Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running | If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. | "}, + { 40 , ALARM_ID_HD_DIAL_OUT_PUMP_OFF_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPo Speed | Dialysis outlet pump failed motor off check. Measured speed while commanded off | If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. | "}, + { 41 , ALARM_ID_HD_DIAL_OUT_PUMP_MC_DIRECTION_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPo Direction | Dialysis outlet pump failed motor direction check. Measured vs. commanded | If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. | "}, + { 42 , ALARM_ID_HD_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK , "HD | Dialysate Pump Error | A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. | HD Fault: DPo Rotor | Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds | If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed | "}, + { 43 , ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Iso ADC Comm | Main primary heater FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, + { 44 , ALARM_ID_HD_UI_COMM_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UI Comm | User interface communication timeout | If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. | "}, + { 45 , ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN CRC Com | Too many bad CRCs detected on received system messages | If too many bad messages CRCs are received within a set period of time | "}, + { 46 , ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_UI , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN ACK UI Com | System message that required acknowledgment was not acknowledged | If no retries left in the pending acknowlegement list. | "}, + { 47 , ALARM_ID_DG_MAX_RO_PUMP_PWM_EXCEEDED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Pump High | DG RO pump out of upper range | If the RO pump > 95% PWM for 12 secs. | "}, + { 48 , ALARM_ID_HD_UF_VOLUME_ACCURACY_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UF Rate Tare Err | Ultrafiltration volume accuracy error during treatment | If the UF volume (refUFvolume - measureVolume) > 100.0 mL. | "}, + { 49 , ALARM_ID_HD_FPGA_COMM_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: FPGA Com | HD FPGA communication down for too long | If there are more than 3 comm errors in a 1 min period. | "}, + { 50 , ALARM_ID_DG_VALVE_CONTROL_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Valve Control | DG FPGA not accepting commanded valve states | If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. | "}, + { 51 , ALARM_ID_HD_BLOOD_PUMP_MOTOR_SPEED_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Speed | Blood pump failed motor speed check. Measured vs. commanded | If measured motor speed versus commanded rotor speed exceeded limits. | "}, + { 52 , ALARM_ID_HD_DIAL_IN_PUMP_MOTOR_SPEED_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPi Speed | Dialysate inlet pump failed motor speed check. Measured vs. commanded | If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. | "}, + { 53 , ALARM_ID_HD_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: DPo Speed | Dialysate outlet pump failed motor speed check. Measured vs. commanded | If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. | "}, + { 54 , ALARM_ID_HD_CRITICAL_DATA_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Critical Data Err | HD critical data integrity check failed | If the critical data failed range check. | "}, + { 55 , ALARM_ID_DG_CRITICAL_DATA_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Critical Data Err | DG critical data integrity check failed | If the critical data failed range check. | "}, + { 56 , ALARM_ID_HD_ACCELEROMETER_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Acc Com | HD accelerometer error (no readings or FPGA reports error) | If no fresh sample of accelerometer data has been detected for more than 100 msec. | "}, + { 57 , ALARM_ID_DG_ACCELEROMETER_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Acc Com | DG accelerometer error (no readings or FPGA reports error) | If no fresh sample of accelerometer data has been detected for more than 100 msec. | "}, + { 58 , ALARM_ID_HD_VALVE_HOMING_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Home | HD valve homing failed | If the number of failed homing > = 3 counts. | "}, + { 59 , ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Transition Err | HD valve transition time out | if the valve's transition time has timed out (3000msec) | "}, + { 60 , ALARM_ID_HD_VALVE_NOT_FUNCTIONAL , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Control | HD valve not functional | If the valve bit masks are not as expected in different valve control mode. | "}, + { 61 , ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Current | HD valve current out of range | If the valve current is over the threshold of 1.0 A for 1000 msec. | "}, + { 62 , ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Position | HD valve position out of target | If the current position has deviated from the expected position for more than 1 second. | "}, + { 63 , ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Pre- Tx Hep. Pump Timeout. | HD syringe pump timeout alarm | If the syringe pump priming > 5 seconds | "}, + { 64 , ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Range | Barometric pressure is out of range | If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. | "}, + { 65 , ALARM_ID_HD_DG_COMMAND_INVALID_PARAMETER_FAULT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Reject) | HD requests DG command with invalid parameter fault | If the latest DG command response indicated invalid parameter | "}, + { 66 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Config Set Point | HD blood leak sensor set point set failure | If the set point of the blood leak sensor does not match the set point in the calibration record. | "}, + { 67 , ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD blood pump occlusion self-test failure alarm | If the blood pump pressure > 5000 | "}, + { 68 , ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE , "HD | Dialysate Generation Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. | Dial. Gen: Dialysate not ready | HD active reservoir recirculation out of range | If the recirculation level has exceeded the max allowed | "}, + { 69 , ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Cal | HD blood leak sensor invalid calibration record | If the BLD data that was received from NVDataMgmt is NOT legitimate. | "}, + { 70 , ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD arterial pressure self-test failure alarm | If arterial pressure < -300 or > 100 mmHg | "}, + { 71 , ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD venous pressure self-test failure alarm | If the venous pressure < -100 or > 600 mmHg | "}, + { 72 , ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (LC) | HD No load cell data message receive at least once every 2 seconds | No new load cell data for 2 seconds. | "}, + { 73 , ALARM_ID_HD_NEW_DIALYSATE_TEMP_DATA_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (TDI/TRO) | HD No dialysate temperature data message received at least once every 2 seconds | No new dialysate temperature data for 2 seconds. | "}, + { 74 , ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVi On No Flow | DG inlet UV reactor is on while there is no flow | If the inlet UV reactor is on while VPi is closed for a certain period of time. | "}, + { 75 , ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe DAC | HD syringe pump self-test failure alarm | If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. | "}, + { 76 , ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Voltage Monitor | HD monitored voltage is out of range | If the HD monitored voltages are out of range for a certain period of time. | "}, + { 77 , ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Voltage Monitor | DG monitored voltage is out of range | If the DG monitored voltages are out of range for a certain period of time. | "}, + { 78 , ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Enc Dir | HD syringe pump direction (from encoder) error | If the measured encoder direction of the syringe pump does not agree with the expected direction | "}, + { 79 , ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Drv Dir | HD syringe pump direction (from controller) error | If the measured direction of the syringe pump motor controller does not agree with the expected direction | "}, + { 80 , ALARM_ID_HD_SYRINGE_PUMP_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Driver | HD syringe pump fault reported by FPGA | If the FPGA reported the syringe pump fault | "}, + { 81 , ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump OT | HD syringe pump over travel error | If the syringe pump position is deemed to have overtraveled | "}, + { 82 , ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump DAC Err | HD syringe pump DAC write failure | If the syringe pump DAC write error is reported by the FPGA | "}, + { 83 , ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump On Err | HD syringe pump is running while the BP is off | If the syringe pump is on while the blood pump is off | "}, + { 84 , ALARM_ID_DG_SET_RTC_YEAR_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: DG RTC Time | DG RTC set year is invalid | If DG RTC set year is < 2021. | "}, + { 85 , ALARM_ID_HD_PUMP_TRACK_LATCH_OPENED , "HD | Cartridge Lever is Open | A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press Resume/OK button to continue. | Blood Stop: Cartridge Lever Open | HD pump track lever is not fully latched | If the pump track lever is detected as unlatched when it should be. | "}, + { 86 , ALARM_ID_HD_SET_RTC_YEAR_INVALID , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: HD RTC Time | HD RTC set year is invalid | If HD RTC set year is < 2021. | "}, + { 87 , ALARM_ID_DG_HEATING_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Heat Cal | DG heating parameters invalid calibration record | If the heating paramters' non-volatile record's CRC does not match the calculated CRC. | "}, + { 88 , ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CP1 CP2 Speed Range | DG concentrate pump hall sensor out of range. | Hall sensor pulse to pulse timing is < 1ms (speed too fast). | "}, + { 89 , ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVo On No Flow | DG outlet UV reactor is on while there is no flow | If the outlet UV reactor is on while the RO flow is less than 200 mL/min. | "}, + { 90 , ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC A1/B1 Comm | DG load cells A1/B1 FPGA fault | If the FPGA does not report fresh data for a certain period of time. | "}, + { 91 , ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (FMD) | HD No dialysate flow data received in the last 3 seconds | If not receiving new dialysate flow readings in expected interval (3 seconds) | "}, + { 92 , ALARM_ID_DG_TEMPERATURE_SENSORS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Temp Snsr Cal | DG temperature sensors invalid calibration | If the calibration date of the temperature sensors is 0. | "}, + { 93 , ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Permeate Cond High | DG outlet primary conductivity sensor out of range | If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. | "}, + { 94 , ALARM_ID_DG_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Pressure Sensor Range | DG pressure sensor out of range | If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. | "}, + { 95 , ALARM_ID_DG_WATCHDOG_EXPIRED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: WD Error | DG watchdog expired | If the watchdog signal is low while not in POST. | "}, + { 96 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE , "DG | Incoming Water Temperature High | Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Temp High | DG inlet water temperature out of range | If the inlet water temperature is > 37 C. | "}, + { 97 , ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Out of Range | Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Dialysate Cond. | DG fill conductivity out of range | If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. | "}, + { 98 , ALARM_ID_HD_BATTERY_COMM_FAULT , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Battery Com | HD battery or battery charger communication fault | If battery I2C communication fault persistence count > 5. | "}, + { 99 , ALARM_ID_HD_SYRINGE_PUMP_STALL , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Stall | HD syringe pump stall alarm | If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. | "}, + { 100 , ALARM_ID_DG_COND_SENSORS_INVALID_TEMP_COMP_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conductivity Cal | DG conductivity sensors temperatrue compensation invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, + { 101 , ALARM_ID_UI_POST_HD_COMM , "UI | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI- Proc Comm | UI POST FW Communication failure | If the HD is not communicating with the UI during POST. | "}, + { 102 , ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Tpo THD Temp Delta | DG heat disinfection temperature gradient out of range | If temperature gradient between Tpo and Thd exceeds maximum 15°C. | "}, + { 103 , ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Cal NVRAM CRC | HD non-volatile calibration group invalid record CRC | If the computed CRC of the calibration group record is not equal to the record CRC. | "}, + { 104 , ALARM_ID_HD_AIR_TRAP_ILLEGAL_LEVELS , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Air Trap Sensor | Air trap level sensors reporting illegal combination of air/fluid | If the air trap level readings are not valid for more than 2 seconds. | "}, + { 105 , ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Cal NVRAM CRC | DG non-volatile system group invalid record CRC | If the DG NVDATAMGMT CAL group calculated CRC does not matched the expected CRC. | "}, + { 106 , ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Flow Sensor Cal | DG flow sensors calibration error | If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. | "}, + { 107 , ALARM_ID_HD_DG_RESTARTED_FAULT , "HD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DG Restarted | HD reports DG restarted fault | If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. | "}, + { 108 , ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump ADC Com | HD syringe pump ADC error | If the ADC read is not fresh after POST init and the ADC read counter has not changed. | "}, + { 109 , ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Vol | HD syringe pump volume check error | If the volume delivered is off by more than 5% or 0.1mL. | "}, + { 110 , ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Speed | HD syringe pump speed check error | If the syringe measured rate is off by more than 5% or 0.1mL. | "}, + { 111 , ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Speed | HD syringe pump not stopped in off state error | If the syringe pump position is not equal to its last position. | "}, + { 112 , ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC A2/B2 Comm | DG load cells A2/B2 FPGA fault | If the FPGA does not report fresh data for a certain period of time. | "}, + { 113 , ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Current | DG drain pump current out of range | If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. | "}, + { 114 , ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE , "HD | Self Test: Venous Air Detector | A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD venous air bubble detector self-test failure | If the venous air bubble self test did not detect bubble for more than 500 msec. | "}, + { 115 , ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Temp Sensor Range | DG temperature sensor out of range | If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. | "}, + { 116 , ALARM_ID_HD_UI_SDCARD_FAILURE , "UI HD | Service Required: Dialysate Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): SD- Card Error | SD card is non-operational | SD card is non-operational | "}, + { 117 , ALARM_ID_DG_CAN_MESSAGE_NOT_ACKED , "DG | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of HD Comm (HD ACK) | System message that required acknowledgment was not acknowledged | If no retries left in the pending acknowlegement list. | "}, + { 118 , ALARM_ID_DG_RTC_CONFIG_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RTC Config | DG RTC configuration error | If the RTC configuration registers report an error. | "}, + { 119 , ALARM_ID_DG_RTC_BATTERY_LOW , "DG | Service Required: Dialysate Device | A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. | DG Alarm (Low): RTC Bat Low | DG RTC battery low | If the RTC battery bit report a low alarm. | "}, + { 120 , ALARM_ID_HD_PRE_TREATMENT_WET_PRIME_TEST_FAILURE , "HD | Self Test: Venous Air Detector | A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD pre-treatment wet prime test failure | If the HD pre-treatment wet self test failed. | "}, + { 121 , ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Prim Heater Lg Control | DG main primary heater voltage out of range | If the main primary heater's voltage is out of range. | "}, + { 122 , ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Prim Heater Sm Control | DG small primary heater voltage out of range | If the small primary heater's voltage is out of range. | "}, + { 123 , ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Trimmer Heater Control | DG trimmer heater voltage out of range | If the trimmer voltage is out of range. | "}, + { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, + { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, + { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. | Blood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, + { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access\n issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, + { 129 , ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "HD | Blood Leak: Bypass | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. | Blood Stop: Dialyzer Blood Leak Bypass | HD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "}, + { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 131 , ALARM_ID_HD_ARTERIAL_PRESSURE_LOW , "HD | Low Arterial Pressure | Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. Low | Arterial pressure too low during treatment | If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 132 , ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH , "HD | High Arterial Pressure | Arterial pressure is high.\n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. High | Arterial pressure too high during treatment | If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 133 , ALARM_ID_DG_FLUID_LEAK_DETECTED , "DG | Dialysate Device Fluid Leak | Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Fluid Leak in DG | DG fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, + { 134 , ALARM_ID_HD_FLUID_LEAK_DETECTED , "HD | Fluid Leak | Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. | Blood Stop: Fluid Leak in HD | HD fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, + { 135 , ALARM_ID_HD_SHOCK , "HD | Blood Pump Stopped: Excessive Movement | Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Blood Stop: HD Shock Detected | HD experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, + { 136 , ALARM_ID_DG_SHOCK , "DG | Dialysate Device: Excessive Movement | Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Dial. Gen: Shock- Vibration | DG experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, + { 137 , ALARM_ID_HD_EXCESSIVE_TILT , "HD | Blood Pump Stopped: Excessive Tilt | Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. | Blood Stop: HD Tilted | HD tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 3 deg for more than 1 sec. | "}, + { 138 , ALARM_ID_DG_EXCESSIVE_TILT , "DG | Dialysate Device: Excessive Tilt | Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. | Dial. Gen: Tilt | DG tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 3 deg for more than 1 sec. | "}, + { 139 , ALARM_ID_HD_AC_POWER_LOST , "HD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. | "}, + { 140 , ALARM_ID_HD_DG_COMM_TIMEOUT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Timeout) | Dialysate generator communication timeout | If DG communication timeout is detected for more than 2 seconds. | "}, + { 141 , ALARM_ID_HD_AIR_TRAP_FILL_DURING_TREATMENT , "HD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Fill | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "}, + { 142 , ALARM_ID_HD_OCCLUSION_BLOOD_PUMP , "HD | Blood Pump Stopped: Occlusion | Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Blood Pump Occlusion | Blood pump occlusion detected | If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall | "}, + { 143 , ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: TRo TDi Delta | DG dialysate temperature sensors drift timeout. | If TDi and TRo have drifted for more 1 C for a certain period of time. | "}, + { 144 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_HIGH , "DG | Disinfection: Water Conductivity High | Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Cond High | DG cleaning mode inlet water conductivity out of high range | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode | "}, + { 145 , ALARM_ID_DG_ACID_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Error During Prime | Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Acid Conc. Cond. | DG Concentrate conductivity after adding acid out of range alarm | If the conductivity value for the acid concentrate is out of range for more than 45 secs. | "}, + { 146 , ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: DG RTC Timer | DG RTC vs. Timer accuracy error | If RTC and timer are not aligned on 1 second elapsed time. | "}, + { 147 , ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT , "DG | Dialysate Generation Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. | Dial. Gen: Dialysate not ready | DG dialysate generation conductivity fault alarm | If dialysate conductivity is not with 5% of the TCD | "}, + { 148 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_HIGH , "DG | Disinfection: Water Temperature High | Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Temp High | Cleaning Mode Inlet water temperature too high | Inlet temperature > 37 C. | "}, + { 149 , ALARM_ID_HD_WATCHDOG_EXPIRED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: WD Error | HD watchdog expired | If the watchdog signal is low while not in POST. | "}, + { 150 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE , "DG | Incoming Water Conductivity High | Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Cond High | Inlet water conductivity too high | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). | "}, + { 151 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE , "DG | Incoming Water Conductivity Low | Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Cond Low | Inlet water conductivity too low | Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). | "}, + { 152 , ALARM_ID_DG_INLET_WATER_PRESSURE_IN_LOW_RANGE , "DG | Incoming Water Pressure Low | Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Pres Low | Inlet water pressure too low | Inlet water pressure < 25 PSI. | "}, + { 153 , ALARM_ID_HD_PRIME_COMPLETED_HIGH , "HD | Pre-Treatment Process Terminated | The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press Resume/OK for continue with next steps to remove \ncartridge from the device. | Process: Pre- Tx Terminated | HD prime completed high priority alarm | In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) | "}, + { 154 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_LOW , "DG | Disinfection: Water Conductivity Low | Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Cond Low | Cleaning mode inlet water conductivity too low | Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode | "}, + { 155 , ALARM_ID_HD_NEW_RESERVOIRS_DATA_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Reservoir) | HD No reservoir data message receive at least once every 2 seconds | No new reservoir data for 2 seconds. | "}, + { 156 , ALARM_ID_HD_DG_NEW_OPERATION_MODE_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Op Mode) | HD No DG op mode message receive at least once every 2 seconds | No DG op mode for 2 seconds. | "}, + { 157 , ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT , "DG | Chemical Disinfectant Not Detected | Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Chem Disinf PAA Prime | Chemical disinfect priming the acid line timeout | If the disinfectant acid line is not primed within a certain period of time. | "}, + { 158 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_LOW_RANGE , "DG | Incoming Water Temperature Low | Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Temp Low | Inlet water temperature low | Inlet temperature < 24 °C for 5 seconds. | "}, + { 159 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_REMOVE_ACID , "DG | Remove Chemical Disinfectant | The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. | Cleaning: Chem Disinf Connect Cap | Close concentrate line be able to flush it | Start of chemical disinfect flush. | "}, + { 160 , ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Rotor Max Speed | Blood pump rotor speed too high | If blood pump motor speed exceeded 100.0 RPM. | "}, + { 161 , ALARM_ID_DG_COMM_TOO_MANY_BAD_CRCS , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: HD- DG Comm | Too many bad CRCs detected on received system messages | If >= 5 bad CRCs detected on received CAN messages within 10 minutes. | "}, + { 162 , ALARM_ID_DG_FPGA_CLOCK_SPEED_CHECK_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Proc- FPGA Clock | FPGA clock speed failure | If the FPGA clock speed is relatively close to the processor clock for a certain period of time. | "}, + { 163 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM , "HD | Ultrafiltration Error | A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. | Dial. Gen: UF Rate Check | HD sees primary load cell for reservoir 1 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, + { 164 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM , "HD | Ultrafiltration Error | A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. | Dial. Gen: UF Rate Check | HD sees primary load cell for reservoir 2 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, + { 165 , ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_ALARM , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | HD in treatment rinseback operation timeout | If the maximum time inrinseback operation has been exceeded. | "}, + { 166 , ALARM_ID_TEST_ALARM_NO_BLOOD_RECIRC_CLR_TOP_ONLY , " | Test Alarm - Alarm 166 | | | | | "}, + { 167 , ALARM_ID_HD_CARTRIDGE_DOOR_OPENED , "HD | Blood Pump Stopped: Door Open | The device door is open.\n Close the device door. \n- Press Resume/OK to continue. | Blood Stop: Door Open | HD cartridge door opened alarm | If the cartridge door is opened as reported by the FPGA. | "}, + { 168 , ALARM_ID_TEST_ALARM_NO_DIAL_RECIRC_CLR_TOP_ONLY , " | Test Alarm - Alarm 168 | | | | | "}, + { 169 , ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Dialysate Flow Sens Rg | Dialysate flow rate out of range | If the dialysate flow rate > 2.00 L/min. | "}, + { 170 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY , "HD | Heparin Syringe Empty | Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. | Heparin Syringe Empty | HD syringe empty alarm | If the syringe pump position is near empty position. | "}, + { 171 , ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION , "HD | Heparin Pump Occlusion | A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. | Process: Pre- Tx Hep. Pump Occ. | HD syringe pump occlusion alarm | If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. | "}, + { 172 , ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM , "HD | Not Enough Heparin | There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. | Process: Pre- Tx Not Enough Hep. | HD syringe pump not enough Heparin alarm | If the estimated syringe volume needed for treatment is not sufficient | "}, + { 173 , ALARM_ID_HD_RTC_CONFIG_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: RTC Config | HD RTC configuration error | If the RTC configuration registers report an error. | "}, + { 174 , ALARM_ID_HD_RTC_OR_TIMER_ACCURACY_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: HD RTC Timer | HD RTC vs. Timer accuracy error | If RTC and timer are not aligned on 1 second elapsed time. | "}, + { 175 , ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Pump Commutation Err | HD pump commutation error | If the set pump direction does not agree with the hall sensor direction. | "}, + { 176 , ALARM_ID_TEST_ALARM_NO_RECIRC_CLR_TOP_ONLY , " | Test Alarm - Alarm 176 | | | | | "}, + { 177 , ALARM_ID_DG_SOFTWARE_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Software Invalid State | DG software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, + { 178 , ALARM_ID_HD_COMM_TIMEOUT , "UI | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UI- Proc Comm | HD communication timeout | If the HD stops communicating with the UI. | "}, + { 179 , ALARM_ID_DG_FPGA_COMM_TIMEOUT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: FPGA Com | DG FPGA communication down for too long | If retries for commands exceeds limit or FPGA reports comm error. | "}, + { 180 , ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Flow Sensor Range | RO flow rate out of range | If the RO flow rate > 2.00 L/min. | "}, + { 181 , ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Tare Out of Range | DG load cells weight out of range for tare | If the load cell weight is out of range for tare. | "}, + { 182 , ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: LC Cal | DG load cells invalid calibration | If the CRC of load cells calibration data does not match the calculated CRC. | "}, + { 183 , ALARM_ID_DG_INVALID_LOAD_CELL_VALUE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Invalid Resv Transfer | DG invalid load cell value | If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. | "}, + { 184 , ALARM_ID_DG_INLET_UV_REACTOR_NOT_HEALTHY , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVi Error | Inlet reactor not healthy | If the inlet reactor are not healthy. | "}, + { 185 , ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. | DG Alarm (Low): Fan Speed | DG fan RPM out of range | If the fan RPM is out of Min/Max range. | "}, + { 186 , ALARM_ID_DG_CONCENTRATE_PUMP_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CP1/CP2 Fault | DG concentrate pump fault | If the concentrate pumps fault is reported by the FPGA. | "}, + { 187 , ALARM_ID_DG_CP1_SPEED_CONTROL_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CP1 Speed Fault | DG concentrate pump CP1 speed control error | If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. | "}, + { 188 , ALARM_ID_DG_CP2_SPEED_CONTROL_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CP2 Speed Fault | DG concentrate pump CP2 speed control error | If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. | "}, + { 189 , ALARM_ID_DG_DRAIN_PUMP_RPM_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Speed Fault | DG drain pump RPM out of range | If the drain pump RPM > 100.0 of the target RPM. | "}, + { 190 , ALARM_ID_DG_DRAIN_PUMP_OFF_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Speed Not Off | DG drain pump off fault | If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. | "}, + { 191 , ALARM_ID_DG_FLOW_RATE_OUT_OF_UPPER_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Flow High | DG flow rate out of upper range | If the RO flow rate > 10% from the target flow rate. | "}, + { 192 , ALARM_ID_DG_FLOW_RATE_OUT_OF_LOWER_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Flow Low | DG flow rate out of lower range | If the RO flow rate < 10% from the target flow rate. | "}, + { 193 , ALARM_ID_HD_PARTIAL_OCCLUSION_BLOOD_PUMP , "HD | Blood Pump Stopped: Partial Occlusion | Partial occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment. | Blood Stop: Blood Pump Partial Occlusion | Blood pump partial occlusion detected | If blood occlusion is detected > PARTIAL_OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionDuringTreatement | "}, + { 194 , ALARM_ID_DG_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Driver Feedback | DG RO pump duty cycle out of range | If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. | "}, + { 195 , ALARM_ID_DG_RO_PUMP_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Pump High Pressure | DG RO pump pressure out of range | If the DG RO pump pressure > 135 psi for a certain period of time. | "}, + { 196 , ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CPi/CPo Comm | DG CPi/CPo conductivity sensors FPGA fault | If the FPGA reports error for a certain period of time. | "}, + { 197 , ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CD1/CD2 Comm | DG CD1/CD2 conductivity sensors FPGA fault | If the FPGA reports error for a certain period of time. | "}, + { 198 , ALARM_ID_DG_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: HP On With No Flow | DG primary heater while RO flow is low | If the RO flow is < 0.2 L/min for a certain period of time. | "}, + { 199 , ALARM_ID_DG_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: HR On With No Flow | DG trimmer heater while dialysate flow is low | If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. | "}, + { 200 , ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Thermistor Range | DG thermistor or temperature sensor out of range | If thermistor > 80.0 C or < 0.0 C. | "}, + { 201 , ALARM_ID_HD_PRE_TREATMENT_WET_FLOW_TEST_FAILURE , "HD | Self Test: Dialysate Flow Rate | A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD pre-treatment wet flow test failure | If the pre-treatment wet flow test has failed. | "}, + { 202 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE , "HD | Self Test: Pressure | A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. | Process: Pre- Tx Dry Self- Test | HD pre-treatment dry pressure normal test failure | If the pre-treatment dry pressure normal test failed. | "}, + { 203 , ALARM_ID_DG_FILL_CONDUCTIVITIES_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Fill Cond Cal | DG fill acid/bicarb invalid calibration record | If the fill target acid/bicarb conductivities' non-volatile record's calibration time is not 0. | "}, + { 204 , ALARM_ID_DG_RO_REJECTION_RATIO_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Rejection | RO rejection ratio out of range | If RO rejection ratio > 0.1 | "}, + { 205 , ALARM_ID_DG_CONDUCTIVITY_SENSOR_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Cond Sensors Error | Conductivity sensor fault | If conductivity sensor reports 0x11 for more than 2 consecutive reads. | "}, + { 206 , ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Fill Timeout | DG dialysate fill runs out of time | If DG dialysate fill > 5 minutes. | "}, + { 207 , ALARM_ID_DG_FLOW_METER_CHECK_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: FMP Flow Rate Error | DG flow meter check failure alarm | If integrated Volume To Load Cell Reading Percent > 0.1. | "}, + { 208 , ALARM_ID_HD_VENOUS_LINE_OCCLUSION , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Venous Pres. High | HD venous line occlusion fault | If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. | "}, + { 209 , ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Flush Circ Error | DG drain circulation line timeout | If the integrated volume < 0.1 for more than 60 secs. | "}, + { 210 , ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Battery Fault | HD battery pack detected an error | If the battery pack status error is detected from the slave device I2C address 0x0B. | "}, + { 211 , ALARM_ID_HD_BLOOD_SITTING_WARNING , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. | "}, + { 212 , ALARM_ID_HD_END_OF_TREATMENT_ALARM , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment end paused state, if the blood is in the line for more than 4 minutes. | "}, + { 213 , ALARM_ID_HD_PRIME_COMPLETED_MEDIUM , "HD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre- Tx Prime Complete | HD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) | "}, + { 214 , ALARM_ID_AVAILABLE_30 , " | | | | | | "}, + { 215 , ALARM_ID_AVAILABLE_31 , " | | | | | | "}, + { 216 , ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM , "HD | Treatment Complete | Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. | Process: End of Dialysis Tx | HD end treatment sub-mode timeout alarm | If the treatment ended successfully. | "}, + { 217 , ALARM_ID_AVAILABLE_32 , " | | | | | | "}, + { 218 , ALARM_ID_HD_SYRINGE_DETECTED , "HD | Syringe Detected | Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. | Process: Remove Syringe | HD syringe detected alarm | If the syringe detection switch > 2.0 volts. | "}, + { 219 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED , "HD | Blood Stopped: Syringe Not Detected | There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. | Blood Stop: Syringe Not Detected | HD syringe pump syringe removed alarm | If the syringe detection switch < 2.0 volts. | "}, + { 220 , ALARM_ID_DG_THD_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: THD RTD Comm | DG THd sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, + { 221 , ALARM_ID_HD_EMPTY_SALINE_BAG , "HD | Blood Pump Stopped: Saline Bag Empty | The saline bag appears empty.\n- Check the saline bag and replace if empty. | Blood Stop: Saline Bag Empty | HD Saline bag is empty | If the measured arterial pressure < -300.0 mmHg for more than 250 msec. | "}, + { 222 , ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: BP Occ Com | HD occlusion sensor FPGA fault | If the FPGA does not report fresh data or indicates an error for a certain period of time. | "}, + { 223 , ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Art. Pres. Com | HD arterial sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, + { 224 , ALARM_ID_HD_TREATMENT_STOPPED_BY_USER , "HD | Blood Pump Stopped: Treatment Paused | Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. | Blood Stopped: Treatment Pause | Treatment stopped by user action - pressed stop button | If the user has pressed the stop button. | "}, + { 225 , ALARM_ID_HD_END_OF_TREATMENT_WARNING , "HD | Treatment Complete | Dialysis treatment is complete.\n- If blood sampling is required, draw blood at\n the arterial sample port\n- Press RINSEBACK to end treatment and rinseback. | Process: End of Treatment | HD end of treatment warning | If the treatment duration has exceeded the prescribed treatment time entered by the user. | "}, + { 226 , ALARM_ID_HD_PRIME_COMPLETED_LOW_PRIORITY , "HD | Priming Complete | Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre- Tx Priming Complete | HD prime completed low priority alarm | Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection | "}, + { 227 , ALARM_ID_AVAILABLE_15 , " | | | | | | "}, + { 228 , ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. | Process: Pre- Tx Prime Timeout | HD time out on prime sub-mode purge air state. | During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. | "}, + { 229 , ALARM_ID_HD_PRIME_DIALYSATE_DIALYZER_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. | Process: Pre- Tx Prime Timeout | HD prime dialysate dialyzer time out alarm | If the priming of the dialyzer dialysate flow path has exceeded the time limit. | "}, + { 230 , ALARM_ID_HD_PRIME_DIALYSATE_BYPASS_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. | Process: Pre- Tx Prime Timeout | HD prime dialysate bypass time out alarm | If the priming of the dialyzer bypass flow path has exceeded the time limit. | "}, + { 231 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "HD | Self Test: Cartridge Check | A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. | Process: Pre- Tx Dry Self- Test | HD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, + { 232 , ALARM_ID_HD_PRE_TREATMENT_WET_LC_TEST_FAILURE , "HD | Self Test: Dialysate Flow Rate | A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. | Process: Pre- Tx Wet Self- Test | HD pre-treatment mode wet self-tests failure | If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState | "}, + { 233 , ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Cal CRC | DG barometric pressure sensor bad coefficients CRC | If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. | "}, + { 234 , ALARM_ID_DG_INLET_WATER_PRESSURE_IN_HIGH_RANGE , "DG | Incoming Water Pressure High | Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Pres High | DG inlet water pressure is too high | If the inlet water pressure is > 80 psig. | "}, + { 235 , ALARM_ID_HD_TREATMENT_STOPPED_AFTER_RINSEBACK , "HD | Rinseback Complete | Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press Resume/OK to continue. | Process: Rinseback Complete | HD in treatment stopped sub-mode after rinseback completed (no escalation) | If there is no active alarm and rinseback has completed. | "}, + { 236 , ALARM_ID_HD_INSTALL_NEW_CARTRIDGE , "HD | Self Test: Fluid Present | A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume. | Process: Pre- Tx Dry Self- Test | HD needs new cartridge to be installed | If the air trap sensor reported AIR and bubble sensor detected bubbles. | "}, + { 237 , ALARM_ID_AVAILABLE_13 , " | | | | | | "}, + { 238 , ALARM_ID_HD_NO_CARTRIDGE_LOADED , "HD | Cartridge Not Detected | A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed. | Process: Pre- Tx Dry Self- Test | HD no cartridge loaded or installed improperly alarm | If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. | "}, + { 239 , ALARM_ID_HD_CARTRIDGE_REMOVAL_FAILURE , "HD | Cartridge Still Detected | A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed. | Process: Post- Tx Remove Cartridge | HD fail to remove cartridge alarm | In the post treatment state, if the cartridge has not been removed. | "}, + { 240 , ALARM_ID_DG_BICARB_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Error During Prime | Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Bicarb Conc. Cond. | Bicarb conductivity out of range during bicarb pump check alarm | If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. | "}, + { 241 , ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Resv Drain Timeout | DG reservoir drain time out | If the DG reservoir drain is not complete in 120 secs. | "}, + { 242 , ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Resv Fill Timeout | DG reservoir fill time out | If the DG reservoir fill is not complete in 120 secs. | "}, + { 243 , ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Fluid Loss Detected | DG reservoir leak time out | If the reservoir level is drifted by 100 mL for more than 30 seconds. | "}, + { 244 , ALARM_ID_DG_CLEANING_MODE_TEMP_SENSORS_DIFF_OUT_OF_RANGE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Temp Sensor Check | DG temperature sensors difference out of range | If any of the temperature sensors are out of the average of the sensors by more than 3 C. | "}, + { 245 , ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Heat Disinfect Temp TO | DG heat disinfect target temperature time out (could not reach to temperature) | If heating up to minimum temperature for heat disinfect failed after 4 hours. | "}, + { 246 , ALARM_ID_DG_CLEANING_MODE_COND_SENSORS_OUT_OF_RANGE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: RO Water Cond Error | DG chemical disinfect inlet conducitivty out of range | If CPo and CD2 conductivity values are above 100 uS/cm. | "}, + { 247 , ALARM_ID_AVAILABLE_7 , " | | | | | | "}, + { 248 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Chem Disinfect Proc | DG chemical disinfect out of target temperature | If TPo > 50 C after 1 second. | "}, + { 249 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE , "DG | Chemical Disinfectant Not Detected | Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Chem Disinf PAA Cond | DG chemical disinfect out of target conductivity | If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. | "}, + { 250 , ALARM_ID_DG_CHEM_DISINFECT_INSERT_ACID , "DG | Connect Chemical Disinfectant | Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press Resume/OK to continue. | Cleaning: Chem Disinf Connect PAA | Insert the acid to start chemical disinfect | Start of chemical disinfect | "}, + { 251 , ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Sys Rec NVRAM CRC | HD invalid system record | If the calculated CRC of the HD system record image does not match the expected CRC. | "}, + { 252 , ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Serv