Index: AlarmDefs.h =================================================================== diff -u -rec1633c6cfefe3d69891852e2470b03ac4b1cef7 -rc47f83bfd98de87f0da34b9853160c6aef9b34f2 --- AlarmDefs.h (.../AlarmDefs.h) (revision ec1633c6cfefe3d69891852e2470b03ac4b1cef7) +++ AlarmDefs.h (.../AlarmDefs.h) (revision c47f83bfd98de87f0da34b9853160c6aef9b34f2) @@ -1381,7 +1381,7 @@ { 214 , ALARM_ID_DG_CONCENTRATE_PUMPS_INVALID_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conc Pump Cal | DG concentrate pump calibration record | If the non-volatile data is invalid raise the alarm. | "}, { 215 , ALARM_ID_HD_AIR_TRAP_UPPER_LEVEL_FLUID_TIME_OUT , "HD | Air Trap Overfill | Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducers are secured.\n- Inspect venous transducer for moisture; change if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. | Air Trap: Air Trap Overfill | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 60 seconds. | "}, { 216 , ALARM_ID_HD_END_TREATMENT_TIMEOUT_ALARM , "HD | Treatment Complete | Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. | Process: End of Dialysis Tx | HD end treatment sub-mode timeout alarm | If the treatment ended successfully. | "}, - { 217 , ALARM_ID_HD_AIR_TRAP_ACTIVITY_TOO_FREQUENT , "HD | Multiple Air Trap Events | Multiple air trap events have been detected. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducers are secured.\n- Inspect venous transducer for moisture; change if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. | Air Trap: Air Trap Overfill | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 5 times in 5 minutes. | "}, + { 217 , ALARM_ID_HD_AIR_TRAP_ACTIVITY_TOO_FREQUENT , "HD | Multiple Air Trap Events | Multiple air trap events have been detected. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducers are secured.\n- Inspect venous transducer for moisture; change if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. | Air Trap: Multiple Air Trap Events | HD air trap upper level fluid detection for too long alarm | If air trap upper level is detecting fluid for more than 5 times in 5 minutes. | "}, { 218 , ALARM_ID_HD_SYRINGE_DETECTED , "HD | Syringe Detected | Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. | Process: Remove Syringe | HD syringe detected alarm | If the syringe detection switch > 2.0 volts. | "}, { 219 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED , "HD | Blood Stopped: Syringe Not Detected | There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. | Blood Stop: Syringe Not Detected | HD syringe pump syringe removed alarm | If the syringe detection switch < 2.0 volts. | "}, { 220 , ALARM_ID_DG_THD_SENSORS_FPGA_FAULT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: THD RTD Comm | DG THd sensor FPGA fault | If the FPGA does not report fresh data for a certain period of time | "},