Index: AlarmDefs.h =================================================================== diff -u -r042ba8de39a1e5109df6d438bcc73a63030daf9b -rcdb6b2a6bfda96b939f45602fd8b64d5ffc78a34 --- AlarmDefs.h (.../AlarmDefs.h) (revision 042ba8de39a1e5109df6d438bcc73a63030daf9b) +++ AlarmDefs.h (.../AlarmDefs.h) (revision cdb6b2a6bfda96b939f45602fd8b64d5ffc78a34) @@ -305,8 +305,8 @@ ALARM_ID_UI_POST_FAILURE_BLUETOOTH = 267, ///< UI POST Bluetooth failure ALARM_ID_UI_POST_FAILURE_ETHERNET = 268, ///< UI POST Ethernet failure ALARM_ID_UI_POST_FAILURE_SOUND = 269, ///< UI POST Sound failure - ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED = 270, ///< HD Watchdog POST failure - ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED = 271, ///< HD Watchdog POST failure + ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED = 270, ///< HD POST Safety Shutdown failure + ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED = 271, ///< HD POST Safety Shutdown failure ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE = 272, ///< HD Fan RPM out of range ALARM_ID_HD_BLOOD_FLOW_OUT_OF_RANGE = 273, ///< HD measured blood pump flow rate is out of range ALARM_ID_HD_DIAL_IN_FLOW_OUT_OF_RANGE = 274, ///< HD measured dialysate inlet pump flow rate is out of range