Index: firmware/App/Controllers/BalancingChamber.c =================================================================== diff -u -rf68d42f6972d7be8e994cc35ba52d72a275800f0 -r12929ac400de74684371b01b962926f868922448 --- firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision f68d42f6972d7be8e994cc35ba52d72a275800f0) +++ firmware/App/Controllers/BalancingChamber.c (.../BalancingChamber.c) (revision 12929ac400de74684371b01b962926f868922448) @@ -404,7 +404,7 @@ F32 bicarbVolume = getF32OverrideValue( &bicarbDoseVolume ); // Check fresh and spent dialysate pressure in range or BC switch only flag set -#ifndef __BC_PRESSURE_ALARM__ +#ifdef __BC_PRESSURE_ALARM__ if ( 1 ) #else if ( ( ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ) ) && @@ -597,7 +597,7 @@ F32 bicarbVolume = getF32OverrideValue( &bicarbDoseVolume ); // Check fresh and spent dialysate pressure in range -#ifndef __BC_PRESSURE_ALARM__ +#ifdef __BC_PRESSURE_ALARM__ if ( 1 ) #else if ( ( ( freshDialPressure >= FRESH_DIAL_PRESSURE_MIN_PSIG ) && ( freshDialPressure <= FRESH_DIAL_PRESSURE_MAX_PSIG ) ) &&