Rec NVRAM CRC | HD invalid service record | If the calculated CRC of the HD service record image does not match the expected CRC. | "}, + { 253 , ALARM_ID_DG_INVALID_SYSTEM_RECORD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Sys Rec NVRAM CRC | DG invalid system record | If the calculated CRC of the DG system record image does not match the expected CRC. | "}, + { 254 , ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Serv Rec NVRAM CRC | DG invalid service record | If the calculated CRC of the DG service record image does not match the expected CRC. | "}, + { 255 , ALARM_ID_HD_UI_COMPATIBILITY_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Compatibility | HD and UI software builds are not compatible | If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. | "}, + { 256 , ALARM_ID_HD_DISINFECT_CHEM_FLUSH , "HD | Chemical Flush in Progress | Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Chem Flush In Progress | DG flush started | If DG flush successfully started from the UI. | "}, + { 257 , ALARM_ID_HD_INVALID_INSTITUTIONAL_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Institutional Rec NVRAM CRC | HD invalid institutional record | If the calculated CRC of the HD institutional record image does not match the expected CRC. | "}, + { 258 , ALARM_ID_HD_UI_POST_FAILURE_OS_VERSION , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: OS Version | UI POST OS Version compatibility failure | If the UI POST reports the OS Version compatibility failure. | "}, + { 259 , ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Temp Sensors Error | HD temperatures out of range | If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. | "}, + { 260 , ALARM_ID_HD_UI_POST_FAILURE_SHASUM , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Image | UI POST Application Integrity (Sha256Sum) failure | If the UI POST reports a shasum failure. | "}, + { 261 , ALARM_ID_HD_UI_POST_FAILURE_CANBUS , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI CAN Driver | UI POST CANBus failure | If the UI POST reports a can bus failure. | "}, + { 262 , ALARM_ID_HD_UI_POST_FAILURE_DISPLAY , " | | | | UI POST Display failure | If the UI POST reports a display failure. | "}, + { 263 , ALARM_ID_HD_UI_POST_FAILURE_TOUCH , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Touch Driver | UI POST Touch failure | If the UI POST reports a touch failure. | "}, + { 264 , ALARM_ID_HD_UI_POST_FAILURE_SDCARD , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI SD- Card | UI POST SD-Card failure | If the UI POST reports SD card failure. | "}, + { 265 , ALARM_ID_HD_UI_POST_FAILURE_RTC , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI RTC | UI POST RTC failure | If the UI POST reports RTC failure. | "}, + { 266 , ALARM_ID_HD_UI_POST_FAILURE_WIFI , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI WiFi | UI POST WiFi failure | If the UI POST reports WiFi failure. | "}, + { 267 , ALARM_ID_HD_UI_POST_FAILURE_BLUETOOTH , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Bluetooth | UI POST Bluetooth failure | If the UI POST reports bluetooth failure. | "}, + { 268 , ALARM_ID_HD_UI_POST_FAILURE_ETHERNET , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Ethernet | UI POST Ethernet failure | If the UI POST reports ethernet failure. | "}, + { 269 , ALARM_ID_HD_UI_POST_FAILURE_SOUND , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI Sound | UI POST Sound failure | If the UI POST reports sound failure. | "}, + { 270 , ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Proc Safestate | HD POST Safety Shutdown failure | If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. | "}, + { 271 , ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Proc Safestate | DG POST Safety Shutdown failure | If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. | "}, + { 272 , ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Fan Speed | HD Fan RPM out of range | If the fan RPM is out of Min/Max range. | "}, + { 273 , ALARM_ID_AVAILABLE_10 , " | | | | | | "}, + { 274 , ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Fault | DG idle reservoir weight out of range | If the reservoir weight > 200.0 of the target mL. | "}, + { 275 , ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Arterial Press Range | HD arterial pressure sensor is reading out of range | If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, + { 276 , ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Venous Press Range | HD venous pressure sensor is reading out of range | If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, + { 277 , ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Occlusion Sens Range | HD BP occlusion sensor is reading out of range | If the blood pump occlusion count is < 2000. | "}, + { 278 , ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "HD | Dialysate Reservoir Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed\n when dialysate is ready for use. | Dial. Gen: Reservoir Overfill | HD active reservoir weight out of range. | If the active reservoir weight is > 1900.0 g. | "}, + { 279 , ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Drain | DG dialysate drain time out | If the DG dialysate drain exceeded 120 seconds. | "}, + { 280 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Direction Fault | DG drain pump invalid direction | If the drain pump's direction is not 1 when draining. | "}, + { 281 , ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME , "DG | Acid Concentrate Low | The acid concentrate is low.\n- Replace acid concentrate container. | Dial. Gen: Acid Conc. Low | DG Acid concentration bottle low volume alarm | If the acid concentration bottle volume > 3240 (3340 - 100). | "}, + { 282 , ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME , "DG | Bicarb Concentrate Low | Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. | Dial. Gen: Bicarb Conc. Low | DG Bicarbonate concentration bottle low volume alarm | If the bicarbonate concentration bottle volume > 3680 (3780 - 100). | "}, + { 283 , ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC Range | DG load cells weight out of range | If the weight of the DG load cells > 4500.0 g or < 0.0 g. | "}, + { 284 , ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Redundant LC Check | DG load cells primary/back up drift out of range | If redundant load cells deviate by more than 60g after tare for 5 seconds. | "}, + { 285 , ALARM_ID_HD_RO_PERMEATE_SAMPLE , "HD | RO Permeate Sample in Progress | RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: RO Permeate Sample In Progress | DG RO permeate sample started | If DG RO permeate sample successfully started from UI. | "}, + { 286 , ALARM_ID_DG_CONCENTRATE_CAP_NOT_IN_PROPER_POSITION , "DG | Concentrate Connection Cap Removed | The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. | Cleaning: Conc Cap Removed | DG cap in wrong state | If the concentrate cap is not in the right position in the cleaning modes. | "}, + { 287 , ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Proc- FPGA Clock | HD processor clock speed checks against FPGA clock failure | If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. | "}, + { 288 , ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE , "HD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC Drift Check | HD load cell inconsistency | If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. | "}, + { 289 , ALARM_ID_DG_DIALYSATE_CAP_NOT_IN_PROPER_POSITION , "DG | Dialysate Connection