Index: AlarmDefs.h =================================================================== diff -u -r9c792fdea5d43a7f540b268fcdc7ad4d5faf77df -r1e017e7e1063b7fdd10829a948773575cd258f7e --- AlarmDefs.h (.../AlarmDefs.h) (revision 9c792fdea5d43a7f540b268fcdc7ad4d5faf77df) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 1e017e7e1063b7fdd10829a948773575cd258f7e) @@ -1,6 +1,6 @@ /**********************************************************************//** * - * Copyright (c) 2019-2020 Diality Inc. - All Rights Reserved. + * Copyright (c) 2019-2020 Diality Inc. - All Rights Reserved. * * THIS CODE MAY NOT BE COPIED OR REPRODUCED IN ANY FORM, IN PART OR IN * WHOLE, WITHOUT THE EXPLICIT PERMISSION OF THE COPYRIGHT OWNER. @@ -73,12 +73,13 @@ ALARM_ID_TEMPERATURE_SENSORS_OUT_OF_RANGE = 48, ///< DG temperature sensors ADC read out of range ALARM_ID_TEMPERATURE_SENSORS_INCONSISTENT = 49, ///< DG temperature sensors values are inconsistent with respect to each other ALARM_ID_HD_COMM_TIMEOUT = 50, ///< HD communication timeout. - ALARM_ID_BLOOD_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 51, ///< Blood pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed. - ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 52, ///< Dialysate inlet pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed. - ALARM_ID_DIAL_OUT_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 53, ///< Dialysate outlet pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed. - ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK = 54, ///< Blood pump failed motor speed check. Measured vs. commanded. - ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK = 55, ///< Dialysate inlet pump failed motor speed check. Measured vs. commanded. - ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK = 56, ///< Dialysate outlet pump failed motor speed check. Measured vs. commanded. + ALARM_ID_VALVE_CONTROL_FAILURE = 51, ///< FPGA not accepting commanded valve states. + ALARM_ID_BLOOD_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 52, ///< Blood pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed. + ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 53, ///< Dialysate inlet pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed. + ALARM_ID_DIAL_OUT_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 54, ///< Dialysate outlet pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed. + ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK = 55, ///< Blood pump failed motor speed check. Measured vs. commanded. + ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK = 56, ///< Dialysate inlet pump failed motor speed check. Measured vs. commanded. + ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK = 57, ///< Dialysate outlet pump failed motor speed check. Measured vs. commanded. NUM_OF_ALARM_IDS ///< Total # of alarms. }; /// Type for alarm list enumeration.