Index: AlarmDefs.h =================================================================== diff -u -rc10d200f70b4fbec28c4ed385ca0aa3ae3909a0b -rc12b4970db3be16f05b60332d0b27a4387aa1bd6 --- AlarmDefs.h (.../AlarmDefs.h) (revision c10d200f70b4fbec28c4ed385ca0aa3ae3909a0b) +++ AlarmDefs.h (.../AlarmDefs.h) (revision c12b4970db3be16f05b60332d0b27a4387aa1bd6) @@ -8,7 +8,7 @@ * @file AlarmDefs.h * * @author (last) Dara Navaei -* @date (last) 20-Apr-2023 +* @date (last) 24-Apr-2023 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -1167,23 +1167,23 @@ { 8 , ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE , "HD | HD POST: Acc Vector | Call service to schedule a repair. | HD accelerometer failed POST | If the calculated vector length of the accelerometer x, y, z is not the nominal length. | "}, { 9 , ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE , "DG | DG POST: Acc Vector | Call service to schedule a repair. | DG accelerometer failed POST | If the calculated vector length of the accelerometer x, y, z is not the nominal length. | "}, { 10 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE , "HD | HD POST: BLD Config Emb Mode | Call service to schedule a repair. | HD blood leak sensor embedded mode communication failed | If the blood leak sensor embedded mode is unresponsive. | "}, - { 11 , ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP , "HD | Dial. Gen: Dialysate Temp Low | Dialysate temperature is low\n- Treatment is paused and will be ready to resume when dialysate is within target range. | Dialysate temperature is below the target temperature | If the dialysate temperature is below the target temperature for a certain period of time. | "}, + { 11 , ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP , "HD | Dial. Gen: Dialysate Temp Low | Dialysate temperature is low\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. | Dialysate temperature is below the target temperature | If the dialysate temperature is below the target temperature for a certain period of time. | "}, { 12 , ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED , "HD | HD POST: HD Image | Call service to schedule a repair. | HD firmware image integrity POST test failed | If the calculated CRC of the HD image does not match the expected CRC. | "}, { 13 , ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED , "DG | DG POST: DG Image | Call service to schedule a repair. | DG firmware image integrity POST test failed | If the calculated CRC of the DG firmware image does not matched the expected CRC | "}, { 14 , ALARM_ID_DG_INVALID_USAGE_RECORD_CRC , "DG | DG POST: Usage Rec NVRAM CRC | Call service to schedule a repair. | DG non-volatile usage information group invalid record CRC | If the DG usage information record's CRC does not match the calculated CRC in POST. | "}, - { 15 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP , "HD | Dial. Gen: Dialysate Temp High | Dialysate temperature is high\n-Treatment is paused and will be ready to resume when dialysate is within target range. | Dialysate temperature is out of high safety range | If the dialysate temperature is > 46 C for a certain period of time. | "}, + { 15 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP , "HD | Dial. Gen: Dialysate Temp High | Dialysate temperature is high\n-Treatment is paused and will be ready to resume when \ndialysate is within target range. | Dialysate temperature is out of high safety range | If the dialysate temperature is > 46 C for a certain period of time. | "}, { 16 , ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE , "HD | HD POST: Primary Alarm Current | Call service to schedule a repair. | HD alarm audio failed POST | If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output | "}, { 17 , ALARM_ID_HD_UI_POST_TIMEOUT , "HD | HD POST: UI POST Not Complete | Call service to schedule a repair. | HD UI POST failed | If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. | "}, { 18 , ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_DG , "HD | Loss of DG Comm (DG ACK) | Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. | HD to DG comm error. | HD message to DG that requires acknowledge was not acknowledged within a certain time. | "}, - { 19 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP , "HD | Dial. Gen: Dialysate Temp High | Dialysate temperature is high\n- Treatment is paused and will be ready to resume when dialysate is within target range. | Dialysate temperature is above the target temperature | If the dialysate temperature is above the target temperature for a certain period of time. | "}, + { 19 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP , "HD | Dial. Gen: Dialysate Temp High | Dialysate temperature is high\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. | Dialysate temperature is above the target temperature | If the dialysate temperature is above the target temperature for a certain period of time. | "}, { 20 , ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD , "DG | DG POST: Conductivity Cal | Call service to schedule a repair. | DG conductivity sensors invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, { 21 , ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD , "DG | DG POST: Drain Line Vol Cal | Call service to schedule a repair. | DG drain line volume invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, { 22 , ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD , "DG | DG POST: Reservoir Vol Cal | Call service to schedule a repair. | DG reservoirs invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, { 23 , ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD , "DG | DG POST: Acid Conc Cal | Call service to schedule a repair. | DG acid concentrate calibration record | If the non-volatile data is valid and if not raise the alarm | "}, { 24 , ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD , "DG | DG POST: Bicarb Conc Cal | Call service to schedule a repair. | DG bicarb concentrate calibration record | If the non-volatile data is valid and if not raise the alarm | "}, { 25 , ALARM_ID_DG_ACCELEROMETERS_INVALID_CAL_RECORD , "DG | DG POST: Acc Cal | Call service to schedule a repair. | DG accelerometer invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, { 26 , ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD , "HD | HD POST: Acc Cal | Call service to schedule a repair. | HD accelerometer invalid calibration record | If the accel calibration data that was received from NVDataMgmt is NOT legitimate. | "}, - { 27 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED , "HD | Process: Pre-Tx Wet Self-Test | Self-test detected has a problem with the blood leak sensor\n- Check cartridge installation on to machine\n- If problem persists terminate pre-treatment and restart with a new cartridge. | HD BLD sensor failed zero sequence | BLD fails to complete zero sequence. | "}, + { 27 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED , "HD | Process: Pre-Tx Wet Self-Test | Self-test detected has a problem with the blood leak sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. | HD BLD sensor failed zero sequence | BLD fails to complete zero sequence. | "}, { 28 , ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT , "DG | DG Fault: Conductivity RTD Comm | Call service to schedule a repair. | DG two wire sensors FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, { 29 , ALARM_ID_HD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD , "HD | HD POST: SP DAC Cal | Call service to schedule a repair. | HD heparin force sensor invalid calibration record | If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. | "}, { 30 , ALARM_ID_HD_SOFTWARE_FAULT , "HD | HD Fault: SW Invalid State | Call service to schedule a repair. | HD Software fault. Software found itself in an unexpected state | If the specific software faults listed in alarm management SW faults has been detected | "}, @@ -1224,7 +1224,7 @@ { 65 , ALARM_ID_HD_DG_COMMAND_INVALID_PARAMETER_FAULT , "HD | Loss of DG Comm (Reject) | Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. | HD requests DG command with invalid parameter fault | If the latest DG command response indicated invalid parameter | "}, { 66 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE , "HD | HD POST: BLD Config Set Point | Call service to schedule a repair. | HD blood leak sensor set point set failure | If the set point of the blood leak sensor does not match the set point in the calibration record. | "}, { 67 , ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE , "HD | Process: Pre-Tx Remove Cartridge | Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume | HD blood pump occlusion self-test failure alarm | If the blood pump pressure > 5000 | "}, - { 68 , ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE , "HD | Dial. Gen: Dialysate not ready | Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when dialysate is ready for use. | HD active reservoir recirculation out of range | If the recirculation level has exceeded the max allowed | "}, + { 68 , ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE , "HD | Dial. Gen: Dialysate not ready | Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. | HD active reservoir recirculation out of range | If the recirculation level has exceeded the max allowed | "}, { 69 , ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD , "HD | HD POST: BLD Cal | Call service to schedule a repair. | HD blood leak sensor invalid calibration record | If the BLD data that was received from NVDataMgmt is NOT legitimate. | "}, { 70 , ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE , "HD | Process: Pre-Tx Remove Cartridge | Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume | HD arterial pressure self-test failure alarm | If arterial pressure < -300 or > 100 mmHg | "}, { 71 , ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE , "HD | Process: Pre-Tx Remove Cartridge | Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume | HD venous pressure self-test failure alarm | If the venous pressure < -100 or > 600 mmHg | "}, @@ -1258,7 +1258,7 @@ { 99 , ALARM_ID_HD_SYRINGE_PUMP_STALL , "HD | HD Fault: Syringe Pump Stall | Call service to schedule a repair. | HD syringe pump stall alarm | If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. | "}, { 100 , ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT , "HD | HD Fault: Timeout No Cart ST | Call service to schedule a repair. | HD no cartridge self-test timeout | If during the no cartridge self-tests state, the elapsed self-test time > maximum no cartridge self test time (30secs) | "}, { 101 , ALARM_ID_UI_POST_HD_COMM , "UI | HD POST: UI-Proc Comm | Call service to schedule a repair. | UI POST FW Communication failure | If the HD is not communicating with the UI during POST. | "}, - { 102 , ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE , "DG | Cleaning: Tpo THD Temp Delta | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG heat disinfection temperature gradient out of range | If temperature gradient between Tpo and Thd exceeds maximum 15°C. | "}, + { 102 , ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE , "DG | Cleaning: Tpo THD Temp Delta | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG heat disinfection temperature gradient out of range | If temperature gradient between Tpo and Thd exceeds maximum 15°C. | "}, { 103 , ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID , "HD | HD POST: Cal NVRAM CRC | Call service to schedule a repair. | HD non-volatile calibration group invalid record CRC | If the computed CRC of the calibration group record is not equal to the record CRC. | "}, { 104 , ALARM_ID_HD_AIR_TRAP_ILLEGAL_LEVELS , "HD | HD Fault: Air Trap Sensor | Call service to schedule a repair. | Air trap level sensors reporting illegal combination of air/fluid | If the air trap level readings are not valid for more than 2 seconds. | "}, { 105 , ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID , "DG | DG POST: Cal NVRAM CRC | Call service to schedule a repair. | DG non-volatile system group invalid record CRC | If the DG NVDATAMGMT CAL group calculated CRC does not mached the expected CRC. | "}, @@ -1270,65 +1270,65 @@ { 111 , ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR , "HD | HD Fault: Syringe Pump Speed | Call service to schedule a repair. | HD syringe pump not stopped in off state error | If the syringe pump position is not equal to its last position. | "}, { 112 , ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT , "DG | DG Fault: LC A2/B2 Comm | Call service to schedule a repair. | DG load cells A2/B2 FPGA fault | If the FPGA does not report fresh data for a certain period of time. | "}, { 113 , ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE , "DG | DG Fault: DRP Current | Call service to schedule a repair. | DG drain pump current out of range | If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. | "}, - { 114 , ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE , "HD | Process: Pre-Tx Wet Self-Test | Self-test has detected a problem with the air bubble sensor\n- Check cartridge installation on to machine\n- If problem persists terminate pre-treatment and restart with a new cartridge. | HD venous air bubble detector self-test failure | If the venous air bubble self test did not detect bubble for more than 500 msec. | "}, + { 114 , ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE , "HD | Process: Pre-Tx Wet Self-Test | Self-test has detected a problem with the air bubble sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. | HD venous air bubble detector self-test failure | If the venous air bubble self test did not detect bubble for more than 500 msec. | "}, { 115 , ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE , "DG | DG Fault: Temp Sensor Range | Call service to schedule a repair. | DG temperature sensor out of range | If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. | "}, - { 116 , ALARM_ID_HD_UI_SDCARD_FAILURE , "UI HD | HD Alarm (Low): SD-Card Error | SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists | SD card is non-operational | SD card is non-operational | "}, + { 116 , ALARM_ID_HD_UI_SDCARD_FAILURE , "UI HD | HD Alarm (Low): SD-Card Error | SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. | SD card is non-operational | SD card is non-operational | "}, { 117 , ALARM_ID_DG_CAN_MESSAGE_NOT_ACKED , "DG | Loss of HD Comm (HD ACK) | Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. | System message that required acknowledgment was not acknowledged | If no retries left in the pending acknowlegement list. | "}, { 118 , ALARM_ID_DG_RTC_CONFIG_ERROR , "DG | DG Fault: RTC Config | Call service to schedule a repair. | DG RTC configuration error | If the RTC configuration registers report an error. | "}, { 119 , ALARM_ID_DG_RTC_BATTERY_LOW , "DG | DG Alarm (Low): RTC Bat Low | Call service to schedule a repair. | DG RTC battery low | If the RTC battery bit report a low alarm. | "}, - { 120 , ALARM_ID_HD_PRE_TREATMENT_WET_PRIME_TEST_FAILURE , "HD | Process: Pre-Tx Wet Self-Test | Self-test has detected a bubble in the venous bubble detector\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists terminate pre-treatment and restart with a new cartridge. | HD pre-treatment wet prime test failure | If the HD pre-treatment wet self test failed. | "}, + { 120 , ALARM_ID_HD_PRE_TREATMENT_WET_PRIME_TEST_FAILURE , "HD | Process: Pre-Tx Wet Self-Test | Self-test has detected a bubble in the venous bubble detector\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. | HD pre-treatment wet prime test failure | If the HD pre-treatment wet self test failed. | "}, { 121 , ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | DG Fault: Prim Heater Lg Control | Call service to schedule a repair. | DG main primary heater voltage out of range | If the main primary heater's voltage is out of range. | "}, { 122 , ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | DG Fault: Prim Heater Sm Control | Call service to schedule a repair. | DG small primary heater voltage out of range | If the small primary heater's voltage is out of range. | "}, { 123 , ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | DG Fault: Trimmer Heater Control | Call service to schedule a repair. | DG trimmer heater voltage out of range | If the trimmer voltage is out of range. | "}, - { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Stop: Blood Clot Risk | Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n-\n- NO RINSEBACK ALLOWED. | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, - { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Stop: Blood Clot Risk | Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n-\n- NO RINSEBACK ALLOWED. | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, + { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Stop: Blood Clot Risk | Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, + { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Stop: Blood Clot Risk | Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Stop: Dialyzer Blood Leak | A blood leak has been detected. Treatment must end. | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, - { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Blood Stop: Venous Pres. Low | Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured properly. | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, - { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Blood Stop: Venous Air Bubble | Venous air bubble detected indicating failure of air trap control system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, - { 129 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK , "HD | Blood Stop: Venous Air Bubble | Venous air bubble detected indicating failure of air trap control system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement | HD venous air bubble detected rinseback alarm | If bubble is detected in the treatment rinseback state. | "}, + { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Blood Stop: Venous Pres. Low | Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured \nproperly. | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Blood Stop: Venous Air Bubble | Venous air bubble detected indicating failure of air trap \ncontrol system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, + { 129 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK , "HD | Blood Stop: Venous Air Bubble | Venous air bubble detected indicating failure of air trap \ncontrol system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. | HD venous air bubble detected rinseback alarm | If bubble is detected in the treatment rinseback state. | "}, { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | Blood Stop: Venous Pres. High | Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 131 , ALARM_ID_HD_ARTERIAL_PRESSURE_LOW , "HD | Blood Stop: Arterial Pres. Low | Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. | Arterial pressure too low during treatment | If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 132 , ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH , "HD | Blood Stop: Arterial Pres. High | Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. | Arterial pressure too high during treatment | If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, - { 133 , ALARM_ID_DG_FLUID_LEAK_DETECTED , "DG | DG Fault: Fluid Leak in DG | Fluid leak detected in Dialysate Generation Delivery Device.\n-Treatment must be terminated.\n- Shut off water source and disconnect water and electrical connections from system.\n- Call service to schedule a repair. | DG fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, - { 134 , ALARM_ID_HD_FLUID_LEAK_DETECTED , "HD | Blood Stop: Fluid Leak in HD | Fluid leak detected in Hemodialysis Delivery device.\n- If in treatment BLOOD STOPPED.\n-Identify and correct the cause of the spill.\n- Clean up the spill and dry leak sensor.\n- If in treatment resume treatment. | HD fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, - { 135 , ALARM_ID_HD_SHOCK , "HD | Blood Stop: HD Shock Detected | Excessive movement detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. | HD experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, - { 136 , ALARM_ID_DG_SHOCK , "DG | Dial. Gen: Shock-Vibration | Excessive movement detected with Dialysis Generation device.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. | DG experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, - { 137 , ALARM_ID_HD_EXCESSIVE_TILT , "HD | Blood Stop: HD Tilted | Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing | HD tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. | "}, + { 133 , ALARM_ID_DG_FLUID_LEAK_DETECTED , "DG | DG Fault: Fluid Leak in DG | Fluid leak detected in Dialysate Generation Delivery Device.\n-Treatment must be terminated.\n- Shut off water source and disconnect water and \nelectrical connections from system.\n- Call service to schedule a repair. | DG fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, + { 134 , ALARM_ID_HD_FLUID_LEAK_DETECTED , "HD | Blood Stop: Fluid Leak in HD | Fluid leak detected in Hemodialysis Delivery device.\n- If in treatment, BLOOD STOPPED.\n-Identify and correct the cause of the spill.\n- Clean up the spill and dry leak sensor.\n- If in treatment, resume treatment. | HD fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, + { 135 , ALARM_ID_HD_SHOCK , "HD | Blood Stop: HD Shock Detected | Excessive movement detected with Hemodialysis Delivery \ndevice.\n- Confirm the device is stationary and level before continuing. | HD experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, + { 136 , ALARM_ID_DG_SHOCK , "DG | Dial. Gen: Shock-Vibration | Excessive movement detected with Dialysis Generation \ndevice.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. | DG experienced large acceleration | If the system experienced excessive shock > 2.5 g. | "}, + { 137 , ALARM_ID_HD_EXCESSIVE_TILT , "HD | Blood Stop: HD Tilted | Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. | HD tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. | "}, { 138 , ALARM_ID_DG_EXCESSIVE_TILT , "DG | Dial. Gen: Tilt | Excessive tilt detected with Dialysis Generation device.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing. | DG tilt exceeds maximum | If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. | "}, - { 139 , ALARM_ID_HD_AC_POWER_LOST , "HD | Loss of AC Power | Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until power returns. | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. | "}, + { 139 , ALARM_ID_HD_AC_POWER_LOST , "HD | Loss of AC Power | Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns. | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. | "}, { 140 , ALARM_ID_HD_DG_COMM_TIMEOUT , "HD | Loss of DG Comm (Timeout) | Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. | Dialysate generator communication timeout | If DG communication timeout is detected for more than 2 seconds. | "}, - { 141 , ALARM_ID_HD_AIR_TRAP_FILL_DURING_TREATMENT , "HD | Blood Stop: Air Trap Fill | Failure to fill the Air Trap.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "}, - { 142 , ALARM_ID_HD_OCCLUSION_BLOOD_PUMP , "HD | Blood Stop: Blood Pump Occlusion | Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the venous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. | Blood pump occlusion detected | If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall | "}, + { 141 , ALARM_ID_HD_AIR_TRAP_FILL_DURING_TREATMENT , "HD | Blood Stop: Air Trap Fill | Failure to fill the Air Trap.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "}, + { 142 , ALARM_ID_HD_OCCLUSION_BLOOD_PUMP , "HD | Blood Stop: Blood Pump Occlusion | Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the \nvenous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. | Blood pump occlusion detected | If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall | "}, { 143 , ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE , "DG | DG Fault: TRo TDi Delta | Call service to schedule a repair. | DG dialysate temperature sensors drift timeout. | If TDi and TRo have drifted for more 1 C for a certain period of time. | "}, { 144 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_HIGH , "DG | Cleaning: Inlet Water Cond High | Cleaning Mode Inlet water conductivity too high.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. | DG cleaning mode inlet water conductivity out of high range | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode | "}, { 145 , ALARM_ID_DG_ACID_CONDUCTIVITY_OUT_OF_RANGE , "DG | Dial. Gen: Acid Conc. Cond. | Acid concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. | DG Concentrate conductivity after adding acid out of range alarm | If the conductivity value for the acid concentrate is out of range for more than 45 secs. | "}, { 146 , ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE , "DG | DG POST: DG RTC Timer | Call service to schedule a repair. | DG RTC vs. Timer accuracy error | If RTC and timer are not aligned on 1 second elapsed time. | "}, - { 147 , ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT , "DG | Dial. Gen: Dialysate not ready | Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when dialysate is ready for use. | DG dialysate generation conductivity fault alarm | If dialysate conductivity is not with 5% of the TCD | "}, - { 148 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_HIGH , "DG | Cleaning: Inlet Water Temp High | Cleaning Mode Inlet water temperature too high.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. | Cleaning Mode Inlet water temperature too high | Inlet temperature > 37 C. | "}, + { 147 , ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT , "DG | Dial. Gen: Dialysate not ready | Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. | DG dialysate generation conductivity fault alarm | If dialysate conductivity is not with 5% of the TCD | "}, + { 148 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_HIGH , "DG | Cleaning: Inlet Water Temp High | Cleaning Mode Inlet water temperature too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. | Cleaning Mode Inlet water temperature too high | Inlet temperature > 37 C. | "}, { 149 , ALARM_ID_HD_WATCHDOG_EXPIRED , "HD | HD Fault: WD Error | Call service to schedule a repair. | HD watchdog expired | If the watchdog signal is low while not in POST. | "}, - { 150 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE , "DG | Dial. Gen: Inlet Water Cond High | Inlet water conductivity is high.