Index: firmware/App/Modes/FPModes/FlushConcentrate.c =================================================================== diff -u -re8a090df33223455312adc5ad494a3b5e2507f00 -r25e44b3d8133ef2d8bed9a2e2dd55696c41e8acf --- firmware/App/Modes/FPModes/FlushConcentrate.c (.../FlushConcentrate.c) (revision e8a090df33223455312adc5ad494a3b5e2507f00) +++ firmware/App/Modes/FPModes/FlushConcentrate.c (.../FlushConcentrate.c) (revision 25e44b3d8133ef2d8bed9a2e2dd55696c41e8acf) @@ -203,13 +203,13 @@ { case CONC_FLUSH_IN_PROGRESS: setValveState( M4_VALV, VALVE_STATE_OPEN ); + setValveState( M12_VALV, VALVE_STATE_OPEN ); setValveState( P6_VALV, VALVE_STATE_CLOSED ); setValveState( P11_VALV, VALVE_STATE_OPEN ); setValveState( P33_VALV, VALVE_STATE_CLOSED ); setValveState( P34_VALV, VALVE_STATE_CLOSED ); - setValveState( P37_VALV, VALVE_STATE_CLOSED ); + setValveState( P37_VALV, VALVE_STATE_OPEN ); setValveState( P39_VALV, VALVE_STATE_OPEN ); - setValveState( P43_VALV, VALVE_STATE_CLOSED ); startPermeateTankControl(); if ( TRUE == isBoostPumpInstalled() ) { @@ -225,13 +225,13 @@ if ( FALSE == isConcentrateFlushComplete() ) { setValveState( M4_VALV, VALVE_STATE_CLOSED ); + setValveState( M12_VALV, VALVE_STATE_CLOSED ); setValveState( P6_VALV, VALVE_STATE_CLOSED ); setValveState( P11_VALV, VALVE_STATE_CLOSED ); setValveState( P33_VALV, VALVE_STATE_CLOSED ); setValveState( P34_VALV, VALVE_STATE_CLOSED ); setValveState( P37_VALV, VALVE_STATE_CLOSED ); setValveState( P39_VALV, VALVE_STATE_CLOSED ); - setValveState( P43_VALV, VALVE_STATE_CLOSED ); if ( TRUE == isBoostPumpInstalled() ) {