Index: AlarmDefs.h =================================================================== diff -u -r6499712ba4bef6cc8fe7be5556644c71de51b1d2 -r4a54c70f6bb275a32a43c483f52222583c73f7a1 --- AlarmDefs.h (.../AlarmDefs.h) (revision 6499712ba4bef6cc8fe7be5556644c71de51b1d2) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 4a54c70f6bb275a32a43c483f52222583c73f7a1) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) Sean Nash -* @date (last) 20-Sep-2022 +* @author (last) Dara Navaei +* @date (last) 25-Oct-2022 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -78,7 +78,7 @@ ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK = 40, ///< Dialysis outlet pump failed motor off check. Measured speed while commanded off ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK = 41, ///< Dialysis outlet pump failed motor direction check. Measured vs. commanded ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK = 42, ///< Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds - ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAULT = 43, // TODO remove when DEN-13460 was merged ///< HD blood leak self test fault + ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT = 43, ///< DG main primary heater FPGA fault ALARM_ID_UI_COMM_TIMEOUT = 44, ///< User interface communication timeout ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS = 45, ///< HD too many bad communications CRC ALARM_ID_CAN_MESSAGE_NOT_ACKED = 46, ///< System message that required acknowledgment was not acknowledged @@ -98,8 +98,8 @@ ALARM_ID_HD_VALVE_NOT_FUNCTIONAL = 60, ///< HD valve not functional ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE = 61, ///< HD valve current out of range ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE = 62, ///< HD valve position out of target - ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT = 63, ///< FPGA reports failure reading arterial pressure sensor - ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT = 64, ///< FPGA reports failure reading venous pressure sensor + ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT = 63, ///< HD blood leak FPGA fault + ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE = 64, ///< DG barometric pressure sensor out of range ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT = 65, ///< HD requests DG command with invalid parameter fault ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE = 66, ///< HD blood leak sensor set point set failure ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE = 67, ///< HD blood pump occlusion self-test failure alarm @@ -109,7 +109,7 @@ ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE = 71, ///< HD venous pressure self-test failure alarm ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE = 72, ///< HD No load cell data message receive at least once every 2 seconds ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE = 73, ///< HD No dialysate temperature data message receive at least once every 2 seconds - ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_SEND_TIMEOUT = 74, // TODO remove when DEN-13460 was merged ///< HD blood leak detector self-test transmit timeout + ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW = 74, ///< DG inlet UV reactor is on with no flow ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE = 75, ///< HD syringe pump self-test failure alarm ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE = 76, ///< HD monitored voltage is out of range ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE = 77, ///< DG monitored voltage is out of range @@ -119,13 +119,13 @@ ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR = 81, ///< HD syringe pump over travel error ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR = 82, ///< HD syringe pump DAC write failure ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR = 83, ///< HD syringe pump is running while the BP is off - ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR = 84, ///< HD venous pressure sensor not being read + ALARM_ID_DG_SET_RTC_YEAR_INVALID = 84, ///< DG set RTC year is invalid ALARM_ID_PUMP_TRACK_LATCH_OPENED = 85, ///< HD pump track latch opened alarm - ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR = 86, ///< HD BP occlusion sensor not being read + ALARM_ID_HD_SET_RTC_YEAR_INVALID = 86, ///< HD set RTC year is invalid ALARM_ID_DG_HEATING_INVALID_CAL_RECORD = 87, ///< DG heating invalid calibration record ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE = 88, ///< DG concentrate pumps hall sensor out of range - ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR = 89, ///< HD BP occlusion sensor error - ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR = 90, ///< DG load cells FPGA read error. + ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW = 89, ///< DG outlet UV reactor on with no flow + ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT = 90, ///< DG load cells A1/B1 FPGA fault ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE = 91, ///< HD No dialysate flow data receive in the last 3 seconds ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS = 92, ///< DG conductivity sensor bad status. ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE = 93, ///< DG outlet primary conductivity out of range @@ -137,18 +137,18 @@ ALARM_ID_HD_SYRINGE_PUMP_STALL = 99, ///< HD syringe pump stall alarm ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT = 100, ///< HD no cartridge self-test timeout ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT = 101, ///< HD dry self-test timeout - ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE = 102, ///< DG heat disinfect temperature gradient out of range + ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE = 102, ///< DG heat disinfect temperature gradient out of range ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID = 103, ///< HD invalid calibration CRC ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS = 104, ///< Air trap level sensors reporting illegal combination of air/fluid ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID = 105, ///< DG invalid calibration CRC - ALARM_ID_DG_DIALYSATE_FLOW_SENSOR_INVALID_CAL_RECORD = 106, ///< DG dialysate flow sensor invalid calibration record + ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD = 106, ///< DG dialysate flow sensor invalid calibration record ALARM_ID_DG_RESTARTED_FAULT = 107, ///< HD reports DG restarted fault - ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR = 108, ///< HD syringe pump ADC error + ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT = 108, ///< HD syringe pump ADC FPGA fault ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR = 109, ///< HD syringe pump volume check error ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR = 110, ///< HD syringe pump speed check error ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR = 111, ///< HD syringe pump not stopped in off state error - ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_SEND_TIMEOUT = 112, // TODO remove when DEN-13460 was merged ///< HD blood leak zero command transmit timeout. - ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_FAULT = 113, // TODO remove when DEN-13460 was merged ///< HD blood leak zero command fault. + ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT = 112, ///< DG load cells A1/B1 FPGA fault + ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE = 113, ///< DG Drain pump current out of range ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 114, ///< HD venous air bubble detector self-test failure ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE = 115, ///< DG temperature sensor out of range ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT = 116, ///< DG temperature sensor ADC fault @@ -204,7 +204,7 @@ ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM = 166, ///< HD in treatment re-circ sub-mode for too long ALARM_ID_CARTRIDGE_DOOR_OPENED = 167, ///< HD cartridge door opened alarm ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT = 168, ///< HD active reservoir depletion timeout. - ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE = 169, ///< Dialysate flow rate out of range + ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE = 169, ///< DG Dialysate flow rate out of maximum range ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY = 170, ///< HD syringe empty alarm ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION = 171, ///< HD syringe pump occlusion alarm ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM = 172, ///< HD syringe pump not enough Heparin alarm @@ -215,7 +215,7 @@ ALARM_ID_DG_SOFTWARE_FAULT = 177, ///< DG software fault. Software found itself in an unexpected state ALARM_ID_HD_COMM_TIMEOUT = 178, ///< HD communication timeout ALARM_ID_DG_FPGA_COMM_TIMEOUT = 179, ///< DG FPGA communication down for too long - ALARM_ID_DG_LOAD_CELL_ADC_ERROR = 180, ///< DG load cell ADC error + ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE = 180, ///< DG RO flow out of maximum range ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE = 181, ///< DG load cells weight out of range for tare ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD = 182, ///< DG load cells invalid calibration ALARM_ID_DG_INVALID_LOAD_CELL_VALUE = 183, ///< DG invalid load cell value @@ -231,8 +231,8 @@ ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE = 193, ///< HD dialysate temperature out of low safety range. ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE = 194, ///< DG RO pump duty cycle out of range ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE = 195, ///< DG RO pump pressure out of range - ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT = 196, ///< DG temperature sensor error bit fault - ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT = 197, // TODO remove when DEN-13460 was merged ///< DG temperature sensors FPGA fault + ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT = 196, ///< DG CPi/CPo sensors FPGA fault + ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT = 197, ///< DG CD1/CD2 sensors FPGA fault ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON = 198, ///< RO flow too low while primary heater is on ALARM_ID_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON = 199, ///< Dialysate flow too low while trimmer heater is on ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE = 200, ///< DG thermistors/sensors temperature out of range @@ -257,8 +257,8 @@ ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED = 219, ///< HD syringe pump syringe removed alarm ALARM_ID_DG_THD_SENSORS_FPGA_FAULT = 220, ///< DG THd sensors FPGA fault ALARM_ID_EMPTY_SALINE_BAG = 221, ///< HD Saline bag is empty - ALARM_ID_DIALYSATE_TEMPERATURE_HIGH = 222, // TODO remove when DEN-13460 merged ///< HD dialysate temperature too high alarm - ALARM_ID_DIALYSATE_TEMPERATURE_LOW = 223, // TODO remove when DEN-13460 merged ///< HD dialysate temperature too low alarm + ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT = 222, ///< HD occlusion sensor FPGA fault + ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT = 223, ///< HD arterial sensor FPGA fault ALARM_ID_TREATMENT_STOPPED_BY_USER = 224, ///< Treatment stopped by user action - pressed stop button ALARM_ID_END_OF_TREATMENT_WARNING = 225, ///< HD end of treatment warning ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY = 226, ///< HD prime completed low priority alarm @@ -315,7 +315,7 @@ ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE = 277, ///< HD BP occlusion sensor is reading out of range ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE = 278, ///< HD active reservoir weight out of range. ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT = 279, ///< DG dialysate drain time out - ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR = 280, ///< HD arterial pressure sensor read timeout error + ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID = 280, ///< DG drain pump direction invalid ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME = 281, ///< Acid concentration bottle low volume alarm ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME = 282, ///< Bicarbonate concentration bottle low volume alarm ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE = 283, ///< DG load cells weight out of range @@ -325,7 +325,7 @@ ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE = 287, ///< HD processor clock speed checks against FPGA clock failure ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE = 288, ///< HD load cells primary/back up drift out of range ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION = 289, ///< DG dialysate or concentrate caps not closed - ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON = 290, // TODO remove when DEN-13460 merged ///< DG flow too low while heater is on + ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT = 290, ///< HD venous pressure sensor FPGA fault ALARM_ID_HD_DISINFECT_FLUSH = 291, ///< HD processor is in Disinfect Flush mode ALARM_ID_HD_DISINFECT_HEAT = 292, ///< HD processor is in Disinfect Heat mode ALARM_ID_HD_DISINFECT_CHEM = 293, ///< HD processor is in Disinfect Chemical mode @@ -334,14 +334,14 @@ ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT = 296, ///< DG TDi sensors FPGA fault ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT = 297, ///< DG TRo sensors FPGA fault ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT = 298, ///< DG baro sensor FPGA fault - ALARM_ID_DG_DISINFECT_HAS_BEEN_EXPIRED = 299, ///< DG disinfect has been expired - ALARM_ID_DG_SERIVCE_TIME_INTERVAL_HAS_ELAPSED = 300, ///< DG service time interval has elapsed + ALARM_ID_DG_INVALID_SERIAL_NUMBER = 299, ///< DG invalid serial number + ALARM_ID_HD_INVALID_SERIAL_NUMBER = 300, ///< HD invalid serial number ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE = 301, ///< DG primary heater power voltage out of range ALARM_ID_DG_CPU_RAM_ERROR = 302, ///< DG processor RAM error - ALARM_ID_HD_SERVICE_TIME_INTERVAL_HAS_ELAPSED = 303, ///< HD service time interval has elapsed + ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT = 303, ///< DG drain pump direction FPGA fault ALARM_ID_HD_INVALID_USAGE_RECORD_CRC = 304, ///< HD invalid usage record ALARM_ID_HD_CPU_RAM_ERROR = 305, ///< HD processor RAM error - ALARM_ID__AVAILABLE_13 = 306, ///< Alarm ID available + ALARM_ID_HD_AIR_TRAP_OVERFILL = 306, ///< HD air trap over-fill alarm ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT = 307, ///< HD AC power lost second alarm ALARM_ID_DG_CPI_COND_SENSOR_INVALID_CHAR = 308, ///< DG CPi conductivity sensor invalid character received ALARM_ID_DG_CPO_COND_SENSOR_INVALID_CHAR = 309, ///< DG CPo conductivity sensor invalid character received @@ -478,15 +478,15 @@ { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP, ALM_ESC_10_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP, ALM_ESC_10_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_USAGE_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_USAGE_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_USAGE_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_UI_POST_FAILED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_PRESSURE_SENSORS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP, ALM_ESC_10_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP, ALM_ESC_10_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, @@ -496,7 +496,7 @@ { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_ACCELEROMETERS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SOFTWARE_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK, 0, ALARM_ID_NO_ALARM }, @@ -511,7 +511,7 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAULT, 0, ALARM_ID_NO_ALARM },//add DG version + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS, 0, ALARM_ID_NO_ALARM },//add DG version // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To @@ -532,18 +532,18 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_NOT_FUNCTIONAL, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_SEND_TIMEOUT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -554,13 +554,13 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_SET_RTC_YEAR_INVALID, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PUMP_TRACK_LATCH_OPENED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SET_RTC_YEAR_INVALID, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -573,28 +573,28 @@ // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FLOW_SENSOR_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_RESTARTED_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_SEND_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CAN_MESSAGE_NOT_ACKED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, ALARM_ID_DG_RTC_CONFIG_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, ALARM_ID_DG_RTC_BATTERY_LOW, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRE_TREATMENT_WET_PRIME_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 4, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_HIGH, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 5, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK, 0, ALARM_ID_NO_ALARM }, @@ -615,7 +615,7 @@ { ALARM_PRIORITY_HIGH, 19, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DG_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 20, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 21, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_OCCLUSION_BLOOD_PUMP, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_RO_FLOW_SENSOR_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE, 0, ALARM_ID_NO_ALARM }, @@ -642,19 +642,19 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_RTC_CONFIG_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_RTC_OR_TIMER_ACCURACY_FAILURE, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_HIGH_SAFETY_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_HIGH_SAFETY_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_SOFTWARE_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_FPGA_COMM_TIMEOUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_ADC_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_LOAD_CELL_VALUE, 0, ALARM_ID_NO_ALARM }, @@ -663,19 +663,19 @@ { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_CP1_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_CP2_SPEED_CONTROL_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DRAIN_PUMP_OFF_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON,0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRE_TREATMENT_WET_FLOW_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_PRESSURE_SENSOR_FAULT, 0, ALARM_ID_NO_ALARM }, @@ -695,10 +695,10 @@ { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_DETECTED, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH , 799, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH , 3, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_THD_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_THD_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 900, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_EMPTY_SALINE_BAG, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_LOW, 901, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_DIALYSATE_TEMPERATURE_HIGH, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, - { ALARM_PRIORITY_LOW, 902, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , ALARM_ID_DIALYSATE_TEMPERATURE_LOW, ALM_ESC_5_MIN, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 903, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_BY_USER, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To { ALARM_PRIORITY_LOW, 904, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_END_OF_TREATMENT_WARNING, 0, ALARM_ID_NO_ALARM }, @@ -709,7 +709,7 @@ { ALARM_PRIORITY_LOW, 909, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 910, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_PRE_TREATMENT_WET_LC_TEST_FAILURE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, TRUE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_RTC_RAM_OPS_ERROR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_INSTALL_NEW_CARTRIDGE, 0, ALARM_ID_NO_ALARM }, @@ -758,7 +758,7 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, @@ -768,23 +768,23 @@ { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DISINFECT_FLUSH, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DISINFECT_HEAT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_DISINFECT_CHEM, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_DISINFECT_CHEM_INSTRUCTION1, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 999, FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_UI_POST_FAILURE_CLOUDSYNC, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_DISINFECT_HAS_BEEN_EXPIRED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_DG_SERIVCE_TIME_INTERVAL_HAS_ELAPSED, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPU_RAM_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_MEDIUM, 899, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_SERVICE_TIME_INTERVAL_HAS_ELAPSED, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_INVALID_SERIAL_NUMBER, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_SERIAL_NUMBER, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPU_RAM_ERROR, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 1, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_HD_INVALID_USAGE_RECORD_CRC, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_CPU_RAM_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_13, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_HD_AIR_TRAP_OVERFILL, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 18, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPI_COND_SENSOR_INVALID_CHAR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CPO_COND_SENSOR_INVALID_CHAR, 0, ALARM_ID_NO_ALARM }, @@ -835,9 +835,11 @@ { 1, ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC }, { 1, ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE }, { 1, ALARM_ID_HD_RTC_OR_TIMER_ACCURACY_FAILURE }, - { 1, ALARM_ID_DG_DIALYSATE_FLOW_SENSOR_INVALID_CAL_RECORD }, + { 1, ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD }, { 1, ALARM_ID_DG_RO_FLOW_SENSOR_INVALID_CAL_RECORD }, { 1, ALARM_ID_HD_INVALID_USAGE_RECORD_CRC }, + { 1, ALARM_ID_DG_SET_RTC_YEAR_INVALID }, + { 1, ALARM_ID_HD_SET_RTC_YEAR_INVALID }, { 3, ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC }, { 3, ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC }, { 3, ALARM_ID_HD_SOFTWARE_FAULT }, @@ -853,7 +855,7 @@ { 3, ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK }, { 3, ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK }, { 3, ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK }, - { 3, ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAULT }, + { 3, ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT }, { 3, ALARM_ID_UI_COMM_TIMEOUT }, { 3, ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS }, { 3, ALARM_ID_CAN_MESSAGE_NOT_ACKED }, @@ -873,8 +875,8 @@ { 3, ALARM_ID_HD_VALVE_NOT_FUNCTIONAL }, { 3, ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE }, { 3, ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE }, - { 3, ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT }, - { 3, ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT }, + { 3, ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT }, + { 3, ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE }, { 3, ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT }, { 3, ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, { 3, ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE }, @@ -883,7 +885,7 @@ { 3, ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, { 3, ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE }, { 3, ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE }, - { 3, ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_SEND_TIMEOUT }, + { 3, ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW }, { 3, ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE }, { 3, ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE }, { 3, ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE }, @@ -893,12 +895,10 @@ { 3, ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR }, - { 3, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR }, - { 3, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR }, { 3, ALARM_ID_DG_HEATING_INVALID_CAL_RECORD }, { 3, ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE }, - { 3, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR }, - { 3, ALARM_ID_DG_LOAD_CELL_FPGA_READ_ERROR }, + { 3, ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW }, + { 3, ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT }, { 3, ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE }, { 3, ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS }, { 3, ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE }, @@ -908,16 +908,16 @@ { 3, ALARM_ID_HD_SYRINGE_PUMP_STALL }, { 3, ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT }, { 3, ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT }, - { 3, ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE }, + { 3, ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANGE }, { 3, ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID }, { 3, ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS }, { 3, ALARM_ID_DG_RESTARTED_FAULT }, - { 3, ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR }, + { 3, ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT }, { 3, ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR }, { 3, ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR }, - { 3, ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_SEND_TIMEOUT }, - { 3, ALARM_ID_HD_BLOOD_LEAK_ZERO_CMD_FAULT }, + { 3, ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT }, + { 3, ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE }, { 3, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE }, { 3, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT }, @@ -940,7 +940,7 @@ { 3, ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE }, { 3, ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE }, { 3, ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT }, - { 3, ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR }, + { 3, ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID }, { 3, ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE }, { 3, ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE }, { 3, ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE }, @@ -961,15 +961,15 @@ { 3, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE }, { 3, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME }, { 3, ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, - { 3, ALARM_ID_DIALYSATE_FLOW_RATE_OUT_OF_RANGE }, + { 3, ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE }, { 3, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE }, { 3, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE }, { 3, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, { 3, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, { 3, ALARM_ID_UV_REACTOR_NOT_HEALTHY }, - { 3, ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT }, - { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT }, - { 3, ALARM_ID_DG_LOAD_CELL_ADC_ERROR }, + { 3, ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT }, + { 3, ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT }, + { 3, ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE }, { 3, ALARM_ID_DG_FPGA_COMM_TIMEOUT }, { 3, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT }, { 3, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT }, @@ -981,7 +981,7 @@ { 3, ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR }, { 3, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH }, { 3, ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK }, - { 3, ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON }, + { 3, ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT }, { 3, ALARM_ID_HD_RTC_CONFIG_ERROR }, { 3, ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT }, { 3, ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT }, @@ -1004,13 +1004,18 @@ { 3, ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT }, { 3, ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT }, { 3, ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE }, + { 3, ALARM_ID_DG_INVALID_SERIAL_NUMBER }, + { 3, ALARM_ID_HD_INVALID_SERIAL_NUMBER }, { 3, ALARM_ID_DG_CPU_RAM_ERROR }, { 3, ALARM_ID_HD_CPU_RAM_ERROR }, + { 3, ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT }, + { 3, ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT }, + { 3, ALARM_ID_HD_AIR_TRAP_OVERFILL }, { 3, ALARM_ID_DG_CPI_COND_SENSOR_INVALID_CHAR }, { 3, ALARM_ID_DG_CPO_COND_SENSOR_INVALID_CHAR }, { 3, ALARM_ID_DG_CD1_COND_SENSOR_INVALID_CHAR }, { 3, ALARM_ID_DG_CD2_COND_SENSOR_INVALID_CHAR }, - { 3, ALARM_ID__AVAILABLE_13 }, + { 3, ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT }, { 4, ALARM_ID_END_OF_TREATMENT_HIGH }, { 5, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, { 6, ALARM_ID_HD_BLOOD_LEAK_DETECTED }, @@ -1074,12 +1079,7 @@ { 899, ALARM_ID_HD_TREATMENT_RECIRC_TIMEOUT_WARNING }, { 899, ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_WARNING }, { 899, ALARM_ID_HD_DISINFECT_CHEM_INSTRUCTION1 }, - { 899, ALARM_ID_DG_DISINFECT_HAS_BEEN_EXPIRED }, - { 899, ALARM_ID_DG_SERIVCE_TIME_INTERVAL_HAS_ELAPSED }, - { 899, ALARM_ID_HD_SERVICE_TIME_INTERVAL_HAS_ELAPSED }, { 900, ALARM_ID_EMPTY_SALINE_BAG }, - { 901, ALARM_ID_DIALYSATE_TEMPERATURE_HIGH }, - { 902, ALARM_ID_DIALYSATE_TEMPERATURE_LOW }, { 903, ALARM_ID_TREATMENT_STOPPED_BY_USER }, { 904, ALARM_ID_END_OF_TREATMENT_WARNING }, { 905, ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY }, Index: MsgDefs.h =================================================================== diff -u -r789dc146333360f6882276dc702c326fd68db12c -r4a54c70f6bb275a32a43c483f52222583c73f7a1 --- MsgDefs.h (.../MsgDefs.h) (revision 789dc146333360f6882276dc702c326fd68db12c) +++ MsgDefs.h (.../MsgDefs.h) (revision 4a54c70f6bb275a32a43c483f52222583c73f7a1) @@ -8,8 +8,8 @@ * * @file MsgDefs.h * -* @author (last) Sean Nash -* @date (last) 26-Sep-2022 +* @author (last) Dara Navaei +* @date (last) 09-Nov-2022 * * @author (original) Sean * @date (original) 27-Feb-2020 @@ -72,7 +72,7 @@ MSG_ID_DG_RESERVOIRS_DATA = 0x28, ///< DG broadcast reservoirs data MSG_ID_DG_SAMPLE_WATER_CMD = 0x29, ///< HD request to DG to allow user to get a water sample MSG_ID_DG_VALVES_STATES = 0x2A, ///< DG broadcast valve states - MSG_ID_DG_START_STOP_TRIMMER_HEATER_CMD = 0x2B, ///< HD request to DG to start/stop trimmer heater + MSG_ID_HD_START_STOP_TRIMMER_HEATER_CMD = 0x2B, ///< HD request to DG to start/stop trimmer heater MSG_ID_DG_HEATERS_DATA = 0x2C, ///< DG broadcast of the heaters data MSG_ID_DG_TEMPERATURE_DATA = 0x2D, ///< DG broadcast of the temperature sensors data MSG_ID_USER_UF_SETTINGS_CHANGE_CONFIRMATION_RESPONSE = 0x2E, ///< HD response to user selection of ultrafiltration settings change option selection @@ -168,9 +168,9 @@ MSG_ID_DG_SERIAL_NUMBER = 0x88, ///< DG serial number MSG_ID_UI_REQUEST_SERVICE_INFO = 0x89, ///< UI request for service schedule information MSG_ID_HD_SERVICE_SCHEDULE_DATA = 0x8A, ///< HD service schedule data - MSG_ID_HD_USAGE_DATA = 0x8B, ///< HD usage data TODO is this needed? + MSG_ID_HD_USAGE_DATA = 0x8B, ///< HD usage data MSG_ID_DG_SERVICE_SCHEDULE_DATA = 0x8C, ///< DG service schedule data - MSG_ID_DG_USAGE_DATA = 0x8D, ///< DG usage data TODO is this needed? + MSG_ID_DG_USAGE_DATA = 0x8D, ///< DG usage data MSG_ID_HD_POST_SINGLE_TEST_RESULT = 0x8E, ///< HD POST single test result MSG_ID_HD_POST_FINAL_TEST_RESULT = 0x8F, ///< HD POST final test result MSG_ID_DG_POST_SINGLE_TEST_RESULT = 0x90, ///< DG POST single test result @@ -195,13 +195,13 @@ MSG_ID_HD_FANS_DATA = 0xA3, ///< HD fans data publish MSG_ID_HD_EVENT = 0xA4, ///< HD general event MSG_ID_DG_EVENT = 0xA5, ///< DG general event - MSG_ID_DG_DIALYSATE_FLOW_METER_DATA = 0xA6, // TODO remove when DEN-13460 merged ///< DG dialysate flow meter data (liter/minute) + MSG_ID____AVAILABLE_4 = 0xA6, ///< This message is available for use MSG_ID_DG_ALARM_INFO = 0xA7, ///< DG alarm information data publish MSG_ID_HD_RESERVOIRS_DATA = 0xA8, ///< HD reservoirs data publish MSG_ID_HD_REQUEST_DG_CONCENTRATE_MIXING_RATIOS = 0xA9, ///< HD request DG concentrate ratios MSG_ID_DG_CONCENTRATE_MIXING_RATIOS_DATA = 0xAA, ///< DG concentrate mixing ratios data MSG_ID_DG_SCHEDULED_RUNS_DATA = 0xAB, ///< DG scheduled runs data - MSG_ID_DG_SCHEDULED_RUNS_INFO = 0xAC, ///< DG scheduled runs info to be sent to the UI and Dialin + MSG_ID____AVAILABLE_10 = 0xAC, ///< This message is available for use MSG_ID_UI_INSTALLATION_CONFIRM_RESPONSE = 0xAD, ///< HD Confirm installation response MSG_ID_DG_FILL_MODE_DATA = 0xAE, ///< DG broadcast of fill mode data such as used acid & bicarb volume MSG_ID_DG_BAD_FILL_SUB_STATE = 0xAF, ///< DG broadcast of idle mode bad fill sub-state @@ -214,6 +214,13 @@ MSG_ID_HD_REQUEST_DG_SERVICE_MODE = 0xB6, ///< HD request for DG to to go to service mode MSG_ID_DG_RTC_EPOCH = 0xB7, ///< DG RTC epoch broadcast message MSG_ID_HD_REQUEST_DG_USAGE_INFO = 0xB8, ///< HD request DG usage info + MSG_ID_HD_REQUEST_DG_SERVICE_RECORD = 0xB9, ///< HD request DG service record + MSG_ID_HD_REQUEST_UI_CONFIRMATION = 0xBA, ///< HD request UI confirmation + MSG_ID_UI_CONFIRMATION_RESULT = 0xBB, ///< UI confirmation result response + MSG_ID_REQUEST_CPLD_STATUS = 0xBC, ///< Dialin request for CPLD Status + MSG_ID_DG_CPLD_STATUS = 0xBD, ///< DG CPLD Status + MSG_ID_HD_BATTERY_MANAGEMENT_DATA = 0xBE, ///< HD battery management logging data + MSG_ID_HD_BATTERY_STATUS_DATA = 0xBF, ///< HD battery status logging data #ifdef EMC_TEST_BUILD MSG_ID_CAN_ERROR_COUNT = 0x999, // test code in support of EMC testing @@ -248,7 +255,7 @@ MSG_ID_PRESSURE_VENOUS_OVERRIDE = 0x8018, ///< Venous pressure measured override request MSG_ID_OCCLUSION_BLOOD_PUMP_OVERRIDE = 0x8019, ///< Blood pump occlusion pressure measured override request MSG_ID_BLOOD_ROTOR_COUNT_OVERRIDE = 0x801A, ///< Blood pump rotor count override request - MSG_ID____AVAILABLE_1 = 0x801B, ///< Unused - this message ID is available for use + MSG_ID____AVAILABLE_11 = 0x801B, ///< Message ID available MSG_ID_PRES_OCCL_SEND_INTERVAL_OVERRIDE = 0x801C, ///< Pressure and occlusion broadcast interval override request MSG_ID_SET_RTC_DATE_TIME = 0x801D, ///< RTC set date/time request MSG_ID_DIAL_OUT_FLOW_SEND_INTERVAL_OVERRIDE = 0x801E, ///< Dialysate outlet flow broadcast interval override request @@ -320,7 +327,7 @@ MSG_ID____AVAILABLE_2 = 0x8061, ///< Unused - this message ID is available for use MSG_ID_HD_SWITCHES_STATUS_OVERRIDE = 0x8062, ///< HD switches status override request MSG_ID_HD_SWITCHES_PUBLISH_INTERVAL_OVERRIDE = 0x8063, ///< HD switches data publish override request - MSG_ID_HD_BATTERY_REMAINING_PERCENT_OVERRIDE = 0x8064, ///< HD battery remaining percent override request + MSG_ID_HD_BATTERY_REMAINING_CAP_MWH_OVERRIDE = 0x8064, ///< HD battery remaining capacity (in mWh) override request MSG_ID_HD_TEMPERATURES_VALUE_OVERRIDE = 0x8065, ///< HD temperatures value override request MSG_ID_HD_TEMPERATURES_PUBLISH_INTERVAL_OVERRIDE = 0x8066, ///< HD temperatures data publish override request MSG_ID_HD_FANS_PUBLISH_INTERVAL_OVERRIDE = 0x8067, ///< HD fans data publish override request @@ -367,7 +374,7 @@ MSG_ID_PRESSURE_OVERRIDE = 0xA006, ///< Pressure sensor override request MSG_ID_PRESSURE_SEND_INTERVAL_OVERRIDE = 0xA007, ///< Pressure data broadcast interval override request MSG_ID_DG_HD_COMMUNICATION_STATUS_OVERRIDE = 0xA008, ///< DG HD Communication Status override - MSG_ID_RO_MEASURED_FLOW_OVERRIDE = 0xA009, ///< RO measured flow rate override request + MSG_ID_DG_SET_PRIMARY_AND_TRIMMER_HEATERS_TARGET_TEMP = 0xA009, ///< DG set primary and trimmer heater target temperature MSG_ID_RO_PUMP_SEND_INTERVAL_OVERRIDE = 0xA00A, ///< RO pump data broadcast interval override request MSG_ID_DRAIN_PUMP_SET_RPM = 0xA00B, ///< Drain pump set RPM override request MSG_ID_DRAIN_PUMP_SEND_INTERVAL_OVERRIDE = 0xA00C, ///< Drain pump data broadcast interval override request @@ -391,7 +398,7 @@ MSG_ID_DG_SWITCHES_PUBLISH_INTERVAL_OVERRIDE = 0xA01E, ///< DG switches data publish interval override MSG_ID_DG_OP_MODE_PUBLISH_INTERVAL_OVERRIDE = 0xA01F, ///< DG operation mode publish interval override MSG_ID_DG_BLOCK_MESSAGE_TRANSMISSION = 0xA020, ///< HD prevent transmission of given message(s) - MSG_ID_DIALYSATE_MEASURED_FLOW_OVERRIDE = 0xA021, ///< DG dialysate measured flow rate override request + MSG_ID_MEASURED_FLOW_SENSORS_OVERRIDE = 0xA021, ///< DG measured flow rate override request MSG_ID_DG_SOFTWARE_RESET_REQUEST = 0xA022, ///< DG reset request MSG_ID_DG_CONCENTRATE_PUMP_MEASURED_SPEED_OVERRIDE = 0xA023, ///< DG concentrate pump measured speed override request MSG_ID_CONCENTRATE_PUMP_TARGET_SPEED_OVERRIDE = 0xA024, ///< Concentrate pumps' target speed override request @@ -405,7 +412,7 @@ MSG_ID_DG_THERMISTORS_DATA_PUBLISH_INTERVAL_OVERRIDE = 0xA02D, ///< DG thermistors data publish interval override MSG_ID_DG_THERMISTORS_VALUE_OVERRIDE = 0xA02E, ///< DG thermistors value override MSG_ID_DG_RO_PUMP_DUTY_CYCLE_OVERRIDE = 0xA02F, ///< DG RO pump duty cycle override - MSG_ID_DG_RO_FLOW_RATE_OVERRIDE = 0xA030, ///< DG RO flow rate override + MSG_ID_DG_VALVES_SENSED_STATE_OVERRIDE = 0xA030, ///< DG valves sensed state override MSG_ID_DG_SET_RO_PUMP_TARGET_FLOW = 0xA031, ///< DG set RO pump target flow rate MSG_ID_DG_RO_PUMP_TARGET_PRESSURE_OVERRIDE = 0xA032, ///< DG RO pump target pressure override MSG_ID_DG_SET_CALIBRATION_RECORD = 0xA033, ///< DG set calibration record that is received from Dialin @@ -425,12 +432,12 @@ MSG_ID_DG_FLUSH_PUBLISH_INTERVAL_OVERRIDE = 0xA041, ///< DG flush mode data publish interval override MSG_ID_FILTER_FLUSH_TIME_PERIOD_OVERRIDE = 0xA042, ///< DG filter flush time period override MSG_ID_DG_FANS_RPM_OVERRIDE = 0xA043, ///< DG fans RPM override - MSG_ID_DIALYSATE_FLOW_SEND_INTERVAL_OVERRIDE = 0xA044, ///< Dialysate flow data broadcast interval override request + MSG_ID_FLOW_DATA_PUBLISH_INTERVAL_OVERRIDE = 0xA044, ///< Flow data broadcast interval override request MSG_ID_DG_STOP_RTC_CLOCK = 0xA045, ///< DG stop RTC clock MSG_ID_DG_DRAIN_PUMP_MEASURED_RPM_OVERRIDE = 0xA046, ///< DG drain pump measured RPM override MSG_ID_DG_SUPER_CLEAR_ALARMS_CMD = 0xA047, ///< Clears all DG alarms (even if non-recoverable or fault) MSG_ID_DG_ALARM_INFO_SEND_INTERVAL_OVERRIDE = 0xA048, ///< DG alarm information broadcast interval override request - MSG_ID____AVAILABLE_3 = 0xA049, ///< Unused - this message ID is available for use + MSG_ID_DG_START_STOP_TRIMMER_HEATER_CMD = 0xA049, ///< DG start/stop trimmer heater command MSG_ID_DG_GET_SW_CONFIG_RECORD = 0xA04A, ///< DG get software configuration record request MSG_ID_DG_SET_SW_CONFIG_RECORD = 0xA04B, ///< DG set software configuration record MSG_ID_DG_SEND_SW_CONFIG_RECORD = 0xA04C, ///< DG send software configuration record @@ -445,6 +452,9 @@ MSG_ID_DG_SEND_USAGE_INFO_RECORD = 0xA055, ///< DG send usage information record MSG_ID_DG_SET_OP_MODE_REQUEST = 0xA056, ///< DG set operation mode request MSG_ID_DG_RESERVOIR_TARE_REQUEST = 0xA057, ///< DG reservoir tare request + MSG_ID_DG_CPLD_PUBLISH_INTERVAL_OVERRIDE = 0xA058, ///< DG CPLD publish interval override + MSG_ID_DG_DRAIN_PUMP_CURRENT_OVERRIDE = 0xA059, ///< DG drain pump current override request + MSG_ID_DG_DRAIN_PUMP_DIRECTION_OVERRIDE = 0xA05A, ///< DG drain pump direction override request MSG_ID_HD_DEBUG_EVENT = 0xFFF1, ///< HD debug event text to be logged in event log MSG_ID_DG_DEBUG_EVENT = 0xFFF2, ///< DG debug event text to be logged in event log @@ -504,6 +514,9 @@ REQUEST_REJECT_REASON_SYRINGE_NOT_PRESENT = 43, ///< Syringe is not present REQUEST_REJECT_REASON_DG_DIALYSATE_CAP_OPEN = 44, ///< Dialysate cap open REQUEST_REJECT_REASON_DG_CONCENTRATE_CAP_OPEN = 45, ///< Concentrate cap open + REQUEST_REJECT_REASON_DG_DISINFECT_HAS_BEEN_EXPIRED = 46, ///< DG disinfect has been expired + REQUEST_REJECT_REASON_DG_SERVICE_IS_DUE = 47, ///< DG service is due + REQUEST_REJECT_REASON_HD_SERVICE_IS_DUE = 48, ///< HD service is due NUM_OF_REQUEST_REJECT_REASONS ///< Number of settings change reject codes }; typedef enum Request_Reject_Reasons REQUEST_REJECT_REASON_CODE_T; ///< Type for settings change rejection reason codes enumeration