Index: AlarmDefs.h =================================================================== diff -u -r3a3ee26761a21ee1446faf59afca978a1a6fcf79 -r1985b08b231747f89d462873eae5ecdf3da35b1b --- AlarmDefs.h (.../AlarmDefs.h) (revision 3a3ee26761a21ee1446faf59afca978a1a6fcf79) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 1985b08b231747f89d462873eae5ecdf3da35b1b) @@ -98,7 +98,7 @@ ALARM_ID_HD_VALVE_NOT_FUNCTIONAL = 60, ///< HD valve not functional ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE = 61, ///< HD valve current out of range ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE = 62, ///< HD valve position out of target - ALARM_ID_AVAILABLE_20 = 63, ///< Alarm ID available + ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT = 63, ///< HD syringe pump prime timeout ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE = 64, ///< DG barometric pressure sensor out of range ALARM_ID_HD_DG_COMMAND_INVALID_PARAMETER_FAULT = 65, ///< HD DG requests DG command with invalid parameter fault ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE = 66, ///< HD blood leak sensor set point set failure @@ -543,7 +543,7 @@ { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_NOT_FUNCTIONAL }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE }, - { ALARM_PRIORITY_LOW, 999, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_20 }, + { ALARM_PRIORITY_LOW, 716, FALSE, FALSE, TRUE , TRUE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 111, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_DG_COMMAND_INVALID_PARAMETER_FAULT }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, @@ -1093,6 +1093,7 @@ { 710, ALARM_ID_HD_NO_CARTRIDGE_LOADED }, { 715, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM }, { 716, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION }, + { 716, ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT }, { 720, ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT }, { 720, ALARM_ID_HD_PRIME_DIALYSATE_DIALYZER_TIME_OUT }, { 720, ALARM_ID_HD_PRIME_DIALYSATE_BYPASS_TIME_OUT }, @@ -1142,7 +1143,6 @@ { 999, ALARM_ID_AVAILABLE_30 }, { 999, ALARM_ID_AVAILABLE_31 }, { 999, ALARM_ID_AVAILABLE_32 }, - { 999, ALARM_ID_AVAILABLE_20 }, { 999, ALARM_ID_NO_ALARM }, }; @@ -1227,7 +1227,7 @@ { 60 , ALARM_ID_HD_VALVE_NOT_FUNCTIONAL , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Control | HD valve not functional | If the valve bit masks are not as expected in different valve control mode. | "}, { 61 , ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Current | HD valve current out of range | If the valve current is over the threshold of 1.0 A for 1000 msec. | "}, { 62 , ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Valve Position | HD valve position out of target | If the current position has deviated from the expected position for more than 1 second. | "}, - { 63 , ALARM_ID_AVAILABLE_20 , " | | | | | | "}, + { 63 , ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT , "HD | Heparin Syringe Pump Prime Timeout | A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. | Process: Pre- Tx Hep. Pump Timeout. | HD syringe pump timeout alarm | If the syringe pump priming > 5 seconds | "}, { 64 , ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Baro Range | Barometric pressure is out of range | If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. | "}, { 65 , ALARM_ID_HD_DG_COMMAND_INVALID_PARAMETER_FAULT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Reject) | HD requests DG command with invalid parameter fault | If the latest DG command response indicated invalid parameter | "}, { 66 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Config Set Point | HD blood leak sensor set point set failure | If the set point of the blood leak sensor does not match the set point in the calibration record. | "},