Index: AlarmDefs.h =================================================================== diff -u -r2ef382945f6ccc9e84b2f1b7992a85df6d31af33 -r4b41179e318c71605d6a4171e9ee6d7b475bd0d2 --- AlarmDefs.h (.../AlarmDefs.h) (revision 2ef382945f6ccc9e84b2f1b7992a85df6d31af33) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 4b41179e318c71605d6a4171e9ee6d7b475bd0d2) @@ -1392,7 +1392,7 @@ { 234 , ALARM_ID_INLET_WATER_PRESSURE_IN_HIGH_RANGE , "DG | Inlet Water Pressure High | Inlet water pressure is high.\n- Check the incoming water for issues. | DG inlet water pressure is too high | If the inlet water pressure is > 80 psig. | "}, { 235 , ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK , "HD | Rinseback Complete | Rinseback has been completed. | HD in treatment stopped sub-mode after rinseback completed (no escalation) | If there is no active alarm and rinseback has completed. | "}, { 236 , ALARM_ID_INSTALL_NEW_CARTRIDGE , "HD | New Cartridge Needed | New cartridge needs to be installed. | HD needs new cartridge to be installed | If the air trap sensor reported AIR and bubble sensor detected bubbles. | "}, - { 237 , ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT , "HD | Prime Error | Call service to schedule a repair. | HD prime saline dialyzer time out alarm | This alarm is available. If timeout, just transition to HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE | "}, + { 237 , ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT , "HD | Prime Error | Call service to schedule a repair. | HD prime saline dialyzer time out alarm | If timeout, just transition to HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE | "}, { 238 , ALARM_ID_NO_CARTRIDGE_LOADED , "HD | Cartridge Needed | Cartridge not loaded or loaded improperly.\n- Check that a cartridge is properly installed. | HD no cartridge loaded or installed improperly alarm | If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. | "}, { 239 , ALARM_ID_CARTRIDGE_REMOVAL_FAILURE , "HD | Cartridge Removal Error | Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. | HD fail to remove cartridge alarm | In the post treatment state, if the cartridge has not been removed. | "}, { 240 , ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE , "DG | Dialysate Conductivity Error | Dialysate conductivity out of range.\n- Check to be sure the correct bicarbonate dialysate concentrate is connected. | Bicarb conductivity out of range during bicarb pump check alarm | If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. | "},