Index: AlarmDefs.h =================================================================== diff -u -rb20e371abf1f3f6a3fa0d953127dc5bdcf737eb3 -rde163bd40f302115eaeaa7e3db8757225c0cc8ba --- AlarmDefs.h (.../AlarmDefs.h) (revision b20e371abf1f3f6a3fa0d953127dc5bdcf737eb3) +++ AlarmDefs.h (.../AlarmDefs.h) (revision de163bd40f302115eaeaa7e3db8757225c0cc8ba) @@ -8,7 +8,7 @@ * @file AlarmDefs.h * * @author (last) Dara Navaei -* @date (last) 21-Mar-2022 +* @date (last) 30-Mar-2022 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -137,7 +137,7 @@ ALARM_ID_HD_SYRINGE_PUMP_STALL = 99, ///< HD syringe pump stall alarm ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT = 100, ///< HD no cartridge self-test timeout ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT = 101, ///< HD dry self-test timeout - ALARM_ID_AVAILABLE_1 = 102, ///< Alarm available for use + ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE = 102, ///< DG heat disinfect temperature gradient out of range ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID = 103, ///< HD non-volatile calibration group invalid record CRC ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS = 104, ///< Air trap level sensors reporting illegal combination of air/fluid ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID = 105, ///< DG non-volatile system group invalid record CRC @@ -393,12 +393,12 @@ ALARM_STATE_FLAG_BIT_POS_NO_RESUME = 3, ///< The "resume" user recovery option is disabled ALARM_STATE_FLAG_BIT_POS_NO_RINSEBACK = 4, ///< The "rinseback" user recovery option is disabled ALARM_STATE_FLAG_BIT_POS_NO_END_TREATMENT = 5, ///< The "end treatment" user recovery option is disabled - ALARM_STATE_FLAG_BIT_POS_UNUSED_1 = 6, ///< Reserved + ALARM_STATE_FLAG_BIT_POS____AVAILABLE = 6, ///< Not used - available for use ALARM_STATE_FLAG_BIT_POS_OK_BUTTON_ONLY = 7, ///< The "ok" user recovery option is enabled ALARM_STATE_FLAG_BIT_POS_ALARMS_TO_ESCALATE = 8, ///< One or more active alarms will escalate in time ALARM_STATE_FLAG_BIT_POS_ALARMS_SILENCED = 9, ///< Alarms have been temporarily silenced by user ALARM_STATE_FLAG_BIT_POS_LAMP_ON = 10, ///< Alarm lamp is currently on (for syncing to UI) - ALARM_STATE_FLAG_BIT_POS_UNUSED_2 = 11, ///< Reserved + ALARM_STATE_FLAG_BIT_POS_UNUSED_1 = 11, ///< Reserved ALARM_STATE_FLAG_BIT_POS_BLOOD_RECIRC = 12, ///< Blood recirculation is disabled ALARM_STATE_FLAG_BIT_POS_DIALYSATE_RECIRC = 13, ///< Dialysate recirculation is disabled ALARM_STATE_FLAG_BIT_POS_NO_MINIMIZE = 14, ///< Prevent user from minimizing alarm window @@ -444,6 +444,7 @@ /// Table of alarms and their static properties. const ALARM_T ALARM_TABLE[] = +/* <<<<<<< HEAD */ { // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_NONE, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NO_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_STUCK_BUTTON_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, @@ -551,7 +552,7 @@ // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_1, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE, ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, @@ -749,7 +750,313 @@ { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON, 0, ALARM_ID_NO_ALARM }, }; // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To - +/*======= +{ // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_NONE, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NO_ALARM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_STUCK_BUTTON_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_WATCHDOG_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_WATCHDOG_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_COMM_POST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RTC_CONFIG_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_RTC_OR_TIMER_ACCURACY_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_BLOOD_FLOW_INVALID_CALIBRATION, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CALIBRATION, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_UI_POST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRESSURE_SENSORS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETERS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_FLOW_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SOFTWARE_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_OFF_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_MC_DIRECTION_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_MC_CURRENT_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_OFF_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_MC_DIRECTION_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_ROTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_MC_CURRENT_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_WATCHDOG_EXPIRED, 0, ALARM_ID_NO_ALARM },//add DG version + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_COMM_TOO_MANY_BAD_CRCS, 0, ALARM_ID_NO_ALARM },//add DG version + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_CAN_MESSAGE_NOT_ACKED, 0, ALARM_ID_NO_ALARM },//add DG version + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_UF_RATE_TOO_HIGH_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_UF_VOLUME_ACCURACY_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_VALVE_CONTROL_FAILURE, 0, ALARM_ID_NO_ALARM },//add DG version + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_CRITICAL_DATA_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CRITICAL_DATA_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETER_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_HOMING_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_NOT_FUNCTIONAL, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_FLOW_STATUS_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_STATUS_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_PUMP_TRACK_LATCH_OPENED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_WATCHDOG_EXPIRED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BATTERY_COMM_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_STALL, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE, ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESTARTED_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_DG_CAN_MESSAGE_NOT_ACKED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_RTC_CONFIG_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_RTC_BATTERY_LOW, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRE_TREATMENT_WET_PRIME_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 4, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_HIGH, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 5, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 6, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE, TRUE, FALSE, TRUE, FALSE, TRUE, ALARM_ID_HD_BLOOD_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 8, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_VENOUS_PRESSURE_LOW, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 9, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_VENOUS_BUBBLE_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 10, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 11, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_VENOUS_PRESSURE_HIGH, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 12, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_ARTERIAL_PRESSURE_LOW, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 13, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_ARTERIAL_PRESSURE_HIGH, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 14, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE, ALARM_ID_DG_FLUID_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 15, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_FLUID_LEAK_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 16, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_SHOCK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 16, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_SHOCK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 17, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_EXCESSIVE_TILT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 17, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_EXCESSIVE_TILT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 18, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE, ALARM_ID_HD_AC_POWER_LOST, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 19, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_DG_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 20, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 21, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_OCCLUSION_BLOOD_PUMP, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE, ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 21, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_DG_SW_CONFIG_RECORD_INVALID_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE , TRUE , FALSE, TRUE, ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_INLET_WATER_HIGH_TEMPERATURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_WATCHDOG_EXPIRED, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 36, FALSE, FALSE, FALSE, TRUE, FALSE, TRUE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_COMPLETED_HIGH, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_EEPROM_OPS_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_DG_COMM_TOO_MANY_BAD_CRCS, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_CLOCK_SPEED_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED_RINSEBACK, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_SOFTWARE_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_LOAD_CELL_VALUE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UV_REACTOR_NOT_HEALTHY, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP1_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CP2_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_OFF_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEATERS_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRE_TREATMENT_WET_FLOW_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_METER_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 800, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_SITTING_WARNING, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 801, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_ALARM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 802, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_COMPLETED_MEDIUM, ALM_ESC_30_MIN, ALARM_ID_PRIME_COMPLETED_HIGH }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_SALINE_BOLUS_VOLUME_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_END_TREATMENT_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_DETECTED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH , 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_BATTERY_PACK_CHARGE_TOO_LOW, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 900, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_EMPTY_SALINE_BAG, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 901, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_DIALYSATE_TEMPERATURE_HIGH, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_LOW, 902, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_DIALYSATE_TEMPERATURE_LOW, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_LOW, 903, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_BY_USER, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_LOW, 904, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_WARNING, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 905, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY, ALM_ESC_90_MIN, ALARM_ID_PRIME_COMPLETED_MEDIUM }, + { ALARM_PRIORITY_LOW, 909, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_OUT_OF_TIME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 909, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 909, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 909, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRE_TREATMENT_WET_LC_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_RTC_BATTERY_LOW, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_RTC_RAM_OPS_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_INSTALL_NEW_CARTRIDGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_NO_CARTRIDGE_LOADED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_REMOVAL_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMP_SENSORS_DIFF_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_AND_TEMP_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_CHEM_DISINFECT_INLET_COND_AND_TEMP_OUT, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_SYSTEM_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_UI_COMPATIBILITY_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DG_COMPATIBILITY_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_POWER_OUT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_POWER_OUT_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SHASUM, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CANBUS, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_DISPLAY, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_TOUCH, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SDCARD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_RTC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_UI_POST_FAILURE_WIFI, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_UI_POST_FAILURE_BLUETOOTH, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_UI_POST_FAILURE_ETHERNET, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_UI_POST_FAILURE_SOUND, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_FLOW_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_FLOW_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_HD_TREATMENT_RECIRC_TIMEOUT_WARNING, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_WARNING, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_ON_WITH_NO_FLOW_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON, 0, ALARM_ID_NO_ALARM }, +}; // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoRcr ClrOnly TxLog ID Escalate In Escalate To +>>>>>>> staging +*/ /// Table of alarms and their rank. const ALARM_RANK_T ALARM_RANK_TABLE[] = { @@ -867,7 +1174,7 @@ { 3, ALARM_ID_HD_SYRINGE_PUMP_STALL }, { 3, ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT }, { 3, ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT }, - { 3, ALARM_ID_AVAILABLE_1 }, + { 3, ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE }, { 3, ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID }, { 3, ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS }, { 3, ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY },