Index: firmware/App/Controllers/DialysatePumps.h =================================================================== diff -u -rc07917caced76b53a0ed8f35167fac6f9d8310a4 -rb83b525617ddffecc06b81c97e2534cdd3446c47 --- firmware/App/Controllers/DialysatePumps.h (.../DialysatePumps.h) (revision c07917caced76b53a0ed8f35167fac6f9d8310a4) +++ firmware/App/Controllers/DialysatePumps.h (.../DialysatePumps.h) (revision b83b525617ddffecc06b81c97e2534cdd3446c47) @@ -7,8 +7,8 @@ * * @file DialysatePumps.h * -* @author (last) Vinayakam Mani * @date (last) 13-May-2025 +* @date (last) 13-Nov-2025 * * @author (original) Vinayakam Mani * @date (original) 07-Oct-2024 @@ -24,15 +24,15 @@ * @defgroup DialysatePumps DialysatePumps * @brief Dialysate Pumps monitor/controller module. Controls and monitors the dialysate pumps. * Dialysate pump manufacturer: Diener Silencer Series Gear Pumps, PN: 01483-PM-3-RA. - * The pump shall produce flow rates of 350-2700 ml/min. + * The pump shall run at 200(10%) -2700(90%)RPM. * * @addtogroup DialysatePumps * @{ */ // ********** public definitions ********** -#define MIN_DIALYSATE_PUMP_RPM 350 ///< Minimum RPM target for dialysate pump (though zero is allowed if turning pump off). +#define MIN_DIALYSATE_PUMP_RPM 200 ///< Minimum RPM target for dialysate pump (though zero is allowed if turning pump off). #define MAX_DIALYSATE_PUMP_RPM 2650 ///< Maximum RPM target for dialysate pump. #define DEGAS_PUMP_TARGET_PRES_ADJ_THRESHOLD -1.0F ///< Dialysate Pump(D12) target pressure threshold adjustment factor.