Index: firmware/App/Controllers/DryBiCart.c =================================================================== diff -u -r37d6f3341f62349589f63975779c0df952c0cfd4 -rb0da135bc589efaf8ad6f3c8c1988dcb672f3ca0 --- firmware/App/Controllers/DryBiCart.c (.../DryBiCart.c) (revision 37d6f3341f62349589f63975779c0df952c0cfd4) +++ firmware/App/Controllers/DryBiCart.c (.../DryBiCart.c) (revision b0da135bc589efaf8ad6f3c8c1988dcb672f3ca0) @@ -457,11 +457,11 @@ void execDryBicart( void ) { //( TRUE == getBicarbChamberFillRequested() ) - //supply dryBicartStartRequest = DRY_BICART_SUPPLY ; + // supply dryBicartStartRequest = DRY_BICART_SUPPLY ; //( TRUE == getBicartFillRequested() ) - //fill dryBicartStartRequest = DRY_BICART_INITIAL_FILL; + // fill dryBicartStartRequest = DRY_BICART_INITIAL_FILL; //( TRUE == getBicartDrainRequested() ) - //drain dryBicartStartRequest = DRY_BICART_DRAIN; + // drain dryBicartStartRequest = DRY_BICART_DRAIN; //else idle dryBicartStartRequest = DRY_BICART_IDLE; updateDrybicartOperation(); @@ -1178,11 +1178,12 @@ //if(FP_MODE_GENP == getCurrentFPOperationMode()) { // TODO: pre gen should start the D12 pump and D14 valve - //setValveState( D14_VALV, VALVE_STATE_OPEN ); + setValveState( D14_VALV, VALVE_STATE_OPEN ); setDialysatePumpTargetRPM( D12_PUMP, FRESH_DIAL_PUMP_INITIAL_RPM_B1_9_B2_0, TRUE ); state = DRY_BICART_FILL_WATER_START_STATE; } + // alrm } return state;