Index: AlarmDefs.h =================================================================== diff -u -re9beb6b7cbe88dfd9c38ef0ca9f4724ca28916a3 -rcdde55b1e1af5712de18cc908b2b2f2ab6588527 --- AlarmDefs.h (.../AlarmDefs.h) (revision e9beb6b7cbe88dfd9c38ef0ca9f4724ca28916a3) +++ AlarmDefs.h (.../AlarmDefs.h) (revision cdde55b1e1af5712de18cc908b2b2f2ab6588527) @@ -1396,7 +1396,7 @@ { 235 , ALARM_ID_HD_TREATMENT_STOPPED_AFTER_RINSEBACK , "HD | Rinseback Complete | Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press OK to continue. \n\nProcess: Rinseback Complete | HD in treatment stopped sub-mode after rinseback completed (no escalation) | If there is no active alarm and rinseback has completed. | "}, { 236 , ALARM_ID_HD_INSTALL_NEW_CARTRIDGE , "HD | Self Test: Fluid Present | A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume.\n\nProcess: Pre- Tx Dry Self- Test | HD needs new cartridge to be installed | If the air trap sensor reported AIR and bubble sensor detected bubbles. | "}, { 237 , ALARM_ID_HD_PRIME_SALINE_DIALYZER_TIME_OUT , "HD | Priming Error | A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre- Tx Prime Timeout | HD prime saline dialyzer time out alarm | If timeout, just transition to HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE | "}, - { 238 , ALARM_ID_HD_NO_CARTRIDGE_LOADED , "HD | Cartridge Not Dectected | A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed.\n\nProcess: Pre- Tx Dry Self- Test | HD no cartridge loaded or installed improperly alarm | If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. | "}, + { 238 , ALARM_ID_HD_NO_CARTRIDGE_LOADED , "HD | Cartridge Not Detected | A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed.\n\nProcess: Pre- Tx Dry Self- Test | HD no cartridge loaded or installed improperly alarm | If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. | "}, { 239 , ALARM_ID_HD_CARTRIDGE_REMOVAL_FAILURE , "HD | Cartridge Still Detected | A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed.\n\nProcess: Post- Tx Remove Cartridge | HD fail to remove cartridge alarm | In the post treatment state, if the cartridge has not been removed. | "}, { 240 , ALARM_ID_DG_BICARB_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Error During Prime | Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range.\n\nDial. Gen: Bicarb Conc. Cond. | Bicarb conductivity out of range during bicarb pump check alarm | If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. | "}, { 241 , ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT , "DG | Disinfection Interrupted | Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Resv Drain Timeout | DG reservoir drain time out | If the DG reservoir drain is not complete in 120 secs. | "},