Index: firmware/App/Controllers/ConcentratePumps.h =================================================================== diff -u -rf2b23abd44f71103b3ef6b0a96f18cc82d263735 -r0a74bcce6674271e35d5154003dd6c25011f80f9 --- firmware/App/Controllers/ConcentratePumps.h (.../ConcentratePumps.h) (revision f2b23abd44f71103b3ef6b0a96f18cc82d263735) +++ firmware/App/Controllers/ConcentratePumps.h (.../ConcentratePumps.h) (revision 0a74bcce6674271e35d5154003dd6c25011f80f9) @@ -31,13 +31,10 @@ // ********** public definitions ********** #ifdef __DIENER_CONC_PUMP__ -// For 100 RPM, Diener pump delivers 40ml/min,so restricing 300 RPM for now to maintain accuracy. We may allow up to 500 RPM -// that yields 200 ml/min, but as speed increases, accuracy may drop. -#define CONCENTRATE_PUMP_MAX_SPEED 120.0F ///< Optimal speed for concentrate pump in mL/min -#define UF_PUMP_SPEED 100.0F ///< 100 ml/min, runs at 250 RPM +// For 150 RPM, Diener pump delivers 60ml/min +#define CONCENTRATE_PUMP_MAX_SPEED 60.0F ///< Optimal speed for concentrate pump in mL/min #else #define CONCENTRATE_PUMP_MAX_SPEED 48.0F ///< Maximum speed for concentrate pump in mL/min -#define UF_PUMP_SPEED 40.0F ///< 40 ml/min, runs at 400 RPM #endif #define DEFAULT_ACID_VOLUME_ML 0.67F ///< Acid concentrate volume in ml. #define DEFAULT_BICARB_VOLUME_ML 1.15F ///< Bicarb concentrate volume in ml.