Index: AlarmDefs.h =================================================================== diff -u -rdef95827e674b2b257211798f16400546ce44dff -re229e4ea37caf0f9a5e86f5e15af262bc90b8e67 --- AlarmDefs.h (.../AlarmDefs.h) (revision def95827e674b2b257211798f16400546ce44dff) +++ AlarmDefs.h (.../AlarmDefs.h) (revision e229e4ea37caf0f9a5e86f5e15af262bc90b8e67) @@ -1291,10 +1291,10 @@ { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. | Blood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, - { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access issues, \nif applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access issues, \nif applicable.\n- Check air trap and venous pressure lines for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, { 129 , ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "HD | Blood Leak: Bypass | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. | Blood Stop: Dialyzer Blood Leak Bypass | HD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "}, - { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap and venous pressure line for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 131 , ALARM_ID_HD_ARTERIAL_PRESSURE_LOW , "HD | Low Arterial Pressure | Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. Low | Arterial pressure too low during treatment | If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 132 , ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH , "HD | High Arterial Pressure | Arterial pressure is high.\n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. High | Arterial pressure too high during treatment | If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 133 , ALARM_ID_DG_FLUID_LEAK_DETECTED , "DG | Dialysate Device Fluid Leak | Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Fluid Leak in DG | DG fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, @@ -1372,16 +1372,16 @@ { 205 , ALARM_ID_DG_CONDUCTIVITY_SENSOR_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Cond Sensors Error | Conductivity sensor fault | If conductivity sensor reports 0x11 for more than 2 consecutive reads. | "}, { 206 , ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Fill Timeout | DG dialysate fill runs out of time | If DG dialysate fill > 5 minutes. | "}, { 207 , ALARM_ID_DG_FLOW_METER_CHECK_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: FMP Flow Rate Error | DG flow meter check failure alarm | If integrated Volume To Load Cell Reading Percent > 0.1. | "}, - { 208 , ALARM_ID_HD_VENOUS_LINE_OCCLUSION , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Venous Pres. High | HD venous line occlusion fault | If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. | "}, + { 208 , ALARM_ID_HD_VENOUS_LINE_OCCLUSION , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap and venous pressure line for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. High | HD venous line occlusion fault | If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. | "}, { 209 , ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT , "DG | Disinfection Interrupted | Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. | Cleaning: Flush Circ Error | DG drain circulation line timeout | If the integrated volume < 0.1 for more than 60 secs. | "}, { 210 , ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Battery Fault | HD battery pack detected an error | If the battery pack status error is detected from the slave device I2C address 0x0B. | "}, { 211 , ALARM_ID_HD_BLOOD_SITTING_WARNING , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. | "}, { 212 , ALARM_ID_HD_END_OF_TREATMENT_ALARM , "HD | Approaching MAX Blood Pump Stop Time | The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. | Blood Stopped: Warning | Blood sitting too long warning (> 4 min) | During treatment end paused state, if the blood is in the line for more than 4 minutes. | "}, { 213 , ALARM_ID_HD_PRIME_COMPLETED_MEDIUM , "HD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre- Tx Prime Complete | HD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) | "}, { 214 , ALARM_ID_DG_CONCENTRATE_PUMPS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conc Pump Cal | DG concentrate pump calibration record | If the non-volatile data is invalid raise the alarm. | "}, - { 215 , ALARM_ID_HD_AIR_TRAP_UPPER_LEVEL_FLUID_TIME_OUT , "HD | Air Trap Overfill | Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment.| Air Trap: Air Trap Overfill | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 60 seconds. | "}, + { 215 , ALARM_ID_HD_AIR_TRAP_UPPER_LEVEL_FLUID_TIME_OUT , "HD | Air Trap Overfill | Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment.| Air Trap: Air Trap Overfill | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 60 seconds. | "}, { 216 , ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM , "HD | Treatment Complete | Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. | Process: End of Dialysis Tx | HD end treatment sub-mode timeout alarm | If the treatment ended successfully. | "}, - { 217 , ALARM_ID_HD_AIR_TRAP_ACTIVITY_TOO_FREQUENT , "HD | Air Trap Overfill | Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment.| Air Trap: Air Trap Overfill | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 60 seconds. | "}, + { 217 , ALARM_ID_HD_AIR_TRAP_ACTIVITY_TOO_FREQUENT , "HD | Multiple Air Trap Events | Multiple Air Trap Events have been detected. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment.| Air Trap: Air Trap Overfill | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 60 seconds. | "}, { 218 , ALARM_ID_HD_SYRINGE_DETECTED , "HD | Syringe Detected | Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. | Process: Remove Syringe | HD syringe detected alarm | If the syringe detection switch > 2.0 volts. | "}, { 219 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED , "HD | Blood Stopped: Syringe Not Detected | There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. | Blood Stop: Syringe Not Detected | HD syringe pump syringe removed alarm | If the syringe detection switch < 2.0 volts. | "}, { 220 , ALARM_ID_DG_THD_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: THD RTD Comm | DG THd sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "},