Index: AlarmDefs.h =================================================================== diff -u -rc8c433400ed045ceebadec7e7b1b121a1a857734 -r2ae479b6b85b6d6ec70547f54edfc32b59c3cef2 --- AlarmDefs.h (.../AlarmDefs.h) (revision c8c433400ed045ceebadec7e7b1b121a1a857734) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 2ae479b6b85b6d6ec70547f54edfc32b59c3cef2) @@ -8,7 +8,7 @@ * @file AlarmDefs.h * * @author (last) Dara Navaei -* @date (last) 31-May-2024 +* @date (last) 29-Oct-2024 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -1211,7 +1211,7 @@ { 44 , ALARM_ID_HD_UI_COMM_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UI Comm | User interface communication timeout | If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. | "}, { 45 , ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN CRC Com | Too many bad CRCs detected on received system messages | If too many bad messages CRCs are received within a set period of time | "}, { 46 , ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_UI , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: CAN ACK UI Com | System message that required acknowledgment was not acknowledged | If no retries left in the pending acknowlegement list. | "}, - { 47 , ALARM_ID_DG_MAX_RO_PUMP_PWM_EXCEEDED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Pump High | DG RO pump out of upper range | If the RO pump > 95% PWM for 12 secs. | "}, + { 47 , ALARM_ID_DG_MAX_RO_PUMP_PWM_EXCEEDED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Pump High | DG RO pump out of upper range | If the RO pump > 95% PWM for 30 secs. | "}, { 48 , ALARM_ID_HD_UF_VOLUME_ACCURACY_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: UF Rate Tare Err | Ultrafiltration volume accuracy error during treatment | If the UF volume (refUFvolume - measureVolume) > 100.0 mL. | "}, { 49 , ALARM_ID_HD_FPGA_COMM_TIMEOUT , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: FPGA Com | HD FPGA communication down for too long | If there are more than 3 comm errors in a 1 min period. | "}, { 50 , ALARM_ID_DG_VALVE_CONTROL_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Valve Control | DG FPGA not accepting commanded valve states | If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. | "}, @@ -1445,8 +1445,8 @@ { 278 , ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "HD | Dialysate Reservoir Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. | Dial. Gen: Reservoir Overfill | HD active reservoir weight out of range. | If the active reservoir weight is > 1900.0 g. | "}, { 279 , ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Drain | DG dialysate drain time out | If the DG dialysate drain exceeded 120 seconds. | "}, { 280 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Direction Fault | DG drain pump invalid direction | If the drain pump's direction is not 1 when draining. | "}, - { 281 , ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME , "DG | Acid Concentrate Low | The acid concentrate is low.\n- Replace acid concentrate container. | Dial. Gen: Acid Conc. Low | DG Acid concentration bottle low volume alarm | If the acid concentration bottle volume > 3240 (3340 - 100). | "}, - { 282 , ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME , "DG | Bicarb Concentrate Low | Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. | Dial. Gen: Bicarb Conc. Low | DG Bicarbonate concentration bottle low volume alarm | If the bicarbonate concentration bottle volume > 3680 (3780 - 100). | "}, + { 281 , ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME , "DG | Acid Concentrate Low | The acid concentrate is low.\n- Replace acid concentrate container. | Dial. Gen: Acid Conc. Low | DG Acid concentration bottle low volume alarm | If the acid concentration bottle volume <= 500 mL. | "}, + { 282 , ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME , "DG | Bicarb Concentrate Low | Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. | Dial. Gen: Bicarb Conc. Low | DG Bicarbonate concentration bottle low volume alarm | If the bicarbonate concentration bottle volume <= 500 mL. | "}, { 283 , ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: LC Range | DG load cells weight out of range | If the weight of the DG load cells > 4500.0 g or < 0.0 g. | "}, { 284 , ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Redundant LC Check | DG load cells primary/back up drift out of range | If redundant load cells deviate by more than 60g after tare for 5 seconds. | "}, { 285 , ALARM_ID_HD_RO_PERMEATE_SAMPLE , "HD | RO Permeate Sample in Progress | RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. | Cleaning: RO Permeate Sample In Progress | DG RO permeate sample started | If DG RO permeate sample successfully started from UI. | "}, @@ -1465,7 +1465,7 @@ { 298 , ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Pres Comm | DG barometric sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "}, { 299 , ALARM_ID_DG_INVALID_SERIAL_NUMBER , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Serial Number Check | DG invalid serial number | If the DG serial number is empty and does not have any characters. | "}, { 300 , ALARM_ID_HD_INVALID_SERIAL_NUMBER , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Serial Number Check | HD invalid serial number | If the HD serial number is empty and does not have any characters. | "}, - { 301 , ALARM_ID_DG_RO_PERMEATE_SAMPLE_REMOVE_DIA_CAP , "DG | Remove Dialysate Cap | Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. | Cleaning: RO Permeate Sample Connect Cap | Open dialysate cap to be able to collecte the dispensing fluid | RO permeate sample collection state | "}, + { 301 , ALARM_ID_DG_RO_PERMEATE_SAMPLE_REMOVE_DIA_CAP , "DG | Remove Dialysate Cap | - Remove the left dialysate cap.\n- Disinfect the bottom dialysate port using a 70% isopropyl\nalcohol wipe and allow to dry completely.\n- Using aseptic technique, connect extension tubing\nto bottom dialysate port.\n- Keep the right concentrate cap on the device. | Cleaning: RO Permeate Sample Connect Cap | Open dialysate cap to be able to collecte the dispensing fluid | RO permeate sample collection state | "}, { 302 , ALARM_ID_DG_CPU_RAM_ERROR , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Processor RAM error | DG_CPU RAM error | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, { 303 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Hall Commutation | DG drain pump direction FPGA fault | If the FPGA indicates an error for a certain period of time. | "}, { 304 , ALARM_ID_HD_INVALID_USAGE_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: Usage Rec NVRAM CRC | HD invalid usage reocord | If the CRC of the HD usage record does not match the calculated CRC. | "},