Index: AlarmDefs.h =================================================================== diff -u -r1d60fe2157e679b1c2e545a7f55c6e25dcd2379e -rf188f2ae79da099284b4df624673cc0b8b0eaefd --- AlarmDefs.h (.../AlarmDefs.h) (revision 1d60fe2157e679b1c2e545a7f55c6e25dcd2379e) +++ AlarmDefs.h (.../AlarmDefs.h) (revision f188f2ae79da099284b4df624673cc0b8b0eaefd) @@ -8,7 +8,7 @@ * @file AlarmDefs.h * * @author (last) Sean Nash -* @date (last) 31-May-2023 +* @date (last) 01-Jun-2023 * * @author (original) Sean * @date (original) 11-Feb-2020 @@ -430,7 +430,7 @@ ALARM_STATE_FLAG_BIT_POS_ALARMS_TO_ESCALATE = 8, ///< One or more active alarms will escalate in time ALARM_STATE_FLAG_BIT_POS_ALARMS_SILENCED = 9, ///< Alarms have been temporarily silenced by user ALARM_STATE_FLAG_BIT_POS_LAMP_ON = 10, ///< Alarm lamp is currently on (for syncing to UI) - ALARM_STATE_FLAG_BIT_POS_UNUSED_1 = 11, ///< Reserved + ALARM_STATE_FLAG_BIT_POS_NO_RETRIGGER = 11, ///< Alarm block re-trigger flag is set ALARM_STATE_FLAG_BIT_POS_BLOOD_RECIRC = 12, ///< Blood recirculation is disabled ALARM_STATE_FLAG_BIT_POS_DIALYSATE_RECIRC = 13, ///< Dialysate recirculation is disabled ALARM_STATE_FLAG_BIT_POS_NO_MINIMIZE = 14, ///< Prevent user from minimizing alarm window