Index: AlarmDefs.h =================================================================== diff -u -rd96e30bdbb76c099bd8253ea4a95df215d5eec7e -r44374c9ae60ba0c9f5e3799d99a789ed187945a9 --- AlarmDefs.h (.../AlarmDefs.h) (revision d96e30bdbb76c099bd8253ea4a95df215d5eec7e) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 44374c9ae60ba0c9f5e3799d99a789ed187945a9) @@ -1261,7 +1261,7 @@ { 94 , ALARM_ID_DG_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Pressure Sensor Range | DG pressure sensor out of range | If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. | "}, { 95 , ALARM_ID_DG_WATCHDOG_EXPIRED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: WD Error | DG watchdog expired | If the watchdog signal is low while not in POST. | "}, { 96 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE , "DG | Incoming Water Temperature High | Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Temp High | DG inlet water temperature out of range | If the inlet water temperature is > 37 C. | "}, - { 97 , ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Out of Range | Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If bubbles are observed in either\nof the concentrate lines replace it with a new concentrate container.\n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Dialysate Cond. | DG fill conductivity out of range | If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. | "}, + { 97 , ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Out of Range | Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates are securely\nconnected.\n- If bubbles are observed in either of the concentrate lines\nreplace it with a new concentrate container.\n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Dialysate Cond. | DG fill conductivity out of range | If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. | "}, { 98 , ALARM_ID_HD_BATTERY_COMM_FAULT , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Battery Com | HD battery or battery charger communication fault | If battery I2C communication fault persistence count > 5. | "}, { 99 , ALARM_ID_HD_SYRINGE_PUMP_STALL , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Stall | HD syringe pump stall alarm | If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. | "}, { 100 , ALARM_ID_DG_COND_SENSORS_INVALID_TEMP_COMP_CAL_RECORD , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Conductivity Cal | DG conductivity sensors temperatrue compensation invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "},