Index: firmware/App/Modes/FPModes/FlushConcentrate.c =================================================================== diff -u -ra0c144ed9f72066d7a7b1d047b01fb76ca56d667 -re8a090df33223455312adc5ad494a3b5e2507f00 --- firmware/App/Modes/FPModes/FlushConcentrate.c (.../FlushConcentrate.c) (revision a0c144ed9f72066d7a7b1d047b01fb76ca56d667) +++ firmware/App/Modes/FPModes/FlushConcentrate.c (.../FlushConcentrate.c) (revision e8a090df33223455312adc5ad494a3b5e2507f00) @@ -203,14 +203,12 @@ { case CONC_FLUSH_IN_PROGRESS: setValveState( M4_VALV, VALVE_STATE_OPEN ); - setValveState( M7_VALV, VALVE_STATE_CLOSED ); setValveState( P6_VALV, VALVE_STATE_CLOSED ); setValveState( P11_VALV, VALVE_STATE_OPEN ); setValveState( P33_VALV, VALVE_STATE_CLOSED ); setValveState( P34_VALV, VALVE_STATE_CLOSED ); setValveState( P37_VALV, VALVE_STATE_CLOSED ); setValveState( P39_VALV, VALVE_STATE_OPEN ); - setValveState( P20_VALV, VALVE_STATE_CLOSED ); setValveState( P43_VALV, VALVE_STATE_CLOSED ); startPermeateTankControl(); if ( TRUE == isBoostPumpInstalled() ) @@ -227,14 +225,12 @@ if ( FALSE == isConcentrateFlushComplete() ) { setValveState( M4_VALV, VALVE_STATE_CLOSED ); - setValveState( M7_VALV, VALVE_STATE_CLOSED ); setValveState( P6_VALV, VALVE_STATE_CLOSED ); setValveState( P11_VALV, VALVE_STATE_CLOSED ); setValveState( P33_VALV, VALVE_STATE_CLOSED ); setValveState( P34_VALV, VALVE_STATE_CLOSED ); setValveState( P37_VALV, VALVE_STATE_CLOSED ); setValveState( P39_VALV, VALVE_STATE_CLOSED ); - setValveState( P20_VALV, VALVE_STATE_CLOSED ); setValveState( P43_VALV, VALVE_STATE_CLOSED ); if ( TRUE == isBoostPumpInstalled() )