Index: AlarmDefs.h =================================================================== diff -u -r08e1637a87576d92ddbfa29e91a852cd26451a70 -rc76f57ef110a0c4b65789683e07bcb42696a03b1 --- AlarmDefs.h (.../AlarmDefs.h) (revision 08e1637a87576d92ddbfa29e91a852cd26451a70) +++ AlarmDefs.h (.../AlarmDefs.h) (revision c76f57ef110a0c4b65789683e07bcb42696a03b1) @@ -348,6 +348,7 @@ ALARM_ID_DG_CD1_COND_SENSOR_INVALID_CHAR = 310, ///< DG CD1 conductivity sensor invalid character received ALARM_ID_DG_CD2_COND_SENSOR_INVALID_CHAR = 311, ///< DG CD2 conductivity sensor invalid character received ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE = 312, ///< HD Dialysate In flow rate to too low. + ALARM_ID_HD_AIR_PUMP_TIMEOUT = 316, ///< HD Air Pump timeout alarm NUM_OF_ALARM_IDS ///< Total number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -791,6 +792,7 @@ { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CD1_COND_SENSOR_INVALID_CHAR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_CD2_COND_SENSOR_INVALID_CHAR, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE , TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, TRUE, FALSE, TRUE , TRUE , TRUE, TRUE , TRUE, TRUE, TRUE, TRUE , FALSE, FALSE, ALARM_ID_HD_AIR_PUMP_TIMEOUT, 0, ALARM_ID_NO_ALARM }, }; // Priority Rank Fault DFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To /// Table of alarms and their rank. @@ -1016,6 +1018,7 @@ { 3, ALARM_ID_DG_CD1_COND_SENSOR_INVALID_CHAR }, { 3, ALARM_ID_DG_CD2_COND_SENSOR_INVALID_CHAR }, { 3, ALARM_ID_DG_DRAIN_PUMP_DIRECTION_FPGA_FAULT }, + { 3, ALARM_ID_HD_AIR_PUMP_TIMEOUT }, { 4, ALARM_ID_END_OF_TREATMENT_HIGH }, { 5, ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK }, { 6, ALARM_ID_HD_BLOOD_LEAK_DETECTED },