Index: AlarmDefs.h =================================================================== diff -u -r0181fb65d627b504c5bedd4c65e59125fc68c70c -r5046c66fef77ab962a17fbc51c308360546391f0 --- AlarmDefs.h (.../AlarmDefs.h) (revision 0181fb65d627b504c5bedd4c65e59125fc68c70c) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 5046c66fef77ab962a17fbc51c308360546391f0) @@ -1232,7 +1232,7 @@ { 65 , ALARM_ID_HD_DG_COMMAND_INVALID_PARAMETER_FAULT , "HD | Communication Error | A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | Loss of DG Comm (Reject) | HD requests DG command with invalid parameter fault | If the latest DG command response indicated invalid parameter | "}, { 66 , ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Config Set Point | HD blood leak sensor set point set failure | If the set point of the blood leak sensor does not match the set point in the calibration record. | "}, { 67 , ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD blood pump occlusion self-test failure alarm | If the blood pump pressure > 5000 | "}, - { 68 , ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE , "HD | Dialysate Generation Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. | Dial. Gen: Dialysate not ready | HD active reservoir recirculation out of range | If the recirculation level has exceeded the max allowed | "}, + { 68 , ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE , "HD | Dialysate Generation Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. | Dial. Gen: Dialysate not ready | HD active reservoir recirculation out of range | If the recirculation level has exceeded the max allowed | "}, { 69 , ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD POST: BLD Cal | HD blood leak sensor invalid calibration record | If the BLD data that was received from NVDataMgmt is NOT legitimate. | "}, { 70 , ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD arterial pressure self-test failure alarm | If arterial pressure < -300 or > 100 mmHg | "}, { 71 , ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE , "HD | Self Test: Cartridge | A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | HD venous pressure self-test failure alarm | If the venous pressure < -100 or > 600 mmHg | "}, @@ -1291,10 +1291,10 @@ { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. | Blood Stop: Blood Clot Risk | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. | Blood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, - { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access\n issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access\n issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, { 129 , ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "HD | Blood Leak: Bypass | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. | Blood Stop: Dialyzer Blood Leak Bypass | HD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "}, - { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 131 , ALARM_ID_HD_ARTERIAL_PRESSURE_LOW , "HD | Low Arterial Pressure | Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. Low | Arterial pressure too low during treatment | If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 132 , ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH , "HD | High Arterial Pressure | Arterial pressure is high.\n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. High | Arterial pressure too high during treatment | If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 133 , ALARM_ID_DG_FLUID_LEAK_DETECTED , "DG | Dialysate Device Fluid Leak | Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Fluid Leak in DG | DG fluid leak detected alarm | If the fluid leak is detected as reported by the fluid leak detector. | "}, @@ -1442,7 +1442,7 @@ { 275 , ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Arterial Press Range | HD arterial pressure sensor is reading out of range | If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, { 276 , ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Venous Press Range | HD venous pressure sensor is reading out of range | If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, { 277 , ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Occlusion Sens Range | HD BP occlusion sensor is reading out of range | If the blood pump occlusion count is < 2000. | "}, - { 278 , ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "HD | Dialysate Reservoir Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed\n when dialysate is ready for use. | Dial. Gen: Reservoir Overfill | HD active reservoir weight out of range. | If the active reservoir weight is > 1900.0 g. | "}, + { 278 , ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "HD | Dialysate Reservoir Error | A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. | Dial. Gen: Reservoir Overfill | HD active reservoir weight out of range. | If the active reservoir weight is > 1900.0 g. | "}, { 279 , ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Drain | DG dialysate drain time out | If the DG dialysate drain exceeded 120 seconds. | "}, { 280 , ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: DRP Direction Fault | DG drain pump invalid direction | If the drain pump's direction is not 1 when draining. | "}, { 281 , ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME , "DG | Acid Concentrate Low | The acid concentrate is low.\n- Replace acid concentrate container. | Dial. Gen: Acid Conc. Low | DG Acid concentration bottle low volume alarm | If the acid concentration bottle volume > 3240 (3340 - 100). | "},