Index: AlarmDefs.h =================================================================== diff -u -rc12b4970db3be16f05b60332d0b27a4387aa1bd6 -r90d4fcc52b9f19f3e1ae3d0bb745bbcaf0adc4f4 --- AlarmDefs.h (.../AlarmDefs.h) (revision c12b4970db3be16f05b60332d0b27a4387aa1bd6) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 90d4fcc52b9f19f3e1ae3d0bb745bbcaf0adc4f4) @@ -1174,7 +1174,7 @@ { 15 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP , "HD | Dial. Gen: Dialysate Temp High | Dialysate temperature is high\n-Treatment is paused and will be ready to resume when \ndialysate is within target range. | Dialysate temperature is out of high safety range | If the dialysate temperature is > 46 C for a certain period of time. | "}, { 16 , ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE , "HD | HD POST: Primary Alarm Current | Call service to schedule a repair. | HD alarm audio failed POST | If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output | "}, { 17 , ALARM_ID_HD_UI_POST_TIMEOUT , "HD | HD POST: UI POST Not Complete | Call service to schedule a repair. | HD UI POST failed | If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. | "}, - { 18 , ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_DG , "HD | Loss of DG Comm (DG ACK) | Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. | HD to DG comm error. | HD message to DG that requires acknowledge was not acknowledged within a certain time. | "}, + { 18 , ALARM_ID_HD_CAN_MESSAGE_NOT_ACKED_BY_DG , " | | | | | "}, { 19 , ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP , "HD | Dial. Gen: Dialysate Temp High | Dialysate temperature is high\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. | Dialysate temperature is above the target temperature | If the dialysate temperature is above the target temperature for a certain period of time. | "}, { 20 , ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD , "DG | DG POST: Conductivity Cal | Call service to schedule a repair. | DG conductivity sensors invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, { 21 , ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD , "DG | DG POST: Drain Line Vol Cal | Call service to schedule a repair. | DG drain line volume invalid calibration record | If the calibration data that was received from NVDataMgmt is NOT legitimate | "}, @@ -1319,7 +1319,7 @@ { 160 , ALARM_ID_HD_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH , "HD | HD Fault: BP Rotor Max Speed | Call service to schedule a repair. | Blood pump rotor speed too high | If blood pump motor speed exceeded 100.0 RPM. | "}, { 161 , ALARM_ID_DG_COMM_TOO_MANY_BAD_CRCS , "DG | DG Fault: HD-DG Comm | Call service to schedule a repair. | Too many bad CRCs detected on received system messages | If >= 5 bad CRCs detected on received CAN messages within 10 minutes. | "}, { 162 , ALARM_ID_DG_FPGA_CLOCK_SPEED_CHECK_FAILURE , "DG | DG Fault: Proc-FPGA Clock | Call service to schedule a repair. | FPGA clock speed failure | If the FPGA clock speed is relatively close to the processor clock for a certain period of time. | "}, - { 163 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM , "HD | Dial. Gen: UF Rate Check | Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system (shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a repair. | HD sees primary load cell for reservoir 1 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, + { 163 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM , "HD | Dial. Gen: UF Rate Check | Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system\n(shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a\nrepair. | HD sees primary load cell for reservoir 1 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, { 164 , ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM , "HD | Dial. Gen: UF Rate Check | Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system \n(shock/vibration).\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a \nrepair. | HD sees primary load cell for reservoir 2 change too much too fast | If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). | "}, { 165 , ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_ALARM , "HD | Blood Stop: Blood Clot Risk | Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. | HD in treatment rinseback operation timeout | If the maximum time inrinseback operation has been exceeded. | "}, { 166 , ALARM_ID_AVAILABLE_3 , " | | | | | "}, @@ -1396,7 +1396,7 @@ { 237 , ALARM_ID_HD_PRIME_SALINE_DIALYZER_TIME_OUT , "HD | Process: Pre-Tx Prime Timeout | Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. | HD prime saline dialyzer time out alarm | If timeout, just transition to HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE | "}, { 238 , ALARM_ID_HD_NO_CARTRIDGE_LOADED , "HD | Process: Pre-Tx Dry Self-Test | Self-test has detected cartridge is not installed.\n- Check that a cartridge is properly installed. | HD no cartridge loaded or installed improperly alarm | If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. | "}, { 239 , ALARM_ID_HD_CARTRIDGE_REMOVAL_FAILURE , "HD | Process: Post-Tx Remove Cartridge | Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. | HD fail to remove cartridge alarm | In the post treatment state, if the cartridge has not been removed. | "}, - { 240 , ALARM_ID_DG_BICARB_CONDUCTIVITY_OUT_OF_RANGE , "DG | Dial. Gen: Bicarb Conc. Cond. | Bicarbonate concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. | Bicarb conductivity out of range during bicarb pump check alarm | If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. | "}, + { 240 , ALARM_ID_DG_BICARB_CONDUCTIVITY_OUT_OF_RANGE , "DG | Dial. Gen: Bicarb Conc. Cond. | Bicarbonate conductivity during prime is out of range.\n- Check correct concentrate is connected and full.\n- Resume to reprime concentrate. | Bicarb conductivity out of range during bicarb pump check alarm | If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. | "}, { 241 , ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT , "DG | Cleaning: Resv Drain Timeout | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG reservoir drain time out | If the DG reservoir drain is not complete in 120 secs. | "}, { 242 , ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT , "DG | Cleaning: Resv Fill Timeout | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG reservoir fill time out | If the DG reservoir fill is not complete in 120 secs. | "}, { 243 , ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT , "DG | Cleaning: Fluid Loss Detected | Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. | DG reservoir leak time out | If the reservoir level is drifted by 100 mL for more than 30 seconds. | "},