Index: AlarmDefs.h =================================================================== diff -u -rd69f391038951ee0157a3159fe38dc7760294736 -rdc19a471519f238d7c1991ae062c58a685cbf82f --- AlarmDefs.h (.../AlarmDefs.h) (revision d69f391038951ee0157a3159fe38dc7760294736) +++ AlarmDefs.h (.../AlarmDefs.h) (revision dc19a471519f238d7c1991ae062c58a685cbf82f) @@ -1451,7 +1451,7 @@ { 313 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_TEMP_TOO_LOW , "DG | Cleaning Mode Inlet Water Temperature Too Low | Inlet water temperature too low.\n- Check the incoming water for issues. | DG cleaning mode inlet water temperature too low | If inlet water temperature is < 24 C. | "}, { 314 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_HIGH , "DG | Cleaning Mode Inlet Water Pressure Too High | Inlet water pressure too high.\n- Check the incoming water for issues. | DG cleaning mode inlet water pressure too high | If inlet water temperature is > 80 psig. | "}, { 315 , ALARM_ID_DG_CLEANING_MODE_INLET_WATER_PRESSURE_TOO_LOW , "DG | Cleaning Mode Inlet Water Pressure Too Low | Inlet water pressure too low.\n- Check the incoming water for issues. | DG cleaning mode inlet water pressure too low | If inlet water pressure is < 14 psig. | "}, - { 315 , ALARM_ID_HD_AIR_PUMP_TIMEOUT , "HD | Air Pump Operation Timeout | Call service to schedule a repair. | HD Air Pump operation timeout | If air pump has been turned on for greater than 10s. | "}, + { 316 , ALARM_ID_HD_AIR_PUMP_TIMEOUT , "HD | Air Pump Operation Timeout | Call service to schedule a repair. | HD Air Pump operation timeout | If air pump has been turned on for greater than 10s. | "}, }; #endif