Index: AlarmDefs.h =================================================================== diff -u -r33e10fc60db6cfa75d1c543a937f2eb08147d458 -rc736e4356576ade478873b1b13fb9cc1aad6388a --- AlarmDefs.h (.../AlarmDefs.h) (revision 33e10fc60db6cfa75d1c543a937f2eb08147d458) +++ AlarmDefs.h (.../AlarmDefs.h) (revision c736e4356576ade478873b1b13fb9cc1aad6388a) @@ -33,243 +33,253 @@ #define ALM_ESC_90_MIN (90 * 60000) ///< Number of ms in 90 minutes. /// Enumeration of alarm IDs. -enum Alarm_List -{ +enum Alarm_List +{ ALARM_ID_NO_ALARM = 0, ///< Indication for when no alarms is a possible situation - ALARM_ID_HD_SOFTWARE_FAULT = 1, ///< HD Software fault. Software found itself in an unexpected state - ALARM_ID_STUCK_BUTTON_TEST_FAILED = 2, ///< Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up - ALARM_ID_HD_FPGA_POST_TEST_FAILED = 3, ///< HD FPGA POST failure - ALARM_ID_WATCHDOG_POST_TEST_FAILED = 4, ///< Watchdog POST failure - ALARM_ID_UI_COMM_POST_FAILED = 5, ///< UI communication POST failure. UI failed to communicate within a reasonable time after power up - ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK = 6, ///< Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running - ALARM_ID_BLOOD_PUMP_OFF_CHECK = 7, ///< Blood pump failed motor off check. Measured speed while commanded off - ALARM_ID_BLOOD_PUMP_MC_DIRECTION_CHECK = 8, ///< Blood pump failed motor direction check. Measured vs. commanded - ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_CHECK = 9, ///< Blood pump failed rotor speed check. Mismatch with rotor and motor speeds - ALARM_ID_DIAL_IN_PUMP_MC_CURRENT_CHECK = 10, ///< Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running - ALARM_ID_DIAL_IN_PUMP_OFF_CHECK = 11, ///< Dialysis inlet pump failed motor off check. Measured speed while commanded off - ALARM_ID_DIAL_IN_PUMP_MC_DIRECTION_CHECK = 12, ///< Dialysis inlet pump failed motor direction check. Measured vs. commanded - ALARM_ID_DIAL_IN_PUMP_ROTOR_SPEED_CHECK = 13, ///< Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds - ALARM_ID_DIAL_OUT_PUMP_MC_CURRENT_CHECK = 14, ///< Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running - ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK = 15, ///< Dialysis outlet pump failed motor off check. Measured speed while commanded off - ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK = 16, ///< Dialysis outlet pump failed motor direction check. Measured vs. commanded - ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK = 17, ///< Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds - ALARM_ID_WATCHDOG_EXPIRED = 18, ///< Watchdog expired error. Watchdog was not "pet" within the appropriate time - ALARM_ID_RTC_COMM_ERROR = 19, ///< Real-time clock communication error - ALARM_ID_RTC_CONFIG_ERROR = 20, ///< Real-time clock configuration error - ALARM_ID_DG_COMM_TIMEOUT = 21, ///< Dialysate generator communication timeout - ALARM_ID_UI_COMM_TIMEOUT = 22, ///< User interface communication timeout - ALARM_ID_COMM_TOO_MANY_BAD_CRCS = 23, ///< Too many bad CRCs detected on received system messages - ALARM_ID_TREATMENT_STOPPED_BY_USER = 24, ///< Treatment stopped by user action - pressed stop button - ALARM_ID_BLOOD_SITTING_WARNING = 25, ///< Blood sitting too long warning (>4 min) - ALARM_ID_BLOOD_SITTING_TOO_LONG_NO_RESUME = 26, ///< Blood sitting too long alarm (>5 min) - ALARM_ID_BLOOD_SITTING_TOO_LONG_NO_RINSEBACK = 27, ///< Blood sitting too long alarm (>10 min) - ALARM_ID_CAN_MESSAGE_NOT_ACKED = 28, ///< System message that required acknowledgment was not acknowledged - ALARM_ID_OCCLUSION_BLOOD_PUMP = 29, ///< Blood pump occlusion detected - ALARM_ID_OCCLUSION_DIAL_IN_PUMP = 30, ///< Dialysate inlet pump occlusion detected - ALARM_ID_OCCLUSION_DIAL_OUT_PUMP = 31, ///< Dialysate outlet pump occlusion detected - ALARM_ID_ARTERIAL_PRESSURE_LOW = 32, ///< Arterial pressure too low during treatment - ALARM_ID_ARTERIAL_PRESSURE_HIGH = 33, ///< Arterial pressure too high during treatment - ALARM_ID_VENOUS_PRESSURE_LOW = 34, ///< Venous pressure too low during treatment - ALARM_ID_VENOUS_PRESSURE_HIGH = 35, ///< Venous pressure too high during treatment - ALARM_ID_UF_RATE_TOO_HIGH_ERROR = 36, ///< Ultrafiltration rate is too high error during treatment - ALARM_ID_UF_VOLUME_ACCURACY_ERROR = 37, ///< Ultrafiltration volume accuracy error during treatment - ALARM_ID_RTC_BATTERY_LOW = 38, ///< RTC battery low - ALARM_ID_RTC_OR_TIMER_ACCURACY_FAILURE = 39, ///< RTC or MCU timer inaccurate - ALARM_ID_RTC_RAM_OPS_ERROR = 40, ///< RTC (i.e read time) or RAM operations failure (read or write) - ALARM_ID_NVDATA_EEPROM_OPS_FAILURE = 41, ///< EEPROM operations (read, write, erase) failure - ALARM_ID_NVDATA_MFG_RECORD_CRC_ERROR = 42, ///< Manufacturing record CRC failure - ALARM_ID_NVDATA_SRVC_RECORD_CRC_ERROR = 43, ///< Service record CRC failure - ALARM_ID_NVDATA_CAL_RECORD_CRC_ERROR = 44, ///< Calibration record CRC failure - ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR = 45, ///< HW usage data (treatment time in HD and total consumed water in DG) failure - ALARM_ID_AVAILABLE_1 = 46, ///< Last disinfection date CRC error - ALARM_ID_AVAILABLE_2 = 47, ///< Pressure at outlet of RO pump is out of range - ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT = 48, ///< DG temperature sensors ADC fault - ALARM_ID_HD_FPGA_COMM_TIMEOUT = 49, ///< HD FPGA communication down for too long - ALARM_ID_HD_COMM_TIMEOUT = 50, ///< HD communication timeout - ALARM_ID_VALVE_CONTROL_FAILURE = 51, ///< FPGA not accepting commanded valve states - ALARM_ID_BLOOD_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 52, ///< Blood pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed - ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK = 53, ///< Dialysate inlet pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed - ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT = 54, ///< HD time out on prime sub-mode purge air state. - ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK = 55, ///< Blood pump failed motor speed check. Measured vs. commanded - ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK = 56, ///< Dialysate inlet pump failed motor speed check. Measured vs. commanded - ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK = 57, ///< Dialysate outlet pump failed motor speed check. Measured vs. commanded - ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH = 58, ///< Blood pump rotor speed too high - ALARM_ID_INLET_WATER_LOW_CONDUCTIVITY = 59, ///< Inlet water conductivity too low - ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY = 60, ///< Inlet water conductivity too high - ALARM_ID_DG_SOFTWARE_FAULT = 61, ///< DG software fault. Software found itself in an unexpected state - ALARM_ID_CONDUCTIVITY_SENSOR_FAULT = 62, ///< Conductivity sensor fault - ALARM_ID_INLET_WATER_LOW_TEMPERATURE = 63, ///< Inlet water temperature too low - ALARM_ID_INLET_WATER_HIGH_TEMPERATURE = 64, ///< Inlet water temperature too high - ALARM_ID_PRESSURE_SENSOR_FAULT = 65, ///< Pressure sensor fault - ALARM_ID_INLET_WATER_LOW_PRESSURE = 66, ///< Inlet water pressure too low - ALARM_ID_HD_CRITICAL_DATA_ERROR = 67, ///< HD critical data integrity check failed - ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE = 68, ///< HD accelerometer failed POST - ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE = 69, ///< DG accelerometer failed POST - ALARM_ID_HD_EXCESSIVE_TILT = 70, ///< HD tilt exceeds maximum - ALARM_ID_DG_EXCESSIVE_TILT = 71, ///< DG tilt exceeds maximum - ALARM_ID_HD_SHOCK = 72, ///< HD experienced large acceleration - ALARM_ID_DG_SHOCK = 73, ///< DG experienced large acceleration - ALARM_ID_HD_ACCELEROMETER_FAILURE = 74, ///< HD accelerometer error (no readings or FPGA reports error) - ALARM_ID_DG_ACCELEROMETER_FAILURE = 75, ///< DG accelerometer error (no readings or FPGA reports error) - ALARM_ID_DG_CRITICAL_DATA_ERROR = 76, ///< DG critical data integrity check failed - ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE = 77, ///< RO rejection ratio out of range - ALARM_ID_INLET_WATER_PRESSURE_FAULT = 78, ///< Inlet water pressure fault - ALARM_ID_HD_VALVE_HOMING_FAILED = 79, ///< HD valve homing failed - ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT = 80, ///< HD valve transition time out - ALARM_ID_HD_VALVE_NOT_FUNCTIONAL = 81, ///< HD valve not functional - ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE = 82, ///< HD valve current out of range - ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE = 83, ///< HD valve position out of target - ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS = 84, ///< Air trap level sensors reporting illegal combination of air/fluid - ALARM_ID_AIR_TRAP_FILL_DURING_PRIME = 85, ///< Air trap fill timeout during priming - ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT = 86, ///< Air trap fill timeout during treatment - ALARM_ID_BLOOD_FLOW_SIGNAL_STRENGTH_TOO_LOW = 87, ///< Blood flow sensor signal strength too low - ALARM_ID_DIALYSATE_FLOW_SIGNAL_STRENGTH_TOO_LOW = 88, ///< Dialysate flow sensor signal strength too low - ALARM_ID_EMPTY_SALINE_BAG = 89, ///< HD Saline bag is empty - ALARM_ID_SALINE_BOLUS_VOLUME_CHECK_FAILURE = 90, ///< HD Saline bolus volume check failure - ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT = 91, ///< FPGA reports failure reading arterial pressure sensor - ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT = 92, ///< FPGA reports failure reading venous pressure sensor - ALARM_ID_UV_REACTOR_NOT_HEALTHY = 93, ///< Inlet or Outlet UV reactor not healthy - ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE = 94, ///< Concentrate conductivity after adding acid out of range alarm - ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE = 95, ///< Concentrate conductivity after adding bicarb out of range alarm - ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE = 96, ///< RO pump flow rate out of range - ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE = 97, ///< DG thermistors/sensors temperature out of range - ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE = 98, ///< DG fan RPM out of range - ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME = 99, ///< DG dialysate fill runs out of time - ALARM_ID_CP1_SPEED_CONTROL_ERROR = 100, ///< DG concentrate pump CP1 speed control error - ALARM_ID_CP2_SPEED_CONTROL_ERROR = 101, ///< DG concentrate pump CP2 speed control error - ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE = 102, ///< DG flow rate out of upper range - ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE = 103, ///< DG RO pump pressure out of range - ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE = 104, ///< DG drain pump RPM out of range - ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE = 105, ///< DG heaters cold junction temperature out of range - ALARM_ID_DG_HEATERS_FAULT = 106, ///< DG heaters fault - ALARM_ID_END_OF_TREATMENT_WARNING = 107, ///< HD end of treatment warning - ALARM_ID_END_OF_TREATMENT_ALARM = 108, ///< HD end of treatment alarm (user not acting to end treatment) - ALARM_ID_END_OF_TREATMENT_HIGH = 109, ///< HD end of treatment alarm (high priority - user still not acting to end treatment) - ALARM_ID_AVAILABLE_3 = 110, ///< DG heat disinfect cycle failed - ALARM_ID_RO_PUMP_RAMP_UP_TO_FLOW_TIMEOUT = 111, ///< DG RO pump ramp up to target flow time out - ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE = 112, ///< DG flow rate out of lower range - ALARM_ID_DRAIN_PUMP_OFF_FAULT = 113, ///< DG drain pump off fault - ALARM_ID_RO_PUMP_OFF_FAULT = 114, ///< DG RO pump off fault - ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT = 115, ///< HD prime dialysate dialyzer time out alarm - ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT = 116, ///< HD prime dialysate bypass time out alarm - ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY = 117, ///< HD prime completed low priority alarm - ALARM_ID_PRIME_COMPLETED_MEDIUM_PRIORITY = 118, ///< HD prime completed medium priority alarm - ALARM_ID_PRIME_COMPLETED_HIGH_PRIORITY = 119, ///< HD prime completed high priority alarm - ALARM_ID_PRIME_OUT_OF_TIME = 120, ///< HD prime out of time alarm - ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT = 121, ///< HD requests DG command with invalid parameter fault - ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM = 122, ///< HD sees primary load cell for reservoir 1 change too much too fast - ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM = 123, ///< HD sees primary load cell for reservoir 1 change too much too fast - ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK = 124, ///< HD in treatment stopped sub-mode after rinseback completed (no escalation) - ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM = 125, ///< HD in treatment rinseback operation timeout - ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM = 126, ///< HD in treatment re-circ sub-mode for too long - ALARM_ID_CARTRIDGE_DOOR_OPENED = 127, ///< HD cartridge door opened alarm - ALARM_ID_AVAILABLE_4 = 128, ///< HD pre-treatment mode no cartridge self-tests failure - ALARM_ID_INSTALL_NEW_CARTRIDGE = 129, ///< HD needs new cartridge to be installed - ALARM_ID_CARTRIDGE_INSTALLED_IMPROPERLY = 130, ///< HD cartridge installed improperly alarm - ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE = 131, ///< HD rinseback volume check failure - ALARM_ID_NO_CARTRIDGE_LOADED = 132, ///< HD no cartridge loaded alarm - ALARM_ID_END_TREATMENT_TIMEOUT_ALARM = 133, ///< HD end treatment sub-mode timeout alarm - ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE = 134, ///< HD blood prime volume check failure - ALARM_ID_DIALYSATE_TEMPERATURE_TOO_HIGH = 135, ///< HD dialysate temperature too high alarm - ALARM_ID_DIALYSATE_TEMPERATURE_TOO_LOW = 136, ///< HD dialysate temperature too low alarm - ALARM_ID_PRE_TREATMENT_DRY_SELF_TEST_FAILURE = 137, ///< HD pre-treatment mode dry self-tests failure - ALARM_ID_PRE_TREATMENT_WET_SELF_TEST_FAILURE = 138, ///< HD pre-treatment mode wet self-tests failure - ALARM_ID_DG_FLOW_METER_CHECK_FAILURE = 139, ///< DG flow meter check failure alarm - ALARM_ID_AVAILABLE_5 = 140, ///< DG RO pump ramp up exceed maximum allowed number of retries - ALARM_ID_NVDATAMGMT_GROUP_RECORD_CRC_INVALID = 141, ///< DG/HD NV data management overall group record CRC invalid - ALARM_ID_NVDATAMGMT_INDIVIDUAL_RECORD_CRC_INVALID = 142, ///< DG/HD NV data management individual record CRC invalid - ALARM_ID_NVDATAMGMT_INDIVIDUAL_RECORD_TIME_INVALID = 143, ///< DG/HD NV data management individual record time invalid - ALARM_ID_HD_AC_POWER_LOST = 144, ///< HD AC power lost alarm - ALARM_ID_DG_RESTARTED_FAULT = 145, ///< HD reports DG restarted fault - ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE = 146, ///< Dialysate conductivity out of range - ALARM_ID_HD_FLUID_LEAK_DETECTED = 147, ///< HD fluid leak detected alarm - ALARM_ID_DG_FLUID_LEAK_DETECTED = 148, ///< DG fluid leak detected alarm - ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR = 149, ///< HD syringe pump ADC error - ALARM_ID_HD_SYRINGE_DETECTED = 150, ///< HD syringe detected alarm - ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR = 151, ///< HD syringe pump volume check error - ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR = 152, ///< HD syringe pump speed check error - ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY = 153, ///< HD syringe empty alarm - ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION = 154, ///< HD syringe pump occlusion alarm - ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED = 155, ///< HD syringe pump syringe removed alarm - ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR = 156, ///< HD syringe pump not stopped in off state error - ALARM_ID_HD_BLOOD_LEAK_DETECTED = 157, ///< HD blood leak detected alarm - ALARM_ID_CARTRIDGE_REMOVAL_FAILURE = 158, ///< HD fail to remove cartridge alarm - ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT = 159, ///< DG reservoir drain time out - ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT = 160, ///< DG reservoir fill time out - ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT = 161, ///< DG reservoir leak time out - ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT = 162, ///< DG heat disinfect target temperature time out (could not reach to temperature) - ALARM_ID_DG_TEMP_SENSORS_DIFF_OUT_OF_RANGE = 163, ///< DG temperature sensors difference out of range - ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT = 164, ///< DG heat disinfect inlet pressure and temperature sensors out of range - ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_AND_TEMP_OUT = 165, ///< DG heat disinfect inlet conductivity and temperature out of range - ALARM_ID_DG_BAD_INLET_WATER_QUALITY = 166, ///< Inlet water quality is bad (temperature, conductivity, pressure) - ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE = 167, ///< HD blood pump occlusion self-test failure alarm - ALARM_ID_HD_DIP_OCCLUSION_SELF_TEST_FAILURE = 168, ///< HD dialysate inlet pump occlusion self-test failure alarm - ALARM_ID_HD_DOP_OCCLUSION_SELF_TEST_FAILURE = 169, ///< HD dialysate outlet pump occlusion self-test failure alarm - ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE = 170, ///< HD arterial pressure self-test failure alarm - ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE = 171, ///< HD venous pressure self-test failure alarm - ALARM_ID_HD_BLOOD_FLOW_STATUS_SELF_TEST_FAILURE = 172, ///< HD blood flow meter status check self-test failure alarm - ALARM_ID_HD_DIALYSATE_FLOW_STATUS_SELF_TEST_FAILURE = 173, ///< HD dialysate flow meter status check self-test failure alarm - ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAILURE = 174, ///< HD blood leak detector self-test failure alarm - ALARM_ID_HD_BLOOD_LEAK_FAULT = 175, ///< HD blood leak detector fault - ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION = 176, ///< DG load cells invalid calibration - ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE = 177, ///< HD monitored voltage is out of range - ALARM_ID_DG_PRIMARY_HEATERS_INTERNAL_TEMP_OUT_OF_RANGE = 178, ///< DG primary heaters internal temperature out of range - ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE = 179, ///< DG trimmer heater internal temperature out of range - ALARM_ID_DG_HEATERS_ON_WITH_NO_FLOW_TIMEOUT = 180, ///< DG heaters on with no flow time out - ALARM_ID_DIALYSATE_CONDUCTIVITY_FAULT = 181, ///< DG dialysate generation conductivity fault alarm - ALARM_ID_DIALYSATE_TEMPERATURE_HIGH = 182, ///< HD dialysate temperature too high alarm - ALARM_ID_DIALYSATE_TEMPERATURE_LOW = 183, ///< HD dialysate temperature too low alarm - ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE = 184, ///< DG monitored voltage is out of range - ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT = 185, ///< DG drain circulation line timeout - ALARM_ID_DG_INVALID_LOAD_CELL_VALUE = 186, ///< DG invalid load cell value - ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR = 187, ///< HD syringe pump direction (from encoder) error - ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR = 188, ///< HD syringe pump direction (from controller) error - ALARM_ID_HD_SYRINGE_PUMP_FAULT = 189, ///< HD syringe pump fault reported by FPGA - ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR = 190, ///< HD syringe pump over travel error - ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM = 191, ///< HD syringe pump not enough Heparin alarm - ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR = 192, ///< HD syringe pump DAC write failure - ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR = 193, ///< HD syringe pump is running while the BP is off - ALARM_ID_DG_CHEM_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT = 194, ///< DG chemical disinfect inlet pressure and temperature sensors out of range - ALARM_ID_DG_CHEM_DISINFECT_INLET_COND_AND_TEMP_OUT = 195, ///< DG chemical disinfect inlet conductivity and temperature out of range - ALARM_ID_DG_FPGA_COMM_TIMEOUT = 196, ///< DG FPGA communication down for too long - ALARM_ID_DG_LOAD_CELL_ADC_ERROR = 197, ///< DG load cell ADC error - ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR = 198, ///< HD venous pressure sensor not being read - ALARM_ID_HD_VENOUS_PRESSURE_SENSOR_TEMP_OUT_OF_RANGE = 199, ///< HD venous pressure sensor temperature out of range error - ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR = 200, ///< HD BP occlusion sensor not being read - ALARM_ID_HD_DPI_OCCLUSON_READ_TIMEOUT_ERROR = 201, ///< HD DPi occlusion sensor not being read - ALARM_ID_HD_DPO_OCCLUSION_READ_TIMEOUT_ERROR = 202, ///< HD DPo occlusion sensor not being read - ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR = 203, ///< HD BP occlusion sensor error - ALARM_ID_HD_DPI_OCCLUSION_SENSOR_ERROR = 204, ///< HD DPi occlusion sensor error - ALARM_ID_HD_DPO_OCCLUSION_SENSOR_ERROR = 205, ///< HD DPo occlusion sensor error - ALARM_ID_DG_CONCENTRATE_PUMP_FAULT = 206, ///< DG concentrate pump fault - ALARM_ID_HD_BP_FLOW_READ_TIMEOUT_ERROR = 207, ///< HD blood flow sensor not being read - ALARM_ID_HD_BP_FLOW_SENSOR_ERROR = 208, ///< HD blood flow sensor error - ALARM_ID_HD_DP_FLOW_READ_TIMEOUT_ERROR = 209, ///< HD dialysate flow sensor not being read - ALARM_ID_HD_DP_FLOW_SENSOR_ERROR = 210, ///< HD dialysate flow sensor error - ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT = 211, ///< DG chemical disinfect could not reach to target temperature - ALARM_ID_DG_FPGA_POST_TEST_FAILED = 212, ///< DG FPGA POST failure - ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE = 213, ///< HD arterial air bubble detector self-test failure - ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED_RINSEBACK = 214, ///< HD arterial air bubble detected rinseback alarm - ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE = 215, ///< DG heaters failed POST - ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 216, ///< HD venous air bubble detector self-test failure - ALARM_ID_AVAILABLE_6 = 217, ///< HD venous air bubble detector treatment fault - ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK = 218, ///< HD venous air bubble detected rinseback alarm - ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED = 219, ///< HD arterial air bubble detected alarm - ALARM_ID_HD_VENOUS_BUBBLE_DETECTED = 220, ///< HD venous air bubble detected alarm - ALARM_ID_HD_BP_FLOW_SLOW_READ_TIMEOUT_ERROR = 221, ///< HD blood flow sensor not being read - slow packet - ALARM_ID_HD_DP_FLOW_SLOW_READ_TIMEOUT_ERROR = 222, ///< HD dialysate flow sensor not being read - slow packet - ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR = 223, ///< HD pump direction status error - ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE = 224, ///< DG load cells weight out of range before tare - ALARM_ID_HD_BATTERY_COMM_FAULT = 225, ///< HD battery communication fault - ALARM_ID_HD_SYRINGE_PUMP_STALL = 226, ///< HD syringe pump stall alarm - ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED = 227, ///< HD firmware image integrity POST test failed - ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED = 228, ///< DG firmware image integrity POST test failed - ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT = 229, ///< HD no cartridge self-test timeout - ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT = 230, ///< HD dry self-test timeout - ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT = 231, ///< DG temperature sensor error flag fault. - ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH = 232, ///< HD active reservoir is full during treatment and DG not ready to switch reservoirs. - ALARM_ID_HD_BLOOD_FLOW_INVALID_CALIBRATION = 233, ///< HD blood flow invalid calibration - ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CALIBRATION = 234, ///< HD dialysate flow invalid calibration + // POST machine fault alarms + ALARM_ID_STUCK_BUTTON_TEST_FAILED, ///< Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up + ALARM_ID_HD_FPGA_POST_TEST_FAILED, ///< HD FPGA POST failure + ALARM_ID_DG_FPGA_POST_TEST_FAILED, ///< DG FPGA POST failure + ALARM_ID_WATCHDOG_POST_TEST_FAILED, ///< Watchdog POST failure + ALARM_ID_UI_COMM_POST_FAILED, ///< UI communication POST failure. UI failed to communicate within a reasonable time after power up + ALARM_ID_RTC_CONFIG_ERROR, ///< Real-time clock configuration error + ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE, ///< HD accelerometer failed POST + ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE, ///< DG accelerometer failed POST + ALARM_ID_RTC_OR_TIMER_ACCURACY_FAILURE, ///< RTC or MCU timer inaccurate + ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE, ///< DG heaters failed POST + ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED, ///< HD firmware image integrity POST test failed + ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED, ///< DG firmware image integrity POST test failed + ALARM_ID_HD_BLOOD_FLOW_INVALID_CALIBRATION, ///< HD blood flow invalid calibration + ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CALIBRATION, ///< HD dialysate flow invalid calibration + + // Service alarms + + // Machine fault alarms + ALARM_ID_HD_SOFTWARE_FAULT, ///< HD Software fault. Software found itself in an unexpected state + ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK, ///< Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running + ALARM_ID_BLOOD_PUMP_OFF_CHECK, ///< Blood pump failed motor off check. Measured speed while commanded off + ALARM_ID_BLOOD_PUMP_MC_DIRECTION_CHECK, ///< Blood pump failed motor direction check. Measured vs. commanded + ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_CHECK, ///< Blood pump failed rotor speed check. Mismatch with rotor and motor speeds + ALARM_ID_DIAL_IN_PUMP_MC_CURRENT_CHECK, ///< Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running + ALARM_ID_DIAL_IN_PUMP_OFF_CHECK, ///< Dialysis inlet pump failed motor off check. Measured speed while commanded off + ALARM_ID_DIAL_IN_PUMP_MC_DIRECTION_CHECK, ///< Dialysis inlet pump failed motor direction check. Measured vs. commanded + ALARM_ID_DIAL_IN_PUMP_ROTOR_SPEED_CHECK, ///< Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds + ALARM_ID_DIAL_OUT_PUMP_MC_CURRENT_CHECK, ///< Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running + ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK, ///< Dialysis outlet pump failed motor off check. Measured speed while commanded off + ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK, ///< Dialysis outlet pump failed motor direction check. Measured vs. commanded + ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK, ///< Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds + ALARM_ID_WATCHDOG_EXPIRED, ///< Watchdog expired error. Watchdog was not "pet" within the appropriate time + ALARM_ID_UI_COMM_TIMEOUT, ///< User interface communication timeout + ALARM_ID_COMM_TOO_MANY_BAD_CRCS, ///< Too many bad CRCs detected on received system messages + ALARM_ID_CAN_MESSAGE_NOT_ACKED, ///< System message that required acknowledgment was not acknowledged + ALARM_ID_UF_RATE_TOO_HIGH_ERROR, ///< Ultrafiltration rate is too high error during treatment + ALARM_ID_UF_VOLUME_ACCURACY_ERROR, ///< Ultrafiltration volume accuracy error during treatment + ALARM_ID_HD_FPGA_COMM_TIMEOUT, ///< HD FPGA communication down for too long + ALARM_ID_VALVE_CONTROL_FAILURE, ///< FPGA not accepting commanded valve states + ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK, ///< Blood pump failed motor speed check. Measured vs. commanded + ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK, ///< Dialysate inlet pump failed motor speed check. Measured vs. commanded + ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK, ///< Dialysate outlet pump failed motor speed check. Measured vs. commanded + ALARM_ID_HD_CRITICAL_DATA_ERROR, ///< HD critical data integrity check failed + ALARM_ID_DG_CRITICAL_DATA_ERROR, ///< DG critical data integrity check failed + ALARM_ID_HD_ACCELEROMETER_FAILURE, ///< HD accelerometer error (no readings or FPGA reports error) + ALARM_ID_DG_ACCELEROMETER_FAILURE, ///< DG accelerometer error (no readings or FPGA reports error) + ALARM_ID_HD_VALVE_HOMING_FAILED, ///< HD valve homing failed + ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT, ///< HD valve transition time out + ALARM_ID_HD_VALVE_NOT_FUNCTIONAL, ///< HD valve not functional + ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE, ///< HD valve current out of range + ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE, ///< HD valve position out of target + ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT, ///< FPGA reports failure reading arterial pressure sensor + ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT, ///< FPGA reports failure reading venous pressure sensor + ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT, ///< HD requests DG command with invalid parameter fault + ALARM_ID_NVDATAMGMT_GROUP_RECORD_CRC_INVALID, ///< DG/HD NV data management overall group record CRC invalid + ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE, ///< HD blood pump occlusion self-test failure alarm + ALARM_ID_HD_DIP_OCCLUSION_SELF_TEST_FAILURE, ///< HD dialysate inlet pump occlusion self-test failure alarm + ALARM_ID_HD_DOP_OCCLUSION_SELF_TEST_FAILURE, ///< HD dialysate outlet pump occlusion self-test failure alarm + ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE, ///< HD arterial pressure self-test failure alarm + ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE, ///< HD venous pressure self-test failure alarm + ALARM_ID_HD_BLOOD_FLOW_STATUS_SELF_TEST_FAILURE, ///< HD blood flow meter status check self-test failure alarm + ALARM_ID_HD_DIALYSATE_FLOW_STATUS_SELF_TEST_FAILURE, ///< HD dialysate flow meter status check self-test failure alarm + ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAILURE, ///< HD blood leak detector self-test failure alarm + ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE, ///< HD monitored voltage is out of range + ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE, ///< DG monitored voltage is out of range + ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR, ///< HD syringe pump direction (from encoder) error + ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR, ///< HD syringe pump direction (from controller) error + ALARM_ID_HD_SYRINGE_PUMP_FAULT, ///< HD syringe pump fault reported by FPGA + ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR, ///< HD syringe pump over travel error + ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR, ///< HD syringe pump DAC write failure + ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, ///< HD syringe pump is running while the BP is off + ALARM_ID_DG_FPGA_COMM_TIMEOUT, ///< DG FPGA communication down for too long + ALARM_ID_DG_LOAD_CELL_ADC_ERROR, ///< DG load cell ADC error + ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, ///< HD venous pressure sensor not being read + ALARM_ID_HD_VENOUS_PRESSURE_SENSOR_TEMP_OUT_OF_RANGE, ///< HD venous pressure sensor temperature out of range error + ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR, ///< HD BP occlusion sensor not being read + ALARM_ID_HD_DPI_OCCLUSON_READ_TIMEOUT_ERROR, ///< HD DPi occlusion sensor not being read + ALARM_ID_HD_DPO_OCCLUSION_READ_TIMEOUT_ERROR, ///< HD DPo occlusion sensor not being read + ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR, ///< HD BP occlusion sensor error + ALARM_ID_HD_DPI_OCCLUSION_SENSOR_ERROR, ///< HD DPi occlusion sensor error + ALARM_ID_HD_DPO_OCCLUSION_SENSOR_ERROR, ///< HD DPo occlusion sensor error + ALARM_ID_DG_CONCENTRATE_PUMP_FAULT, ///< DG concentrate pump fault + ALARM_ID_HD_BP_FLOW_READ_TIMEOUT_ERROR, ///< HD blood flow sensor not being read + ALARM_ID_HD_BP_FLOW_SENSOR_ERROR, ///< HD blood flow sensor error + ALARM_ID_HD_DP_FLOW_READ_TIMEOUT_ERROR, ///< HD dialysate flow sensor not being read + ALARM_ID_HD_DP_FLOW_SENSOR_ERROR, ///< HD dialysate flow sensor error + ALARM_ID_HD_BP_FLOW_SLOW_READ_TIMEOUT_ERROR, ///< HD blood flow sensor not being read - slow packet + ALARM_ID_HD_DP_FLOW_SLOW_READ_TIMEOUT_ERROR, ///< HD dialysate flow sensor not being read - slow packet + ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE, ///< DG load cells weight out of range before tare + ALARM_ID_HD_BATTERY_COMM_FAULT, ///< HD battery communication fault + ALARM_ID_HD_SYRINGE_PUMP_STALL, ///< HD syringe pump stall alarm + ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, ///< HD no cartridge self-test timeout + ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT, ///< HD dry self-test timeout + ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT, ///< DG temperature sensor error flag fault. + ALARM_ID_DG_SOFTWARE_FAULT, ///< DG software fault. Software found itself in an unexpected state + ALARM_ID_RTC_COMM_ERROR, ///< Real-time clock communication error + ALARM_ID_DG_INVALID_LOAD_CELL_VALUE, ///< DG invalid load cell value + ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT, ///< DG drain circulation line timeout + ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT, ///< DG temperature sensors ADC fault + ALARM_ID_DG_FLOW_METER_CHECK_FAILURE, ///< DG flow meter check failure alarm + ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE, ///< RO pump flow rate out of range + ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE, ///< DG flow rate out of upper range + ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE, ///< DG flow rate out of lower range + ALARM_ID_RO_PUMP_OFF_FAULT, ///< DG RO pump off fault + ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE, ///< DG RO pump pressure out of range + ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE, ///< DG drain pump RPM out of range + ALARM_ID_DRAIN_PUMP_OFF_FAULT, ///< DG drain pump off fault + ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE, ///< DG heaters cold junction temperature out of range + ALARM_ID_DG_HEATERS_FAULT, ///< DG heaters fault + ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE, ///< DG thermistors/sensors temperature out of range + ALARM_ID_CP1_SPEED_CONTROL_ERROR, ///< DG concentrate pump CP1 speed control error + ALARM_ID_CP2_SPEED_CONTROL_ERROR, ///< DG concentrate pump CP2 speed control error + ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE, ///< RO rejection ratio out of range + ALARM_ID_INLET_WATER_PRESSURE_FAULT, ///< Inlet water pressure fault + ALARM_ID_PRESSURE_SENSOR_FAULT, ///< Pressure sensor fault + ALARM_ID_CONDUCTIVITY_SENSOR_FAULT, ///< Conductivity sensor fault + ALARM_ID_NVDATA_MFG_RECORD_CRC_ERROR, ///< Manufacturing record CRC failure + ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS, ///< Air trap level sensors reporting illegal combination of air/fluid + ALARM_ID_NVDATAMGMT_INDIVIDUAL_RECORD_CRC_INVALID, ///< DG/HD NV data management individual record CRC invalid + ALARM_ID_NVDATAMGMT_INDIVIDUAL_RECORD_TIME_INVALID, ///< DG/HD NV data management individual record time invalid + ALARM_ID_DG_RESTARTED_FAULT, ///< HD reports DG restarted fault + ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR, ///< HD syringe pump ADC error + ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR, ///< HD syringe pump volume check error + ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR, ///< HD syringe pump speed check error + ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR, ///< HD syringe pump not stopped in off state error + ALARM_ID_HD_BLOOD_LEAK_FAULT, ///< HD blood leak detector fault + ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION, ///< DG load cells invalid calibration + ALARM_ID_DG_PRIMARY_HEATERS_INTERNAL_TEMP_OUT_OF_RANGE, ///< DG primary heaters internal temperature out of range + ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE, ///< DG trimmer heater internal temperature out of range + ALARM_ID_DG_HEATERS_ON_WITH_NO_FLOW_TIMEOUT, ///< DG heaters on with no flow time out + ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE, ///< HD arterial air bubble detector self-test failure + ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE, ///< HD venous air bubble detector self-test failure + + // Ranked high priority alarms + ALARM_ID_END_OF_TREATMENT_HIGH, ///< HD end of treatment alarm (high priority - user still not acting to end treatment) + ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK, ///< Blood sitting too long after treatment stopped by user alarm (>10 min) + ALARM_ID_HD_BLOOD_LEAK_DETECTED, ///< HD blood leak detected alarm + ALARM_ID_VENOUS_PRESSURE_LOW, ///< Venous pressure too low during treatment + ALARM_ID_HD_VENOUS_BUBBLE_DETECTED, ///< HD venous air bubble detected alarm + ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK, ///< HD venous air bubble detected rinseback alarm + ALARM_ID_VENOUS_PRESSURE_HIGH, ///< Venous pressure too high during treatment + ALARM_ID_ARTERIAL_PRESSURE_LOW, ///< Arterial pressure too low during treatment + ALARM_ID_ARTERIAL_PRESSURE_HIGH, ///< Arterial pressure too high during treatment + ALARM_ID_DG_FLUID_LEAK_DETECTED, ///< DG fluid leak detected alarm + ALARM_ID_HD_FLUID_LEAK_DETECTED, ///< HD fluid leak detected alarm + ALARM_ID_HD_SHOCK, ///< HD experienced large acceleration + ALARM_ID_DG_SHOCK, ///< DG experienced large acceleration + ALARM_ID_HD_EXCESSIVE_TILT, ///< HD tilt exceeds maximum + ALARM_ID_DG_EXCESSIVE_TILT, ///< DG tilt exceeds maximum + ALARM_ID_HD_AC_POWER_LOST, ///< HD AC power lost alarm + ALARM_ID_DG_COMM_TIMEOUT, ///< Dialysate generator communication timeout + ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT, ///< Air trap fill timeout during treatment + ALARM_ID_OCCLUSION_BLOOD_PUMP, ///< Blood pump occlusion detected + ALARM_ID_OCCLUSION_DIAL_IN_PUMP, ///< Dialysate inlet pump occlusion detected + ALARM_ID_OCCLUSION_DIAL_OUT_PUMP, ///< Dialysate outlet pump occlusion detected + ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE, ///< Concentrate conductivity after adding acid out of range alarm + ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE, ///< Dialysate conductivity out of range + ALARM_ID_DIALYSATE_CONDUCTIVITY_FAULT, ///< DG dialysate generation conductivity fault alarm + // General inlet water high temperature > 50 degree C + ALARM_ID_INLET_WATER_HIGH_TEMPERATURE, ///< Inlet water temperature too high + // Pre-Treatment, Flush, Chemical Disinfection, Heat Disinfection inlet water high temperature > 39 degree C + ALARM_ID_INLET_WATER_LOW_TEMPERATURE, ///< Inlet water temperature too low + // Pre-Treatment, Flush, Chemical Disinfection, Heat Disinfection inlet water low temperature < 22 degree C + ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY, ///< Inlet water conductivity too high + ALARM_ID_INLET_WATER_LOW_CONDUCTIVITY, ///< Inlet