Index: firmware/App/Controllers/BalancingChamber.c =================================================================== diff -u -rd748813399d38ef5b71d760e327e368cc82d7a38 -rf554e391302359aa2b252c6c13603d831c11cb77 --- firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision d748813399d38ef5b71d760e327e368cc82d7a38) +++ firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision f554e391302359aa2b252c6c13603d831c11cb77) @@ -526,8 +526,16 @@ freshDialPressure = getFilteredPressure( D18_PRES ); spentDialPressure = getFilteredPressure( D51_PRES ); - //Set Trimmer heater Target temp every BC cycle to catch up fresh dialysate temp - setHeaterTargetTemperature( D45_HEAT, getD4AverageTemperature() ); + if ( getTestConfigStatus( TEST_CONFIG_DD_FP_ENABLE_BETA_1_0_HW ) == TRUE ) + { + //Set Trimmer heater Target temp every BC cycle to catch up fresh dialysate temp + setHeaterTargetTemperature( D45_HEAT, getD4AverageTemperature() ); + } + else + { + //Set Trimmer heater Target temp every BC cycle to catch up fresh dialysate temp + setHeaterTargetTemperature( D45_HEAT, getD99AverageTemperature() ); + } BOOL isFreshDialysatePressureInRange = ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ); BOOL isSpentDialysatePressureInRange = ( spentDialPressure >= SPENT_DIAL_PRESSURE_MIN_PSIG ) && ( spentDialPressure <= SPENT_DIAL_PRESSURE_MAX_PSIG ); @@ -740,8 +748,16 @@ freshDialPressure = getFilteredPressure( D18_PRES ); spentDialPressure = getFilteredPressure( D51_PRES ); - //Set Trimmer heater Target temp every BC cycle to catch up fresh dialysate temp - setHeaterTargetTemperature( D45_HEAT, getD4AverageTemperature() ); + if ( getTestConfigStatus( TEST_CONFIG_DD_FP_ENABLE_BETA_1_0_HW ) == TRUE ) + { + //Set Trimmer heater Target temp every BC cycle to catch up fresh dialysate temp + setHeaterTargetTemperature( D45_HEAT, getD4AverageTemperature() ); + } + else + { + //Set Trimmer heater Target temp every BC cycle to catch up fresh dialysate temp + setHeaterTargetTemperature( D45_HEAT, getD99AverageTemperature() ); + } BOOL isFreshDialysatePressureInRange = ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ); BOOL isSpentDialysatePressureInRange = ( spentDialPressure >= SPENT_DIAL_PRESSURE_MIN_PSIG ) && ( spentDialPressure <= SPENT_DIAL_PRESSURE_MAX_PSIG );