Index: AlarmDefs.h =================================================================== diff -u -rcd74aefce2b308e6d27ae8c9e88e9d9dcb7a177d -raabd24dd2ce2ba4c87f5c81025413996e20dc57f --- AlarmDefs.h (.../AlarmDefs.h) (revision cd74aefce2b308e6d27ae8c9e88e9d9dcb7a177d) +++ AlarmDefs.h (.../AlarmDefs.h) (revision aabd24dd2ce2ba4c87f5c81025413996e20dc57f) @@ -1284,10 +1284,10 @@ { 123 , ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Trimmer Heater Control | DG trimmer heater voltage out of range | If the trimmer voltage is out of range. | "}, { 124 , ALARM_ID_HD_END_OF_TREATMENT_HIGH , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk | HD end of treatment alarm (high priority - user still not acting to end treatment) | In treatment end paused state for > 5 minutes. | "}, { 125 , ALARM_ID_HD_TREATMENT_STOPPED_NO_RINSEBACK , "HD | Blood Pump Stop Time Exceeded | The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk | Blood sitting too long after treatment stopped by user alarm (> 5 min) | If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. | "}, - { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Treatment must be terminated without rinseback.\n\nBlood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, + { 126 , ALARM_ID_HD_BLOOD_LEAK_DETECTED , "HD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press OK to restart blood pump. Dialysate will remain in bypass.\n\nBlood Stop: Dialyzer Blood Leak | HD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, { 127 , ALARM_ID_HD_VENOUS_PRESSURE_LOW , "HD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- If there has been an access disconnection or needle\n dislodgement, follow facility policy to address and\nterminate treatment without rinseback.\n- If access connections are properly secured, \npress RESUME to continue treatment.\n\nBlood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 128 , ALARM_ID_HD_VENOUS_BUBBLE_DETECTED , "HD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, - { 129 , ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , " | | \n\n | | | "}, + { 129 , ALARM_ID_HD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "HD | Blood Leak | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection,\n press END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using test strip.\n- Follow facility policy to RESUME or END TREATMENT.\n\nBlood Stop: Dialyzer Blood Leak | HD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "}, { 130 , ALARM_ID_HD_VENOUS_PRESSURE_HIGH , "HD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 131 , ALARM_ID_HD_ARTERIAL_PRESSURE_LOW , "HD | Low Arterial Pressure | Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. Low | Arterial pressure too low during treatment | If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, { 132 , ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH , "HD | High Arterial Pressure | Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. High | Arterial pressure too high during treatment | If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "},