Index: AlarmDefs.h =================================================================== diff -u -rdc7bbe902bd7b646fd636e041790915a24fbb046 -r1d60fe2157e679b1c2e545a7f55c6e25dcd2379e --- AlarmDefs.h (.../AlarmDefs.h) (revision dc7bbe902bd7b646fd636e041790915a24fbb046) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 1d60fe2157e679b1c2e545a7f55c6e25dcd2379e) @@ -1317,7 +1317,7 @@ { 156 , ALARM_ID_HD_DG_NEW_OPERATION_MODE_MESSAGE_NOT_RECEIVE , "HD | Communication Error | A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Op Mode) | HD No DG op mode message receive at least once every 2 seconds | No DG op mode for 2 seconds. | "}, { 157 , ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT , "DG | Chemical Disinfectant Not Detected | Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the concentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Prime | Chemical disinfect priming the acid line timeout | If the disinfectant acid line is not primed within a certain period of time. | "}, { 158 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_LOW_RANGE , "DG | Incoming Water Temperature Low | Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the temperature is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp Low | Inlet water temperature low | Inlet temperature < 24 °C for 5 seconds. | "}, - { 159 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_REMOVE_ACID , "DG | Remove Chemical Disinfectant | The chemical disinfectant is no longer needed. \n-Disconnect chemical disinfectant from the device. \n-Place the concentrate connection cap on the device.\n\nCleaning: Chem Disinf Connect Cap | Close concentrate line be able to flush it | Start of chemical disinfect flush. | "}, + { 159 , ALARM_ID_DG_CHEM_DISINFECT_FLUSH_REMOVE_ACID , "DG | Remove Chemical Disinfectant | The chemical disinfectant is no longer needed. \n-Disconnect chemical disinfectant from the device. \n-Place the concentrate connection cap on the device.\n-Keep dialysate cap on the device.\n\nCleaning: Chem Disinf Connect Cap | Close concentrate line be able to flush it | Start of chemical disinfect flush. | "}, { 160 , ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Rotor Max Speed | Blood pump rotor speed too high | If blood pump motor speed exceeded 100.0 RPM. | "}, { 161 , ALARM_ID_DG_COMM_TOO_MANY_BAD_CRCS , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HD-DG Comm | Too many bad CRCs detected on received system messages | If >= 5 bad CRCs detected on received CAN messages within 10 minutes. | "}, { 162 , ALARM_ID_DG_FPGA_CLOCK_SPEED_CHECK_FAILURE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Proc-FPGA Clock | FPGA clock speed failure | If the FPGA clock speed is relatively close to the processor clock for a certain period of time. | "}, @@ -1408,7 +1408,7 @@ { 247 , ALARM_ID_AVAILABLE_7 , " | | \n\n | | | "}, { 248 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_OUT_OF_RANGE , "DG | Disinfection Interrupted | Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinfect Proc | DG chemical disinfect out of target temperature | If TPo > 50 C after 1 second. | "}, { 249 , ALARM_ID_DG_CHEM_DISINFECT_TARGET_COND_OUT_OF_RANGE , "DG | Chemical Disinfectant Not Detected | Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the concentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Cond | DG chemical disinfect out of target conductivity | If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. | "}, - { 250 , ALARM_ID_DG_CHEM_DISINFECT_INSERT_ACID , "DG | Connect Chemical Disinfectant | Connect chemical disinfectant to the device. \n-Place disinfectant straw in the disinfectant bottle. \n-Attach straw to concentrate connection and ensure connection is secure. \n-Press OK to continue. \n\nCleaning: Chem Disinf Connect PAA | Insert the acid to start chemical disinfect | Start of chemical disinfect | "}, + { 250 , ALARM_ID_DG_CHEM_DISINFECT_INSERT_ACID , "DG | Connect Chemical Disinfectant | Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n-Keep dialysate cap on the device.\n-Place disinfectant straw in the disinfectant bottle. \n-Attach straw to concentrate connection and ensure connection is secure. \n-Press OK to continue.\n\nCleaning: Chem Disinf Connect PAA | Insert the acid to start chemical disinfect | Start of chemical disinfect | "}, { 251 , ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Sys Rec NVRAM CRC | HD invalid system record | If the calculated CRC of the HD system record image does not match the expected CRC. | "}, { 252 , ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Serv Rec NVRAM CRC | HD invalid service record | If the calculated CRC of the HD service record image does not match the expected CRC. | "}, { 253 , ALARM_ID_DG_INVALID_SYSTEM_RECORD_CRC , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Sys Rec NVRAM CRC | DG invalid system record | If the calculated CRC of the DG system record image does not match the expected CRC. | "},