Index: firmware/App/Controllers/BalancingChamber.c =================================================================== diff -u -r12929ac400de74684371b01b962926f868922448 -r2245e17875668a9326a5f5155a56747378e139e9 --- firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision 12929ac400de74684371b01b962926f868922448) +++ firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision 2245e17875668a9326a5f5155a56747378e139e9) @@ -404,7 +404,7 @@ F32 bicarbVolume = getF32OverrideValue( &bicarbDoseVolume ); // Check fresh and spent dialysate pressure in range or BC switch only flag set -#ifdef __BC_PRESSURE_ALARM__ +#ifndef __BC_PRESSURE_ALARM__ if ( 1 ) #else if ( ( ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ) ) && @@ -597,7 +597,7 @@ F32 bicarbVolume = getF32OverrideValue( &bicarbDoseVolume ); // Check fresh and spent dialysate pressure in range -#ifdef __BC_PRESSURE_ALARM__ +#ifndef __BC_PRESSURE_ALARM__ if ( 1 ) #else if ( ( ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ) ) &&