Index: firmware/App/Controllers/ConcentratePumps.h =================================================================== diff -u -rd0d24b086ffdfa00838570e24ffdf2194beb1e17 -rfd897db8177752330ad08d877e0a13620513dbdc --- firmware/App/Controllers/ConcentratePumps.h (.../ConcentratePumps.h) (revision d0d24b086ffdfa00838570e24ffdf2194beb1e17) +++ firmware/App/Controllers/ConcentratePumps.h (.../ConcentratePumps.h) (revision fd897db8177752330ad08d877e0a13620513dbdc) @@ -7,8 +7,8 @@ * * @file ConcentratePumps.h * -* @author (last) Jashwant Gantyada -* @date (last) 17-Mar-2026 +* @author (last) Sameer Kalliadan Poyil +* @date (last) 15-Apr-2026 * * @author (original) Vinayakam Mani * @date (original) 19-Sep-2024 @@ -32,11 +32,11 @@ // ********** public definitions ********** // For 150 RPM, Diener pump delivers 60ml/min #define CONCENTRATE_PUMP_MAX_SPEED 200.0F ///< Maximum Diener pump speed for concentrate pump in mL/min -#define DOSING_CONCENTRATE_PUMP_SPEED 60.0F ///< Diener pump speed for acid and bicarb dosing pump speed in mL/min +#define DOSING_CONCENTRATE_PUMP_SPEED 100.0F ///< Diener pump speed for acid and bicarb dosing pump speed in mL/min #define DRAIN_BICART_PUMP_SPEED 200.0F ///< Diener pump speed for cartridge drains at maximum speed in mL/min -#define DEFAULT_ACID_VOLUME_ML 0.67F ///< Acid concentrate volume in ml. -#define DEFAULT_BICARB_VOLUME_ML 1.15F ///< Bicarb concentrate volume in ml. +#define DEFAULT_ACID_VOLUME_ML 0.666666667F ///< Acid concentrate volume in ml. +#define DEFAULT_BICARB_VOLUME_ML 1.146666667F ///< Bicarb concentrate volume in ml. #define DOSING_CONT_VOLUME 0xFFFFFFFF ///< Volume set to 0xFFFFFFFF enables continuous delivery based on the speed set.