\n- Treatment is paused\n- Check the incoming water supply. | Inlet water conductivity too high | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). | "}, - { 151 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE , "DG | Dial. Gen: Inlet Water Cond Low | Inlet water conductivity is low.\n- Treatment is paused\n- Check the incoming water supply. | Inlet water conductivity too low | Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). | "}, - { 152 , ALARM_ID_DG_INLET_WATER_PRESSURE_IN_LOW_RANGE , "DG | Dial. Gen: Inlet Water Pres Low | Inlet water pressure is low.\n- Treatment is paused\n- Check the incoming water supply. | Inlet water pressure too low | Inlet water pressure < 25 PSI. | "}, - { 153 , ALARM_ID_HD_PRIME_COMPLETED_HIGH , "HD | Process: Pre-Tx Terminated | Pre-Treatment timeout.\n- Patient was not connected to device\n- Pre-Treatment has been terminated\n- Proceed to end treatment to remove cartridge from system | HD prime completed high priority alarm | In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) | "}, + { 150 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE , "DG | Dial. Gen: Inlet Water Cond High | Inlet water conductivity is high.\n- Treatment is paused.\n- Check the incoming water supply. | Inlet water conductivity too high | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). | "}, + { 151 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE , "DG | Dial. Gen: Inlet Water Cond Low | Inlet water conductivity is low.\n- Treatment is paused.\n- Check the incoming water supply. | Inlet water conductivity too low | Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). | "}, + { 152 , ALARM_ID_DG_INLET_WATER_PRESSURE_IN_LOW_RANGE , "DG | Dial. Gen: Inlet Water Pres Low | Inlet water pressure is low.\n- Treatment is paused.\n- Check the incoming water supply. | Inlet water pressure too low | Inlet water pressure < 25 PSI. | "}, + { 153 , ALARM_ID_HD_PRIME_COMPLETED_HIGH , "HD | Process: Pre-Tx Terminated | Pre-Treatment timeout.\n- Patient was not connected to device\n- Pre-Treatment has been terminated\n- Proceed to end treatment to remove cartridge from system. | HD prime completed high priority alarm | In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) | "}, { 154 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_LOW , "DG | Cleaning: Inlet Water Cond Low | Cleaning Mode Inlet water conductivity too low.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. | Cleaning mode inlet water conductivity too low | Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode | "}, { 155 , ALARM_ID_HD_NEW_RESERVOIRS_DATA_MESSAGE_NOT_RECEIVE , "HD | Loss of DG Comm (Reservoir) | Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. | HD No reservoir data message receive at least once every 2 seconds | No new reservoir data for 2 seconds. | "}, { 156 , ALARM_ID_HD_DG_NEW_OPERATION_MODE_MESSAGE_NOT_RECEIVE , "HD | Loss of DG Comm (Op Mode) | Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. | HD No DG op mode message receive at least once every 2 seconds | No DG op mode for 2 seconds. | "}, - { 157 , ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT , "DG | Cleaning: Chem Disinf PAA Prime | Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the bicarb port.\n- Check chemical disinfectant bottle is not empty. | Chemical disinfect priming the acid line timeout | If the disinfectant acid line is not primed within a certain period of time. | "}, - { 158 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_LOW_RANGE , "DG | Dial. Gen: Inlet Water Temp Low | Inlet water temperature is low.\n- Treatment is paused\n- Check the incoming water supply. | Inlet water temperature low | Inlet temperature < 24 °C for 5 seconds. | "}, + { 157 , ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT , "DG | Cleaning: Chem Disinf PAA Prime | Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. | Chemical disinfect priming the acid line timeout | If the disinfectant acid line is not primed within a certain period of time. | "}, + { 158 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_LOW_RANGE , "DG | Dial. Gen: Inlet Water Temp Low | Inlet water temperature is low.\n- Treatment is paused.\n- Check the incoming water supply. | Inlet water temperature low | Inlet temperature < 24 °C for 5 seconds. | "}, { 159 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_REMOVE_ACID , "DG | Cleaning: Chem Disinf Connect Cap | Remove chemical disinfectant connection and close cap. | Close concentrate line be able to flush it | Start of chemical disinfect flush. | "}, { 160 , ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH , "HD | HD Fault: BP Rotor Max Speed | Call service to schedule a repair. | Blood pump rotor speed too high | If blood pump motor speed exceeded 100.0 RPM. | "}, { 161 , ALARM_ID_DG_COMM_TOO_MANY_BAD_CRCS , "DG | DG Fault: HD-DG Comm | Call service to schedule a repair. | Too many bad CRCs detected on received system messages | If >= 5 bad CRCs detected on received CAN messages within 10 minutes. | "}, { 162 , ALARM_ID_DG_FPGA_CLOCK_SPEED_CHECK_FAILURE , "DG | DG Fault: Proc-FPGA Clock | Call service to schedule a repair. | FPGA clock speed failure | If the FPGA clock speed is relatively close to the processor clock for a certain period of time. | "}, - { 163 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM , "HD | Dial. Gen: UF Rate Check | Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system (shock/vibration)\n- Check system is stable and level\n- Resume and if problem persists call service to schedule a repair. | HD sees primary load cell for reservoir 1 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, - { 164 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM , "HD | Dial. Gen: UF Rate Check | Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system (shock/vibration)\n- Check system is stable and level\n- Resume and if problem persists call service to schedule a repair. | HD sees primary load cell for reservoir 2 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, - { 165 , ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_ALARM , "HD | Blood Stop: Blood Clot Risk | Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n-\n- NO RINSEBACK ALLOWED. | HD in treatment rinseback operation timeout | If the maximum time inrinseback operation has been exceeded. | "}, + { 163 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM , "HD | Dial. Gen: UF Rate Check | Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system (shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a repair. | HD sees primary load cell for reservoir 1 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, + { 164 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM , "HD | Dial. Gen: UF Rate Check | Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system \n(shock/vibration).\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a \nrepair. | HD sees primary load cell for reservoir 2 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, + { 165 , ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_ALARM , "HD | Blood Stop: Blood Clot Risk | Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. | HD in treatment rinseback operation timeout | If the maximum time inrinseback operation has been exceeded. | "}, { 166 , ALARM_ID_AVAILABLE_3 , " | | | | | "}, { 167 , ALARM_ID_HD_CARTRIDGE_DOOR_OPENED , "HD | Blood Stop: Door Open | Device door is open.\n - Close door and press RESUME button to continue. | HD cartridge door opened alarm | If the cartridge door is opened as reported by the FPGA. | "}, { 168 , ALARM_ID__AVAILABLE_70 , " | | | | | "}, { 169 , ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE , "DG | DG Fault: Dialysate Flow Sens Rg | Call service to schedule a repair. | Dialysate flow rate out of range | If the dialysate flow rate > 2.00 L/min. | "}, - { 170 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY , "HD | Heparin Syringe Empty | Heparin syringe empty.\n- Treatment will continue\n- If additional heparin is required to complete treatment admisinster manually | HD syringe empty alarm | If the syringe pump position is near empty position. | "}, + { 170 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY , "HD | Heparin Syringe Empty | Heparin syringe empty.