Index: firmware/App/Controllers/ConcentratePumps.c =================================================================== diff -u -r5ab03489929d8cc5bbf47d8533d624fd8e4136c3 -ra9f55b2dd2aad22db9676faf661b81866b47affa --- firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision 5ab03489929d8cc5bbf47d8533d624fd8e4136c3) +++ firmware/App/Controllers/ConcentratePumps.c (.../ConcentratePumps.c) (revision a9f55b2dd2aad22db9676faf661b81866b47affa) @@ -53,7 +53,7 @@ #define CONCENTRATE_PUMP_VOLUME_TO_REVOLUTION ( ( 1 / CONCENTRATE_PUMP_VOLUME_PER_REV ) * \ ( CONCENTRATE_PUMP_STEP_PER_REV * \ CONCENTRATE_PUMP_MICRO_STEPS_PER_STEP ) ) ///< Convert volume in to number of revolutions needed. -#define CONCENTRATE_PUMP_STEP_PERIOD_RESOLUTION ( 0.25F / ( US_PER_SECOND * SEC_PER_MIN ) ) ///< Convert step period resolution (0.25 us) to minute. +#define CONCENTRATE_PUMP_STEP_PERIOD_RESOLUTION ( 0.50F / ( US_PER_SECOND * SEC_PER_MIN ) ) ///< Convert step period resolution (0.50 us) to minute. /// Volume output per pulse. #define CONCENTRATE_PUMP_VOLUME_PER_PULSE ( CONCENTRATE_PUMP_VOLUME_PER_REV / CONCENTRATE_PUMP_PULSE_PER_REV ) @@ -769,14 +769,14 @@ if ( CONCENTRATEPUMPS_CP1_ACID == pumpId ) { setFPGAAcidPumpSetStepSpeed( CONCENTRATE_PUMP_ZERO_FLOW_RATE ); - concentratePumps[ CONCENTRATEPUMPS_CP1_ACID ].controlSet |= CONCENTRATE_PUMP_CONTROL_ENABLE_MASK; - setFPGAAcidPumpControl( concentratePumps[ CONCENTRATEPUMPS_CP1_ACID ].controlSet ); + //concentratePumps[ CONCENTRATEPUMPS_CP1_ACID ].controlSet |= CONCENTRATE_PUMP_CONTROL_ENABLE_MASK; + //setFPGAAcidPumpControl( concentratePumps[ CONCENTRATEPUMPS_CP1_ACID ].controlSet ); } else { setFPGABicarbSetStepSpeed( CONCENTRATE_PUMP_ZERO_FLOW_RATE ); - concentratePumps[ CONCENTRATEPUMPS_CP2_BICARB ].controlSet |= CONCENTRATE_PUMP_CONTROL_ENABLE_MASK; - setFPGABicarbPumpControl( concentratePumps[ CONCENTRATEPUMPS_CP2_BICARB ].controlSet ); + //concentratePumps[ CONCENTRATEPUMPS_CP2_BICARB ].controlSet |= CONCENTRATE_PUMP_CONTROL_ENABLE_MASK; + //setFPGABicarbPumpControl( concentratePumps[ CONCENTRATEPUMPS_CP2_BICARB ].controlSet ); } // Park concentrate pump too if requested