Index: AlarmDefs.h =================================================================== diff -u -rf95c44da2702286c6608cea0f6c9825cb120f841 -rf3b5ec41095d3066636c24149fafbf1b2ee8ebfc --- AlarmDefs.h (.../AlarmDefs.h) (revision f95c44da2702286c6608cea0f6c9825cb120f841) +++ AlarmDefs.h (.../AlarmDefs.h) (revision f3b5ec41095d3066636c24149fafbf1b2ee8ebfc) @@ -1438,7 +1438,7 @@ { 271 , ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG POST: Proc Safestate | DG POST Safety Shutdown failure | If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. | "}, { 272 , ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. | HD Alarm (Low): Fan Speed | HD Fan RPM out of range | If the fan RPM is out of Min/Max range. | "}, { 273 , ALARM_ID_AVAILABLE_10 , " | | | | | | "}, - { 274 , ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Fault | DG idle reservoir out of range | If the reservoir weight > 200.0 of the target mL. | "}, + { 274 , ALARM_ID_DG_INACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE , "DG | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DG Fault: Reservoir Fault | DG idle reservoir weight out of range | If the reservoir weight > 200.0 of the target mL. | "}, { 275 , ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Arterial Press Range | HD arterial pressure sensor is reading out of range | If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, { 276 , ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Venous Press Range | HD venous pressure sensor is reading out of range | If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, { 277 , ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE , "HD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Occlusion Sens Range | HD BP occlusion sensor is reading out of range | If the blood pump occlusion count is < 2000. | "},