\n- Treatment will continue\n- If additional heparin is required to complete treatment, \nadminister manually. | HD syringe empty alarm | If the syringe pump position is near empty position. | "}, { 171 , ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION , "HD | Process: Pre-Tx Hep. Pump Occ. | Heparin pump occlusion detected.\n- Check heparin line is unclamped.\n- Check there are no kinks in the heparin line. | HD syringe pump occlusion alarm | If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. | "}, - { 172 , ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM , "HD | Process: Pre-Tx Not Enough Hep. | Not enough heparin is available for treatment. Check if syringe pump is touching syringe plunger. If syringe pump is touching syringe plunger, end treatment. If syringe pump is not touching syringe plunger, reload syringe with enough heparin. | HD syringe pump not enough Heparin alarm | If the estimated syringe volume needed for treatment is not sufficient | "}, + { 172 , ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM , "HD | Process: Pre-Tx Not Enough Hep. | Not enough heparin is available for treatment.\n- Check if syringe pump is touching syringe plunger.\n- If syringe pump is touching syringe plunger, end treatment.\n- If syringe pump is not touching syringe plunger, reload \nsyringe with enough heparin. | HD syringe pump not enough Heparin alarm | If the estimated syringe volume needed for treatment is not sufficient | "}, { 173 , ALARM_ID_HD_RTC_CONFIG_ERROR , "HD | HD Fault: RTC Config | Call service to schedule a repair. | HD RTC configuration error | If the RTC configuration registers report an error. | "}, { 174 , ALARM_ID_HD_RTC_OR_TIMER_ACCURACY_FAILURE , "HD | HD POST: HD RTC Timer | Call service to schedule a repair. | HD RTC vs. Timer accuracy error | If RTC and timer are not aligned on 1 second elapsed time. | "}, { 175 , ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR , "HD | HD Fault: Pump Commutation Err | Call service to schedule a repair. | HD pump commutation error | If the set pump direction does not agree with the hall sensor direction. | "}, @@ -1339,7 +1339,7 @@ { 180 , ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE , "DG | DG Fault: RO Flow Sensor Range | Call service to schedule a repair. | RO flow rate out of range | If the RO flow rate > 2.00 L/min. | "}, { 181 , ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE , "DG | DG Fault: Tare Out of Range | Call service to schedule a repair. | DG load cells weight out of range for tare | If the load cell weight is out of range for tare. | "}, { 182 , ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD , "DG | DG POST: LC Cal | Call service to schedule a repair. | DG load cells invalid calibration | If the CRC of load cells calibration data does not match the calculated CRC. | "}, - { 183 , ALARM_ID_DG_INVALID_LOAD_CELL_VALUE , "DG | Cleaning: Invalid Resv Transfer | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG invalid load cell value | If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. | "}, + { 183 , ALARM_ID_DG_INVALID_LOAD_CELL_VALUE , "DG | Cleaning: Invalid Resv Transfer | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG invalid load cell value | If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. | "}, { 184 , ALARM_ID_DG_INLET_UV_REACTOR_NOT_HEALTHY , "DG | DG Fault: UVi Error | Call service to schedule a repair. | Inlet reactor not healthy | If the inlet reactor are not healthy. | "}, { 185 , ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE , "DG | DG Alarm (Low): Fan Speed | Fan failure.\n- Check that the fans inlets are not blocked or obstructed. | DG fan RPM out of range | If the fan RPM is out of Min/Max range. | "}, { 186 , ALARM_ID_DG_CONCENTRATE_PUMP_FAULT , "DG | DG Fault: CP1/CP2 Fault | Call service to schedule a repair. | DG concentrate pump fault | If the concentrate pumps fault is reported by the FPGA. | "}, @@ -1357,38 +1357,38 @@ { 198 , ALARM_ID_DG_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON , "DG | DG Fault: HP On With No Flow | Call service to schedule a repair. | DG primary heater while RO flow is low | If the RO flow is < 0.2 L/min for a certain period of time. | "}, { 199 , ALARM_ID_DG_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON , "DG | DG Fault: HR On With No Flow | Call service to schedule a repair. | DG trimmer heater while dialysate flow is low | If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. | "}, { 200 , ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE , "DG | DG Fault: Thermistor Range | Call service to schedule a repair. | DG thermistor or temperature sensor out of range | If thermistor > 80.0 C or < 0.0 C. | "}, - { 201 , ALARM_ID_HD_PRE_TREATMENT_WET_FLOW_TEST_FAILURE , "HD | Process: Pre-Tx Wet Self-Test | Self-test has detected a problem in the dialysate flow measurement or there is a leak in the dialysate in/out lines\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists terminate pre-treatment and restart with a new cartridge. | HD pre-treatment wet flow test failure | If the pre-treatment wet flow test has failed. | "}, - { 202 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE , "HD | Process: Pre-Tx Dry Self-Test | Self-test has detected blood circuit pressure cannot be relieved\n- Check PBo connection to HD for kinks\n -Check PBo transducer protector for occlusion\n- If problem persists install a new cartridge and PBo transducer protector. | HD pre-treatment dry pressure normal test failure | If the pre-treatment dry pressure normal test failed. | "}, + { 201 , ALARM_ID_HD_PRE_TREATMENT_WET_FLOW_TEST_FAILURE , "HD | Process: Pre-Tx Wet Self-Test | Self-test has detected a problem in the dialysate flow \nmeasurement or there is a leak in the dialysate in/out lines.\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. | HD pre-treatment wet flow test failure | If the pre-treatment wet flow test has failed. | "}, + { 202 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE , "HD | Process: Pre-Tx Dry Self-Test | Self-test has detected blood circuit pressure cannot be \nrelieved.\n- Check PBo connection to HD for kinks\n -Check PBo transducer protector for occlusion\n- If problem persists, install a new cartridge and \nPBo transducer protector. | HD pre-treatment dry pressure normal test failure | If the pre-treatment dry pressure normal test failed. | "}, { 203 , ALARM_ID_AVAILABLE_6 , " | | | | | "}, { 204 , ALARM_ID_DG_RO_REJECTION_RATIO_OUT_OF_RANGE , "DG | DG Fault: RO Rejection | Call service to schedule a repair. | RO rejection ratio out of range | If RO rejection ratio > 0.1 | "}, { 205 , ALARM_ID_DG_CONDUCTIVITY_SENSOR_FAULT , "DG | DG Fault: Cond Sensors Error | Call service to schedule a repair. | Conductivity sensor fault | If conductivity sensor reports 0x11 for more than 2 consecutive reads. | "}, { 206 , ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME , "DG | DG Fault: Reservoir Fill Timeout | Call service to schedule a repair. | DG dialysate fill runs out of time | If DG dialysate fill > 5 minutes. | "}, { 207 , ALARM_ID_DG_FLOW_METER_CHECK_FAILURE , "DG | DG Fault: FMP Flow Rate Error | Call service to schedule a repair. | DG flow meter check failure alarm | If integrated Volume To Load Cell Reading Percent > 0.1. | "}, { 208 , ALARM_ID_HD_VENOUS_LINE_OCCLUSION , "HD | Blood Stop: Venous Pres. High | Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. | HD venous line occlusion fault | If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. | "}, - { 209 , ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT , "DG | Cleaning: Flush Circ Error | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG drain circulation line timeout | If the integrated volume < 0.1 for more than 60 secs. | "}, + { 209 , ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT , "DG | Cleaning: Flush Circ Error | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG drain circulation line timeout | If the integrated volume < 0.1 for more than 60 secs. | "}, { 210 , ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED , "HD | HD Alarm (Low): Battery Fault | Call service to schedule a repair. | HD battery pack detected an error | If the battery pack status error is detected from the slave device I2C address 0x0B. | "}, { 211 , ALARM_ID_HD_BLOOD_SITTING_WARNING , "HD | Blood Stopped: Warning | Blood pump is stopped.\n- Resume treatment before countdown timer expires. | Blood sitting too long warning (> 4 min) | During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. | "}, { 212 , ALARM_ID_HD_END_OF_TREATMENT_ALARM , "HD | Blood Stopped: Warning | Blood pump is stopped.\n- Resume treatment before countdown timer expires. | Blood sitting too long warning (> 4 min) | During treatment end paused state, if the blood is in the line for more than 4 minutes. | "}, { 213 , ALARM_ID_HD_PRIME_COMPLETED_MEDIUM , "HD | Process: Pre-Tx Prime Complete | Prime of the cartridge is complete.