Index: AlarmDefs.h =================================================================== diff -u -raa6c87589ba194305fc105811d8f6ba571c83339 -r6058ce086760b8cb84cb023869068008ed61b9e3 --- AlarmDefs.h (.../AlarmDefs.h) (revision aa6c87589ba194305fc105811d8f6ba571c83339) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 6058ce086760b8cb84cb023869068008ed61b9e3) @@ -53,6 +53,13 @@ ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CALIBRATION, ///< HD dialysate flow invalid calibration ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE, ///< HD alarm audio failed POST ALARM_ID_HD_UI_POST_FAILED, ///< HD UI POST failed + ALARM_ID_DG_PRESSURE_SENSORS_INVALID_CAL_RECORD, ///< DG load cells invalid calibration record + ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD, ///< DG flow sensors invalid calibration record + ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD, ///< DG conductivity sensors invalid calibration record + ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD, ///< DG drain line volume invalid calibration record + ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD, ///< DG reservoirs invalid calibration record + ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD, ///< DG acid concentrate calibration record + ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD, ///< DG bicarb concentrate calibration record ALARM_ID_HD_SOFTWARE_FAULT, ///< HD Software fault. Software found itself in an unexpected state ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK, ///< Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running ALARM_ID_BLOOD_PUMP_OFF_CHECK, ///< Blood pump failed motor off check. Measured speed while commanded off @@ -266,13 +273,6 @@ ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT, ///< DG chemical disinfect could not reach to target temperature ALARM_ID_DG_CHEM_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT, ///< DG chemical disinfect inlet pressure and temperature sensors out of range ALARM_ID_DG_CHEM_DISINFECT_INLET_COND_AND_TEMP_OUT, ///< DG chemical disinfect inlet conductivity and temperature out of range - ALARM_ID_DG_PRESSURE_SENSORS_INVALID_CAL_RECORD, ///< DG load cells invalid calibration record - ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD, ///< DG flow sensors invalid calibration record - ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD, ///< DG conductivity sensors invalid calibration record - ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD, ///< DG drain line volume invalid calibration record - ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD, ///< DG reservoirs invalid calibration record - ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD, ///< DG acid concentrate calibration record - ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD, ///< DG bicarb concentrate calibration record NUM_OF_ALARM_IDS ///< Total number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration Index: MsgDefs.h =================================================================== diff -u -re508294f3a5857c9e9ab4d7c4b0dce27c23c3b36 -r6058ce086760b8cb84cb023869068008ed61b9e3 --- MsgDefs.h (.../MsgDefs.h) (revision e508294f3a5857c9e9ab4d7c4b0dce27c23c3b36) +++ MsgDefs.h (.../MsgDefs.h) (revision 6058ce086760b8cb84cb023869068008ed61b9e3) @@ -155,7 +155,7 @@ MSG_ID_HD_VOLTAGES_DATA = 0x7B, ///< HD voltages data publish MSG_ID_HD_ALARM_AUDIO_VOLUME_SET_RESPONSE = 0x7C, ///< HD response to UI set alarm audio volume request MSG_ID_HD_ALARM_INFORMATION = 0x7D, ///< HD alarm information broadcast message - MSG_ID_HD_STANDBY_STATE = 0x7E, ///< HD broadcast of standby state + MSG_ID_HD_DISINFECT_STANDBY_DATA = 0x7E, ///< HD broadcast of standby disinfect data MSG_ID_UI_DISINFECT_REQUEST = 0x7F, ///< UI disinfection selection MSG_ID_HD_DISINFECT_RESPONSE = 0x80, ///< HD response to disinfection selection MSG_ID_UI_CHEM_DISINFECT_CONFIRM = 0x81, ///< UI confirmation of chemical disinfection start