Index: AlarmDefs.h =================================================================== diff -u -r1d702021ecc4d4cfbc487f8904212001c6070f80 -r6db5724424b780b1bbcc297b6389a7f077324ef3 --- AlarmDefs.h (.../AlarmDefs.h) (revision 1d702021ecc4d4cfbc487f8904212001c6070f80) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 6db5724424b780b1bbcc297b6389a7f077324ef3) @@ -1254,7 +1254,7 @@ { 93 , ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: RO Permeate Cond High | DG outlet primary conductivity sensor out of range | If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. | "}, { 94 , ALARM_ID_DG_PRESSURE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Pressure Sensor Range | DG pressure sensor out of range | If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. | "}, { 95 , ALARM_ID_DG_WATCHDOG_EXPIRED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: WD Error | DG watchdog expired | If the watchdog signal is low while not in POST. | "}, - { 96 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE , "DG | Incoming Water Temperature High | Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nRsume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Temp High | DG inlet water temperature out of range | If the inlet water temperature is > 37 C. | "}, + { 96 , ALARM_ID_DG_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE , "DG | Incoming Water Temperature High | Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | Dial. Gen: Inlet Water Temp High | DG inlet water temperature out of range | If the inlet water temperature is > 37 C. | "}, { 97 , ALARM_ID_DG_FILL_CONDUCTIVITY_OUT_OF_RANGE , "DG | Conductivity Out of Range | Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. | Dial. Gen: Dialysate Cond. | DG fill conductivity out of range | If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. | "}, { 98 , ALARM_ID_HD_BATTERY_COMM_FAULT , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Battery Com | HD battery or battery charger communication fault | If battery I2C communication fault persistence count > 5. | "}, { 99 , ALARM_ID_HD_SYRINGE_PUMP_STALL , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Syringe Pump Stall | HD syringe pump stall alarm | If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. | "},