\n- Proceed to connect to patient. | HD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) | "}, { 214 , ALARM_ID_AVAILABLE_30 , " | | | | | "}, { 215 , ALARM_ID_AVAILABLE_31 , " | | | | | "}, - { 216 , ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM , "HD | Process: End of Dialysis Tx | Dialysis Treatment has completed\n- Proceed to end treatment and rinseback | HD end treatment sub-mode timeout alarm | If the treatment ended successfully. | "}, + { 216 , ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM , "HD | Process: End of Dialysis Tx | Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. | HD end treatment sub-mode timeout alarm | If the treatment ended successfully. | "}, { 217 , ALARM_ID_AVAILABLE_32 , " | | | | | "}, { 218 , ALARM_ID_HD_SYRINGE_DETECTED , "HD | Process: Remove Syringe | Syringe not removed from machine.\n- Remove heparin syringe. | HD syringe detected alarm | If the syringe detection switch > 2.0 volts. | "}, - { 219 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED , "HD | Blood Stop: Syringe Not Detected | Heparin is needed for treatment but syringe has been removed from syringe pump.\n- Install Heparin syringe. | HD syringe pump syringe removed alarm | If the syringe detection switch < 2.0 volts. | "}, + { 219 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED , "HD | Blood Stop: Syringe Not Detected | Heparin is needed for treatment but syringe has been removed \nfrom syringe pump.\n- Install Heparin syringe. | HD syringe pump syringe removed alarm | If the syringe detection switch < 2.0 volts. | "}, { 220 , ALARM_ID_DG_THD_SENSORS_FPGA_FAULT , "DG | DG Fault: THD RTD Comm | Call service to schedule a repair. | DG THd sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 221 , ALARM_ID_HD_EMPTY_SALINE_BAG , "HD | Blood Stop: Saline Bag Empty | Saline bag appears empty.\n- Check saline bag and line and replace if saline bag is empty. | HD Saline bag is empty | If the measured arterial pressure < -300.0 mmHg for more than 250 msec. | "}, + { 221 , ALARM_ID_HD_EMPTY_SALINE_BAG , "HD | Blood Stop: Saline Bag Empty | Saline bag appears empty.\n- Check saline bag and line and replace if saline bag is \nempty. | HD Saline bag is empty | If the measured arterial pressure < -300.0 mmHg for more than 250 msec. | "}, { 222 , ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT , "HD | HD Fault: BP Occ Com | Call service to schedule a repair. | HD occlusion sensor FPGA fault | If the FPGA does not report fresh data or indicates an error for a certain period of time. | "}, { 223 , ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT , "HD | HD Fault: Art. Pres. Com | Call service to schedule a repair. | HD arterial sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, { 224 , ALARM_ID_HD_TREATMENT_STOPPED_BY_USER , "HD | Blood Stopped: Treatment Pause | Treatment paused by user. | Treatment stopped by user action - pressed stop button | If the user has pressed the stop button. | "}, - { 225 , ALARM_ID_HD_END_OF_TREATMENT_WARNING , "HD | Process: End of Treatment | Dialysis Treatment has completed\n- Proceed to end treatment and rinseback | HD end of treatment warning | If the treatment duration has exceeded the prescribed treatment time entered by the user. | "}, + { 225 , ALARM_ID_HD_END_OF_TREATMENT_WARNING , "HD | Process: End of Treatment | Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. | HD end of treatment warning | If the treatment duration has exceeded the prescribed treatment time entered by the user. | "}, { 226 , ALARM_ID_HD_PRIME_COMPLETED_LOW_PRIORITY , "HD | Process: Pre-Tx Priming Complete | Prime of the cartridge is complete.\n- Proceed to connect to patient. | HD prime completed low priority alarm | Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection | "}, { 227 , ALARM_ID_AVAILABLE_15 , " | | | | | "}, { 228 , ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT , "HD | Process: Pre-Tx Prime Timeout | Blood line prime with saline timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. | HD time out on prime sub-mode purge air state. | During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. | "}, { 229 , ALARM_ID_HD_PRIME_DIALYSATE_DIALYZER_TIME_OUT , "HD | Process: Pre-Tx Prime Timeout | Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. | HD prime dialysate dialyzer time out alarm | If the priming of the dialyzer dialysate flow path has exceeded the time limit. | "}, { 230 , ALARM_ID_HD_PRIME_DIALYSATE_BYPASS_TIME_OUT , "HD | Process: Pre-Tx Prime Timeout | Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. | HD prime dialysate bypass time out alarm | If the priming of the dialyzer bypass flow path has exceeded the time limit. | "}, - { 231 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "HD | Process: Pre-Tx Dry Self-Test | Self-test has detected a leak in the circuit\n- Check blood line connections\n- Check dialysate line connections\n- Check PBo and PBa transducer protectors for occlusion\n- Check clamps\n- If problem persists install a new cartridge | HD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, - { 232 , ALARM_ID_HD_PRE_TREATMENT_WET_LC_TEST_FAILURE , "HD | Process: Pre-Tx Wet Self-Test | Self-test has detected a problem in the load cell measurement or there is a leak in the dialysate in/out lines\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists terminate pre-treatment and restart with a new cartridge. | HD pre-treatment mode wet self-tests failure | If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState | "}, + { 231 , ALARM_ID_HD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "HD | Process: Pre-Tx Dry Self-Test | Self-test has detected a leak in the circuit\n- Check blood line connections\n- Check dialysate line connections\n- Check PBo and PBa transducer protectors for occlusion.\n- Check clamps.\n- If problem persists, install a new cartridge. | HD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, + { 232 , ALARM_ID_HD_PRE_TREATMENT_WET_LC_TEST_FAILURE , "HD | Process: Pre-Tx Wet Self-Test | Self-test has detected a problem in the load cell \nmeasurement or there is a leak in the dialysate in/out lines\n- The cartridge should be primed\n- Check cartridge installation onto machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. | HD pre-treatment mode wet self-tests failure | If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState | "}, { 233 , ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC , "DG | DG Fault: Baro Cal CRC | Call service to schedule a repair. | DG barometric pressure sensor bad coefficients CRC | If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. | "}, { 234 , ALARM_ID_DG_INLET_WATER_PRESSURE_IN_HIGH_RANGE , "DG | Dial. Gen: Inlet Water Pres High | Inlet water pressure is high.\n- Treatment is paused\n- Check the incoming water supply. | DG inlet water pressure is too high | If the inlet water pressure is > 80 psig. | "}, { 235 , ALARM_ID_HD_TREATMENT_STOPPED_AFTER_RINSEBACK , "HD | Process: Rinseback Complete | Rinseback has been completed. | HD in treatment stopped sub-mode after rinseback completed (no escalation) | If there is no active alarm and rinseback has completed. | "}, @@ -1397,30 +1397,30 @@ { 238 , ALARM_ID_HD_NO_CARTRIDGE_LOADED , "HD | Process: Pre-Tx Dry Self-Test | Self-test has detected cartridge is not installed.\n- Check that a cartridge is properly installed. | HD no cartridge loaded or installed improperly alarm | If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. | "}, { 239 , ALARM_ID_HD_CARTRIDGE_REMOVAL_FAILURE , "HD | Process: Post-Tx Remove Cartridge | Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. | HD fail to remove cartridge alarm | In the post treatment state, if the cartridge has not been removed. | "}, { 240 , ALARM_ID_DG_BICARB_CONDUCTIVITY_OUT_OF_RANGE , "DG | Dial. Gen: Bicarb Conc. Cond. | Bicarbonate concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. | Bicarb conductivity out of range during bicarb pump check alarm | If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. | "}, - { 241 , ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT , "DG | Cleaning: Resv Drain Timeout | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG reservoir drain time out | If the DG reservoir drain is not complete in 120 secs. | "}, - { 242 , ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT , "DG | Cleaning: Resv Fill Timeout | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG reservoir fill time out | If the DG reservoir fill is not complete in 120 secs. | "}, - { 243 , ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT , "DG | Cleaning: Fluid Loss Detected | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG