Index: firmware/App/Controllers/BalancingChamber.c =================================================================== diff -u -ra569e34b86e389ce6d6f4c971ea86d2690dbaaed -rc9c9b1bade37555a0f226b9cc24529410325e3ed --- firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision a569e34b86e389ce6d6f4c971ea86d2690dbaaed) +++ firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision c9c9b1bade37555a0f226b9cc24529410325e3ed) @@ -57,7 +57,8 @@ #define SPENT_DIFF_COUNT_ZERO 0 ///< Zero count difference for spent side fill comparing target count #define D48_SPEED_ADJUST_FACTOR 0.5F ///< D48 speed adjustment factor ( 50% of speed adjustment = 0.5) #define D48_SPEED_RANGE_LIMIT 0.25F ///< D48 speed adjustment range check limit ( D48 speed can vary +/-25% of initial calculated speed) -#define BICARB_CHAMBER_PERIODIC_FILL_TIME ( 1 * SEC_PER_MIN * ( MS_PER_SECOND / TASK_GENERAL_INTERVAL ) ) ///< Periodic bicarb chamber fill request 60 sec x 20 = 1200 +#define BICARB_CHAMBER_PERIODIC_FILL_TIME ( 1 * SEC_PER_MIN * \ + ( MS_PER_SECOND / TASK_GENERAL_INTERVAL ) ) ///< Periodic bicarb chamber fill request 60 sec x 20 = 1200 #define BAL_CHAMBER_FILL_TIMEOUT_FACTOR 1.5 ///< Balancing Chamber fill timeout factor (150% of observed fill count) /// Payload record structure for balancing chamber switch only request @@ -259,7 +260,7 @@ currentBalChamberSwitchingCounter += 1; currentBalChamberFillCounter += 1; - if ( getTestConfigStatus( TEST_CONFIG_DD_ENABLE_DRY_BICARB ) == TRUE ) + if ( getTestConfigStatus( TEST_CONFIG_DD_ENABLE_DRY_BICARB ) == TRUE && balChamberExecState > BAL_CHAMBER_STATE_IDLE) { // Increment counter for dry bicarb chamber fill bicarbChamberPeriodicFillCounter += 1; @@ -507,6 +508,8 @@ *************************************************************************/ static BAL_CHAMBER_EXEC_STATE_T handleBalChamberState1FillStart( void ) { + F32 acidVolume; + F32 bicarbVolume; BAL_CHAMBER_EXEC_STATE_T state = BAL_CHAMBER_STATE1_FILL_START; balChamberSWState = BAL_CHAMBER_SW_STATE1; balChamberFillTimeoutCount = 0; @@ -521,8 +524,17 @@ spentFillRiseMissCounter = 0; isSpentFillComplete = FALSE; - F32 acidVolume = getF32OverrideValue( &acidDoseVolume ); - F32 bicarbVolume = getF32OverrideValue( &bicarbDoseVolume ); + if ( getTestConfigStatus( TEST_CONFIG_DD_ENABLE_DRY_BICARB ) == TRUE ) + { + bicarbVolume = getBicarbMixVol(); + acidVolume = getAcidMixVol(); + } + else // wet bicarb + { + acidVolume = getF32OverrideValue( &acidDoseVolume ); + bicarbVolume = getF32OverrideValue( &bicarbDoseVolume ); + } + freshDialPressure = getFilteredPressure( D18_PRES ); spentDialPressure = getFilteredPressure( D51_PRES ); @@ -551,8 +563,8 @@ if ( TRUE != getBalChamberSwitchingOnlyStatus() ) { // start acid and bicarb pump with the expected quantity - setConcentratePumpTargetSpeed( D11_PUMP, CONCENTRATE_PUMP_MAX_SPEED, acidVolume ); - setConcentratePumpTargetSpeed( D10_PUMP, CONCENTRATE_PUMP_MAX_SPEED, bicarbVolume ); + setConcentratePumpTargetSpeed( D11_PUMP, DOSING_CONCENTRATE_PUMP_SPEED, acidVolume ); + setConcentratePumpTargetSpeed( D10_PUMP, DOSING_CONCENTRATE_PUMP_SPEED, bicarbVolume ); requestConcentratePumpOn( D11_PUMP ); requestConcentratePumpOn( D10_PUMP ); } @@ -799,6 +811,8 @@ *************************************************************************/ static BAL_CHAMBER_EXEC_STATE_T handleBalChamberState2FillStart( void ) { + F32 acidVolume; + F32 bicarbVolume; BAL_CHAMBER_EXEC_STATE_T state = BAL_CHAMBER_STATE2_FILL_START; isBalChamberFillInProgress = FALSE; isPressureStabilizedDuringFill = FALSE; @@ -813,8 +827,16 @@ spentFillRiseMissCounter = 0; isSpentFillComplete = FALSE; - F32 acidVolume = getF32OverrideValue( &acidDoseVolume ); - F32 bicarbVolume = getF32OverrideValue( &bicarbDoseVolume ); + if ( getTestConfigStatus( TEST_CONFIG_DD_ENABLE_DRY_BICARB ) == TRUE ) + { + bicarbVolume = getBicarbMixVol(); + acidVolume = getAcidMixVol(); + } + else + { + acidVolume = getF32OverrideValue( &acidDoseVolume ); + bicarbVolume = getF32OverrideValue( &bicarbDoseVolume ); + } freshDialPressure = getFilteredPressure( D18_PRES ); spentDialPressure = getFilteredPressure( D51_PRES ); @@ -843,8 +865,8 @@ if ( TRUE != getBalChamberSwitchingOnlyStatus() ) { // start acid and bicarb pump with the expected quantity - setConcentratePumpTargetSpeed( D11_PUMP, CONCENTRATE_PUMP_MAX_SPEED, acidVolume ); - setConcentratePumpTargetSpeed( D10_PUMP, CONCENTRATE_PUMP_MAX_SPEED, bicarbVolume ); + setConcentratePumpTargetSpeed( D11_PUMP, DOSING_CONCENTRATE_PUMP_SPEED, acidVolume ); + setConcentratePumpTargetSpeed( D10_PUMP, DOSING_CONCENTRATE_PUMP_SPEED, bicarbVolume ); requestConcentratePumpOn( D11_PUMP ); requestConcentratePumpOn( D10_PUMP ); }