Index: firmware/App/Controllers/MixingControl.c =================================================================== diff -u -r337209eec0c5a2a879f2df587e989f19f5a1effe -rcabab4c56b6840c94126b71696d2c95011cb01a0 --- firmware/App/Controllers/MixingControl.c (.../MixingControl.c) (revision 337209eec0c5a2a879f2df587e989f19f5a1effe) +++ firmware/App/Controllers/MixingControl.c (.../MixingControl.c) (revision cabab4c56b6840c94126b71696d2c95011cb01a0) @@ -97,8 +97,8 @@ #define STD_DILUTION_BICARB_POST_MIX_NA 37.0F ///< Bicarb post mix Sodium #define STD_DILUTION_BICARB_POST_MIX_CHO3 33.0F ///< Post mix bicarbonate -#define DIALYSATE_TARGET_STD_NA_SETTINGS 137.0F ///< Standard target sodium -#define DIALYSATE_TARGET_STD_BICARB_SETTING 33.0F ///< Standard bicarb +//#define DIALYSATE_TARGET_STD_NA_SETTINGS 137.0F ///< Standard target sodium +//#define DIALYSATE_TARGET_STD_BICARB_SETTING 33.0F ///< Standard bicarb #define DIALYSATE_TARGET_USER_ADJ_SODIUM_SETTING getTDSodiumConcentration() ///< User Sodium #define DIALYSATE_TARGET_USER_ADJ_BICARB_SETTING getTDBicarbonateConcentration() ///< User bicarb @@ -724,7 +724,7 @@ // Control based on the measured and target conductivity PI_CONTROLLER_SIGNALS_DATA debugBicarbControl; F32 measuredBicarbConductivity = getFilteredConductivity( D17_COND ); - F32 bicarbConductivity = getBicarbConductivityPre();//getBicarbTargetConductivity(); + F32 bicarbConductivity = getBicarbTargetConductivity();//getBicarbConductivityPre(); if ( fabs ( bicarbConductivity - measuredBicarbConductivity ) > BICARB_DEADBAND_CONTROL ) { @@ -755,7 +755,7 @@ // TODO: use the calculated API F32 measuredDialysateConductivity = getFilteredConductivity( D27_COND ); - F32 DialysateConductivity = getTotalConductivity();//getDialysateTargetConductivity(); + F32 DialysateConductivity = getDialysateTargetConductivity();//getTotalConductivity(); PI_CONTROLLER_SIGNALS_DATA debugAcidControl; if ( fabs ( DialysateConductivity - measuredDialysateConductivity ) > ACID_DEADBAND_CONTROL )