reservoir leak time out | If the reservoir level is drifted by 100 mL for more than 30 seconds. | "}, - { 244 , ALARM_ID_DG_CLEANING_MODE_TEMP_SENSORS_DIFF_OUT_OF_RANGE , "DG | Cleaning: Temp Sensor Check | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG temperature sensors difference out of range | If any of the temperature sensors are out of the average of the sensors by more than 3 C. | "}, - { 245 , ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT , "DG | Cleaning: Heat Disinfect Temp TO | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG heat disinfect target temperature time out (could not reach to temperature) | If heating up to minimum temperature for heat disinfect failed after 4 hours. | "}, - { 246 , ALARM_ID_DG_CLEANING_MODE_COND_SENSORS_OUT_OF_RANGE , "DG | Cleaning: RO Water Cond Error | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG chemical disinfect inlet conducitivty out of range | If CPo and CD2 conductivity values are above 100 uS/cm. | "}, + { 241 , ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT , "DG | Cleaning: Resv Drain Timeout | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG reservoir drain time out | If the DG reservoir drain is not complete in 120 secs. | "}, + { 242 , ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT , "DG | Cleaning: Resv Fill Timeout | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG reservoir fill time out | If the DG reservoir fill is not complete in 120 secs. | "}, + { 243 , ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT , "DG | Cleaning: Fluid Loss Detected | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG reservoir leak time out | If the reservoir level is drifted by 100 mL for more than 30 seconds. | "}, + { 244 , ALARM_ID_DG_CLEANING_MODE_TEMP_SENSORS_DIFF_OUT_OF_RANGE , "DG | Cleaning: Temp Sensor Check | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG temperature sensors difference out of range | If any of the temperature sensors are out of the average of the sensors by more than 3 C. | "}, + { 245 , ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT , "DG | Cleaning: Heat Disinfect Temp TO | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG heat disinfect target temperature time out (could not reach to temperature) | If heating up to minimum temperature for heat disinfect failed after 4 hours. | "}, + { 246 , ALARM_ID_DG_CLEANING_MODE_COND_SENSORS_OUT_OF_RANGE , "DG | Cleaning: RO Water Cond Error | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG chemical disinfect inlet conducitivty out of range | If CPo and CD2 conductivity values are above 100 uS/cm. | "}, { 247 , ALARM_ID_AVAILABLE_7 , " | | | | | "}, - { 248 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE , "DG | Cleaning: Chem Disinfect Proc | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists call service to schedule a repair. | DG chemical disinfect out of target temperature | If TPo > 50 C after 1 second. | "}, - { 249 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE , "DG | Cleaning: Chem Disinf PAA Cond | Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the bicarb port.\n- Check chemical disinfectant bottle is not empty. | DG chemical disinfect out of target conductivity | If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. | "}, + { 248 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE , "DG | Cleaning: Chem Disinfect Proc | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG chemical disinfect out of target temperature | If TPo > 50 C after 1 second. | "}, + { 249 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE , "DG | Cleaning: Chem Disinf PAA Cond | Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. | DG chemical disinfect out of target conductivity | If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. | "}, { 250 , ALARM_ID_DG_CHEM_DISINFECT_INSERT_ACID , "DG | Cleaning: Chem Disinf Connect PAA | Connect chemical disinfectant bottle to bicarb port. | Insert the acid to start chemical disinfect | Start of chemical disinfect | "}, { 251 , ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC , "HD | HD POST: Sys Rec NVRAM CRC | Call service to schedule a repair. | HD invalid system record | If the calculated CRC of the HD system record image does not match the expected CRC. | "}, { 252 , ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC , "HD | HD POST: Serv Rec NVRAM CRC | Call service to schedule a repair. | HD invalid service record | If the calculated CRC of the HD service record image does not match the expected CRC. | "}, { 253 , ALARM_ID_DG_INVALID_SYSTEM_RECORD_CRC , "DG | DG POST: Sys Rec NVRAM CRC | Call service to schedule a repair. | DG invalid system record | If the calculated CRC of the DG system record image does not match the expected CRC. | "}, { 254 , ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC , "DG | DG POST: Serv Rec NVRAM CRC | Call service to schedule a repair. | DG invalid service record | If the calculated CRC of the DG service record image does not match the expected CRC. | "}, { 255 , ALARM_ID_HD_UI_COMPATIBILITY_ERROR , "HD | HD POST: UI Compatibility | Call service to schedule a repair. | HD and UI software builds are not compatible | If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. | "}, - { 256 , ALARM_ID_HD_DISINFECT_CHEM_FLUSH , "HD | Cleaning: Chem Flush In Progress | Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started untill process completes\n- Process may be terminated by pressing the HD stop button | DG flush started | If DG flush successfully started from the UI. | "}, + { 256 , ALARM_ID_HD_DISINFECT_CHEM_FLUSH , "HD | Cleaning: Chem Flush In Progress | Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. | DG flush started | If DG flush successfully started from the UI. | "}, { 257 , ALARM_ID_AVAILABLE_8 , " | | | | | "}, { 258 , ALARM_ID_AVAILABLE_9 , " | | | | | "}, { 259 , ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE , "HD | HD Fault: Temp Sensors Error | Call service to schedule a repair. | HD temperatures out of range | If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. | "}, { 260 , ALARM_ID_HD_UI_POST_FAILURE_SHASUM , "UI HD | HD POST: UI Image | Call service to schedule a repair. | UI POST Application Integrity (Sha256Sum) failure | If the UI POST reports a shasum faillure. | "}, { 261 , ALARM_ID_HD_UI_POST_FAILURE_CANBUS , "UI HD | HD POST: UI CAN Driver | Call service to schedule a repair. | UI POST CANBus failure | If the UI POST reports a can bus failure. | "}, { 262 , ALARM_ID_HD_UI_POST_FAILURE_DISPLAY , "UI HD | HD POST: UI Disp Driver | Call service to schedule a repair. | UI POST Display failure | If the UI POST reports a display failure. | "}, { 263 , ALARM_ID_HD_UI_POST_FAILURE_TOUCH , "UI HD | HD POST: UI Touch Driver | Call service to schedule a repair. | UI POST Touch failure | If the UI POST reports a touch failure. | "}, - { 264 , ALARM_ID_HD_UI_POST_FAILURE_SDCARD , "UI HD | HD POST: UI SD-Card | SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists | UI POST SD-Card failure | If the UI POST reports SD card failure. | "}, + { 264 , ALARM_ID_HD_UI_POST_FAILURE_SDCARD , "UI HD | HD POST: UI SD-Card | SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. | UI POST SD-Card failure | If the UI POST reports SD card failure. | "}, { 265 , ALARM_ID_HD_UI_POST_FAILURE_RTC , "UI HD | HD POST: UI RTC | Call service to schedule a repair. | UI POST RTC failure | If the UI POST reports RTC failure. | "}, { 266 , ALARM_ID_HD_UI_POST_FAILURE_WIFI , "UI HD | HD POST: UI WiFi | Call service to schedule a repair. | UI POST WiFi failure | If the UI POST reports WiFi failure. | "}, { 267 , ALARM_ID_HD_UI_POST_FAILURE_BLUETOOTH , "UI HD | HD POST: UI Bluetooth | Call service to schedule a repair. | UI POST Bluetooth failure | If the UI POST reports bluetooth failure. | "}, @@ -1447,9 +1447,9 @@ { 288 , ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE , "HD | DG Fault: LC Drift Check | Call service to schedule a repair. | HD load cell inconsistency | If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. | "}, { 289 , ALARM_ID_DG_DIALYSATE_CAP_NOT_IN_PROPER_POSITION , "DG | Cleaning: Dialysate Cap Removed | Dialysate cap removed from DG during cleaning.