Index: AlarmDefs.h =================================================================== diff -u -rdb777bdaacdb2f32cc547ef0e424e9642c0df8c4 -r465af3c06f9a495fada40e297b178a1482eb11df --- AlarmDefs.h (.../AlarmDefs.h) (revision db777bdaacdb2f32cc547ef0e424e9642c0df8c4) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 465af3c06f9a495fada40e297b178a1482eb11df) @@ -1389,8 +1389,8 @@ { 228 , ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT , "HD | Maximum Prime Time Exceeded | Priming failure.\n- Check blood and dialysate lines for clamps and kinks. | HD time out on prime sub-mode purge air state. | During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. | "}, { 229 , ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT , "HD | Maximum Prime Time Exceeded | Priming failure.\n- Check blood and dialysate lines for clamps and kinks. | HD prime dialysate dialyzer time out alarm | If the priming of the dialyzer dialysate flow path has exceeded the time limit. | "}, { 230 , ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT , "HD | Maximum Prime Time Exceeded | Priming failure.\n- Check blood and dialysate lines for clamps and kinks. | HD prime dialysate bypass time out alarm | If the priming of the dialyzer bypass flow path has exceeded the time limit. | "}, - { 231 , ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "HD | Self Test Error | Self Test Failure. Please ensure the disposable cartridge is completely installed and the handle moved to the down position. | HD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, - { 232 , ALARM_ID_PRE_TREATMENT_WET_LC_TEST_FAILURE , "HD | Self Test Error | Self Test Failure. Please ensure the disposable cartridge is completely installed and the handle moved to the down position. Check for closed clamps or kinks in the tubing. | HD pre-treatment mode wet self-tests failure | If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState | "}, + { 231 , ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "HD | Self Test Error | Self Test Failure.\n Please ensure the disposable cartridge is completely installed and\nthe handle moved to the down position. | HD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, + { 232 , ALARM_ID_PRE_TREATMENT_WET_LC_TEST_FAILURE , "HD | Self Test Error | Self Test Failure.\n Please ensure the disposable cartridge is completely installed and\nthe handle moved to the down position.\nCheck for closed clamps or kinks in the tubing. | HD pre-treatment mode wet self-tests failure | If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState | "}, { 233 , ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC , "DG | Pressure Sensors Error | Call service to schedule a repair. | DG barometric pressure sensor bad coefficients CRC | If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. | "}, { 234 , ALARM_ID_INLET_WATER_PRESSURE_IN_HIGH_RANGE , "DG | Inlet Water Pressure High | Inlet water pressure is high.\n- Check the incoming water for issues. | DG inlet water pressure is too high | If the inlet water pressure is > 80 psig. | "}, { 235 , ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK , "HD | Rinseback Complete | Rinseback has been completed. | HD in treatment stopped sub-mode after rinseback completed (no escalation) | If there is no active alarm and rinseback has completed. | "},