Index: firmware/App/Controllers/DryBiCart.c =================================================================== diff -u -r50aa4c2d630370fa62f871b82db3afc70d763dcb -r3c3ce491ef31e52b19e5166a04467ea359fbfdae --- firmware/App/Controllers/DryBiCart.c (.../DryBiCart.c) (revision 50aa4c2d630370fa62f871b82db3afc70d763dcb) +++ firmware/App/Controllers/DryBiCart.c (.../DryBiCart.c) (revision 3c3ce491ef31e52b19e5166a04467ea359fbfdae) @@ -56,7 +56,7 @@ DRY_BICART_WCID2_FILL_COMPLETE_PRESSURE_PSI : DRY_BICART_WCID1_FILL_COMPLETE_PRESSURE_PSI ) ///< Get fill complete pressure based on the WCID configuration #define DRY_BICART_FILL_INITIATE_PRESSURE_PSI 1.5F ///< Minimum pressure required to initiate the dry bicart fill process. #define DRY_BICART_FILL_COMPLETE_SUPPLY_PRESSURE_PSI 10.0F ///< Maximum pressure allowed at bicart fill during supply process/state -#define DRY_BICART_FILL_INITIATE_SUPPLY_PRESSURE_PSI 3.0F ///< Minimum pressure required to initiate the dry bicart fill during supply process/state +#define DRY_BICART_FILL_INITIATE_SUPPLY_PRESSURE_PSI 4.0F ///< Minimum pressure required to initiate the dry bicart fill during supply process/state #define DRY_BICART_MAX_FILL_CYCLE_CNT 10 ///< Max fill cycle allowed (by override) for dry bicart fill/mix with water. #define DRY_BICART_MINIMUM_FILL_CYCLE_CNT 3 ///< Minimum fill cycle count for filling dry bicart // Dry Bicart Fill Vent