\n- Cleaning Process Terminated. | DG cap in wrong state | If the dialysate cap is not in the right position in the cleaning modes. | "}, { 290 , ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT , "HD | HD Fault: Ven. Pres. Com | Call service to schedule a repair. | HD venous sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, - { 291 , ALARM_ID_HD_DISINFECT_FLUSH , "HD | Cleaning: Flush In Progress | System flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started untill process completes\n- Process may be terminated by pressing the HD stop button | DG flush started | If DG flush successfully started from the UI. | "}, - { 292 , ALARM_ID_HD_DISINFECT_HEAT , "HD | Cleaning: Heat Disinf In Progress | Heat disinfect in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started untill process completes\n- Process may be terminated by pressing the HD stop button | DG heat disinfect started | If DG heat disinfect successfully started from the UI. | "}, - { 293 , ALARM_ID_HD_DISINFECT_CHEM , "HD | Cleaning: Chem Disinf In Progress | Chemical disinfect in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started untill process completes\n- Process may be terminated by pressing the HD stop button | DG chemical disinfect started | If DG chemical disinfect successfully started from the UI. | "}, + { 291 , ALARM_ID_HD_DISINFECT_FLUSH , "HD | Cleaning: Flush In Progress | System flush in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. | DG flush started | If DG flush successfully started from the UI. | "}, + { 292 , ALARM_ID_HD_DISINFECT_HEAT , "HD | Cleaning: Heat Disinf In Progress | Heat disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. | DG heat disinfect started | If DG heat disinfect successfully started from the UI. | "}, + { 293 , ALARM_ID_HD_DISINFECT_CHEM , "HD | Cleaning: Chem Disinf In Progress | Chemical disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. | DG chemical disinfect started | If DG chemical disinfect successfully started from the UI. | "}, { 294 , ALARM_ID_AVAILABLE_13 , " | | | | | "}, { 295 , ALARM_ID_HD_UI_POST_FAILURE_CLOUDSYNC , "UI HD | HD POST: UI CloudSync | Call service to schedule a repair. | UI cloud sync POST failure | If the UI POST reports cloud sync failure. | "}, { 296 , ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT , "DG | DG Fault: TDi RTD Comm | Call service to schedule a repair. | DG TDi sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, @@ -1463,21 +1463,21 @@ { 304 , ALARM_ID_HD_INVALID_USAGE_RECORD_CRC , "HD | HD POST: Usage Rec NVRAM CRC | Call service to schedule a repair. | HD invalid usage reocord | If the CRC of the HD usage record does not match the calculated CRC. | "}, { 305 , ALARM_ID_HD_CPU_RAM_ERROR , "HD | HD Fault: Processor RAM error | Call service to schedule a repair. | HD_CPU RAM error | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, { 306 , ALARM_ID_DG_CD1_CD2_CHECK_FAULT , "DG | DG Fault: CD1 CD2 Check | Call service to schedule a repair. | | CD1 and CD2 conductivity measurement during acid check is not within 5% of each other | "}, - { 307 , ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT , "HD | Blood Stop: Loss of AC Power | HD power has been lost. \n- Treatment must be terminated.\n-Check all electrical connections.\n- The display and CPU will run off of the battery until power returns.\n-If power is returned prior to blood stopped countdown elapses rinseback will be allowed. | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. | "}, + { 307 , ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT , "HD | Blood Stop: Loss of AC Power | HD power has been lost. \n- Treatment must be terminated.\n-Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns.\n-If power is returned prior to blood stopped countdown elapses, \nrinseback will be allowed. | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. | "}, { 308 , ALARM_ID_DG_CPI_COND_SENSOR_INVALID_CHAR , "DG | DG Fault: CPi Comm | Call service to schedule a repair. | DG CPi conductivity sensor transmits invalid characters | If the CPi conductivity sensor transmits invalid characters for a certain period of time. | "}, { 309 , ALARM_ID_DG_CPO_COND_SENSOR_INVALID_CHAR , "DG | DG Fault: CPo Comm | Call service to schedule a repair. | DG CPo conductivity sensor transmits invalid characters | If the CPo conductivity sensor transmits invalid characters for a certain period of time. | "}, { 310 , ALARM_ID_DG_CD1_COND_SENSOR_INVALID_CHAR , "DG | DG Fault: CD1 Comm | Call service to schedule a repair. | DG CD1 conductivity sensor transmits invalid characters | If the CD1 conductivity sensor transmits invalid characters for a certain period of time. | "}, { 311 , ALARM_ID_DG_CD2_COND_SENSOR_INVALID_CHAR , "DG | DG Fault: CD2 Comm | Call service to schedule a repair. | DG CD2 conductivity sensor transmits invalid characters | If the CD2 conductivity sensor transmits invalid characters for a certain period of time. | "}, { 312 , ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE , "HD | Dial. Gen: Dialysate In Flow | Dialysate in flow rate is low\n- Check dialysate line connection to the DG. | The dialysate inlet flow is too low. | If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. | "}, - { 313 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_LOW , "DG | Cleaning: Inlet Water Temp Low | Cleaning Mode Inlet water temperature too low.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. | DG cleaning mode inlet water temperature too low | If inlet water temperature is < 24 C. | "}, - { 314 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_HIGH , "DG | Cleaning: Inlet Water Press High | Cleaning Mode Inlet water pressure too high.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. | DG cleaning mode inlet water pressure too high | If inlet water temperature is > 80 psig. | "}, - { 315 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_LOW , "DG | Cleaning: Inlet Water Press Low | Cleaning Mode Inlet water pressure too low.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. | DG cleaning mode inlet water pressure too low | If inlet water pressure is < 14 psig. | "}, + { 313 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_LOW , "DG | Cleaning: Inlet Water Temp Low | Cleaning Mode Inlet water temperature too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. | DG cleaning mode inlet water temperature too low | If inlet water temperature is < 24 C. | "}, + { 314 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_HIGH , "DG | Cleaning: Inlet Water Press High | Cleaning Mode Inlet water pressure too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. | DG cleaning mode inlet water pressure too high | If inlet water temperature is > 80 psig. | "}, + { 315 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_LOW , "DG | Cleaning: Inlet Water Press Low | Cleaning Mode Inlet water pressure too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. | DG cleaning mode inlet water pressure too low | If inlet water pressure is < 14 psig. | "}, { 316 , ALARM_ID_DG_ACID_CONCENTRATE_PUMP_PARK_FAULT , "DG | DG Fault: Acid CP Pump Park | Call service to schedule a repair. | Acid concentrate pump did not park successfully | If the park took more than 1 second. | "}, { 317 , ALARM_ID_DG_BICARB_CONCENTRATE_PUMP_PARK_FAULT , "DG | DG Fault: Bicarb CP Pump Park | Call service to schedule a repair. | Bicarb concentrate pump did not park successfully | If the park took more than 1 second. | "}, { 318 , ALARM_ID_HD_AIR_PUMP_TIMEOUT , "HD | HD Fault: Air Pump Timeout | Unable to reduce fluid level.\n- Air trap may be overfilled.\n- Pressure transducer protector may be contaminated.\n- Treatment must be terminated without option for rinseback.\n- Call service to schedule a repair. | HD Air Pump operation timeout | If air pump has been turned on for greater than 10s. | "}, { 319 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_FLUSH_SAMPLE , "DG | Cleaning: Chem Residual Sampling | Collect chemical residuals water sample from drain line. | Collect sample to check the chemical disinfectant acid | Chemical disinfect flush sample alarm is prompted. | "}, { 320 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_SAMPLE_TIMEOUT , "DG | Cleaning: Chem Residual Timeout | During Chemical Disinfection Flush\n-user residual sampling timed out. | The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes | Chemical disinfect flush sample was not collected form more than 30 minutes. | "}, - { 321 , ALARM_ID_DG_OUTLET_UV_REACTOR_NOT_HEALTHY , "DG | DG Fault: UVo Error | Call service to schedule a repair. | Outlet reactor not healthy | If the outlet reactor are not healthy. | "}, + { 321 , ALARM_ID_DG_OUTLET_UV_REACTOR_NOT_HEALTHY , "DG | DG Fault: UVo Error | Call service to schedule a repair. | Outlet reactor not healthy | If the outlet reactor are not healthy. | "}, // Num Alarm ID Device Display Title Instructions Description Trigger Condition { 322 , ALARM_ID_HD_UI_POST_FAILURE_INVALID_YEAR , "UI HD | HD POST Alarm: UI Invalid Year | Call service to schedule a repair. | UI POST Invalid Year | If the UI POST reports a invalid year. | "}, // Num Alarm ID Device Display Title Instructions Description Trigger Condition };