Cap Removed | The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. | Cleaning: Dialysate Cap Removed | DG cap in wrong state | If the dialysate cap is not in the right position in the cleaning modes. | "}, + { 290 , ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Ven. Pres. Com | HD venous sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, + { 291 , ALARM_ID_HD_DISINFECT_FLUSH , "HD | Flush In Progress | System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Flush In Progress | DG flush started | If DG flush successfully started from the UI. | "}, + { 292 , ALARM_ID_HD_DISINFECT_HEAT , "HD | Heat Disinfection in Progress | Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Heat Disinf In Progress | DG heat disinfect started | If DG heat disinfect successfully started from the UI. | "}, + { 293 , ALARM_ID_HD_DISINFECT_CHEM , "HD | Chemical Disinfection in Progress | Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Chem Disinfect In Progress | DG chem disinfect started | If DG flush successfully started from the UI. | "}, + { 294 , ALARM_ID_HD_DISINFECT_HEAT_COOL , "HD | Active Cool in Progress | Active Cool is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: Active Cool In Progress | DG active cool started | If DG active cool successfully started. | "}, + { 295 , ALARM_ID_HD_UI_POST_FAILURE_CLOUDSYNC , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: UI CloudSync | UI cloud sync POST failure | If the UI POST reports cloud sync failure. | "}, + { 296 , ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: TDi RTD Comm | DG TDi sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, + { 297 , ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: TRo RTD Comm | DG TRo sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, + { 298 , ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Pres Comm | DG barometric sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, + { 299 , ALARM_ID_DG_INVALID_SERIAL_NUMBER , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Serial Number Check | DG invalid serial number | If the DG serial number is empty and does not have any characters. | "}, + { 300 , ALARM_ID_HD_INVALID_SERIAL_NUMBER , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Serial Number Check | HD invalid serial number | If the HD serial number is empty and does not have any characters. | "}, + { 301 , ALARM_ID_DG_RO_PERMEATE_SAMPLE_REMOVE_DIA_CAP , "DG | Remove Dialysate Cap | Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. | Cleaning: RO Permeate Sample Connect Cap | Open dialysate cap to be able to collecte the dispensing fluid | RO permeate sample collection state | "}, + { 302 , ALARM_ID_DG_CPU_RAM_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Processor RAM error | DG_CPU RAM error | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, + { 303 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Hall Commutation | DG drain pump direction FPGA fault | If the FPGA indicates an error for a certain period of time. | "}, + { 304 , ALARM_ID_HD_INVALID_USAGE_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Usage Rec NVRAM CRC | HD invalid usage reocord | If the CRC of the HD usage record does not match the calculated CRC. | "}, + { 305 , ALARM_ID_HD_CPU_RAM_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Processor RAM error | HD_CPU RAM error | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, + { 306 , ALARM_ID_AVAILABLE_50 , " | | | | | | "}, + { 307 , ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT , "HD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. | "}, + { 308 , ALARM_ID_DG_CPI_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CPi Comm | DG CPi conductivity sensor transmits invalid characters | If the CPi conductivity sensor transmits invalid characters for a certain period of time. | "}, + { 309 , ALARM_ID_DG_CPO_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CPo Comm | DG CPo conductivity sensor transmits invalid characters | If the CPo conductivity sensor transmits invalid characters for a certain period of time. | "}, + { 310 , ALARM_ID_DG_CD1_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CD1 Comm | DG CD1 conductivity sensor transmits invalid characters | If the CD1 conductivity sensor transmits invalid characters for a certain period of time. | "}, + { 311 , ALARM_ID_DG_CD2_COND_SENSOR_INVALID_CHAR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: CD2 Comm | DG CD2 conductivity sensor transmits invalid characters | If the CD2 conductivity sensor transmits invalid characters for a certain period of time. | "}, + { 312 , ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE , "HD | Low Dialysate Flow | Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nResume/OK button will be available.\n- Press Resume/OK to continue. | Dial. Gen: Dialysate In Flow | The dialysate inlet flow is too low. | If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. | "}, + { 313 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_LOW , "DG | Disinfection: Water Temperature Low | Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Temp Low | DG cleaning mode inlet water temperature too low | If inlet water temperature is < 24 C. | "}, + { 314 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_HIGH , "DG | Disinfection: Water Pressure High | Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. | Cleaning: Inlet Water Press High | DG cleaning mode inlet water pressure too high | If inlet water temperature is > 80 psig. | "}, + { 315 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_LOW , "DG | Disinfection: Water Pressure Low | Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water pressure\n before initiating disinfection again. | Cleaning: Inlet Water Press Low | DG cleaning mode inlet water pressure too low | If inlet water pressure is < 14 psig. | "}, + { 316 , ALARM_ID_AVAILABLE_20 , " | | | | | | "}, + { 317 , ALARM_ID_AVAILABLE_21 , " | | | | | | "}, + { 318 , ALARM_ID_AVAILABLE_2 , " | | | | | | "}, + { 319 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_FLUSH_SAMPLE , "DG | Residual Chemical Sampling | A water sample is needed to confirm the \nabsence of residual chemical following disinfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. | Cleaning: Chem Residual Sampling | Collect sample to check the chemical disinfectant acid | Chemical disinfect flush sample alarm is prompted. | "}, + { 320 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_SAMPLE_TIMEOUT , "DG | Chemical Residual Timeout | A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. | Cleaning: Chem Residual Timeout | The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes | Chemical disinfect flush sample was not collected form more than 30 minutes. | "}, + { 321 , ALARM_ID_DG_OUTLET_UV_REACTOR_NOT_HEALTHY , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVo Error | Outlet reactor not healthy | If the outlet reactor are not healthy. | "}, + { 322 , ALARM_ID_HD_UI_POST_FAILURE_INVALID_YEAR , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST Alarm: UI Invalid Year | UI POST Invalid Year | If the UI POST reports a invalid year. | "}, + { 323 , ALARM_ID_HD_UI_POST_FAILURE_SETTINGS_BAD , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST Alarm: UI Bad Settings | UI POST Bad Settings | If the UI POST reports a bad settings. (e.g. no settings read, empty settings file, invalid/empty parms, CRC failure) | "}, // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition }; #endif