Index: AlarmDefs.h =================================================================== diff -u -r6239d92af0c24bc2f95033723b9c8c58f0213c64 -r08f22add5d45eaba518dd1b85a2d8cf0a72d0ec5 --- AlarmDefs.h (.../AlarmDefs.h) (revision 6239d92af0c24bc2f95033723b9c8c58f0213c64) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 08f22add5d45eaba518dd1b85a2d8cf0a72d0ec5) @@ -87,7 +87,7 @@ ALARM_ID_VALVE_CONTROL_FAILURE = 51, ///< FPGA not accepting commanded valve states ALARM_ID_BLOOD_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 52, ///< Blood pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 53, ///< Dialysate inlet pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed - ALARM_ID____AVAILABLE___ = 54, ///< This alarm ID is available - currently unused + ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT = 54, ///< HD time out on prime sub-mode purge air state. ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK = 55, ///< Blood pump failed motor speed check. Measured vs. commanded ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK = 56, ///< Dialysate inlet pump failed motor speed check. Measured vs. commanded ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK = 57, ///< Dialysate outlet pump failed motor speed check. Measured vs. commanded @@ -287,7 +287,7 @@ { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_VALVE_CONTROL_FAILURE { ALARM_PRIORITY_HIGH, 3, ALM_ESC_4_MIN, ALARM_ID_BLOOD_SITTING_WARNING, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_BLOOD_PUMP_FLOW_VS_MOTOR_SPEED_CHECK { ALARM_PRIORITY_HIGH, 3, ALM_ESC_4_MIN, ALARM_ID_BLOOD_SITTING_WARNING, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // *** Available *** + { ALARM_PRIORITY_HIGH, 5, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK