Index: AlarmDefs.h =================================================================== diff -u -rc380d984b75193bbfd762e5eab843a50112f402f -r7dc1ed31f1a5a5f471fc2f35346921fc93a29a02 --- AlarmDefs.h (.../AlarmDefs.h) (revision c380d984b75193bbfd762e5eab843a50112f402f) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 7dc1ed31f1a5a5f471fc2f35346921fc93a29a02) @@ -1483,7 +1483,7 @@ { 316 , ALARM_ID_DG_ACID_CONCENTRATE_PUMP_PARK_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Acid CP Pump Park | Acid concentrate pump did not park successfully | If the park took more than 1 second. | "}, { 317 , ALARM_ID_DG_BICARB_CONCENTRATE_PUMP_PARK_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Bicarb CP Pump Park | Bicarb concentrate pump did not park successfully | If the park took more than 1 second. | "}, { 318 , ALARM_ID_HD_AIR_PUMP_TIMEOUT , "HD | Air Trap Control Error | A problem has been detected with the air trap level\ncontrol system.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. | HD Fault: Air Pump Timeout | HD Air Pump operation timeout | If air pump has been turned on for greater than 10s. | "}, - { 319 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_FLUSH_SAMPLE , "DG | Residual Chemical Sampling | A water sample is needed to confirm the \nabsence of residual chemical following disinfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. | Cleaning: Chem Residual Sampling | Collect sample to check the chemical disinfectant acid | Chemical disinfect flush sample alarm is prompted. | "}, + { 319 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_FLUSH_SAMPLE , "DG | Residual Chemical Sampling | A water sample is needed to confirm the \nabsence of residual chemical following disinfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. | Cleaning: Chem Residual Sampling | Collect sample to check the chemical disinfectant acid | Chemical disinfect flush sample alarm is prompted. | "}, { 320 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_SAMPLE_TIMEOUT , "DG | Chemical Residual Timeout | A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. | Cleaning: Chem Residual Timeout | The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes | Chemical disinfect flush sample was not collected form more than 30 minutes. | "}, { 321 , ALARM_ID_DG_OUTLET_UV_REACTOR_NOT_HEALTHY , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: UVo Error | Outlet reactor not healthy | If the outlet reactor are not healthy. | "}, { 322 , ALARM_ID_HD_UI_POST_FAILURE_INVALID_YEAR , "UI HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST Alarm: UI Invalid Year | UI POST Invalid Year | If the UI POST reports a invalid year. | "},