Index: AlarmDefs.h =================================================================== diff -u -rebd5ad978a2c6fc04690a29be8c50407484d2e27 -r6a6a8e5f0574c4d4a7d48cf6feba3a3cc0e9d3da --- AlarmDefs.h (.../AlarmDefs.h) (revision ebd5ad978a2c6fc04690a29be8c50407484d2e27) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 6a6a8e5f0574c4d4a7d48cf6feba3a3cc0e9d3da) @@ -192,7 +192,7 @@ ALARM_ID_NVDATA_EEPROM_OPS_FAILURE = 154, ///< EEPROM operations (read, write, erase) failure ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC = 155, ///< DG software configuration record invalid CRC ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR = 156, ///< HW usage data (treatment time in HD and total consumed water in DG) failure - ALARM_ID____AVAILABLE_23 = 157, ///< This alarm ID is available for use + ALARM_ID_INLET_WATER_TEMPERATURE_OUT_OF_RANGE = 157, ///< Inlet water temperature is not within specified range ALARM_ID____AVAILABLE_24 = 158, ///< This alarm ID is available for use ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 159, ///< Dialysate inlet pump failed flow vs. motor speed check. ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH = 160, ///< Blood pump rotor speed too high @@ -605,7 +605,7 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_EEPROM_OPS_FAILURE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_LOW, 799, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID____AVAILABLE_23, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_INLET_WATER_TEMPERATURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID____AVAILABLE_24, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH, 0, ALARM_ID_NO_ALARM }, @@ -933,7 +933,7 @@ { 799, ALARM_ID_NVDATA_EEPROM_OPS_FAILURE }, { 799, ALARM_ID_HD_SW_CONFIG_RECORD_INVALID_CRC }, { 799, ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR }, - { 799, ALARM_ID____AVAILABLE_23 }, + { 799, ALARM_ID_INLET_WATER_TEMPERATURE_OUT_OF_RANGE }, { 799, ALARM_ID____AVAILABLE_24 }, { 799, ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK }, { 799, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH },