Index: firmware/App/Controllers/MixingControl.c =================================================================== diff -u -r7ec4b7db95ed007bb1ea2565dde4d03c877f245a -ra6ead936473ef260bbd330050fcf7e1b10120175 --- firmware/App/Controllers/MixingControl.c (.../MixingControl.c) (revision 7ec4b7db95ed007bb1ea2565dde4d03c877f245a) +++ firmware/App/Controllers/MixingControl.c (.../MixingControl.c) (revision a6ead936473ef260bbd330050fcf7e1b10120175) @@ -59,7 +59,7 @@ #define BICARB_TARGET_CONDUCTIVITY 2714.0F ///< Target bicarb conductivity #define BICARB_DELTA_CONDUCTIVITY 500.0F ///< Delta bicarb conductivity -#define DIALYSATE_TARGET_CONDUCTIVITY 13734.87F ///< Target dialysate conductivity +#define DIALYSATE_TARGET_CONDUCTIVITY 13613.8F ///< Target dialysate conductivity for 1K #define DIALYSATE_DELTA_CONDUCTIVITY 700.0F ///< Delta dialysate conductivity // this is for reference only , it can be removed later @@ -357,7 +357,7 @@ *************************************************************************/ void execMixingControl( void ) { - if( getCurrentBalancingChamberExecState() > BAL_CHAMBER_STATE_IDLE ) + if ( getCurrentBalancingChamberExecState() > BAL_CHAMBER_STATE_IDLE ) { // closed loop bicarb and acid mixing controller handleDialysateCompositionMixing();