water conductivity too low + // Pre-Treatment, Flush, Chemical Disinfection, Heat Disinfection inlet water low conductivity + ALARM_ID_INLET_WATER_LOW_PRESSURE, ///< Inlet water pressure too low + // Pre-Treatment, Flush, Chemical Disinfection, Heat Disinfection inlet water low pressure + ALARM_ID_PRIME_COMPLETED_HIGH, ///< HD prime completed high priority alarm + + // Unranked high priority alarms + ALARM_ID_UV_REACTOR_NOT_HEALTHY, ///< Inlet or Outlet UV reactor not healthy + ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE, ///< DG fan RPM out of range + ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME, ///< DG dialysate fill runs out of time + ALARM_ID_NVDATA_EEPROM_OPS_FAILURE, ///< EEPROM operations (read, write, erase) failure + ALARM_ID_NVDATA_SRVC_RECORD_CRC_ERROR, ///< Service record CRC failure + ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR, ///< HW usage data (treatment time in HD and total consumed water in DG) failure + ALARM_ID_HD_COMM_TIMEOUT, ///< HD communication timeout + ALARM_ID_BLOOD_PUMP_FLOW_VS_MOTOR_SPEED_CHECK, ///< Blood pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed + ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK, ///< Dialysate inlet pump failed flow vs. motor speed check. Mismatch with flow rate and rate implied by motor speed + ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH, ///< Blood pump rotor speed too high + ALARM_ID_BLOOD_FLOW_SIGNAL_STRENGTH_TOO_LOW, ///< Blood flow sensor signal strength too low + ALARM_ID_DIALYSATE_FLOW_SIGNAL_STRENGTH_TOO_LOW, ///< Dialysate flow sensor signal strength too low + ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM, ///< HD sees primary load cell for reservoir 1 change too much too fast + ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM, ///< HD sees primary load cell for reservoir 1 change too much too fast + ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM, ///< HD in treatment rinseback operation timeout + ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM, ///< HD in treatment re-circ sub-mode for too long + ALARM_ID_CARTRIDGE_DOOR_OPENED, ///< HD cartridge door opened alarm + ALARM_ID_DIALYSATE_TEMPERATURE_TOO_HIGH, ///< HD dialysate temperature too high alarm + ALARM_ID_DIALYSATE_TEMPERATURE_TOO_LOW, ///< HD dialysate temperature too low alarm + ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY, ///< HD syringe empty alarm + ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, ///< HD syringe pump occlusion alarm + ALARM_ID_DG_BAD_INLET_WATER_QUALITY, ///< Inlet water quality is bad (temperature, conductivity, pressure) + ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM, ///< HD syringe pump not enough Heparin alarm + ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED, ///< HD arterial air bubble detected alarm + ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED_RINSEBACK, ///< HD arterial air bubble detected rinseback alarm + ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR, ///< HD pump direction status error + ALARM_ID_NVDATA_CAL_RECORD_CRC_ERROR, ///< Calibration record CRC failure + ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH, ///< HD active reservoir is full during treatment and DG not ready to switch reservoirs. + + // Medium priority alarms + ALARM_ID_BLOOD_SITTING_WARNING, ///< Blood sitting too long warning (>4 min) + ALARM_ID_END_OF_TREATMENT_ALARM, ///< HD end of treatment alarm (user not acting to end treatment) + ALARM_ID_PRIME_COMPLETED_MEDIUM, ///< HD prime completed medium priority alarm + ALARM_ID_SALINE_BOLUS_VOLUME_CHECK_FAILURE, ///< HD Saline bolus volume check failure + ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE, ///< HD rinseback volume check failure + ALARM_ID_END_TREATMENT_TIMEOUT_ALARM, ///< HD end treatment sub-mode timeout alarm + ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE, ///< HD blood prime volume check failure + ALARM_ID_HD_SYRINGE_DETECTED, ///< HD syringe detected alarm + ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED, ///< HD syringe pump syringe removed alarm + + // Low priority alarms + ALARM_ID_EMPTY_SALINE_BAG, ///< HD Saline bag is empty + ALARM_ID_DIALYSATE_TEMPERATURE_HIGH, ///< HD dialysate temperature too high alarm + ALARM_ID_DIALYSATE_TEMPERATURE_LOW, ///< HD dialysate temperature too low alarm + ALARM_ID_TREATMENT_STOPPED_BY_USER, ///< Treatment stopped by user action - pressed stop button + ALARM_ID_END_OF_TREATMENT_WARNING, ///< HD end of treatment warning + ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY, ///< HD prime completed low priority alarm + // HD Door open, latch open + ALARM_ID_PRIME_OUT_OF_TIME, ///< HD prime out of time alarm + ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT, ///< HD time out on prime sub-mode purge air state. + ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT, ///< HD prime dialysate dialyzer time out alarm + ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT, ///< HD prime dialysate bypass time out alarm + ALARM_ID_PRE_TREATMENT_DRY_SELF_TEST_FAILURE, ///< HD pre-treatment mode dry self-tests failure + ALARM_ID_PRE_TREATMENT_WET_SELF_TEST_FAILURE, ///< HD pre-treatment mode wet self-tests failure + ALARM_ID_RTC_BATTERY_LOW, ///< RTC battery low + ALARM_ID_RTC_RAM_OPS_ERROR, ///< RTC (i.e read time) or RAM operations failure (read or write) + ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK, ///< HD in treatment stopped sub-mode after rinseback completed (no escalation) + ALARM_ID_INSTALL_NEW_CARTRIDGE, ///< HD needs new cartridge to be installed + ALARM_ID_CARTRIDGE_INSTALLED_IMPROPERLY, ///< HD cartridge installed improperly alarm + ALARM_ID_NO_CARTRIDGE_LOADED, ///< HD no cartridge loaded alarm + ALARM_ID_CARTRIDGE_REMOVAL_FAILURE, ///< HD fail to remove cartridge alarm + ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE, ///< Bicarb conductivity out of range during bicarb pump check alarm + ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT, ///< DG reservoir drain time out + ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT, ///< DG reservoir fill time out + ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT, ///< DG reservoir leak time out + ALARM_ID_DG_TEMP_SENSORS_DIFF_OUT_OF_RANGE, ///< DG temperature sensors difference out of range + ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT, ///< DG heat disinfect target temperature time out (could not reach to temperature) + ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT, ///< DG heat disinfect inlet pressure and temperature sensors out of range + ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_AND_TEMP_OUT, ///< DG heat disinfect inlet conductivity and temperature out of range + ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT, ///< DG chemical disinfect could not reach to target temperature + ALARM_ID_DG_CHEM_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT, ///< DG chemical disinfect inlet pressure and temperature sensors out of range + ALARM_ID_DG_CHEM_DISINFECT_INLET_COND_AND_TEMP_OUT, ///< DG chemical disinfect inlet conductivity and temperature out of range NUM_OF_ALARM_IDS ///< Total number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -355,267 +365,266 @@ /// Record defining the properties of each individual alarm. struct AlarmsStateRecord { - ALARM_PRIORITY_T alarmPriority; ///< priority of alarm - U32 alarmSubRank; ///< alarm's rank within it's priority - U32 alarmEscalatesAfter; ///< time (s) after start when alarm will escalate if not cleared (zero indicates no escalation) + ALARM_PRIORITY_T alarmPriority; ///< Priority of alarm + U32 alarmSubRank; ///< Alarm's rank within it's priority + U32 alarmEscalatesAfter; ///< Time (s) after start when alarm will escalate if not cleared (zero indicates no escalation) ALARM_ID_T alarmEscalatesTo; ///< ID of alarm that this alarm will escalate to (ALARM_ID_NO_ALARM indicates no escalation) - BOOL alarmIsFault; ///< alarm is a system fault? - BOOL alarmStops; ///< alarm activation should cause a controlled stop - BOOL alarmConditionClearImmed; ///< alarm condition is cleared immediately (does not persist after alarm triggered) - BOOL alarmNoClear; ///< alarm cannot be cleared (unrecoverable)? - BOOL alarmNoResume; ///< alarm prevents treatment resume - BOOL alarmNoRinseback; ///< alarm prevents rinseback - BOOL alarmNoEndTreatment; ///< alarm prevents ending treatment - BOOL alarmNoNewTreatment; ///< alarm prevents any new treatments - BOOL alarmNoDialysateRecirc; ///< alarm prevents dialysate re-circulation - BOOL alarmUserAckRequired; ///< alarm requires user acknowledgment + BOOL alarmIsFault; ///< Alarm is a system fault? + BOOL alarmStops; ///< Alarm activation should cause a controlled stop + BOOL alarmConditionClearImmed; ///< Alarm condition is cleared immediately (does not persist after alarm triggered) + BOOL alarmNoClear; ///< Alarm cannot be cleared (unrecoverable)? + BOOL alarmNoResume; ///< Alarm prevents treatment resume + BOOL alarmNoRinseback; ///< Alarm prevents rinseback + BOOL alarmNoEndTreatment; ///< Alarm prevents ending treatment + BOOL alarmNoNewTreatment; ///< Alarm prevents any new treatments + BOOL alarmNoDialysateRecirc; ///< Alarm prevents dialysate re-circulation + BOOL alarmUserAckRequired; ///< Alarm requires user acknowledgment + ALARM_ID_T alarmID; ///< Alarm ID }; typedef struct AlarmsStateRecord ALARM_T; -#define ALARM_NO_RESUME ALARM_ID_BLOOD_SITTING_TOO_LONG_NO_RESUME -#define ALARM_NO_RINSEBACK ALARM_ID_BLOOD_SITTING_TOO_LONG_NO_RINSEBACK -#define ALARM_PRIME_DONE_MED ALARM_ID_PRIME_COMPLETED_MEDIUM_PRIORITY -#define ALARM_PRIME_DONE_HIGH ALARM_ID_PRIME_COMPLETED_HIGH_PRIORITY +#define ALARM_NO_RINSEBACK ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK /// Table of alarms and their static properties. const ALARM_T ALARM_TABLE[] = -{ // Priority Rank Escalate In Escalate To Fault Stops ClrIm NoClr NoRes NoRin NoEnd NoNew NoRcr UsrAck - { ALARM_PRIORITY_NONE, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_NO_ALARM - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SOFTWARE_FAULT - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_STUCK_BUTTON_TEST_FAILED - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_FPGA_POST_TEST_FAILED - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_WATCHDOG_POST_TEST_FAILED - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_UI_COMM_POST_FAILED - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_BLOOD_PUMP_OFF_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_BLOOD_PUMP_MC_DIRECTION_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_IN_PUMP_MC_CURRENT_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_IN_PUMP_OFF_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_IN_PUMP_MC_DIRECTION_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_IN_PUMP_ROTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_OUT_PUMP_MC_CURRENT_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_WATCHDOG_EXPIRED - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_RTC_COMM_ERROR - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_RTC_CONFIG_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DG_COMM_TIMEOUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_UI_COMM_TIMEOUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_COMM_TOO_MANY_BAD_CRCS - { ALARM_PRIORITY_LOW, 31, ALM_ESC_4_MIN, ALARM_ID_BLOOD_SITTING_WARNING, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE }, // ALARM_ID_TREATMENT_STOPPED_BY_USER - { ALARM_PRIORITY_MEDIUM, 25, ALM_ESC_1_MIN, ALARM_NO_RINSEBACK, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_BLOOD_SITTING_WARNING - { ALARM_PRIORITY_MEDIUM, 7, ALM_ESC_5_MIN, ALARM_NO_RINSEBACK, FALSE, TRUE , TRUE , FALSE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_BLOOD_SITTING_TOO_LONG_NO_RESUME TODO - maybe obsolete per PRS 388 (not really a specific alarm - refers to all alarms that stop treatment) - { ALARM_PRIORITY_HIGH, 5, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE, TRUE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_BLOOD_SITTING_TOO_LONG_NO_RINSEBACK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_CAN_MESSAGE_NOT_ACKED - { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_OCCLUSION_BLOOD_PUMP - { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE }, // ALARM_ID_OCCLUSION_DIAL_IN_PUMP - { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE }, // ALARM_ID_OCCLUSION_DIAL_OUT_PUMP - { ALARM_PRIORITY_HIGH, 13, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_ARTERIAL_PRESSURE_LOW - { ALARM_PRIORITY_HIGH, 14, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_ARTERIAL_PRESSURE_HIGH - { ALARM_PRIORITY_HIGH, 8, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_VENOUS_PRESSURE_LOW - { ALARM_PRIORITY_HIGH, 12, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_VENOUS_PRESSURE_HIGH - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_UF_RATE_TOO_HIGH_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_UF_VOLUME_ACCURACY_ERROR - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, FALSE, FALSE }, // ALARM_ID_RTC_BATTERY_LOW - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_RTC_OR_TIMER_ACCURACY_FAILURE - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_RTC_RAM_OPS_ERROR - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_NVDATA_EEPROM_OPS_FAILURE - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_NVDATA_MFG_RECORD_CRC_ERROR - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_NVDATA_SRVC_RECORD_CRC_ERROR - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_NVDATA_CAL_RECORD_CRC_ERROR - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_AVAILABLE_1 - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_AVAILABLE_2 - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_HD_FPGA_COMM_TIMEOUT - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_COMM_TIMEOUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE }, // ALARM_ID_VALVE_CONTROL_FAILURE - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_BLOOD_PUMP_FLOW_VS_MOTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 5, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH - { ALARM_PRIORITY_MEDIUM, 22, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_INLET_WATER_LOW_CONDUCTIVITY - { ALARM_PRIORITY_MEDIUM, 21, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_SOFTWARE_FAULT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_CONDUCTIVITY_SENSOR_FAULT - { ALARM_PRIORITY_HIGH, 20, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_INLET_WATER_LOW_TEMPERATURE - { ALARM_PRIORITY_HIGH, 19, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_INLET_WATER_HIGH_TEMPERATURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_PRESSURE_SENSOR_FAULT - { ALARM_PRIORITY_HIGH, 23, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_INLET_WATER_LOW_PRESSURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_CRITICAL_DATA_ERROR - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 16, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_EXCESSIVE_TILT - { ALARM_PRIORITY_HIGH, 16, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DG_EXCESSIVE_TILT - { ALARM_PRIORITY_HIGH, 15, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_SHOCK - { ALARM_PRIORITY_HIGH, 15, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DG_SHOCK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_ACCELEROMETER_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_ACCELEROMETER_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_CRITICAL_DATA_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_INLET_WATER_PRESSURE_FAULT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VALVE_HOMING_FAILED - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VALVE_NOT_FUNCTIONAL - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 38, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS - { ALARM_PRIORITY_HIGH, 35, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_AIR_TRAP_FILL_DURING_PRIME - { ALARM_PRIORITY_HIGH, 35, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_BLOOD_FLOW_SIGNAL_STRENGTH_TOO_LOW - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DIALYSATE_FLOW_SIGNAL_STRENGTH_TOO_LOW - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_EMPTY_SALINE_BAG - { ALARM_PRIORITY_MEDIUM, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_SALINE_BOLUS_VOLUME_CHECK_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_UV_REACTOR_NOT_HEALTHY - { ALARM_PRIORITY_LOW, 30, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_POST_ACID_CONDUCTIVITY_OUT_OF_RANGE - { ALARM_PRIORITY_LOW, 30, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_POST_BICARB_CONDUCTIVITY_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_CP1_SPEED_CONTROL_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_CP2_SPEED_CONTROL_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_FLOW_RATE_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_HEATERS_FAULT - { ALARM_PRIORITY_LOW, 32, ALM_ESC_5_MIN, ALARM_ID_END_OF_TREATMENT_ALARM, FALSE, FALSE, TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE }, // ALARM_ID_END_OF_TREATMENT_WARNING - { ALARM_PRIORITY_MEDIUM, 26, ALM_ESC_5_MIN, ALARM_ID_END_OF_TREATMENT_HIGH, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE }, // ALARM_ID_END_OF_TREATMENT_ALARM - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE }, // ALARM_ID_END_OF_TREATMENT_HIGH - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_AVAILABLE_3 - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_RO_PUMP_RAMP_UP_TO_FLOW_TIMEOUT - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DRAIN_PUMP_OFF_FAULT - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_RO_PUMP_OFF_FAULT - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT - { ALARM_PRIORITY_LOW, 999, ALM_ESC_90_MIN, ALARM_PRIME_DONE_MED, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY - { ALARM_PRIORITY_MEDIUM, 999, ALM_ESC_30_MIN, ALARM_PRIME_DONE_HIGH, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_PRIME_COMPLETED_MEDIUM_PRIORITY - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_PRIME_COMPLETED_HIGH_PRIORITY - { ALARM_PRIORITY_LOW, 35, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_PRIME_OUT_OF_TIME - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE }, // ALARM_ID_CARTRIDGE_DOOR_OPENED - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_AVAILABLE_4 - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE }, // ALARM_ID_INSTALL_NEW_CARTRIDGE - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE }, // ALARM_ID_CARTRIDGE_INSTALLED_IMPROPERLY - { ALARM_PRIORITY_MEDIUM, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE }, // ALARM_ID_NO_CARTRIDGE_LOADED - { ALARM_PRIORITY_MEDIUM, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_END_TREATMENT_TIMEOUT_ALARM - { ALARM_PRIORITY_MEDIUM, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE - { ALARM_PRIORITY_HIGH, 28, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE }, // ALARM_ID_DIALYSATE_TEMPERATURE_TOO_HIGH - { ALARM_PRIORITY_HIGH, 29, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DIALYSATE_TEMPERATURE_TOO_LOW - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_PRE_TREATMENT_DRY_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_PRE_TREATMENT_WET_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_FLOW_METER_CHECK_FAILURE - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_AVAILABLE_5 - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_NVDATAMGMT_GROUP_RECORD_CRC_INVALID - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_NVDATAMGMT_INDIVIDUAL_RECORD_CRC_INVALID - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_NVDATAMGMT_INDIVIDUAL_RECORD_TIME_INVALID - { ALARM_PRIORITY_HIGH, 15, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_AC_POWER_LOST - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_RESTARTED_FAULT - { ALARM_PRIORITY_LOW, 30, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 36, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_FLUID_LEAK_DETECTED - { ALARM_PRIORITY_HIGH, 36, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DG_FLUID_LEAK_DETECTED - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR - { ALARM_PRIORITY_MEDIUM, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_SYRINGE_DETECTED - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION - { ALARM_PRIORITY_MEDIUM, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR - { ALARM_PRIORITY_HIGH, 6, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, TRUE, TRUE, FALSE, FALSE, TRUE, FALSE }, // ALARM_ID_HD_BLOOD_LEAK_DETECTED - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE, FALSE, FALSE }, // ALARM_ID_CARTRIDGE_REMOVAL_FAILURE - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_TEMP_SENSORS_DIFF_OUT_OF_RANGE - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT - { ALARM_PRIORITY_LOW, 999, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_AND_TEMP_OUT - { ALARM_PRIORITY_HIGH, 36, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DG_BAD_INLET_WATER_QUALITY - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DIP_OCCLUSION_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DOP_OCCLUSION_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BLOOD_FLOW_STATUS_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DIALYSATE_FLOW_STATUS_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_BLOOD_LEAK_FAULT - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_DG_PRIMARY_HEATERS_INTERNAL_TEMP_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_DG_HEATERS_ON_WITH_NO_FLOW_TIMEOUT - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_DIALYSATE_CONDUCTIVITY_FAULT - { ALARM_PRIORITY_LOW, 28, ALM_ESC_5_MIN, ALARM_ID_DIALYSATE_TEMPERATURE_TOO_HIGH, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DIALYSATE_TEMPERATURE_HIGH - { ALARM_PRIORITY_LOW, 29, ALM_ESC_5_MIN, ALARM_ID_DIALYSATE_TEMPERATURE_TOO_LOW, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_DIALYSATE_TEMPERATURE_LOW - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_INVALID_LOAD_CELL_VALUE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_FAULT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_CHEM_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_CHEM_DISINFECT_INLET_COND_AND_TEMP_OUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_FPGA_COMM_TIMEOUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_LOAD_CELL_ADC_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VENOUS_PRESSURE_SENSOR_TEMP_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DPI_OCCLUSON_READ_TIMEOUT_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DPO_OCCLUSION_READ_TIMEOUT_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DPI_OCCLUSION_SENSOR_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DPO_OCCLUSION_SENSOR_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_CONCENTRATE_PUMP_FAULT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BP_FLOW_READ_TIMEOUT_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BP_FLOW_SENSOR_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DP_FLOW_READ_TIMEOUT_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DP_FLOW_SENSOR_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT - { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_FPGA_POST_TEST_FAILED - { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED_RINSEBACK - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE - { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_AVAILABLE_6 - { ALARM_PRIORITY_HIGH, 10, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK - { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED - { ALARM_PRIORITY_HIGH, 10, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_VENOUS_BUBBLE_DETECTED - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_BP_FLOW_SLOW_READ_TIMEOUT_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE }, // ALARM_ID_HD_DP_FLOW_SLOW_READ_TIMEOUT_ERROR - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_BATTERY_COMM_FAULT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_SYRINGE_PUMP_STALL - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT - { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT - { ALARM_PRIORITY_HIGH, 999, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE }, // ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_BLOOD_FLOW_INVALID_CALIBRATION - { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE }, // ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CALIBRATION - // Priority Rank Escalate In Escalate To Fault Stops ClrIm NoClr NoRes NoRin NoEnd NoNew NoRcr UsrAck +{ // Priority Rank Escalate In Escalate To Fault Stops ClrIm NoClr NoRes NoRin NoEnd NoNew NoRcr UsrAck ID + { ALARM_PRIORITY_NONE, 999, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NO_ALARM }, + // POST machine fault high priority alarms + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_STUCK_BUTTON_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_FPGA_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_FPGA_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_WATCHDOG_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_UI_COMM_POST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_RTC_CONFIG_ERROR }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_RTC_OR_TIMER_ACCURACY_FAILURE }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_HEATERS_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_HD_BLOOD_FLOW_INVALID_CALIBRATION }, + { ALARM_PRIORITY_HIGH, 1, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CALIBRATION }, + // Machine fault high priority alarms + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SOFTWARE_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_BLOOD_PUMP_OFF_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_BLOOD_PUMP_MC_DIRECTION_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_IN_PUMP_MC_CURRENT_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_IN_PUMP_OFF_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_IN_PUMP_MC_DIRECTION_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_IN_PUMP_ROTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_OUT_PUMP_MC_CURRENT_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_WATCHDOG_EXPIRED }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_UI_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_COMM_TOO_MANY_BAD_CRCS }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_CAN_MESSAGE_NOT_ACKED }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_UF_RATE_TOO_HIGH_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_UF_VOLUME_ACCURACY_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_HD_FPGA_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_VALVE_CONTROL_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_CRITICAL_DATA_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_CRITICAL_DATA_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_ACCELEROMETER_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_ACCELEROMETER_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VALVE_HOMING_FAILED }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VALVE_NOT_FUNCTIONAL }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_ARTERIAL_PRESSURE_SENSOR_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_VENOUS_PRESSURE_SENSOR_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_NVDATAMGMT_GROUP_RECORD_CRC_INVALID }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DIP_OCCLUSION_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DOP_OCCLUSION_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BLOOD_FLOW_STATUS_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DIALYSATE_FLOW_STATUS_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_FPGA_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_LOAD_CELL_ADC_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VENOUS_PRESSURE_SENSOR_TEMP_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BP_OCCLUSION_READ_TIMEOUT_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DPI_OCCLUSON_READ_TIMEOUT_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DPO_OCCLUSION_READ_TIMEOUT_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BP_OCCLUSION_SENSOR_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DPI_OCCLUSION_SENSOR_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DPO_OCCLUSION_SENSOR_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_CONCENTRATE_PUMP_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BP_FLOW_READ_TIMEOUT_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BP_FLOW_SENSOR_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DP_FLOW_READ_TIMEOUT_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DP_FLOW_SENSOR_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BP_FLOW_SLOW_READ_TIMEOUT_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DP_FLOW_SLOW_READ_TIMEOUT_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_BATTERY_COMM_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_STALL }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_SOFTWARE_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_RTC_COMM_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_INVALID_LOAD_CELL_VALUE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_FLOW_METER_CHECK_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_RO_PUMP_FLOW_RATE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_RO_PUMP_OFF_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_HEATERS_NEGATIVE_COLD_JUNCTION_TEMPERATURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_HEATERS_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_CP1_SPEED_CONTROL_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_CP2_SPEED_CONTROL_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_INLET_WATER_PRESSURE_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_PRESSURE_SENSOR_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_NVDATA_MFG_RECORD_CRC_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_NVDATAMGMT_INDIVIDUAL_RECORD_CRC_INVALID }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_NVDATAMGMT_INDIVIDUAL_RECORD_TIME_INVALID }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_RESTARTED_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_ADC_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_HD_BLOOD_LEAK_FAULT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_LOAD_CELLS_INVALID_CALIBRATION }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_PRIMARY_HEATERS_INTERNAL_TEMP_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_TRIMMER_HEATER_INTERNAL_TEMP_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DG_HEATERS_ON_WITH_NO_FLOW_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_ARTERIAL_BUBBLE_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 3, 0, ALARM_ID_NO_ALARM, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, + + // Ranked high priority alarms + { ALARM_PRIORITY_HIGH, 4, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, ALARM_ID_END_OF_TREATMENT_HIGH }, + { ALARM_PRIORITY_HIGH, 5, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, + { ALARM_PRIORITY_HIGH, 6, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, TRUE, TRUE, FALSE, FALSE, TRUE, FALSE, ALARM_ID_HD_BLOOD_LEAK_DETECTED }, + { ALARM_PRIORITY_HIGH, 8, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_VENOUS_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 9, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_VENOUS_BUBBLE_DETECTED }, + { ALARM_PRIORITY_HIGH, 10, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK }, + { ALARM_PRIORITY_HIGH, 11, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_VENOUS_PRESSURE_HIGH }, + { ALARM_PRIORITY_HIGH, 12, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_ARTERIAL_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 13, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_ARTERIAL_PRESSURE_HIGH }, + { ALARM_PRIORITY_HIGH, 14, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_FLUID_LEAK_DETECTED }, + { ALARM_PRIORITY_HIGH, 15, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_FLUID_LEAK_DETECTED }, + { ALARM_PRIORITY_HIGH, 16, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SHOCK }, + { ALARM_PRIORITY_HIGH, 16, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_SHOCK }, + { ALARM_PRIORITY_HIGH, 17, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_EXCESSIVE_TILT }, + { ALARM_PRIORITY_HIGH, 17, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_EXCESSIVE_TILT }, + { ALARM_PRIORITY_HIGH, 18, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_AC_POWER_LOST }, + { ALARM_PRIORITY_HIGH, 19, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 20, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT }, + { ALARM_PRIORITY_HIGH, 21, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_OCCLUSION_BLOOD_PUMP }, + { ALARM_PRIORITY_HIGH, 21, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_OCCLUSION_DIAL_IN_PUMP }, + { ALARM_PRIORITY_HIGH, 21, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_OCCLUSION_DIAL_OUT_PUMP }, + { ALARM_PRIORITY_HIGH, 22, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 23, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIALYSATE_CONDUCTIVITY_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 24, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, TRUE, FALSE, ALARM_ID_DIALYSATE_CONDUCTIVITY_FAULT }, + { ALARM_PRIORITY_HIGH, 26, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_HIGH_TEMPERATURE }, + { ALARM_PRIORITY_HIGH, 28, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_LOW_TEMPERATURE }, + { ALARM_PRIORITY_HIGH, 30, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY }, + { ALARM_PRIORITY_HIGH, 32, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_LOW_CONDUCTIVITY }, + { ALARM_PRIORITY_HIGH, 34, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_INLET_WATER_LOW_PRESSURE }, + { ALARM_PRIORITY_HIGH, 36, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE, FALSE, TRUE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_COMPLETED_HIGH }, + + // Unranked high priority alarms + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_UV_REACTOR_NOT_HEALTHY }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_EEPROM_OPS_FAILURE }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_SRVC_RECORD_CRC_ERROR }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NVDATA_HW_USAGE_DATA_CRC_ERROR }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_PUMP_FLOW_VS_MOTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_FLOW_SIGNAL_STRENGTH_TOO_LOW }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIALYSATE_FLOW_SIGNAL_STRENGTH_TOO_LOW }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_DOOR_OPENED }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_DIALYSATE_TEMPERATURE_TOO_HIGH }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIALYSATE_TEMPERATURE_TOO_LOW }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DG_BAD_INLET_WATER_QUALITY }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE, TRUE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_ARTERIAL_BUBBLE_DETECTED_RINSEBACK }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE, TRUE, TRUE, TRUE, TRUE, TRUE , FALSE, ALARM_ID_NVDATA_CAL_RECORD_CRC_ERROR }, + { ALARM_PRIORITY_HIGH, 37, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH }, + + // Medium priority alarms + { ALARM_PRIORITY_MEDIUM, 50, ALM_ESC_1_MIN, ALARM_NO_RINSEBACK, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_SITTING_WARNING }, + { ALARM_PRIORITY_MEDIUM, 51, ALM_ESC_5_MIN, ALARM_ID_END_OF_TREATMENT_HIGH, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, ALARM_ID_END_OF_TREATMENT_ALARM }, + { ALARM_PRIORITY_MEDIUM, 52, ALM_ESC_30_MIN, ALARM_ID_PRIME_COMPLETED_HIGH, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_COMPLETED_MEDIUM }, + { ALARM_PRIORITY_MEDIUM, 53, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_SALINE_BOLUS_VOLUME_CHECK_FAILURE }, + { ALARM_PRIORITY_MEDIUM, 53, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE }, + { ALARM_PRIORITY_MEDIUM, 53, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_END_TREATMENT_TIMEOUT_ALARM }, + { ALARM_PRIORITY_MEDIUM, 53, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE }, + { ALARM_PRIORITY_MEDIUM, 53, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_DETECTED }, + { ALARM_PRIORITY_MEDIUM, 53, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED }, + + // Low priority alarms + { ALARM_PRIORITY_LOW, 70, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_EMPTY_SALINE_BAG }, + { ALARM_PRIORITY_LOW, 71, ALM_ESC_5_MIN, ALARM_ID_DIALYSATE_TEMPERATURE_TOO_HIGH, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIALYSATE_TEMPERATURE_HIGH }, + { ALARM_PRIORITY_LOW, 72, ALM_ESC_5_MIN, ALARM_ID_DIALYSATE_TEMPERATURE_TOO_LOW, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DIALYSATE_TEMPERATURE_LOW }, + { ALARM_PRIORITY_LOW, 73, ALM_ESC_4_MIN, ALARM_ID_BLOOD_SITTING_WARNING, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_TREATMENT_STOPPED_BY_USER }, + { ALARM_PRIORITY_LOW, 74, ALM_ESC_5_MIN, ALARM_ID_END_OF_TREATMENT_ALARM, FALSE, FALSE, TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_END_OF_TREATMENT_WARNING }, + { ALARM_PRIORITY_LOW, 75, ALM_ESC_90_MIN, ALARM_ID_PRIME_COMPLETED_MEDIUM, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY }, + { ALARM_PRIORITY_LOW, 79, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_OUT_OF_TIME }, + { ALARM_PRIORITY_LOW, 79, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT }, + { ALARM_PRIORITY_LOW, 79, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT }, + { ALARM_PRIORITY_LOW, 79, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT }, + { ALARM_PRIORITY_LOW, 80, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRE_TREATMENT_DRY_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_LOW, 80, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_PRE_TREATMENT_WET_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE, FALSE, FALSE, ALARM_ID_RTC_BATTERY_LOW }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_RTC_RAM_OPS_ERROR }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_INSTALL_NEW_CARTRIDGE }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_CARTRIDGE_INSTALLED_IMPROPERLY }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_NO_CARTRIDGE_LOADED }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE, FALSE, FALSE, ALARM_ID_CARTRIDGE_REMOVAL_FAILURE }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_TEMP_SENSORS_DIFF_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_AND_TEMP_OUT }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_CHEM_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT }, + { ALARM_PRIORITY_LOW, 81, 0, ALARM_ID_NO_ALARM, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, ALARM_ID_DG_CHEM_DISINFECT_INLET_COND_AND_TEMP_OUT }, + // Priority Rank Escalate In Escalate To Fault Stops ClrIm NoClr NoRes NoRin NoEnd NoNew NoRcr UsrAck ID }; #endif // included by AlarmMgmt.c