Index: AlarmDefs.h =================================================================== diff -u -raec79b5976528cca7e9041b778be367e4758e7a3 -r3330b2c90ed2732c32cf78c47c78c08da71cf7dd --- AlarmDefs.h (.../AlarmDefs.h) (revision aec79b5976528cca7e9041b778be367e4758e7a3) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 3330b2c90ed2732c32cf78c47c78c08da71cf7dd) @@ -308,7 +308,7 @@ ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED = 270, ///< HD POST Safety Shutdown failure ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED = 271, ///< DG POST Safety Shutdown failure ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE = 272, ///< HD Fan RPM out of range - ALARM_ID_AVAILABLE_10 = 273, ///< Available for use + ALARM_ID_AVAILABLE_30 = 273, ///< Alarm available. ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE = 274, ///< DG inactive reservoir weight out of range ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE = 275, ///< HD arterial pressure sensor is reading out of range ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE = 276, ///< HD venous pressure sensor is reading out of range @@ -351,7 +351,7 @@ ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_LOW = 313, ///< DG cleaning mode inlet water temperature too low. ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_HIGH = 314, ///< DG cleaning mode inlet water pressure too high. ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_LOW = 315, ///< DG cleaning mode inlet water pressure too low. - ALARM_ID_AVAILABLE_20 = 316, ///< Alarm available + ALARM_ID_AVAILABLE_20 = 316, ///< Alarm available. ALARM_ID_AVAILABLE_21 = 317, ///< Alarm available. ALARM_ID_AVAILABLE_2 = 318, ///< Alarm available for use ALARM_ID_DG_CHEM_DISINFECT_FLUSH_FLUSH_SAMPLE = 319, ///< DG chemical disinfect flush flush sample. @@ -514,29 +514,29 @@ { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_PUMP_OFF_CHECK }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_PUMP_MC_DIRECTION_CHECK }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_CHECK }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_MC_CURRENT_CHECK }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_MC_CURRENT_CHECK }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_OFF_CHECK }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_MC_DIRECTION_CHECK }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_ROTOR_SPEED_CHECK }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_MC_CURRENT_CHECK }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_MC_DIRECTION_CHECK }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_ROTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_MC_CURRENT_CHECK }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_OFF_CHECK }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_MC_DIRECTION_CHECK }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_MC_DIRECTION_CHECK }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_UI_COMM_TIMEOUT }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET BTSRB BTSET NoBRcr NoDRcr ClrOnly TxLog ID { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_UI }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_MAX_RO_PUMP_PWM_EXCEEDED }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_UF_VOLUME_ACCURACY_ERROR }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_UF_VOLUME_ACCURACY_ERROR }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_COMM_TIMEOUT }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_VALVE_CONTROL_FAILURE }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_PUMP_MOTOR_SPEED_CHECK }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_MOTOR_SPEED_CHECK }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_PUMP_MOTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_CRITICAL_DATA_ERROR }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CRITICAL_DATA_ERROR }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_FAILURE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_FAILURE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETER_FAILURE }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_HOMING_FAILED }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT }, @@ -612,7 +612,7 @@ { ALARM_PRIORITY_HIGH, 150, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , ALARM_ID_HD_BLOOD_LEAK_DETECTED }, { ALARM_PRIORITY_HIGH, 200, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_HD_VENOUS_PRESSURE_LOW }, { ALARM_PRIORITY_HIGH, 12, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED, }, - { ALARM_PRIORITY_MEDIUM, 301, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT, }, + { ALARM_PRIORITY_MEDIUM, 698, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT, }, { ALARM_PRIORITY_HIGH, 201, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_HD_VENOUS_PRESSURE_HIGH }, { ALARM_PRIORITY_HIGH, 202, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_HD_ARTERIAL_PRESSURE_LOW }, { ALARM_PRIORITY_HIGH, 203, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH }, @@ -630,7 +630,7 @@ { ALARM_PRIORITY_MEDIUM, 603, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_HIGH }, { ALARM_PRIORITY_MEDIUM, 410, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , ALARM_ID_DG_ACID_CONDUCTIVITY_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 2, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE }, - { ALARM_PRIORITY_MEDIUM, 421, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT }, + { ALARM_PRIORITY_MEDIUM, 699, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT }, { ALARM_PRIORITY_MEDIUM, 601, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_HIGH }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_WATCHDOG_EXPIRED }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET BTSRB BTSET NoBRcr NoDRcr ClrOnly TxLog ID @@ -761,7 +761,7 @@ { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED }, { ALARM_PRIORITY_HIGH, 2, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED }, { ALARM_PRIORITY_LOW, 850, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE }, - { ALARM_PRIORITY_LOW, 999, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_10 }, + { ALARM_PRIORITY_LOW, 999, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_30 }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET BTSRB BTSET NoBRcr NoDRcr ClrOnly TxLog ID { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE }, @@ -888,25 +888,15 @@ { 10, ALARM_ID_HD_BLOOD_PUMP_OFF_CHECK }, { 10, ALARM_ID_HD_BLOOD_PUMP_MC_DIRECTION_CHECK }, { 10, ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_CHECK }, - { 10, ALARM_ID_HD_DIAL_IN_PUMP_MC_CURRENT_CHECK }, { 10, ALARM_ID_HD_DIAL_IN_PUMP_OFF_CHECK }, - { 10, ALARM_ID_HD_DIAL_IN_PUMP_MC_DIRECTION_CHECK }, - { 10, ALARM_ID_HD_DIAL_IN_PUMP_ROTOR_SPEED_CHECK }, { 10, ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT }, - { 10, ALARM_ID_HD_DIAL_OUT_PUMP_MC_CURRENT_CHECK }, { 10, ALARM_ID_HD_DIAL_OUT_PUMP_OFF_CHECK }, - { 10, ALARM_ID_HD_DIAL_OUT_PUMP_MC_DIRECTION_CHECK }, - { 10, ALARM_ID_HD_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK }, { 10, ALARM_ID_HD_UI_COMM_TIMEOUT }, { 10, ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS }, { 10, ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_UI }, - { 10, ALARM_ID_HD_UF_VOLUME_ACCURACY_ERROR }, { 10, ALARM_ID_HD_FPGA_COMM_TIMEOUT }, { 10, ALARM_ID_HD_BLOOD_PUMP_MOTOR_SPEED_CHECK }, - { 10, ALARM_ID_HD_DIAL_IN_PUMP_MOTOR_SPEED_CHECK }, - { 10, ALARM_ID_HD_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK }, { 10, ALARM_ID_HD_CRITICAL_DATA_ERROR }, - { 10, ALARM_ID_HD_ACCELEROMETER_FAILURE }, { 10, ALARM_ID_HD_VALVE_HOMING_FAILED }, { 10, ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT }, { 10, ALARM_ID_HD_VALVE_NOT_FUNCTIONAL }, @@ -1015,6 +1005,16 @@ { 110, ALARM_ID_TEST_ALARM_NO_DIAL_RECIRC_CLR_TOP_ONLY }, { 110, ALARM_ID_DG_MAX_RO_PUMP_PWM_EXCEEDED }, { 110, ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE }, + { 110, ALARM_ID_HD_DIAL_IN_PUMP_MC_CURRENT_CHECK }, + { 110, ALARM_ID_HD_DIAL_IN_PUMP_ROTOR_SPEED_CHECK }, + { 110, ALARM_ID_HD_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK }, + { 110, ALARM_ID_HD_DIAL_IN_PUMP_MC_DIRECTION_CHECK }, + { 110, ALARM_ID_HD_DIAL_OUT_PUMP_MC_CURRENT_CHECK }, + { 110, ALARM_ID_HD_DIAL_OUT_PUMP_MC_DIRECTION_CHECK }, + { 110, ALARM_ID_HD_UF_VOLUME_ACCURACY_ERROR }, + { 110, ALARM_ID_HD_DIAL_IN_PUMP_MOTOR_SPEED_CHECK }, + { 110, ALARM_ID_HD_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK }, + { 110, ALARM_ID_HD_ACCELEROMETER_FAILURE }, { 111, ALARM_ID_HD_DG_COMM_TIMEOUT }, { 111, ALARM_ID_HD_NEW_RESERVOIRS_DATA_MESSAGE_NOT_RECEIVE }, { 111, ALARM_ID_HD_DG_NEW_OPERATION_MODE_MESSAGE_NOT_RECEIVE }, @@ -1046,7 +1046,6 @@ { 291, ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM }, { 292, ALARM_ID_HD_PRIME_COMPLETED_HIGH }, { 300, ALARM_ID_HD_TREATMENT_STOPPED_BY_USER }, - { 301, ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT }, { 400, ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE }, { 401, ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP }, { 401, ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP }, @@ -1058,7 +1057,6 @@ { 420, ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE }, { 421, ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE }, { 421, ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE }, - { 421, ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT }, { 430, ALARM_ID_DG_INLET_WATER_PRESSURE_IN_HIGH_RANGE }, { 431, ALARM_ID_DG_INLET_WATER_PRESSURE_IN_LOW_RANGE }, { 432, ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE }, @@ -1092,6 +1090,8 @@ { 604, ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_LOW }, { 605, ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_HIGH }, { 606, ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_LOW }, + { 698, ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT }, + { 699, ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT }, { 700, ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE }, { 700, ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, { 700, ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, @@ -1136,11 +1136,11 @@ { 999, ALARM_ID_HD_UI_POST_FAILURE_DISPLAY }, { 999, ALARM_ID_AVAILABLE_2 }, { 999, ALARM_ID_AVAILABLE_7 }, - { 999, ALARM_ID_AVAILABLE_10 }, { 999, ALARM_ID_AVAILABLE_13 }, { 999, ALARM_ID_AVAILABLE_15 }, { 999, ALARM_ID_AVAILABLE_20 }, { 999, ALARM_ID_AVAILABLE_21 }, + { 999, ALARM_ID_AVAILABLE_30 }, { 999, ALARM_ID_AVAILABLE_31 }, { 999, ALARM_ID_AVAILABLE_32 }, { 999, ALARM_ID_NO_ALARM }, @@ -1437,7 +1437,7 @@ { 270 , ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Proc Safestate | HD POST Safety Shutdown failure | If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. | "}, { 271 , ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Proc Safestate | DG POST Safety Shutdown failure | If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. | "}, { 272 , ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Fan Speed | HD Fan RPM out of range | If the fan RPM is out of Min/Max range. | "}, - { 273 , ALARM_ID_AVAILABLE_10 , " | | | | | | "}, + { 273 , ALARM_ID_AVAILABLE_30 , " | | | | | | "}, { 274 , ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Fault | DG idle reservoir weight out of range | If the reservoir weight > 200.0 of the target mL. | "}, { 275 , ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Arterial Press Range | HD arterial pressure sensor is reading out of range | If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, { 276 , ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Venous Press Range | HD venous pressure sensor is reading out of range | If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. | "},