Index: AlarmDefs.h =================================================================== diff -u -rcb4e5be0ef260f2ceb1a97de18cce2bfb3830090 -r5aee21793442263257bdfdcefb06c5d03f0b0944 --- AlarmDefs.h (.../AlarmDefs.h) (revision cb4e5be0ef260f2ceb1a97de18cce2bfb3830090) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 5aee21793442263257bdfdcefb06c5d03f0b0944) @@ -1242,7 +1242,7 @@ { 81 , ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump OT | HD syringe pump over travel error | If the syringe pump position is deemed to have overtraveled | "}, { 82 , ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump DAC Err | HD syringe pump DAC write failure | If the syringe pump DAC write error is reported by the FPGA | "}, { 83 , ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump On Err | HD syringe pump is running while the BP is off | If the syringe pump is on while the blood pump is off | "}, - { 84 , ALARM_ID_DG_SET_RTC_YEAR_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: HD RTC Time | DG RTC set year is invalid | If DG RTC set year is < 2021. | "}, + { 84 , ALARM_ID_DG_SET_RTC_YEAR_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: DG RTC Time | DG RTC set year is invalid | If DG RTC set year is < 2021. | "}, { 85 , ALARM_ID_HD_PUMP_TRACK_LATCH_OPENED , "HD | Cartridge Lever is Open | A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press RESUME button to continue.\n\nBlood Stop: Cartridge Lever Open | HD pump track lever is not fully latched | If the pump track lever is detected as unlatched when it should be. | "}, { 86 , ALARM_ID_HD_SET_RTC_YEAR_INVALID , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: HD RTC Time | HD RTC set year is invalid | If HD RTC set year is < 2021. | "}, { 87 , ALARM_ID_DG_HEATING_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Heat Cal | DG heating parameters invalid calibration record | If the heating paramters' non-volatile record's CRC does not match the calculated CRC. | "}, @@ -1305,7 +1305,7 @@ { 144 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_COND_TOO_HIGH , "DG | Disinfection: Water Conductivity High | Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond High | DG cleaning mode inlet water conductivity out of high range | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode | "}, { 145 , ALARM_ID_DG_ACID_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Error During Prime | Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range.\n\nDial. Gen: Acid Conc. Cond. | DG Concentrate conductivity after adding acid out of range alarm | If the conductivity value for the acid concentrate is out of range for more than 45 secs. | "}, { 146 , ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: DG RTC Timer | DG RTC vs. Timer accuracy error | If RTC and timer are not aligned on 1 second elapsed time. | "}, - { 147 , ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT , "DG | Dialysate Generation Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \nthe dialysate is ready for use. \n\nDial. Gen: Dialysate not ready | DG dialysate generation conductivity fault alarm | If dialysate conductivity is not with 5% of the TCD | "}, + { 147 , ALARM_ID_DG_CREATING_DIALYSATE_PLEASE_WAIT , "DG | Dialysate Generation Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. \n\nDial. Gen: Dialysate not ready | DG dialysate generation conductivity fault alarm | If dialysate conductivity is not with 5% of the TCD | "}, { 148 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_HIGH , "DG | Disinfection: Water Temperature High | Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp High | Cleaning Mode Inlet water temperature too high | Inlet temperature > 37 C. | "}, { 149 , ALARM_ID_HD_WATCHDOG_EXPIRED , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: WD Error | HD watchdog expired | If the watchdog signal is low while not in POST. | "}, { 150 , ALARM_ID_DG_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE , "DG | Incoming Water Conductivity High | Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nRESUME button will be available.\n- Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond High | Inlet water conductivity too high | Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). | "},