Index: AlarmDefs.h =================================================================== diff -u -rd634fa44a46a100dd9754b9507324564166a6667 -r27411437b5d50269ceabec90234b998fd696a098 --- AlarmDefs.h (.../AlarmDefs.h) (revision d634fa44a46a100dd9754b9507324564166a6667) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 27411437b5d50269ceabec90234b998fd696a098) @@ -1328,8 +1328,8 @@ { 168 , ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT , "DG | Dialysate Recirculation Error | Dialysate recirculation error. | HD active reservoir depletion timeout. | If the active reservoir's depletion time has elapsed. | "}, { 169 , ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE , "DG | Dialysate Generation Error | Call service to schedule a repair. | Dialysate flow rate out of range | If the dialysate flow rate > 2.00 L/min. | "}, { 170 , ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY , "HD | Syringe Empty | Heparin syringe empty.\n- Option to resume treatment without heparin. | HD syringe empty alarm | If the syringe pump position is near empty position. | "}, - { 171 , ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION , "HD | Heparin Occlusion | Heparin pump occlusion detected. Check heparin line is unclamped and there are no kinks in the heparin line. | HD syringe pump occlusion alarm | If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. | "}, - { 172 , ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM , "HD | Not Enough Heparin | Not enough heparin is available for treatment. Check if syringe pump is touching syringe plunger. If syringe pump is touching syringe plunger, end treatment. If syringe pump is not touching syringe plunger, reload syringe with enough heparin. | HD syringe pump not enough Heparin alarm | If the estimated syringe volume needed for treatment is not sufficient | "}, + { 171 , ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION , "HD | Heparin Occlusion | Heparin pump occlusion detected.\nCheck heparin line is unclamped and\nthere are no kinks in the heparin line. | HD syringe pump occlusion alarm | If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. | "}, + { 172 , ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM , "HD | Not Enough Heparin | Not enough heparin is available for treatment.\nCheck if syringe pump is touching syringe plunger.\nIf syringe pump is touching syringe plunger, end treatment.\nIf syringe pump is not touching syringe plunger, reload syringe with enough heparin. | HD syringe pump not enough Heparin alarm | If the estimated syringe volume needed for treatment is not sufficient | "}, { 173 , ALARM_ID_HD_RTC_CONFIG_ERROR , "HD | RTC Error | Call service to schedule a repair. | HD RTC configuration error | If the RTC configuration registers report an error. | "}, { 174 , ALARM_ID_HD_RTC_OR_TIMER_ACCURACY_FAILURE , "HD | Self-Test Error | Call service to schedule a repair. | HD RTC vs. Timer accuracy error | If RTC and timer are not aligned on 1 second elapsed time. | "}, { 175 , ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR , "HD | Blood Pump Error | Blood pump failure. \n- Check the pump rotor for secure connection. | HD pump direction status error | If the set pump direction does not agree with the hall sensor direction. | "},