Index: AlarmDefs.h =================================================================== diff -u -rd705e032a38453dc02d2e448fb931566d8f8a0a0 -ra6cd968d7eda4dfe5f4c6980810166b9881d7657 --- AlarmDefs.h (.../AlarmDefs.h) (revision d705e032a38453dc02d2e448fb931566d8f8a0a0) +++ AlarmDefs.h (.../AlarmDefs.h) (revision a6cd968d7eda4dfe5f4c6980810166b9881d7657) @@ -1374,7 +1374,7 @@ { 209 , ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Flush Circ Error | DG drain circulation line timeout | If the integrated volume < 0.1 for more than 60 secs. | "}, { 210 , ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Battery Fault | HD battery pack detected an error | If the battery pack status error is detected from the slave device I2C address 0x0B. | "}, { 211 , ALARM_ID_HD_BLOOD_SITTING_WARNING , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. | "}, - { 212 , ALARM_ID_HD_END_OF_TREATMENT_ALARM , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OKd before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment end paused state, if the blood is in the line for more than 4 minutes. | "}, + { 212 , ALARM_ID_HD_END_OF_TREATMENT_ALARM , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment end paused state, if the blood is in the line for more than 4 minutes. | "}, { 213 , ALARM_ID_HD_PRIME_COMPLETED_MEDIUM , "HD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre- Tx Prime Complete | HD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) | "}, { 214 , ALARM_ID_AVAILABLE_30 , " | | | | | | "}, { 215 , ALARM_ID_AVAILABLE_31 , " | | | | | | "},