Index: AlarmDefs.h =================================================================== diff -u -r2025da4435440ca8787c189516e4a0b062661980 -rf64f3f0b5d3a015a41d7c784b549d9b46667e70b --- AlarmDefs.h (.../AlarmDefs.h) (revision 2025da4435440ca8787c189516e4a0b062661980) +++ AlarmDefs.h (.../AlarmDefs.h) (revision f64f3f0b5d3a015a41d7c784b549d9b46667e70b) @@ -231,7 +231,7 @@ ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE = 193, ///< HD dialysate temperature out of low safety range. ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE = 194, ///< DG RO pump duty cycle out of range ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE = 195, ///< DG RO pump pressure out of range - ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT = 196, ///< DG temperature sensor error bit fault + ALARM_ID__AVAILABLE_1 = 196, ///< Alarm ID available for use ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT = 197, // TODO remove when DEN-13460 was merged ///< DG temperature sensors FPGA fault ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON = 198, ///< RO flow too low while primary heater is on ALARM_ID_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON = 199, ///< Dialysate flow too low while trimmer heater is on @@ -662,7 +662,7 @@ { ALARM_PRIORITY_HIGH, 799, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 799, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE, 0, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT, 0, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID__AVAILABLE_1, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT, 0, ALARM_ID_NO_ALARM }, { ALARM_PRIORITY_HIGH, 3, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON, 0, ALARM_ID_NO_ALARM }, // Priority Rank Fault DGFault Stops ClrIm NoClr NoRes NoRin NoEnd NoBRcr NoDRcr ClrOnly TxLog ID Escalate In Escalate To @@ -951,7 +951,7 @@ { 3, ALARM_ID_DRAIN_PUMP_OFF_FAULT }, { 3, ALARM_ID_UV_REACTOR_NOT_HEALTHY }, { 3, ALARM_ID_DG_TEMPERATURE_SENSORS_FPGA_FAULT }, - { 3, ALARM_ID_DG_TEMPERATURE_SENSOR_ERROR_BIT_FAULT }, + { 3, ALARM_ID__AVAILABLE_1 }, { 3, ALARM_ID_DG_LOAD_CELL_ADC_ERROR }, { 3, ALARM_ID_DG_FPGA_COMM_TIMEOUT }, { 3, ALARM_ID_CONDUCTIVITY_SENSOR_FAULT },